← Product Code [MUE](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE) · K211215

# SenoBright HD (K211215)

_GE Healthcare · MUE · Jun 22, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE/K211215

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [MUE](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE.md)
- **Decision Date:** Jun 22, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1715
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

SenoBright HD is an extension of the existing indication for diagnostic mammography with Senographe Pristina. The SenoBright HD application shall enable contrast enhanced breast imaging using a dual energy technique. This imaging technique can be used as an adjunct following mammography and ultrasound exams to help localize a known or suspected lesion.

## Device Story

SenoBright HD is a software application for the Senographe Pristina FFDM system. It performs Contrast Enhanced Spectral Mammography (CESM) by acquiring two sequential images (high energy and low energy) under breast compression following iodinated contrast media injection. The device uses the New Image Recombination Algorithm (NIRA) to process these inputs. NIRA performs local breast thickness estimation to account for non-uniformity and compensates for patient motion between acquisitions. The output is a recombined image used by radiologists to localize lesions. The device is used in clinical settings; operated by trained imaging staff. The improved image quality aids clinical decision-making by increasing lesion visibility and reducing artifacts.

## Clinical Evidence

Clinical evidence includes an evaluation by 3 independent MQSA-qualified radiologists on 50 clinical images comparing SenoBright HD with NIRA to the predicate. Results: contrast uptake visibility was equivalent or better in 97% of cases; artifact visibility was equivalent or lower in 99% of cases; overall clinical image quality was superior in >98% of cases. Additional non-clinical phantom testing confirmed improved image quality and artifact reduction.

## Technological Characteristics

Software-only update for Senographe Pristina FFDM system. Implements New Image Recombination Algorithm (NIRA) for dual-energy CESM. Features local breast thickness estimation and motion compensation. Operates independently of other digital mammography acquisition components. Compliant with 21 CFR 820 and ISO 13485.

## Regulatory Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

## Predicate Devices

- SenoBright HD ([K172404](/device/K172404.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 22, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

GE Healthcare % Florian Akpakpa Regulatory Affairs Program Manager 283 Rue de la Minière Buc, 78530 FRANCE

Re: K211215

Trade/Device Name: SenoBright HD Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: Class II Product Code: MUE Dated: April 22, 2021 Received: April 23, 2021

Dear Florian Akpakpa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K211215

Device Name

SenoBright HD

#### Indications for Use (Describe)

SenoBright HD is an extension of the existing indication for diagnostic mammography with Senographe Pristina. The SenoBright HD application shall enable contrast enhanced breast imaging a dual energy technique. This imaging technique can be used as an adjunct following mammography and ultrasound exams to help localize a known or suspected lesion.

| Type of Use (Select one or both, as applicable)                                                                                                                                        |                                                                                                                                |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span> <b>                         ☑                     </b> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div style="display:flex; align-items:center;"> <span>□</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Premarket Notification Submission

Image /page/4/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has a white border and four white teardrop shapes evenly spaced around the perimeter, giving the impression of motion or energy.

Section 5: 510(k) Summary

#### NIRA for SENOBRIGHT HD

510(k) Number: K211215

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the text "510(k) Premarket Notification Submission". The text is written in a clear, sans-serif font and is horizontally aligned. The text appears to be the title or heading of a document or presentation related to medical device regulation.

Image /page/5/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has small, droplet-like shapes around it, giving it a dynamic and fluid appearance. The logo is presented in a blue color scheme.

#### 510(k) Summary

K211215

In accordance with 21 CFR 807.92 the following summary of information is provided:

| Date:                                         | June 17, 2021                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                    | GE Healthcare<br>GE Medical Systems SCS<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Primary Contact Person:                       | Florian Akpakpa<br>Regulatory Affairs Program Manager<br>283 rue de la Minière<br>78530 Buc - FRANCE<br>Phone : +33 1 30 70 40 40<br>Email : florian.akpakpa@ge.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Secondary Contact Person:                     | Gregory Pessato,<br>Regulatory Affairs Senior Manager<br>GE Medical Systems SCS<br>283 RUE DE LA MINIERE<br>78530 BUC – FRANCE<br>Phone : +33 1 30 70 40 40<br>Email : gregorypessato@ge.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Trade Name:                            | SenoBright HD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Common/Usual Name:                            | Contrast Enhanced Spectral Mammography                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification Names:<br>Product Code:        | 21 CFR 892.1715, Class II<br>MUE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate Device:                             | Predicate device: SenoBright HD (K172404)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Description:                           | This submission is proposing a software update to SenoBright HD<br>consisting of an improvement of the existing "recombination" algorithm<br>with the New Image Recombination Algorithm (NIRA) by adding a local<br>estimation of breast thickness in the images recombination to account<br>for the non-uniformity of the breast thickness, and by compensating for<br>potential patient movement between the 2 CESM acquisitions (Low<br>Energy and High Energy).                                                                                                                                                                                                                                                                                                                                                |
|                                               | SenoBright HD (K172404) is the name of Senographe Pristina FFDM<br>system allowing to perform Contrast Enhanced Spectral Mammography<br>(CESM) application.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                               | The CESM acquisition technique consists in acquiring two images (one<br>High Energy and one Low Energy) in sequence and under the same<br>breast compression after patient injection with an iodinated contrast<br>media. The two images are then recombined through a post-processing<br>algorithm.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                               | This design change is a software and labeling only option, compatible<br>with SenoBright HD installed base and does not require any hardware<br>modification on the Senographe Pristina platform.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Intended Use:                                 | SenoBright HD is an extension of the existing indication for diagnostic<br>mammography with Senographe Pristina. The SenoBright HD<br>application shall enable contrast enhanced breast imaging using a dual<br>energy technique. This imaging technique can be used as an adjunct<br>following mammography and ultrasound exams to help localize a<br>known or suspected lesion.<br>Note: The intended use of SenoBright HD cleared in K172404 is not<br>changed.                                                                                                                                                                                                                                                                                                                                                 |
| Indications for Use                           | Refer to Intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Technology:                                   | The update of the existing recombination algorithm to <b>New Image<br/>Recombination Algorithm (NIRA)</b> for SenoBright HD is completely<br>independent from all other parts of the digital mammographic<br>equipment Senographe Pristina. It therefore does not affect the dual-<br>energy image acquisition.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                               | The changes introduced with <b>NIRA</b> are specific to the recombination<br>algorithm and can affect exclusively the recombined image. Thus, it also<br>does not affect the acquisition, processing and display of the low-<br>energy image when used as a regular FFDM image.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                               | <b>NIRA</b> is an evolution of the SenoBright HD Recombination algorithm<br>bringing the following improvements:<br>- Use of a local estimation of breast thickness in the image<br>recombination to account for the non-uniformity of the breast<br>thickness.<br>- Compensation for potential patient movement between the 2<br>CESM acquisitions (Low Energy and High Energy).                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Substantial Equivalence /<br>Predicate Device | The update of the existing recombination algorithm to the <b>New Image<br/>Recombination Algorithm (NIRA)</b> for SenoBright HD does not impact<br>the intended use/indications for use of the legally marketed device<br>SenoBright HD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                               | The fundamental principles of operation, functionalities, specifications<br>and technological characteristics of SenoBright HD with NIRA are                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                               | identical to those of the Predicate device SenoBright HD.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                               | The image processing of SenoBright HD is improved with the New                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                               | Image Recombination Algorithm by adding a local estimation of breast                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                               | thickness in the images recombination to account for the non-                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                               | uniformity of the breast thickness, and by compensating for potential                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                               | patient movement between the 2 CESM acquisitions (LE and HE).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                               | Image Quality Performance Testing results demonstrated that the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                               | image quality of SenoBright HD with NIRA performs similarly as its                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                               | legally marketed predicate device, SenoBright HD (K172404).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Determination of Substantial<br>Equivalence:  | NIRA for SenoBright HD successfully completed required design control<br>testing per GE Healthcare's quality management system. No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                               | unexpected test results were obtained. The device is designed and will                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                               | be manufactured in compliance with the Quality System Regulations of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                               | 21CFR 820 and ISO 13485. The following quality assurance measures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                               | were applied to the development of the system:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                               | - Risk Analysis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                               | - Design Reviews                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                               | - Software Development Lifecycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                               | - Testing on unit level (Module verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                               | - Integration testing (System verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                               | - Performance testing (Verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                               | - Safety testing (Verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                               | - Simulated use testing (Validation)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                               | The safety and performance of NIRA for SenoBright HD was                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                               | demonstrated through full verification testing and additional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                               | performance testing such as:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                               | 1. A Non-clinical Performance Testing on phantoms, including                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                               | Image quality testing to demonstrate that SenoBright HD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                               | with NIRA performs at least as well as the cleared device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                               | SenoBright HD (K172404) and brings Image Quality                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                               | improvements. The performance testing also demonstrates                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                               | reduction of artifacts in case of either patient motion or                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                               | breast thickness non-uniformity to increased lesion visibility.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                               | 2. A clinical image evaluation performed by 3 independent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                               | MQSA-qualified radiologists on 10 images with objective                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                               | criteria defined. It illustrated the clinical image acceptability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                               | of images generated by SenoBright HD with NIRA.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                               | 3. A clinical image evaluation comparing <b>SenoBright HD with NIRA</b> to the predicate device SenoBright HD, performed by 3 independent MQSA-qualified radiologists on 50 clinical images, showing results including:<br>The contrast uptake visibility assessed equivalent or better in 97% of the cases. The visibility of artifacts assessed equivalent or lower in 99% of the cases. The overall clinical image quality assessed superior in more than 98% of the cases.              The results of verification and performance testing demonstrate the safe and effective use of <b>SenoBright HD with NIRA</b> .<br><b>SenoBright HD with NIRA</b> is substantially equivalent to the predicate device SenoBright HD (K172404) and there is no new question of safety and effectiveness that was raised. |
| Conclusion:                                   | The fundamental principles of operation, functionalities, specifications and technological characteristics of <b>SenoBright HD with NIRA</b> are identical to those of the Predicate device SenoBright HD.<br>Furthermore, results of successful verification activities and additional bench performance testing and clinical image evaluations do not raise any new issue regarding the safety and effectiveness of the device.<br>Based on this information, GE Healthcare believe that <b>SenoBright HD with NIRA</b> is substantially equivalent to its predicate device SenoBright HD (K172404).                                                                                                                                                                                                             |

{6}------------------------------------------------

# GE Healthcare

______________________________________________________________________________________________________________________________________________________________________________

Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font inside. There are decorative swirls around the letters, giving the logo a classic and recognizable appearance. The blue color is consistent throughout the logo.

# 510(k) Premarket Notification Submission

{7}------------------------------------------------

# GE Healthcare

Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a circle. There are decorative flourishes around the circle, resembling water droplets or stylized leaves. The logo is presented in a blue color.

# 510(k) Premarket Notification Submission

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The letters and the frame are in a light blue color. There are also some white curved lines around the letters.

# 510(k) Premarket Notification Submission

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE/K211215](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE/K211215)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
