MODIFICATION TO COMPUTED ORAL RADIOLOGY SYSTEM

K022953 · Schick Technologies, Inc. · MUH · Oct 2, 2002 · Dental

Device Facts

Record IDK022953
Device NameMODIFICATION TO COMPUTED ORAL RADIOLOGY SYSTEM
ApplicantSchick Technologies, Inc.
Product CodeMUH · Dental
Decision DateOct 2, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.1800
Device ClassClass 2

Intended Use

The Computed Oral Radiology System is indicated for patients undergoing an intra-oral dental x-ray examination. It produces instant, digital, intra-oral x-ray images of a patient's mouth while reducing the necessary x-ray dosage.

Device Story

Digital dental x-ray imaging system; modification of existing wired system to wireless interface. Sensor captures intra-oral x-ray signals; converts to digital images. Wireless transmission via radio frequency (RF) transmitter; sensor powered by battery. Hardware-based image acquisition replaces software polling to minimize idle power consumption. Sensor encapsulation material changed from aluminum to thermoplastic resin. Firmware updated on both sensor and remote module to support unidirectional communication of image and status data. Used in dental clinics; operated by dental professionals. Provides instant digital images for diagnostic review; reduces patient x-ray dosage.

Clinical Evidence

Bench testing only. Risk analysis and validation studies performed in conformance with ISO9001 quality program to evaluate the effect of wireless modification and material changes.

Technological Characteristics

Solid state x-ray imager; wireless RF transmission; battery-powered sensor. Sensor housing: thermoplastic resin. Firmware-controlled image acquisition and data transmission. Operates as digital imaging receptor for dental x-ray systems.

Indications for Use

Indicated for patients undergoing intra-oral dental x-ray examinations.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # OCT 02 2002 # Exhibit IV: 510(k) Summary K022953 #### Schick Computed Oral Radiology System #### Common/Classification Name: Solid State X-ray Imager 21 CFR892.1650 Schick Technologies, Inc. 30-00 47th Avenue Long Island Citv. NY 11101 718-937-5765, 718-937-5962 (FAX) Contact: Daniel Michaeli, Prepared: September 4, 2002 #### A. Legally Marketed Predicate Devices The original Computed Oral Radiology System notification was cleared on January 31st, 1994 (K933455). The device and its predicate are small digital imaging receptors that may be used in place of dental x-ray film. #### B. Modification Description The proposed device modification alters the method of communication between the sensor and computer from a wired to wireless interface. Thus, there is an introduction of a new energy type, a radio frequency transmitter. In this mode, the wireless sensor is powered via a battery rather than a DC-supply derived from AC mains voltage. The operating principle differs in that image acquisition is mediated through hardware rather than software components. Whereas software in the predicate device "polled" the sensor to investigate whether acquisition had commenced, the modified device utilizes a hardware component. This modification is necessary to reduce the sensor's idle state power to a minimum. To facilitate RF transmission, the sensor encapsulation material has been modified from aluminum to a thermoplastic resin. Thermoplastic resins are utilized in other FDA-approved dental sensors. The specific thermoplastic resin chosen for this application has been approved for incidental food contact. The duration and nature of body contact with materials in that discipline is greater than what a patient would endure with a sheath encapsulated sensor. The existing firmware has been altered to support the modified and additional hardware. Whereas the predicate device had firmware installed on the Remote module, the new device has firmware installed both on the remote module and {1}------------------------------------------------ on the sensor. The firmware on the sensor facilitates unidirectional communication of image and sensor status data to the remote module. ## C. Indications for Use The Computed Oral Radiology System is indicated for patients undergoing an intra-oral dental x-ray examination. It produces instant, digital, intra-oral x-ray images of a patient's mouth while reducing the necessary x-ray dosaqe. This device modification in no way alters the indications for use of this machine bevond what was originally approved in K933455. ## D. Substantial Equivalence Summary The intended use of the Computed Oral Radiology System remains unchanged. The new product transmits data through a wireless rather than wired interface to the host image processing system. This and the concomitant modifications in no way alter the fundamental scientific technology governing image acquisition. Those imaging parameters that could be potentially affected by the modification as are outlined generally in the document, "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices" were found to be substantially equivalent to the predicate. ## E. Technological Characteristics Part B of this section describes the technological characteristics of the modified device. The technological characteristics of this machine have been examined through a risk analysis. The fundamental technology remains unchanged from the predicate. ## F. Risk Analysis and Validation study As a result of risk analysis activities performed in conformance with our ISO9001 quality program along with supporting bench-lab measurements, Schick Technologies has determined the effect of the proposed changes to be negligible. ## G. Conclusions Schick Technologies has demonstrated through a risk analysis and validation studies that the device modification is substantially equivalent to the already cleared and marketed device. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle. #### Public Health Service # OCT 02 2002 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Daniel Michaeli, M.S. Product Manager Schick Technologies, Inc. 30-00 47" Avenue LONG ISLAND CITY NY 11101 Re: K022953 Trade/Device Name: Computed Oral Radiology System Regulation Number: 21 CFR §892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Product Code: 90 MQB Regulation Number: 21 CFR §872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: 90 MUH Dated: September 3, 2002 Received: September 5, 2002 Dear Mr. Michaeli: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Exhibit III: Indications for Use Page_1__of_1 510(k) Number (if known):____________________________________________________________________________________________________________________________________________________ Computed Oral Radiology System Device Name: Indications For Use: The Computed Oral Radiology System is indicated for patients undergoing an intra-oral dental x-ray examination. It produces instant, digital, intra-oral x-ray images of a patient's mouth while reducing the necessary x-ray dosage. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) David A. Siggons (Division Sign-Off) Division of Reproduc and Radiological Dev 510(k) Number ﻪ Prescription Use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...