← Product Code [MUH](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH) · K130585

# PCH-2500 DIGITAL EXTRAORAL SOURCE X-RAY SYSTEM, PAX-I DIGITAL EXTRAORAL SOURCE X-RAY SYSTEM (K130585)

_VATECH Co., Ltd. · MUH · Jul 15, 2013 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K130585

## Device Facts

- **Applicant:** VATECH Co., Ltd.
- **Product Code:** [MUH](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH.md)
- **Decision Date:** Jul 15, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.1800
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Pediatric

## Indications for Use

PCH-2500 is a digital extra oral source X-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, and X-ray technicians and other professionals licensed by the law of the states in which the device is used.

## Device Story

PCH-2500 is a digital extra-oral dental X-ray system; captures panoramic and cephalometric images of teeth and jaws. Input: X-ray radiation; Output: digital radiographic images. Operation: dentist or X-ray technician in clinical setting. Device utilizes X-ray generator and solid-state detectors (CMOS photodiode arrays or amorphous silicon TFT with scintillator). Upgraded version features new detectors (1417PGA, Xmaru3001CF) with different active areas; new power board for universal input voltage (100-240V); and new X-ray generator options with increased tube voltage/current capacity. Images viewed via EasyDent software. Output assists clinicians in oral/maxillofacial diagnosis. Benefits: provides diagnostic information for treatment planning.

## Clinical Evidence

Clinical image evaluation performed for each new X-ray generator option to assess radiographic image quality. Non-clinical performance testing included MTF, DQE, and NPS measurements for new detectors (Xmaru3001CF, 1417PGA). Electrical/mechanical safety testing conducted per IEC 60601-1, 60601-1-1, 60601-1-3, 60601-2-7, 60601-2-28, and 60601-2-32. EMC testing per IEC 60601-1-2. Acceptance testing per IEC 61223-3-4. All results were satisfactory, demonstrating substantial equivalence.

## Technological Characteristics

Digital extra-oral X-ray system. Detectors: CMOS photodiode arrays or amorphous silicon TFT with scintillator. Tube voltage: 50-99 kV; Tube current: 4-16 mA. Focal spot: 0.5 mm. Filtration: min 2.8 mmAl. Power: 100-240 V AC. Connectivity: DICOM 3.0 compatible. Software: EasyDent viewing program. Standards: IEC 60601 series, NEMA PS 3.1-3.18 (DICOM), 21 CFR 1020.30/31.

## Regulatory Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

## Predicate Devices

- PCH-2500 ([K122155](/device/K122155.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 872.1800.

Date 510K summary prepared : February 28, 2013

Submitter's Name, address, telephone number, a contact person:

| Submitter's Name :                                 | Vatech Co., Ltd.                                                             |
|----------------------------------------------------|------------------------------------------------------------------------------|
| Submitter's Address:                               | 23-4, Seogu-Dong, Hwaseong-Si,<br>Gyeonggi-Do, 445-170,<br>Republic of Korea |
| Submitter's Telephone:                             | +82-31-379-9585                                                              |
| Contact person:                                    | Mr. Sung-Hee Park                                                            |
| Official Correspondent:<br>(U.S. Designated agent) | Dave Kim (davekim@mtech-inc.net)                                             |
| Address:                                           | 12946 Kimberley Ln, Houston, TX 77079                                        |
| Telephone:                                         | +713-467-2607                                                                |
| Fax:                                               | +713-464-8880                                                                |

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

| Trade/Proprietary Name: | PCH-2500                                                                |
|-------------------------|-------------------------------------------------------------------------|
| Common Name:            | Digital X-ray Imaging System                                            |
| Classification Name:    | System, X-Ray, Extra oral Source, Digital (21CFR<br>872.1800, class II) |
| Product Code:           | MUH                                                                     |

{1}------------------------------------------------

## Predicate Device:

| Manufacturer:  | Vatech Co., Ltd |
|----------------|-----------------|
| Device Name:   | PCH-2500        |
| 510(k) Number: | K130585         |

### Description:

PCH-2500 is a dental digital radiographic imaging system which is available in two different image acquisition modes. Specifically designed for dental radiography of the teeth or jaws. PCH-2500 can be equipped with four dedicated sensors for two different X-ray modalities : one panoramic (Xmaru1501CF), two cephalometric scan type (Xmaru2301CF and Xmaru3001CF) and three flat panel one shot ceph sensors (1210SGA, 910SGA and 1417PGA). The proposed device is available with two X-ray generator options.

PCH-2500 offers the digital panoramic and cephalometric X-ray modality for dental radiographs. The multi platforms of PCH-2500 imaging mode provides a wide range of imaging options based on the patient diagnostic needs.

### Indication for use:

PCH-2500 is a digital extra oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, x-ray technicians and other professionals licensed by the law of the state in which the device is used.

#### Summary of the technological characteristics of the device compared to the predicate device:

The new device described in this Traditional 510(k) submission is an upgraded version of the predicate device with the same model name, the same indications for use and technical characteristics. Table I summarizes the technological characteristics of the new vs. the predicate device.

{2}------------------------------------------------

# 510(k) Submission – PCH-2500

| Table 1. Comparison of new PCH-2500 and the predicate device |                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                             |              |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Characteristic                                               | Proposed<br>Vatech Co., Ltd.<br>PCH-2500                                                                                                                                                                                                                                                                                                                                                                                    | Predicate<br>Vatech Co., Ltd.<br>PCH-2500                                                                                                                                                                                                                                                                                                                                                                                   |              |
| 510(k) number                                                | K130585                                                                                                                                                                                                                                                                                                                                                                                                                     | K122155, dated on<br>10/4/2012                                                                                                                                                                                                                                                                                                                                                                                              |              |
| Indications<br>for use                                       | PCH-2500 is a digital<br>extra oral source x-ray<br>system intended to take<br>panoramic and<br>cephalometric images of<br>the oral and maxillofacial<br>anatomy to provide<br>diagnostic information<br>for adult and pediatric<br>patients. The device<br>should be operated and<br>used by dentists, x-ray<br>technicians and other<br>professionals licensed by<br>the law of the state in<br>which the device is used. | PCH-2500 is a digital<br>extra oral source x-ray<br>system intended to take<br>panoramic and<br>cephalometric images of<br>the oral and maxillofacial<br>anatomy to provide<br>diagnostic information<br>for adult and pediatric<br>patients. The device<br>should be operated and<br>used by dentists, x-ray<br>technicians and other<br>professionals licensed by<br>the law of the state in<br>which the device is used. |              |
| Performance<br>Specification                                 | Panoramic and<br>cephalometric                                                                                                                                                                                                                                                                                                                                                                                              | Panoramic and<br>cephalometric                                                                                                                                                                                                                                                                                                                                                                                              |              |
| Input Voltage                                                | AC 100-240 V                                                                                                                                                                                                                                                                                                                                                                                                                | AC 100-120 / 200-240 V                                                                                                                                                                                                                                                                                                                                                                                                      |              |
| Tube Voltage                                                 | 50-99 kV                                                                                                                                                                                                                                                                                                                                                                                                                    | 50-90 kV                                                                                                                                                                                                                                                                                                                                                                                                                    |              |
| Tube Current                                                 | 4-16 mA                                                                                                                                                                                                                                                                                                                                                                                                                     | 4-10 mA                                                                                                                                                                                                                                                                                                                                                                                                                     |              |
| Exposure Time                                                | Max. 20.2 s                                                                                                                                                                                                                                                                                                                                                                                                                 | Max. 20.2 s                                                                                                                                                                                                                                                                                                                                                                                                                 |              |
| X-ray Source                                                 | D-052SB<br>OPX/105                                                                                                                                                                                                                                                                                                                                                                                                          | D-052SB                                                                                                                                                                                                                                                                                                                                                                                                                     |              |
| X-ray Generator                                              | DG-07D11T2 (for D-<br>052SB)<br>DG-07D11C1 (for<br>OPX/105)                                                                                                                                                                                                                                                                                                                                                                 | HDG-07B10T2                                                                                                                                                                                                                                                                                                                                                                                                                 |              |
| Focal Spot Size                                              | 0.5 mm                                                                                                                                                                                                                                                                                                                                                                                                                      | 0.5 mm                                                                                                                                                                                                                                                                                                                                                                                                                      |              |
| Slice Width                                                  | 0.1 mm min.                                                                                                                                                                                                                                                                                                                                                                                                                 | 0.1 mm min.                                                                                                                                                                                                                                                                                                                                                                                                                 |              |
| Total Filtration                                             | Min. 2.8 mmAl                                                                                                                                                                                                                                                                                                                                                                                                               | Min. 2.8 mmAl                                                                                                                                                                                                                                                                                                                                                                                                               |              |
| Chin Rest                                                    | Equipped Headrest                                                                                                                                                                                                                                                                                                                                                                                                           | Equipped Headrest                                                                                                                                                                                                                                                                                                                                                                                                           |              |
| Mechanical                                                   | Compact design                                                                                                                                                                                                                                                                                                                                                                                                              | Compact design                                                                                                                                                                                                                                                                                                                                                                                                              |              |
| Electrical                                                   | LDCP logic circuit                                                                                                                                                                                                                                                                                                                                                                                                          | LDCP logic circuit                                                                                                                                                                                                                                                                                                                                                                                                          |              |
| Software                                                     | DICOM 3.0 Format<br>compatible                                                                                                                                                                                                                                                                                                                                                                                              | DICOM 3.0 Format<br>compatible                                                                                                                                                                                                                                                                                                                                                                                              |              |
| 2D Image Viewing<br>Program                                  | EasyDent                                                                                                                                                                                                                                                                                                                                                                                                                    | EasyDent                                                                                                                                                                                                                                                                                                                                                                                                                    |              |
| Anatomical Sites                                             | Maxillofacial                                                                                                                                                                                                                                                                                                                                                                                                               | Maxillofacial                                                                                                                                                                                                                                                                                                                                                                                                               |              |
| Image<br>Receptor                                            | panoramic<br>(CMOS<br>photodiode<br>array)                                                                                                                                                                                                                                                                                                                                                                                  | Xmaru1501CF                                                                                                                                                                                                                                                                                                                                                                                                                 | Xmaru1501CF  |
|                                                              | Cephalometric<br>(CMOS<br>photodiode<br>array)                                                                                                                                                                                                                                                                                                                                                                              | Xmaru2301CF                                                                                                                                                                                                                                                                                                                                                                                                                 | Xmaru2301CF  |
|                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                             | Xmaru3001CF                                                                                                                                                                                                                                                                                                                                                                                                                 | -            |
|                                                              | Cephalometric<br>(Amorphous<br>silicon TFT<br>with<br>scintillator)                                                                                                                                                                                                                                                                                                                                                         | 1210SGA                                                                                                                                                                                                                                                                                                                                                                                                                     | 1210SGA      |
|                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                             | 910SGA                                                                                                                                                                                                                                                                                                                                                                                                                      | 910SGA       |
|                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                             | 1417PGA                                                                                                                                                                                                                                                                                                                                                                                                                     | -            |
| Pixel<br>Resolution                                          | Xmaru1501CF                                                                                                                                                                                                                                                                                                                                                                                                                 | 5 lp/mm                                                                                                                                                                                                                                                                                                                                                                                                                     | 5 lp/mm      |
|                                                              | Xmaru2301CF                                                                                                                                                                                                                                                                                                                                                                                                                 | 5 lp/mm                                                                                                                                                                                                                                                                                                                                                                                                                     | 5 lp/mm      |
|                                                              | Xmaru3001CF                                                                                                                                                                                                                                                                                                                                                                                                                 | 5 lp/mm                                                                                                                                                                                                                                                                                                                                                                                                                     | -            |
|                                                              | 1210SGA                                                                                                                                                                                                                                                                                                                                                                                                                     | 3.9 lp/mm                                                                                                                                                                                                                                                                                                                                                                                                                   | 3.9 lp/mm    |
|                                                              | 910SGA                                                                                                                                                                                                                                                                                                                                                                                                                      | 3.9 lp/mm                                                                                                                                                                                                                                                                                                                                                                                                                   | 3.9 lp/mm    |
|                                                              | 1417PGA                                                                                                                                                                                                                                                                                                                                                                                                                     | 3.9 lp/mm                                                                                                                                                                                                                                                                                                                                                                                                                   | -            |
| Pixel Size                                                   | Xmaru1501CF                                                                                                                                                                                                                                                                                                                                                                                                                 | 100 x 100 um                                                                                                                                                                                                                                                                                                                                                                                                                | 100 x 100 um |
|                                                              | Xmaru2301CF                                                                                                                                                                                                                                                                                                                                                                                                                 | 100 x 100 um                                                                                                                                                                                                                                                                                                                                                                                                                | 100 x 100 µm |
|                                                              | Xmaru3001CF                                                                                                                                                                                                                                                                                                                                                                                                                 | 100 x 100 um                                                                                                                                                                                                                                                                                                                                                                                                                | -            |
|                                                              | 1210SGA                                                                                                                                                                                                                                                                                                                                                                                                                     | 127 x 127 μm                                                                                                                                                                                                                                                                                                                                                                                                                | 127 x 127 μm |
|                                                              | 910SGA                                                                                                                                                                                                                                                                                                                                                                                                                      | 127 x 127 um                                                                                                                                                                                                                                                                                                                                                                                                                | 127 x 127 μm |
|                                                              | 1417PGA                                                                                                                                                                                                                                                                                                                                                                                                                     | 127 x 127 um                                                                                                                                                                                                                                                                                                                                                                                                                | -            |

Table 1. Comparison of new PCH-2500 and the predicate device

{3}------------------------------------------------

-

.

{4}------------------------------------------------

-----

{5}------------------------------------------------

# Summary of Performance Testing:

The PCH-2500 dental digital radiographic imaging system described in this 510(k) is identical to the predicate device in its indications for use, performance, materials, safety characteristics, image viewing program and accessory components

Furthermore, the following information further substantiates the substantial equivalence between two devices:

- The fundamental technological characteristics of the subject and predicate . device were the same.
- Laboratory and clinical performance testing using the same test protocols . as used for the cleared detectors was evaluated by qualified individuals employed by the sponsor to demonstrate that adequate design controls (according to 21 CFR 820.30) were in place.
- . The intended use of the modified device, as described in the labeling, has not changed as a result of the labeling modification(s).

For both devices, the differences are as follows.

1. New SSXI detectors, 1417PGA (One Shot Ceph mode) and Xmaru3001CF (Scan Ceph mode) for the newly upgraded PCH-2500 have different active areas compared with k122155, the predicate device.

2. Change to Free Input Voltage: For the predicate device, changing the input voltage from 110V to 200V would require separate tools and electrical works whereas the new device is equipped with a newly designed power board which is capable of handling the input power between 100 V and 240 V without a separate tool or electrical modification.

3. The proposed PCH-2500 is available with two different X-ray generator options. To evaluate the safety, safety test is conducted for each generator according to the IEC Standard. Moreover, separate image evaluation is performed for each Xray generator which is considered as one of critical components affecting the quality of radiographic images and imaging performance of the device.

4. For proposed PCH-2500, a new generator for X-ray tube has the capacity to generate more tube current and tube voltage than the predicate device. Moreover, the maximum irradiation condition for each capture mode is defined differently by diversifying the operating range of the generator specifications.

{6}------------------------------------------------

The non-clinical performance (including MTF, DQE, NPS test) and clinical consideration report for the new image receptor, Xmaru3001CF and 1417PGA, are provided separately in this submission. Based on the non-clinical and clinical consideration and the outcome of an expert review of image comparisons for both devices, we can claim the substantial equivalence of PCH-2500 in comparison with its predicate device, PCH-2500(K122155), in terms of safety and effectiveness.

## Safety, EMC and Performance Data:

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1(A1+A2, 1995), IEC 60601-1-1 (2000), IEC 60601-1-3 (1994), IEC 60601-2-7 (1998), IEC 60601-2-28 (Ed. 1, 1993) and IEC 60601-2-32 (Ed. 1, 1994) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2 (2007).

The manufacturing facility is in conformance with the relevant EPRC standards as specified in 21 CFR 1020.30, 31 and the records are available for review.

PCH-2500 also meets the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.

Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" were performed. For clinical consideration, separate clinical image evaluation is performed for each X-ray generator which is considered as one of critical components affecting the quality of radiographic images of the device.

Acceptance test according to IEC 61223-3-4 was performed. The separate imaging evaluation is performed for each X-ray generator which is considered as one of critical components affecting the imaging performance of the device.

DICOM Conformance Statement, image viewing SW validation reports, biocompatibility evaluation report and nonclinical consideration for detectors same with predicate device are not provide in this submission, because DICOM standard 

{7}------------------------------------------------

conformance, accessory components, image viewing program and the detectors between proposed and predicate device are same.

All test results were satisfactory.

#### Conclusion:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Vatech Co., Ltd. concludes that PCH-2500 is safe and effective and substantially equivalent to predicate device as described herein.

END

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 15, 2013

VaTech Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 12946 Kimberley Lane HOUSTON TX 77079

Re: K130585

Trade/Device Name: PCH-2500 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: June 7, 2013 Received: June 11, 2013

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{9}------------------------------------------------

Page 2 - Mr. Kim

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

![](https://i.imgur.com/99999999999999999999999999999999.png)

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

# Indications for Use

510(k) Number (if known): K130585

Device Name: PCH-2500

Indications for Use:

PCH-2500 is a digital extra oral source X-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, and X-ray technicians and other professionals licensed by the law of the states in which the device is used.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Smh.7)

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

K130585 510(k)

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K130585](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K130585)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
