← Product Code [MUE](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE) · K192317

# Planmed Clarity 2D and Clarity S (K192317)

_Planmed OY · MUE · Oct 23, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE/K192317

## Device Facts

- **Applicant:** Planmed OY
- **Product Code:** [MUE](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE.md)
- **Decision Date:** Oct 23, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1715
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Planmed Clarity 2D and Planmed Clarity S mammography units acquire digital 2D mammographic images. The Planmed Clarity 2D/S systems are intended to be used for screening and diagnosis of breast cancer. The Clarity 2D/S systems may also be used for additional diagnostic workup of the breast.

## Device Story

Full-field digital mammography (FFDM) systems; acquire 2D mammographic images using amorphous silicon flat-panel detectors with CsI:Tl scintillator. Input: X-ray photons; Output: digital mammographic images. Operated by clinicians in clinical settings. Workflow: patient info entry via HIS/RIS/MIS; image acquisition; processing; preview; soft-copy review or printing. Clarity 2D includes side touch-panel controls; Clarity S is a value-oriented system controlled via workstation. Benefits: screening/diagnosis of breast cancer. Software (Clarity Manager) manages acquisition, communication, and image display. Output used by radiologists for clinical decision-making regarding breast health.

## Clinical Evidence

Clinical image evaluation conducted with 6 patients undergoing routine screening/diagnostic mammography. Images (BI-RADS 1 or 2) reviewed by two MQSA-qualified US radiologists. All images rated 'good' or 'excellent'; image quality deemed acceptable for mammographic usage.

## Technological Characteristics

FFDM system; amorphous silicon (a-Si) flat-panel detector with CsI:Tl scintillator; 83 μm pixel size; 16-bit ADC. X-ray generator and tube identical to predicate. Connectivity: DICOM (encrypted). Standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-45, IEC 60601-1-2 (EMC), ISO 10993-1. Software: Clarity Manager 2.3 on Windows 10.

## Regulatory Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

## Predicate Devices

- Planmed Clarity Full Field Digital Mammography X-ray System ([K163328](/device/K163328.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 23, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Planmed Oy % Lars Moring Regulatory Affairs Manager Sorvaajankatu 7 Helsinki. 00880 FINLAND

Re: K192317

Trade/Device Name: Planmed Clarity 2D and Clarity S Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: Class II Product Code: MUE Dated: September 22, 2020 Received: September 25, 2020

Dear Lars Moring:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192317

Device Name

Planmed Clarity 2D and Clarity S

#### Indications for Use (Describe)

The Planmed Clarity 2D and Planmed Clarity S mammography units acquire digital 2D mammographic images. The Planmed Clarity 2D/S systems are intended to be used for screening and diagnosis of breast cancer. The Clarity 2D/S systems may also be used for additional diagnostic workup of the breast.

| Type of Use (Select one or both, as applicable)                                                     |                                                                                                    |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

(updated version 3.) K192317

#### I. SUBMITTER

# Manufacturer

Planmed Oy Sorvaajankatu 7 FI-00880 Helsinki, Finland Phone: +358 20 7795 300 Fax: +358 20 7795 396 Contact person: Lars Moring

### U.S. designated agent

Planmed USA Inc. 100 North Gary Avenue, Suite A Roselle, IL 60172 Phone: (630) 894 2200 Fax: (630) 894 4271 Contact person : Brett Hines

Date Prepared: October 23, 2020

#### II. DEVICE

| Name of Device:       | Planmed Clarity 2D and Clarity S                                  |
|-----------------------|-------------------------------------------------------------------|
| Common or Usual Name: | Full Field Digital Mammography (FFDM) System                      |
| Classification Name:  | Full Field Digital Mammography (FFDM) System<br>(21 CFR 892.1715) |
| Regulatory Class:     | II                                                                |
| Product Code:         | MUE                                                               |

#### III. PREDICATE DEVICE

Planmed Clarity Full Field Digital Mammography X-ray System, K163328 This predicate system has not been subject to a design-related recall. No reference devices were used in this submission.

#### IV. DEVICE DESCRIPTION

The Planmed Clarity 2D and Clarity S are Full Field Digital Mammography (FFDM) systems for generating mammographic x-ray images that can be used for screening and diagnosis of breast cancer. Planmed Clarity 2D and Clarity S utilize an amorphous silicon based digital image receptor to capture images. The receptor directly converts the incoming X-ray photons to digital image data.

The workflow with Clarity 2D is controlled by the side displays/touch panels and the workflow of Clarity S is controlled from the acquisition workstation and Clarity Manager acquisition and communications software. The patient information is entered manually or received from the hospital, radiology, or mammography information systems (HIS, RIS, or MIS. respectively), as a format of modality worklist. Subsequently, the images are acquired. processed, and displayed for preview. After initial evaluation by the user, the images are either printed or transferred for soft-copy review.

{4}------------------------------------------------

#### V. INDICATIONS FOR USE

The Planmed Clarity 2D and S mammography units acquire digital 2D mammographic images. Both systems are intended to be used for screening and diagnosis of breast cancer. The Clarity 2D and S systems may also be used for additional diagnostic workup of the breast.

The Indications for Use (IFU) statement are the same for both subject and predicate systems.

#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

### General

Both predicate and subject devices are using the same basic operating principles and are technically very similar. The Clarity 2D system is similar to the predicate Clarity. The 2D system can be technically upgraded to a DBT system in the field (not for the US market). The Clarity S system is a value system derived from Clarity and it cannot be upgraded to DBT. The dual touchscreen user interfaces of Clarity and Clarity 2D are omitted from Clarity S. X-ray generation and control are identical in both systems. The compression system and AEC technology are the same.

### Integrated detector

The digital full field flat panel detector is a new version of the Clarity detector. The new detector has essentially identical imaging characteristics and signal processing as the predicate device. Quality assurance with pixel defect acceptance criteria comparison is similar. Pixel matrixes are identical, pixel width 83 um. ADC bit depth is 16 bit for the subject system vs. 14 bit for the predicate system. Software interface to communicate with the detector has not been changed.

### X-ray unit

Dimensions of the units are similar.

#### X-rav tube

Units use the same X-ray tube.

### X-ray generator

Units use the same X-ray generator.

#### X-rav anti-scatter grid

Same grid is used on both systems.

### Software

The Clarity Manager workstation software was changed to run on Windows 10 instead of Windows 7 of the predicate system. The software of the system has been improved to include encrypted DICOM communication. Imaging sequence has been made shorter by improving the software interface with the detector and with some optimization of the x-ray generator timing. The software has been adapted for the higher dynamic range of the new detector version (from 14 bit to 16 bit). Also some new hardware version related changes of the user interfaces have been made. A detailed software description document is included with this submission along with a comprehensive description of changes between software versions of the predicate device to the subject device.

### Risks

Cyber security risks and their mitigations have been addressed more closely and comprehensively with the latest software release for the subject systems. New hazards in comparison to predicate device have been identified and addressed.

{5}------------------------------------------------

#### Table 1: Comparison of device specifications

| Device                                                     | Planmed Clarity 2D                                                           | Planmed Clarity S                                                                           | Planmed Clarity                                                                 |
|------------------------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
|                                                            | Subject device                                                               | Subject device                                                                              | Predicate device                                                                |
| General comparison                                         |                                                                              |                                                                                             |                                                                                 |
| Description                                                | Full Field Digital<br>Mammography system                                     | Full Field Digital<br>Mammography system                                                    | Full Field Digital<br>Mammography system                                        |
| System configuration                                       | X-ray stand and<br>Workstation PC                                            | X-ray stand and<br>Workstation PC                                                           | X-ray stand and<br>Workstation PC                                               |
| Application software                                       | Clarity Manager 2.3 on<br>Windows 10                                         | Clarity Manager 2.3 on<br>Windows 10                                                        | Clarity Manager 1.3 on<br>Windows 7                                             |
| User interfaces                                            | Touch screen user<br>interface side displays on<br>both sides of x-ray stand | No side displays on x-ray<br>stand, user interface<br>included in workstation<br>PC display | Touch screen user<br>interface side displays<br>on both sides of x-ray<br>stand |
| 'Side access' feature<br>for easier patient<br>positioning | yes, standard                                                                | not available                                                                               | yes, standard                                                                   |
| Control switches                                           | additional C-arm control<br>switches, compression<br>fine adjustment         | no additional switches                                                                      | additional C-arm control<br>switches, compression<br>fine adjustment            |
| Upgradeability                                             | DBT with s2D (not in US)<br>stereotactic device (not<br>in US) optional      | not upgradeable to 3D<br>not upgradeable with<br>stereotactic device                        | DBT with s2D (not in US)<br>stereotactic device (not<br>in US) optional         |
|                                                            |                                                                              |                                                                                             |                                                                                 |
| Detector comparison                                        |                                                                              |                                                                                             |                                                                                 |
| Detector type                                              | Varex Paxscan 3024MX                                                         | Varex Paxscan 3024MX                                                                        | Varex Paxscan 3024M                                                             |
| Detector technology                                        | Amorphous silicon (a-Si)<br>with Csl:TI scintillator<br>TFT array            | Amorphous silicon (a-Si)<br>with Csl:TI scintillator<br>TFT array                           | Amorphous silicon (a-Si)<br>with Csl:TI scintillator<br>TFT array               |
| Pixel matrix                                               | 2816x3584                                                                    | 2816x3584                                                                                   | 2816x3584                                                                       |
| Size of active area                                        | 23.1x29.1cm                                                                  | 23.1x29.1cm                                                                                 | 23.1x29.1cm                                                                     |
| Pixel size                                                 | 83 μm                                                                        | 83 μm                                                                                       | 83 μm                                                                           |
| ADC bit depth                                              | 16 bit                                                                       | 16 bit                                                                                      | 14 bit                                                                          |
| Pixel fill factor                                          | 0.52                                                                         | 0.52                                                                                        | 0.70                                                                            |
| Readout time                                               | 236 ms                                                                       | 236 ms                                                                                      | 495 ms                                                                          |

(all other imaging effecting specifications are substantially equal for all these systems)

#### VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

### Biocompatibility testing

Since the subject systems are almost identical to the predicate device, no new patient contacting mechanical parts have been added to the subject system. Hence no new biocompatibility testing was necessary. Thus, there is no risk or concern to the patient's safety from contact with the materials of construction of this new mammography unit. Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Planmed Clarity mammography system. The system complies with IEC 60601-1, IEC 60601-1-6, IEC 60601-2-28, IEC

{6}------------------------------------------------

60601-2-45, ISO 10993-1 for safety, and IEC 60601-1-2 for EMC. The only electrical difference is the flat panel detector type which has been separately emc tested by TÜV Rheinland (please see attached emc testing report in VOL17.)

## Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern.

## Physical laboratory testing

The performance of the system was tested according to Class II Special Controls Guidance Document: Full-Field Digital Mammography System Document issued on: March 27, 2012.

Summary of physical laboratory testing (bench testing) results:

- 1. Sensitometric response: Both detectors respond linearly to radiation exposure
- Spatial resolution: Both systems perform similar in terms of MTF since the detector 2. scintillator material or thickness are the same and detector element pixel size is the same.
- 3. Noise analysis: The subject device has slightly better noise performance. Also SNR and CNR are slightly higher.
- 4. Dynamic range: The subject device produces higher DQE than the predicate system.
- 5. Repeated exposures: Ghost tolerance was similar.
- 6. Automatic Exposure Control (AEC) performance: With automatic filtration change the organ and entrance doses are well within generally accepted levels. With a 40 mm PMMA phantom corresponding to an average breast thickness and glandularity, organ dose was measured 1.23 mGy with W/Ag beam quality.
- 7. Phantom test: RMI phantom scores for all attributes are similar for subject and predicate devices. CDMAM test passes with both systems.
- 8. Patient radiation dose: With varving phantom thickness and varying adiposity and glandularity factors, both systems achieve radiation dose levels well within generally acceptable limits.
- 9. Breast compression system: Powered compression pressure test results are similar.

### Clinical image evaluation

Clinical image evaluation was performed with the subject Clarity 2D system. Purpose was to determine if the image quality were judged to be of sufficiently acceptable quality for mammographic usage when reviewed by MQSA qualified experienced radiologists. The images were taken in one site in Bulgaria where 6 patients participated to routine breast cancer screening. Images assigned BI-RADS score 1 or 2 were selected for the evaluation. Some cases also included diagnostic mammograms, i.e. spot and/or magnification images. The images were then reviewed by two MQSA qualified experienced US radiologists independently. All images were rated good or excellent and thus the overall image quality was acceptable for all cases and image types.

# VIII. CONCLUSIONS

The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the Planmed Clarity 2D and S full field digital mammography systems perform as intended in the specified use conditions. The clinical image evaluation also shows that the devices equipped with the new detector version perform comparably to the predicate device that is currently marketed for the same intended use.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE/K192317](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MUE/K192317)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
