K131693 · Ray Co., Ltd. · MUH · Nov 1, 2013 · Dental
Device Facts
Record ID
K131693
Device Name
RAYSCAN A-EXPERT
Applicant
Ray Co., Ltd.
Product Code
MUH · Dental
Decision Date
Nov 1, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
The RAYSCAN α- Expert Dental X-Ray System is an extraoral source dental panoramic and optional cephalometric X-ray system intended to produce X-rays for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures.
Device Story
RAYSCAN α-Expert is a digital dental X-ray system for panoramic and optional cephalometric imaging. It uses an X-ray source and flat panel sensors (or CdTe direct flat panel for scan-type cephalometric) to capture images of teeth, jaw, and oral structures. The system includes optional 'SMARTDent' software for processing and archiving. Operated by dental professionals in clinical settings, the device produces digital radiographs to assist in diagnosing oral diseases. Modifications from the predicate include an additional scan-type cephalometric sensor and updated software protocols for segmentation, bitewing, and orthogonal imaging. The output is viewed by clinicians to inform dental treatment planning.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical performance testing, including software verification and validation, and expert review of image comparisons.
Technological Characteristics
Extraoral dental X-ray system; 60-90kVp voltage; 4-17mA current; 0.5mm focal size; 2.6mm Al equivalent filtration. Detectors: Flat panel X-ray sensor (Pano/One-shot Ceph) and CdTe direct flat panel sensor (Scan-type Ceph). Connectivity: Digital imaging with optional SMARTDent software. Standards: IEC 60601-1, 60601-1-1, 60601-1-3, 60601-2-7, 60601-2-28, 60601-2-32, 60601-1-2.
Indications for Use
Indicated for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures in patients requiring extraoral panoramic or cephalometric imaging.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
# NOV 0 1 2013
K131693
Page 1 of 5
510(k) Submission- RAYSCAN a-Expert
# Special 510(k) Summary
The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
Date:
| APPLICANT | RAY Co.,Ltd |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ADDRESS | #362-43 (218 Maeyoung St.) 3rd & 4th Floor,<br>Wonchun-dong, Youngtong-gu, Suwon-si, Gyeonggi-do, Korea |
| Manufacturer | RAY Co.,Ltd<br>#362-43 (218 Maeyoung St.) 3rd & 4th Floor,<br>Wonchun-dong, Youngtong-gu, Suwon-si, Gyeonggi-do, Korea<br>TEL : +82-31-605-1000<br>FAX : +82-2-6280-5534 |
| Contact Person | Yun-Jung. HA/Manager<br>e-mail : yunjung.ha@raymedical.co.kr |
## Device Name
Trade/Proprietary Name : RAYSCAN α – Expert Common Name: Dental panoramic and cephalometric X-ray system
## Classification
Extraoral source dental X-ray system (21 CFR 872.1800) Class : II Product code : MUH Panel : Radiology
### Predicate device
RAYSCAN a-Expert (K122918)
{1}------------------------------------------------
## Description
Modified RAYSCAN α- Expert is designed for panoramic scanning of teeth, jaw and oral cavity, used to create and control the X-ray beam. And as a dental digital panoramic X-ray system with X-ray located on outer part of the oral cavity, includes the Cephalometric scanning function, as an option, for acquiring images of the head.
The modifications are as following:
- Hardware: Added option CEPH sensor (Scan type). In addition to the one-shot type CEPH sensor of original device (K122918), the modified device offers an additional Scan type CEPH sensor.
- · Updated Software including:
- Additional protocol for new CEPH sensor (Scan type): Lateral wide
- Additional PANO protocols: Segmentation (Individual Tooth), Bitewing, and Orthogonal
RAYSCAN α- Expert offers digital imaging with or without the optional One-shot type & Scan type cephalometric attachment.
Detector Options:
Base: RAYSCAN α-P: PANO Option: RAYSCAN α-OC: PANO+One-shot CEPH Option: RAYSCAN α-SC: PANO+SCAN CEPH
The system includes processing, and archiving "SMARTDent "software(Optional)
#### Indication for use
The RAYSCAN α- Expert Dental X-Ray System is an extraoral source dental panoramic and optional cephalometric X-ray system intended to produce X-rays for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures.
{2}------------------------------------------------
ﺎ ﺳ
| Parameter | RAYSCAN α-Expert<br>[Modified] | RAYSCANα-Expert<br>[K122918] | Remarks |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Common Name | Dental panoramic and<br>cephalometric X-ray system | Dental panoramic and<br>cephalometric X-ray system | Same |
| Indications<br>for use | The RAYSCAN α- Expert Dental<br>X-Ray System is an extraoral<br>source dental panoramic and<br>optional cephalometric X-ray<br>system intended to produce<br>X-rays for dental radiographic<br>examination and diagnosis of<br>diseases of the teeth, jaw,<br>and oral structures. | The RAYSCAN α- Expert Dental<br>X-Ray System is an extraoral<br>source dental panoramic and<br>optional cephalometric X-ray<br>system intended to produce<br>X-rays for dental radiographic<br>examination and diagnosis of<br>diseases of the teeth, jaw, and<br>oral structures. | Same |
| Performance<br>Specification | Panoramic<br>Cephalometric(optional)<br>- One_shot type<br>- Scan type | Panoramic<br>Cephalometric(optional)<br>-One_shot type | Additional<br>Scan type<br>Cephalometric<br>(Optional) |
| Detector type | α-OC<br>: PANO+One-shot CEPH(option)<br>α -P : PANO<br>α -SC: PANO+SCAN CEPH(option) | α-OC<br>:PANO+One-shotCEPH(option)<br>α -P: PANO | Additional<br>Scan type<br>Cephalometric<br>(Optional) |
| Detector<br>Type | Pano: Flat panel X-ray sensor<br>Ceph(Optional)<br>- Flat panel X-ray sensor<br>[One-shot type]<br>- CdTe Direct flat panel<br>sensor[Scan type] | Pano: Flat panel X-ray sensor<br>Ceph(Optional)<br>- Flat panel X-ray sensor | Additional<br>Scan type<br>Cephalometric<br>(Optional) |
| Focal size | 0.5mm | 0.5mm | Same |
| X-ray Voltage | 60~90kVp | 60~90kVp | Same |
| X-ray Current | 4~17mA | 4~17mA | Same |
| Total Filtration | 2.6 mm Al equivalent | 2.6 mm Al equivalent | Same |
| Magnification | Pano : 1.31<br>Ceph[One-shot type] : 1.13 | Pano : 1.31<br>Ceph[One-shot type] : 1.13 | Same |
| | Ceph[Scan type] : 1.11 | | Additional<br>specification<br>of Scan type<br>Cephalometric<br>(Optional) |
| Scan time | Pano : 14sec | Pano : 14sec | Same |
| | Ceph[One-shot type] : | Ceph | Same |
| | 0.3sec~3.0sec | : 0.3sec~3.0sec | |
| | Ceph[Scan type] : below 18sec | | Additional<br>specification<br>of Scan type<br>Cephalometric<br>(Optional) |
| Applicable<br>Standards | IEC 60601-1 | IEC 60601-1 | Same |
| | IEC 60601-1-1 | IEC 60601-1-1 | |
| | IEC 60601-1-3 | IEC 60601-1-3 | |
| | IEC 60601-2-7 | IEC 60601-2-7 | |
| | IEC 60601-2-28 | IEC 60601-2-28 | |
| | IEC 60601-2-32 | IEC 60601-2-32 | |
| | IEC 60601-1-2 | IEC 60601-1-2 | |
| Certificate<br>Product | CE0120(MDD93/42/EEC) | CE0120(MDD93/42/EEC) | Same |
# Statement of Substantial Equivalence
·
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
3 / 5
.
{3}------------------------------------------------
#### 510(k) Submission- RAYSCAN a-Expert
Safety details, for instance the non-clinical performance, in regards to intended use, safety characteristics, PANORAMA sensor (Detector) and One-shot CEPH sensor(Detector) are equivalent. The only difference is the additional option of Scan type CEPH sensor. The safety & effectiveness reports for the added Scan type cephalometric sensor is provided separately.
Remaining sensors are the same, the non-clinical considerations thereof are also equivalent, and the report regarding non-clinical & clinical consideration for the added Scan CEPH sensor is provided separately.
It shares the same technological characteristics as the predicate devices. Minor technological differences do not raise any new questions regarding safety or effectiveness of the device.
Based on the non-clinical and clinical consideration and the outcome of an expert review of image comparisons for both devices, new RAYSCAN α-Expert is substantially equivalent, in terms of safety and effectiveness, with RAYSCAN α- Expert[K122918].
4 / 5
{4}------------------------------------------------
510(k) Submission- RAYSCAN a-Ex
K131693
Page 5 of 5
### Safety and Effectiveness Information
Software verification testing and validation testing was performed to confirm that the modified device performed as intended and that changes made to the hardware and software had no adverse impact on the functionality of the system.
All tests met requirements demonstrating that the modified device performed as expected.
Electrical, mechanical, environmental safety and performance testing according to standards IEC 60601-1, IEC 60601-1-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28 and IEC 60601-2-32 was performed, and EMC testing was conducted in accordance with the standard IEC 60601-1-2.
Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" were performed.
All test results were satisfactory.
### Conclusions
Based on a comparison of intended use, indications, construction materials, principal of Operation, features and technical data, the RAYSCAN &-Expert system are safe and effective to perform its intended use as well as substantially equivalent to the predicate device
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized emblem of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem. The text is in all capital letters and is in a sans-serif font.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 1, 2013
Ray Co., Ltd. % Mr. Andrew Paeng Consultant 4747 Hoen Avenue SANTA ROSA CA 95405
Re: K131693
Trade/Device Name: RAYSCAN a-Expert Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: October 1. 2013 Received: October 11, 2013
Dear Mr. Paeng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Andrew Paeng
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours.
Mekael D. D'Hara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
K131693 510(k) Number (if known):
Device Name: RAYSCAN α-Expert
Indications For Use:
The RAYSCAN α- Expert Dental X-Ray System is an extraoral source dental panoramic and optional cephalometric X-ray system intended to produce X-rays for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures.
Prescription Use X AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Michael D. O'Hara
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiolgical Health Page 1 of 510(k) ======================================================================================================================================================================= K131693
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.