← Product Code [MUH](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH) · K061284

# I-CAT SCANNER (K061284)

_Imaging Sciences Intl., Inc. · MUH · Jul 3, 2006 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K061284

## Device Facts

- **Applicant:** Imaging Sciences Intl., Inc.
- **Product Code:** [MUH](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH.md)
- **Decision Date:** Jul 3, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.1800
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The Imaging Sciences International Inc. i-CAT Scanner constructs a three dimensional model from images taken during a rotational X-ray sequence. The Imaging Sciences i-CAT Scanner is intended to be used whenever a dentist, oral surgeon, or other physician needs 3D information of high contrast objects. The system is designed for imaging of TM Joint studies, mandible & maxilla for implant planning, sinuses, and other areas of the maxillofacial complex.

## Device Story

i-CAT Scanner is an extraoral X-ray system; acquires 360-degree rotational X-ray sequence; reconstructs 3D matrix of examined volume; produces 2D views. Operated by dentists, oral surgeons, or physicians in clinical settings. Software transforms 2D detector images into 3D models; displays on workstation monitor. Allows distance/thickness measurements on 2D images. Output used for diagnostic planning (e.g., implants, TMJ, sinuses). Benefits include high-contrast 3D visualization of maxillofacial structures for improved clinical decision-making.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by design, verification, and validation processes including risk management.

## Technological Characteristics

Extraoral X-ray system; motorized gantry with X-ray source and image detector. Performs 360-degree rotational acquisition. Reconstructs 3D volumes from 2D image sequences. Software-based processing for 3D visualization and measurement. Standalone workstation configuration.

## Regulatory Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

## Predicate Devices

- NewTom QR - DVT 9000 ([K003787](/device/K003787.md))
- 3D Accu-I-tomo XYZ Slice View Tomograph ([K030450](/device/K030450.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K061284

# JUL - 3 2006

# 510 (k) Summary

### Submitters Information

Name:

Address:

Phone Number:

Fax Number:

Person To Contact:

Date Of Summary:

Trade Name Of The Device:

Common Or Usual Name:

Classification Name:

Imaging Sciences International Inc.

1910 North Penn Road Hatfield PA, 19440

215-997-5666

215-997-5665

David W. Cowan Vice President of Engineering, Quality Assurance and Government Compliance

May 05, 2006

i-CAT Scanner

Extraoral Source X-ray System

Extraoral Source X-ray System

{1}------------------------------------------------

Substantial Equivalence Claim: The Imaging Sciences International Inc. i-CAT Scanner is substantially equivalent to the devices listed below:

| Device:         | NewTom QR - DVT 9000                                                                   |
|-----------------|----------------------------------------------------------------------------------------|
| Manufacturer:   | NIM s.r.1.<br>Via Silverstrini, 20<br>37135 Verona<br>Italy                            |
| 510 (k) Number: | K003787                                                                                |
| Device:         | 3D Accu-I-tomo XYZ Slice View Tomograph                                                |
| Manufacturer:   | J. Morita Manufacturing Corporation<br>680 Higashihama Minami-cho, Fushimi-ku<br>Japan |
| 510 (k) Number: | K030450                                                                                |

Description Of The Device: The i-CAT Scanner is a dedicated X-Ray imaging device that acquires a 360 degree rotational X-ray sequence, reconstructs a three-dimensional matrix of the examined volume and produces two dimensional views of this volume. The i-CAT Scanner can measure distances and thickness on two dimensional images. Images produced by the i-CAT Scanner can be printed or exported on magnetic and optical media.

The i-CAT Scanner gantry is comprised of an X-ray source, image detector, and motorized gantry. The gantry facilitates the acquisition of a full X-ray sequence by the software. The software receives the two dimensional images acquired by the detector, transforms them into three dimensional images and displays them on the workstation monitor for viewing.

Intended Use Of The Device: The Imaging Sciences International Inc. i-CAT Scanner constructs a three dimensional model from images taken during a rotational X-ray sequence. The Imaging Sciences i-CAT Scanner is intended to be used whenever a dentist, oral surgeon, or other physician needs 3D information of high contrast objects. The system is designed for imaging of TM Joint studies, mandible & maxilla for implant planning, sinuses, and other areas of the maxillofacial complex.

Comparison with Predicate Devices: The i-CAT Scanner reconstructs a three dimensional model from X-ray images similar to the model obtained using the predicate devices. It displays either two-dimensional cross-sections or three dimensional views and allows the user to take measurements on reconstructed images.

Conclusions: The i-CAT Scanner Acquires an X-ray rotational sequence and provides three-dimensional information on the analyzed volume. The potential hazards (e.g. electrical, mechanical, thermal, radiation, incorrect measurements or misdiagnosis) are controlled by the design development, verification and validation process which includes a risk management system.

{2}------------------------------------------------

The i-CAT Scanner complies with the requirements of 21 CFR 807.87 and does not
pose any new safety risks or effectiveness issues.

:

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

JUL - 3 2006

Mr. David W. Cowan Vice President of Engineering, Quality Assurance and Government Compliance Imaging Sciences International, Inc. 1910 North Penn Road HATFIELD PA 19440

Re: K061284

Trade/Device Name: i-CAT Scanner Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: May 5, 2006 Received: May 9, 2006

#### Dear Mr. Cowan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate croomerse prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Nossmatic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations offecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/3/Picture/9 description: The image is a circular logo or emblem. The text "1906-2006" is at the top of the logo. The letters "FDA" are in the center of the logo. The word "Centennial" is below the letters "FDA". There are three stars below the word "Centennial".

oting Public S

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the sectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legal|y marketed predicate device results in a classification for your device and thus, permitts your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistence at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): _ K061284

Device Name: i-CAT Scanner

Indications For Use: The Imaging Sciences International Inc. i-CAT Scanner constructs a three dimensional model from images taken during a rotational X-ray sequence. The Imaging Sciences i-CAT Scanner is intended to be used whenever a dentist, oral surgeon, or other physician needs 3D information of high contrast objects. The system is designed for imaging of TM Joint studies, mandible & maxilla for implant planning, sinuses, and other areas of the maxillofacial complex.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number
K061284

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K061284](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUH/K061284)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
