PAPAYA PLUS, DIGITAL EXTRAORAL SOURCE X-RAY SYSTEM
Applicant
Genoray Co., Ltd.
Product Code
MUH · Dental
Decision Date
Oct 31, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
PAPAYA Plus is a digital extraoral source X-ray system intended to produce panoramic and cephalometric(option) images of the oral and craniofacial anatomy for a precise treatment planning in adult and pediatric care. The system is used for dental & skull radiographic examination and diagnosis of teeth, jaw, oral structure, and skull by exposing an X-ray image receptor to ionizing radiation, with a digital imaging capability for taking both panoramic and cephalometric images. And this system can be equipped CUST (Tomographic) option, which is capable of taking cross-sectional radiographic images. These images provide dimensional information for dental implant planning and information about location of impacted teeth.
Device Story
Digital extraoral X-ray system; produces panoramic, cephalometric, and cross-sectional (CUST) images of oral/craniofacial anatomy. Input: ionizing radiation exposure to digital image receptor. Transformation: CUST option reconstructs tomographic images from pre-acquired projection radiographic images. Output: digital radiographic images for dental/skull diagnosis. Used in dental clinics; operated by trained dental professionals. Healthcare providers use images for treatment planning, implant positioning, and identifying impacted teeth. Benefits: provides dimensional anatomical information for precise clinical decision-making.
Clinical Evidence
Bench testing and clinical evaluation. Bench testing included electrical/mechanical safety (IEC 60601 series), EMC (IEC 60601-1-2), and imaging performance (MTF, DQE, dynamic range) per FDA guidance for solid-state X-ray devices. Clinical evaluation involved a licensed dentist reviewing panoramic, cephalometric, and CUST images to confirm diagnostic quality. No specific sensitivity/specificity metrics provided; qualitative assessment deemed images sufficient for diagnosis.
Indicated for dental and skull radiographic examination and diagnosis of teeth, jaw, oral structure, and skull in adult and pediatric patients, including dental implant planning and localization of impacted teeth.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized depiction of a human figure embracing a bird. The emblem is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 31, 2014
GENORAY Co., Ltd. % Ms. Kaitlynn Min Business Development Manager GENORAY America, Inc. 3002 Dow Avenue, Suite 420 TUSTIN CA 92780
Re: K141700
Trade/Device Name: PAPAYA Plus, Digital Extraoral Source X-ray System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: October 2, 2014 Received: October 6, 2014
Dear Ms. Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141700
Device Name
Digital X-ray Imaging System (Model: PAPAYA Plus)
# Indications for Use (Describe)
PAPAYA Plus is a digital extraoral source X-ray system intended to produce panoramic and cephalometric(option) images of the oral and craniofacial anatomy for a precise treatment planning in adult and pediatric care. The system is used for dental & skull radiographic examination and diagnosis of teeth, jaw, oral structure, and skull by exposing an X-ray image receptor to ionizing radiation, with a digital imaging capability for taking both panoramic and cephalometric images. And this system can be equipped CUST (Tomographic) option, which is capable of taking cross-sectional radiographic images. These images provide dimensional information for dental implant planning and information about location of impacted teeth.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Exhibit 5 510(k) Summary
Date of Summary Preparation: July 25, 2014
- 1. Submitter and US Official Correspondent
| Submitter: | GENORAY Co., Ltd. |
|----------------|-------------------------------------------------------------------------------------------------------------------|
| Address: | #512, Byucksan Technopia<br>434-6, Sangdaewon 1-dong,<br>Jungwon-gu, Seongnam-city<br>Gyeonggi-do, 462-716, Korea |
| Telephone No.: | +82-31-740-4100 |
| Fax: | +82-31-737-8025 |
Official Correspondent (U.S): Kaitlynn Min - Business Manager Correspondent: GENORAY America Inc. Address: 3002 Dow Avenue, Suite 420, Tustin, CA 92780 Telephone No.: 714-289-8020 Fax: 714-786-8919 Email: kaitlynn@genorayamerica.com
- Establishment Registration Number 2. 3005843418
- 3. Device Information
| Proprietary/Trade Name: | PAPAYA Plus |
|-------------------------|-----------------------------------------|
| Common/Usual Name: | Digital X-ray Imaging System |
| Classification Name: | Extraoral Source x-ray system |
| Product Code: | MUH |
| Device Class: | Class II per regulation 21 CFR 872.1800 |
- 4. Predicate Device (Equivalent Legally Marketed Device) PAPAYA Plus (K130419, GENORAY Co., Ltd) Volumetric Tomography (K063773, Instrumentarium Dental) Orthopantomograph OP 200 (K046312, Instrumentarium Dental)
- ડ . Description of the Device
PAPAYA Plus is a diagnostic imaging system which consists of multiple image acquisition modes; panorama and cephalometric. And proposed PAPAYA Plus has the CUST imaging option which is used to reconstruct tomographic images from a set of preacquired projection radiographic images of the object.
CUST algorithm has already been used for existing panoramic units as the predicate device OP200 (k046312) use the VT (k063773) imaging method. VT imaging method which is similar with CUST is described in the following scientific publication.
CEDERLUND, A; KALKE, M.; WELANDER, U. Volumetric tomography-a new tomographic technique for panoramic units. Dentomaxillofacial Radiology, 2009, 38: 104-111
{4}------------------------------------------------
#### Indications for use (intended use) 6.
PAPAYA Plus is a digital extraoral source X-ray system intended to produce panoramic and cephalometric(option) images of the oral and craniofacial anatomy for a precise treatment planning in adult and pediatric care. The system is used for dental & skull radiographic examination and diagnosis of teeth, jaw, oral structure, and skull by exposing an X-ray image receptor to ionizing radiation, with a digital imaging capability for taking both panoramic and cephalometric images. And this system can be equipped CUST (Tomographic) option, which is capable of taking cross sectional radiographic images. These images provide dimensional information for dental implant planning and information about location of impacted teeth.
| Name | Proposed device<br>PAPAYA Plus | Predicate device<br>PAPAYA Plus | Volumetric Tomography | Orthopantomograph®<br>OP200, OP200D,<br>Orthoceph®OC200,<br>OC200D |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | GENORAY Co., Ltd. | GENORAY Co., Ltd. | Instrumentarium Dental | Instrumentarium Dental |
| 510(k)<br>No. | K141700 | K130419 | K063773 | K046312 |
| Figure | Image: PAPAYA Plus | Image: PAPAYA Plus | *The Volumetric<br>Tomography imaging<br>option is used for<br>Orthopantomograph®<br>OP100 and OP200 product<br>families. | Image: Orthopantomograph |
| Indications<br>for use | PAPAYA Plus is a digital<br>extraoral source X-ray<br>system intended to<br>produce panoramic and<br>cephalometric(option)<br>images of the oral and<br>craniofacial anatomy for a<br>precise treatment planning<br>in adult and pediatric care.<br>The system is used for<br>dental & skull<br>radiographic examination<br>and diagnosis of teeth, jaw,<br>oral structure, and skull by<br>exposing an X-ray image<br>receptor to ionizing<br>radiation, with a digital<br>imaging capability for<br>taking both panoramic and<br>cephalometric images.<br>And This system can be<br>equipped<br>Cust(Tomographic) option,<br>which is capable of taking<br>cross-sectional | PAPAYA Plus is a digital<br>extraoral source X-ray<br>system intended to produce<br>panoramic and<br>cephalometric(option)<br>images of the oral and<br>craniofacial anatomy for a<br>precise treatment planning<br>in adult and pediatric care.<br>The system is used for<br>dental & skull radiographic<br>examination and diagnosis<br>of teeth, jaw, oral<br>structure, and skull by<br>exposing an X-ray image<br>receptor to ionizing<br>radiation, with a digital<br>imaging capability for<br>taking both panoramic and<br>cephalometric images. | Volumetric Tomography is<br>intended to be used for<br>producing cross-sectional<br>(tomographic)<br>radiographic images from<br>the edentulous or dentate<br>area of the jaws. The<br>cross-sectional images<br>provide dimensional<br>information for dental<br>implant planning and<br>information about location<br>of impacted teeth. | Orthopantomograph®<br>OP200 (film unit) and<br>OP200 D (digital unit)<br>devices are intended to be<br>used for producing X-ray<br>radiographs of dentition,<br>TM-joints and other oral<br>structures. The units are<br>capable of taking<br>panoramic, TM-joint and<br>maxillary sinus<br>radiographs from patients.<br>When the units are<br>equipped with<br>cephalometric option<br>Orthoceph® OC200 (film<br>unit) or OC200 D (digital<br>unit) units can be used for<br>cephalometric radiography<br>and examinations related<br>thereto. OP200 or OC200<br>units can also be<br>equipped with Ortho<br>Trans (OT) option, which<br>is capable of taking both |
| | radiographic<br>images.<br>These images provide<br>dimensional information<br>for dental implant planning<br>and information about<br>location of impacted teeth. | | | cross and longitudinal<br>slices of region of interest.<br>Ortho Trans uses linear<br>tomography imaging<br>principle. |
| Perform<br>ance<br>Specific<br>ation | Panoramic and<br>cephalometric | Panoramic and<br>cephalometric | - | Panoramic and<br>cephalometric |
| Input<br>Voltage | 120 V~ | 120 V~ | - | 110 V~, 50/60 Hz,<br>1.65kVA |
| Tube<br>Voltage | 60-90 kV | 60-90 kV | - | 57 - 85 kV |
| Tube<br>Current | 4-12 mA | 4-12 mA | - | 2-16 mA |
| Focal<br>Spot<br>Size | 0.5 mm | 0.5 mm | - | 0.5 mm |
| Exposur<br>e Time | Standard panorama: 12 sec<br>Cephalo (Normal): 8 sec | Standard panorama: 12 sec<br>Cephalo (Normal): 8 sec | - | Panorama: 2.7-16 sec<br>Cephalo: 0.2-19 sec |
| Exposu<br>re<br>mode | Panorama:<br>Panoramic mode<br>TMJ mode<br>SINUS mode<br>CUST mode<br>Cephalo: Cephalo mode | Panorama:<br>Panoramic mode<br>TMJ mode<br>SINUS mode<br>Cephalo: Cephalo mode | Volumetric Tomography | Panorama:<br>Panoramic mode<br>TMJ mode<br>SINUS mode<br>Linear Tomography<br>Volumetric Tomography<br>Cephalo: Cephalo mode |
| Image<br>Receptor | Panoramic &<br>cephalometric sensor :<br>CdTe Sensor | Panoramic &<br>cephalometric sensor :<br>CdTe Sensor | - | Panoramic, Cephalometric<br>sensor: CCD Detector |
#### 7. Substantial equivalence chart
{5}------------------------------------------------
The proposed PAPAYA Plus has the same specification for mechanism as predicate device, PAPAYA Plus (k130419). The primary difference is to add the CUST mode in exposure modes which shows the cross-sectional images through reconstruction. CUST mode is similar with Volumetric Tomography (k063773) which reconstructs tomographic images.
The image performances and dosimetry of CUST mode have evaluated through clinical evaluation report. Evaluation results of dosimetry and images for safe and effective use of PAPAYA Plus are satisfied. We also had a licensed dentist review of panoramic, cephalometric and CUST images. She found them to have enough quality for diagnosis.
Therefore, PAPAYA Plus is deemed to be substantially equivalent to the predicate devices in safety and effectiveness.
# 8. Safety, EMC and Performance data comparison to Predicate
Device Safety, EMC and Performance data is same to predicate device which has been established in 510(k) submission K130419 as below. All test results were satisfactory.
- Electrical, mechanical, environmental safety and performance testing according to standard IEC i 60601-1, IEC 60601-1-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28 and IEC 60601-2-32, and IEC 61223-3-4 were performed.
{6}------------------------------------------------
- EMC testing was conducted in accordance with standard IEC 60601-1-2. i
- FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed. The tests include the MTF, DQE and the dynamic range of the panoramic sensor and the cephalometric sensor. Both the MTF of the two detectors shows the resolution more than 80% at 21p / mm and the DQE of them shows the performance of about 80% at 01p / mm. The dynamic range of them shows more than 72dB.
- PAPAYA Plus meets the EPRC standards (21 CFR 1020.30. 31) i
- PAPAYA Plus also meets the provisions of NEMA PS 3.1-3.18. Digital Imaging and । Communications in Medicine (DICOM) Set.
- i PAPAYA Plus was tested for safety and effectiveness related of the CUST in Clinical Evaluation report.
# 9. Conclusion
In reference to the comparison information provided in substantial equivalence chart, most of functions and electronic features are similar with predicate devices. The addition of CUST option was evaluated through the performance data. Thus PAPAYA Plus is substantially equivalent as compared with the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.