← Product Code [MUH](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH) · K112380

# CLEARVISION (K112380)

_Midmark Corp. · MUH · Nov 22, 2011 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K112380

## Device Facts

- **Applicant:** Midmark Corp.
- **Product Code:** [MUH](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH.md)
- **Decision Date:** Nov 22, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.1800
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

ClearVision is intended to be used by dentists and other qualified professionals for producing diagnostic x-ray radiographs of dentition, jaws and other oral structures.

## Device Story

ClearVision is a digital dental radiographic imaging system. It utilizes CMOS sensors (size 1 or 2) to capture intraoral x-ray images. The sensor connects directly to a PC via USB without intermediate electrical interfaces. Operation is automatic: the sensor detects x-ray exposure, captures the image, and transfers it to PC-based imaging software. Used in dental clinics by dentists or qualified professionals. Disposable sheaths prevent cross-contamination. The system provides diagnostic images of dentition and jaws, aiding clinical decision-making for routine dental examinations. Benefits include digital image acquisition and storage.

## Clinical Evidence

Bench testing only. Imaging performance compared against Schick CDR and Gendex GXS-700 using line pair phantom, aluminum step wedge, and tooth phantom. ClearVision demonstrated equivalence to Gendex GXS-700 and superiority to Schick CDR. Additional testing included electrical safety (IEC 60601-1), EMI/EMC (IEC 60601-1-2), and hermetic classification (IP67 per IEC 60529).

## Technological Characteristics

CMOS sensor technology with CsI scintillator. Pixel size 19.0 µm. Sensor sizes: 37x24 mm and 43x30 mm. USB interface to PC. No intermediate electrical interface assembly. IP67 hermetic rating. Compliant with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMI/EMC).

## Regulatory Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

## Predicate Devices

- Schick CDR ([K072134](/device/K072134.md))
- Gendex GXS-700 ([K090458](/device/K090458.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 1123 80 Page 1 of 3

. . . . . . .

## 510(k) Summary

| Date:                  | June 9, 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:             | Midmark Corporation<br>675 Heathrow Drive<br>Lincolnshire, IL 60069 USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Contact Person:        | Lisa Bartakovics<br>Director of Quality and Regulatory Affairs<br>(847) 415-9800                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Name:           | ClearVision Digital Sensor System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate Devices:     | Schick CDR (K072134)<br>Gendex GXS-700 (K090458), approved under Dexis name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Description of Device: | ClearVision is a digital imaging system for dental radiographic<br>application.  The product is to be used for routine dental radiographic<br>examinations such as bitewings, periapicals, etc.  Two different sized<br>sensors (size 1 and size 2) are utilized to image different anatomy and<br>for different patient sizes.  The CMOS sensor connects directly to a USB<br>connection in a PC without the need for an intermediate electrical<br>interface.  ClearVision works with a standard dental intraoral x-ray<br>source without any connection to the x-ray source.  ClearVision captures<br>an image automatically upon sensing the production of x-ray and after<br>the x-ray is complete, transfers the image to an imaging software<br>program on the PC. Disposable sheaths are used with each use to<br>prevent cross-contamination between patients. |
| Intended Use:          | ClearVision is intended to be used by dentists and other qualified<br>professionals for producing diagnostic x-ray radiographs of dentition,<br>jaws and other oral structures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{1}------------------------------------------------

K112380
Page 2 of 3

## Substantial Equivalence:

| Technical<br>Characteristic          | ClearVision                                                                                                                                                                                       | Schick CDR                                                                                                                                                                                                                                                     | Gendex GXS-700 (Dexis)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sensor Sizes (mm)                    | 37 x 24<br>43 x 30                                                                                                                                                                                | 31 x 22<br>37 x 24<br>43 x 30                                                                                                                                                                                                                                  | 37 x 25<br>42 x 31                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Sensor Technology                    | CMOS                                                                                                                                                                                              | CMOS                                                                                                                                                                                                                                                           | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Pixel Size (µm)                      | 19.0                                                                                                                                                                                              | 40.0                                                                                                                                                                                                                                                           | 19.5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Scintillator                         | CsI                                                                                                                                                                                               | CsI                                                                                                                                                                                                                                                            | CsI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Interface to PC                      | USB                                                                                                                                                                                               | USB                                                                                                                                                                                                                                                            | USB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Electronic<br>Interface<br>Assembly? | No                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                            | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Intended Use                         | ClearVision is<br>intended to be used<br>by dentists and other<br>qualified professionals<br>for producing<br>diagnostic x-ray<br>radiographs of<br>dentition, jaws and<br>other oral structures. | The Computed Oral<br>Radiology System is<br>intended for intra-oral<br>examinations and indicated<br>for dental patients. It<br>produces instant, digital,<br>intra-oral x-ray images of a<br>patient's mouth while<br>reducing the necessary x-ray<br>dosage. | The DEXIS sensor is a USB-<br>driven digital sensor which<br>is intended to acquire<br>dental intra-oral<br>radiography images. The<br>DEXIS sensor shall be<br>operated by healthcare<br>professionals, who are<br>educated and competent<br>to perform the acquisition<br>of dental intra-oral<br>radiographs. The DEXIS<br>sensor can be used either<br>in combination with<br>special positioning devices<br>to facilitate positioning<br>and alignment with the x-<br>ray beam or it may also be<br>positioned by hand with<br>the assistance of the. |

Tests Conducted:

An imaging performance comparison was done of the ClearVision sensor versus the Schick CDR and Gendex GXS-700 sensors. The details of this testing can be found in Section 18. This testing consisted of using each sensor to image a line pair phantom, an aluminum step wedge, and a tooth phantom. In this testing, the ClearVision sensor was shown to be equivalent to the Gendex GXS-700 in all three tests and superior to the Schick sensor in all three imaging tests.

{2}------------------------------------------------

K112380
Page 3 of 3

Significant electrical, mechanical, and imaging tests were performed on the sensors as part of engineering verification. The full details of this testing is found in Section 18. In all cases, the sensor was found to meet the requirements identified earlier in the design phase. Extensive imaging performance testing was done in various configurations of connections of the sensor to a PC workstation (varying lengths of cable and the presence of USB hubs and extenders). The sensor showed itself to be completely reliable in capturing an image and transferring the image to the workstation over an extended life period. Image quality of the sensor was shown to meet requirements and to be consistent over the expected lifetime exposures to radiation. The sensor was put through extensive mechanical testing that tested the durability of the sensor housing and cable. Once again, the sensor met all specified requirements. An external test laboratory confirmed the ClearVision system meets the electrical safety and EMI/EMC provisions of IEC 60601-1 and 60601-1-2. Testing by the sensor supplier by an external laboratory confirmed the sensor met hermetic classification IP67 per IEC 60529.

Conclusion:

The ClearVision sensor system is substantially equivalent to other legally marketed devices in the United States. The ClearVision sensor system is substantially equivalent in technical characteristics and intended use to the CDR sensor marketed by Schick Technologies and the GXS-700 sensor marketed by Gendex Dental Systems (approved under the Dexis name).

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three curved lines that resemble a bird in flight or a symbolic representation of human services.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

NOV 2 2 2011

Ms. Lisa Bartakovics Director, Quality and Regulatory Affairs Progency Dental Midmark Corporation 675 Heathrow Drive LINCOLNSHIRE IL 60069

Re: K112380

Trade/Device Name: Progeny ClearVision Digital Sensor System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: October 18, 2011 Received: October 18, 2011

Dear Ms. Bartakovics:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use Form

510(k) Number (if known): _ K \\2380__________________________________________________________________________________________________________________________________________ 

Device Name: Progeny ClearVision Digital Sensor System

Indications for Use: ClearVision is intended to be used by dentists and other qualified professionals for producing diagnostic x-ray radiographs of dentition, jaws and other oral structures.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over the Counter Use__________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Signature

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K112380

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K112380](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K112380)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
