K122806 · Mic International · MUH · Mar 8, 2013 · Dental
Device Facts
Record ID
K122806
Device Name
IMAGE WORKS PANOURA
Applicant
Mic International
Product Code
MUH · Dental
Decision Date
Mar 8, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
The IMAGEWORKS PANOURA dental panoramic and cephalometric device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D X-ray images as well as X-ray projection images of examined volume for the reconstruction of 3D view. The device must only be operated and used by dentists and other legally qualified professionals.
Device Story
Extraoral dental X-ray system; produces 2D panoramic/cephalometric images and 3D volumetric reconstructions. Device comprises X-ray generator, sensor unit, column, and sliding body. During operation, unit rotates around patient's head, irradiating and capturing multiple X-ray projection data points. Data transferred to image processing unit; superimposed with shift values based on rotation speed to reconstruct images. Used in dental clinics by dentists/qualified professionals. Output viewed by clinicians to assist in diagnosis of dental, jaw, and TMJ conditions. Benefits include enhanced diagnostic visualization via 3D imaging capabilities compared to 2D-only predecessors.
Clinical Evidence
Bench testing only. Performance testing, software verification/validation, and risk analysis (ISO 14971:2007) performed. Compliance with IEC 60601-1, 60601-1-1, 60601-1-2, 60601-1-3, 60601-2-7, 60601-2-28, 60601-2-32, and IEC 60825-1 confirmed.
Technological Characteristics
Digital panoramic/cephalometric X-ray system. CMOS sensor technology. X-ray tube: 0.5mm x 0.5mm focal spot, oil cooling. Tube voltage: 58-82kV; Tube current: 2.0-10mA. Connectivity: Digital image processing unit. Laser positioning (Class 2). Sterilization: N/A (non-invasive). Software: Validation per FDA guidance.
Indications for Use
Indicated for dental radiographic examinations of teeth, jaw, and TMJ areas in patients requiring 2D or 3D imaging. For use by dentists and qualified professionals only.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
## Section 6- 510(k) Summary
## MAR 0 8 2013
## a. Owner/Company name, address
THE YOSHIDA DENTAL MFG CO., LTD. 1-3-6, Kotobashi, Sumida-ku Tokyo 、130-8516, Japan
Michizo Yamanaka President
· Contact person Hidenori Watanabe International Regulatory Affairs Phone: 011-81-3-3631-2165 011-81- 3-3633-9420 Fax: Email: hi-watanabe@yoshida-net.co.jp
## b. Contact/Application Correspondent
Izumi Maruo MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan
011-81-3-3818-8577 Phone: Fax: 011-81-3-3818-8573 Email: maruo@mici.co.jp
#### ﻦ Date prepared
September 5, 2012
## d. Name of device
| Trade Name: | IMAGE WORKS PANOURA |
|----------------------------|------------------------------------------|
| Common Name: | Extraoral source x-ray system |
| Classification Name: | System, x-ray, extraoral source, digital |
| Classification Regulation: | 21 CFR 872.1800 |
THE YOSHIDA DENTAL MFG. CO., LTD. SEPTEMBER 5, 2012
SECTION 6 6- 2 OF 6-9
{1}------------------------------------------------
## ·· e. Predicate devices
The IMAGEWORKS PANOURA is substantially equivalent to the following legally marketed device:
| 510(k): | K111231 |
|---------------|-------------|
| Trade name: | PANOURA 18S |
| Product code: | MUH |
510(k): K093683 Trade name: ORTHOPANTOMOGRAPH OP300 Product code: MUH
The predicate devices are hereinafter called "the PANOURA 18S (k111231)" or "the ORTHOPANTOMOGRAPH (k093683)" , respectively, in this application.
## f. Description of the device
The IMAGEWORKS PANOURA is dental panoramic and cephalometric device and intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D X-ray images as well as X-ray projection images of examined volume for the reconstruction of 3D view. The IMAGEWORKS PANOURA is equipped with an X-ray generator and a Sensor unit at Arm unit supported by Column unit and Sliding body unit. While rotating around the patient's teeth and jaw, the IMAGEWORKS PANOURA irradiates X-ray and detects X-ray absorbed data at the Sensor unit multiple times. Detected multiple data are transferred to an image processing unit and the data are superimposed with appropriate shift value according to the X-ray moving speed from the arm rotation to acquire image.
## g. Statement of Intended Use
The IMAGEWORKS PANOURA dental panoramic and cephalometric device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D X-ray images as well as X-ray projection images of examined volume for the reconstruction of 3D view. The device must only be operated and used by dentists and other legally qualified professionals.
## h. Statement of substantial equivalence
The IMAGEWORKS PANOURA is modified from the PANOURA 18S (k111231) by adding 3D imaging. The characteristics except 3D imaging of the IMAGEWORKS PANOURA are identical to those of the PANOURA 18S (k111231).
Regarding intended use of the IMAGEWORKS PANOURA, 3D imaging is added to intended use of the PANOURA 18S (k111231). However, patient population and the fundamental technologies of the IMAGEWORKS PANOURA are identical to those of the PANOURA 18S (k111231).
3D imaging of the IMAGEWORKS PANOURA is similar to that of the ORTHOPANTOMOGRAPH (k093683). 3D image is produced by reconstruction of X-ray projection images. Intended Use of the IMAGEWORKS PANOURA is similar to that of the ORTHOPANTOMOGRAPH
THE YOSHIDA DENTAL MFG. CO., LTD. SEPTEMBER 5, 2012
SECTION 6 6- 3 OF 6-9
{2}------------------------------------------------
(k093683) because the characteristics of the IMAGEWORKS PANOURA are similar to the y ORTHOPANTOMOGRAPH (k093683). The similarities of the IMAGEWORKS PANOURA to the ORTHOPANTOMOGRAPH (k093683) are;
- Intended use ●
ب ته دارات م
- Operational characteristics .
- . lonizing radiation
- Cephalometric radiogram .
- Panoramic images .
- . 3D Imaging
In order to evaluate safety and effectiveness of the IMAGEWORKS PANOURA, software verification/validation, performance testing, and risk analysis were performed. In conclusion, those testing and analysis demonstrated that the IMAGEWORKS PANOURA did not raise any new safety or effectiveness concerns compared to the PANOURA 18S (k111231) and the ORTHOPANTOMOGRAPH (k093683).
#### i. Comparison table
Table 6-1. compares the characteristics between the IMAGEWORKS PANOURA, the PANOURA 18S (k111231), and the ORTHOPANTOMOGRAPH (k093683).
SECTION 6 6- 4 OF 6-9
{3}------------------------------------------------
| | IMAGEWORKS PANOURA<br>PREMARKET NOTIFICATION 510(k)<br>Table 6-1. Comparison Table | IMAGEWORKS PANOURA | PANOURA 18S (k111231) | ORTHOPANTOMOGRAPH (k093683) | |
|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Device Characteristics | | | | | |
| Intended Use | | The IMAGEWORKS PANOURA dental panoramic and cephalometric device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D X-ray images as well as X-ray projection images of examined volume for the reconstruction of 3D view. The device must only be operated and used by dentists and other legally qualified professionals. | The Panoura 18S dental panoramic and cephalometric device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D X-ray images as well as X-ray projection images. The device must only be operated and used by dentists and other legally qualified professionals. | The Orthopantomograph OP300 dental panoramic, cephalometric and cone beam computed tomography x-ray device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as X-ray projection images of an examined volume for the reconstruction of a 3D view. The device must only be operated and used by dentists and other legally qualified professionals. | |
| Equipment type | | Digital panoramic x-ray equipment | Digital panoramic x-ray equipment | Digital panoramic x-ray equipment | |
| Mode of operation | | Continuous operation with intermittent load | Continuous operation with intermittent load | Continuous operation with intermittent load | |
| X-ray tube focal point | | 0.5 mm×0.5 mm | 0.5 mm×0.5 mm | 0.5 mm×0.5 mm | |
| X-ray tube cooling method | | Oil cooling | Oil cooling | Oil cooling | |
| Nominal maximum electric power<br>(combination of X-ray tube voltage<br>and tube current at maximum output) | | 0.82kW (82kV, 10mA) | 0.82kW (82kV, 10mA) | 1.44kW (90kV, 16mA) | |
| Tube voltage | | 58 - 82kV | 58 - 82kV | 57 - 90kV | |
| Tube current | | 2.0 - 10mA | 2.0 - 10mA | 4 - 16mA | |
| Radiation time | | Adult | | | |
| | | Panoramic | 8, 14, 16s | 8, 14, 16s | 16.4s |
| | | Child | 6.4, 11.2, 12.8s | 6.4, 11.2, 12.8s | 14.4s |
| | | TMJ | 8s | 8s | 10.6s |
| | | Cephalo / Carpus image acquisition | 8 - 10 s | 8 - 10 s | 10-20s |
THE YOSHIDA DENTAL MFG. CO., LTD
SEPTEMBER 5, 2012
# SECTION 6
6- 5 OF 6-9
{4}------------------------------------------------
| IMAGEWORKS PANOURA<br>PREMARKET NOTIFICATION 510(k)<br>Device Characteristics | | 3D oral | 3D | 3D dent | | | IMAGEWORKS PANOURA | PANOURA 18S (k111231) | | | | ORTHOPANTOMOGRAPH (k093683) | | | | | | | |
|-------------------------------------------------------------------------------|---------|-----------|-----------|---------|--|----------------------------------|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|-------------------|--|--|--|
| Radiation time | | 11.5s x 2 | | 11.5s | | Electric power supply resistance | | Maximum 0.2 Ω | | | | Small FOV standard resolution: 2.3 s<br>Large FOV standard resolution: 4.9s<br>Small FOV high resolution: 6.1 s<br>Large FOV high resolution: 12.6 s | | | | | | | |
| Total filtration | | | | | | | | Maximum 0.2 Ω<br>2.5mmAl equivalent or over | | | | Maximum 0.2 Ω<br>3.2mm Al equivalent or over | | | | | | | |
| Leakage dose | | | | | | | | 1.0 mGy/h or less | | | | 1.0 mGy/h or less | | | | | | | |
| Leakage dose calculation standards | | | | | | | | Tube voltage 82kV, tube current 10mA | | | | Tube voltage 90kV, tube current 4mA | | | | | | | |
| Image magnification | | | Panoramic | | | | | 1.2 to 1.29 | | | | 1.3 | | | | | | | |
| | | | TMJ | | | | | 1.2 to 1.29 | | | | 1.23 | | | | | | | |
| | | | Cephalo | | | | | 1.1 | | | | 1.14 | | | | | | | |
| | | | 3D | | | | | 1.0 | | | | 1.0 | | | | | | | |
| Number of phases | | | | | | | | Single phase | | | | Single phase | | | | | | | |
| Frequency | | | | | | | | 50/60 Hz | | | | 50/60Hz | | | | | | | |
| Rated power | Voltage | | | | | | | AC100V-120V / AC220V - 240V | | | | AC100 - 240V (Selectable) | | | | | | | |
| Input | 110VAC | 2.0kVA | | | | | | AC100V-120V / AC220V - 240V<br>2.0kVA | | | | 1.65kVA | | | | | | | |
| | 230VAC | 2.0kVA | | | | | | 2.0kVA | | | | 2.3kVA | | | | | | | |
| Classification | | | | | | | | Class I, Type B | | | | Class I, Type B | | | | | | | |
| Up-and-down stroke | | | | | | | | 800 mm (short type: 400 mm) | | | | 800 mm | | | | | | | |
| Weight | | | | | | | | 140kg Standing position, wall mount<br>(with 3D detector)<br>135kg Standing position, wall mount<br>(short type) (with 3D detector)<br>165 kg Standing position, base<br>mount with an optional base (in the<br>case of wide base: +5kg) (with 3D<br>detector)<br>160kg Standing position, base mount<br>(short type) with an optional base (in<br>the case of wide base: +5kg) (with | | | | 125kg Standing position wall mount<br>120kg Standing position wall mount<br>(short type)<br>150 kg (with an optional base)<br>145kg Standing position base mount<br>short type with an optional base/ 150kg<br>with wide base | | | | 200kg (Panoramic) | | | |
THE YOSHIDA DENTAL MFG. CO., LTD.
SEPTEMBER 5, 2012
SECTION 6
6- 6 OF 6-9
{5}------------------------------------------------
| | IMAGEWORKS PANOURA<br>PREMARKET NOTIFICATION 510(k) | | PANOURA 18S (k111231) | ORTHOPANTOMOGRAPH (k093683) |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Device Characteristics | | | | |
| | IMAGEWORKS PANOURA | | | |
| Weight | 180kg Standing position wall mount<br>(with Cephalo) | | 190kg Standing position wall mount<br>(with Cephalo) | 240kg (Cephalo) |
| | 175kg Standing position, base mount<br>(short type) (with Cephalo) (with 3D<br>detector) | | 185kg Standing position wall mount<br>short type (with Cephalo) | |
| | 205kg Standing position, base mount<br>with an optional base (with Cephalo)<br>(in the case of wide base: +5kg)<br>(with 3D detector) | | 215kg Standing position base mount<br>with Cephalo and optional base | |
| | 200kg Standing position, base mount<br>(short type) (with Cephalo) (in the<br>case of wide base: +5kg) (with 3D<br>detector) | | 210kg Standing position base mount<br>short type with Cephalo and optional<br>base/ 215kg with wide base | |
| Size | 2209 x 846 x 1192mm(Standing<br>position wall mount) | | 2209 x 849 x 1192mm(Standing<br>position wall mount) | 2410 x 830 x 1126mm (standard<br>column) |
| | 2209 x 1835 x 1192mm(Standing<br>position wall-mount with Cephalo) | | 2209 x 1759 x 1192mm(Standing<br>position wall-mount with Cephalo) | 2410 x 1931 x 1193 mm (with<br>Cephalo) |
| ID/SOD (Panoramic) | 485mm/350mm | | 485mm/350mm | 500mm(SID) |
| ID/SOD (Cephalo) | 1650mm/1500mm | | 1650mm/1500mm | 1745mm/1520mm |
| ID/SOD (3D) | 570mm/350mm | | | unknown |
| Operating<br>environment | Temperature | 10 to 40°C | 10 to 40°C | 10 to 35°C |
| | Relative humidity | 30 to 75% (no condensation) | 30 to 75% (no condensation) | max 85% |
| EMC Classification | | ClassA | ClassA | Class B |
| Target angle | | 15 degrees | 15 degrees | 5 degrees |
| Positioning laser lights | | laser light (CLASS 2 LASER<br>PRODUCT) | laser light (CLASS 2 LASER<br>PRODUCT) | laser light (CLASS 1 LASER<br>PRODUCT) |
| | | IEC<br>60825-1:1993+A1:1997+A2:2001 | IEC 60825-1:1993+A1:1997+A2:2001 | IEC<br>60825-1:1993+A1:1997+A2:2001 |
| Cephalometric<br>radiogram | Scanning method | Horizontal scan, synchronized<br>sensor and secondary slot motion | Horizontal scan, synchronized<br>sensor and secondary slot motion | Horizontal scan, synchronized<br>sensor and secondary slot motion |
| | Scanning time | 8-10s | 8-10s | 10-20s |
| Panoramic<br>image receptor | Sensor unit | Pan sensor or interchangeable<br>Cephalo sensor | Pan sensor or interchangeable Cephalo<br>sensor | Pan sensor or interchangeable<br>Cephalo sensor |
| | Technology | CMOS | CMOS | CMOS |
THE YOSHIDA DENTAL MFG. CO., LTD
SEPTEMBER 5, 2012
SECTION 6
6- 7 OF 6-9
{6}------------------------------------------------
| | IMAGEWORKS PANOURA<br>PREMARKET NOTIFICATION 510(k) | IMAGEWORKS PANOURA | PANOURA-18S (k111231) | ORTHOPANTOMOGRAPH (k093683) |
|---------------------------------|-----------------------------------------------------|--------------------------------|--------------------------------|----------------------------------|
| Device Characteristics | | | | |
| Panoramic<br>Image receptor | Image pixel size | 100 x 100 μm | 100 x 100 μm | 100 x 100 μm |
| | Image field height | 151mm / 1510pixels | 151mm / 1510pixels | 151mm / 1480pixels |
| | Sensor unit | interchangeable Cephalo sensor | interchangeable Cephalo sensor | interchangeable Cephalo sensor |
| | Technology | CMOS | CMOS | CMOS |
| Cephalometric<br>Image receptor | Image pixel size | 100 x 100 μm | 100 x 100 μm | 100 x 100 μm |
| | Image field height | 226.6mm / 2266pixels | 226.6mm / 2266pixels | 226mm / 2232pixels |
| | Image field width in<br>LA view | 254mm | 254mm | 260mm, maximum<br>170mm, minimum |
| | Image field width<br>in PA view | 203.2mm | 203.2mm | 200mm |
| | Sensor unit | - | - | 3D sensor |
| | Technology | - | - | CMOS |
| 3D image receptor | Image pixel size | 100 x 100 μm | - | 200μm |
| | Image field height | 99.2mm / 992pixels | - | 61 mm |
THE YOSHIDA DENTAL MFG. CO., LT
SEPTEMBER 5, 2012
# SECTION 6
6- 8 OF 6-9
{7}------------------------------------------------
#### Risk Analysis ... j.
The IMAGEWORKS PANOURA was evaluated in accordance with ISO14971:2007. The risk management of the device was deemed satisfactory.
## k. Bench Testing
THE YOSHIDA DENTAL MFG. CO., LTD has performed bench tests regarding laser safety to ensure safety and effectiveness, verify conformity with IEC 60825-1.
The laser system of the IMAGEWORKS PANOURA is identical to that of the PANOURA 18S (K111231). Therefore, the test report for IEC 60825-1 for the PANOURA 18S (K111231) is used as the test report for the IMAGEWORKS PANOURA.
The software of the IMAGEWORKS PANOURA has been validated according to "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
EMC, Electric safety, and X-ray radiation safety are confirmed in accordance with IEC60601-1. IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, and IEC 60601-2-32.
#### ﺎﺕ Conclusion
The IMAGEWORKS PANOURA is modified from the PANOURA 18S (k111231) by adding 3D imaging. The characteristics except 3D imaging of the IMAGEWORKS PANOURA are identical to those of the PANOURA 18S (k111231). The IMAGEWORKS PANOURA has similar intended use and operational and technological characteristics to the ORTHOPANTOMOGRAPH (k093683). The performance test results indicate that the IMAGEWORKS PANOURA meets the requirements of recognized consensus or voluntary standard. Based on the information presented above regarding substantial equivalence to the PANOURA 18S (k111231) and the ORTHOPANTOMOGRAPH (k093683), THE YOSHIDA DENTAL MFG CO., LTD. concludes that the IMAGEWORKS PANOURA is substantially equivalent to the PANOURA 18S (k111231) and the ORTHOPANTOMOGRAPH (k093683), and does not raise any new questions regarding safety or effectiveness.
THE YOSHIDA DENTAL MFG. CO., LTD. SEPTEMBER 5, 2012
SECTION 6 6- 9 OF 6-9
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows a circular logo with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring an eagle-like figure with outstretched wings, rendered in a bold, black design. The emblem is a simplified representation, focusing on the wing structure and overall form rather than detailed features.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2013
Ms. Izumi Maruo MIC International 4-1-17 Hongo, Bunkyo-ku TOKYO 113-0033 JAPAN
Re: K122806
Trade/Device Name: ImageWorks Panoura Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: February 25, 2013 Received: March 1, 2013
Dear Ms. Maruo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device lo contract to contract mability
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (7) & FR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.
{9}------------------------------------------------
### Page 2-Ms. Maruo
i If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
MichFDA Hara.
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use
510(k) Number (if known): K122806
Device Name: IMAGEWORKS PANOURA
Indications for Use:
The IMAGEWORKS PANOURA dental panoramic and cephalometric device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D X-ray images as well as X-ray projection images of examined volume for the reconstruction of 3D view. The device must only be operated and used by dentists and other legally qualified professionals.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
MichaELON'Hara
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) K122806
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.