K123520 · Siemens Medical Solutions USA, Inc. · MUE · Jun 11, 2013 · Radiology
Device Facts
Record ID
K123520
Device Name
MAMMOMAT INSPIRATION PRIME
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
MUE · Radiology
Decision Date
Jun 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1715
Device Class
Class 2
Indications for Use
The MAMMOMAT Inspiration PRIME system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.
Device Story
Floor-mounted mammography system for screening, diagnostic, and biopsy procedures; used on standing, seated, or recumbent patients. System includes examination stand, high-frequency generator, radiation shield, and Acquisition Workstation (AWS). Swivel arm holds X-ray tube and detector. Features optional gridless acquisition mode; replaces physical scatter radiation grid with progressive image reconstruction algorithm. Algorithm identifies scatter-causing structures to calculate corrected image, allowing primary radiation usage without grid absorption. Reduces radiation dose up to 30% compared to conventional grid-based systems. Operated by healthcare professionals. Output viewed via hard copy film or soft copy workstation to assist clinical diagnosis.
Clinical Evidence
Clinical testing performed to quantify potential dose savings. Bench testing included phantom studies and clinical image review to verify equivalent image quality compared to the predicate device.
Technological Characteristics
Full-field digital mammography system. Components: high-frequency generator, X-ray tube, TFT detector. Gridless acquisition mode utilizes progressive image reconstruction algorithm to replace physical scatter radiation grid. Connectivity: Acquisition Workstation (AWS). Safety features: visual/audible warnings, system monitoring with error blocking.
Indications for Use
Indicated for mammography exams, screening, diagnosis, and stereotactic biopsies in patients requiring such procedures, performed under the supervision of medical professionals.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
{0}------------------------------------------------
K123520
Page 1 of 3
### 510(k) Summary: Mammomat Inspiration PRIME
Company:
Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355.
Date Prepared: June 11, 2013
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### 1. General Information:
SIEMENS
Importer / Distributor: Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Establishment Registration Number: 2240869 Manufacturing Site: SIEMENS AG Sector Healthcare Henkestraße 127 91050 Erlangen Establishment Registration Number:
3004977335
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 Malvern, PA 19355 Phone: (610) 448 -3536 Fax: (610) 448-1787 Email: patricia.d.jones@siemens.com
#### 3. Device Name and Classification:
Trade Name: Mammomat Inspiration PRIME Device: Full Field Digital, system, X-Ray Mammographic Requlation: Medical Specialty Radiology Review Panel Radiology Product Code MUE Submission Type 510(k) Regulation Number 892.1715 Device Class 2
#### 4. Legally Marketed Predicate Device
| Trade Name: | Mammomat Inspiration VB30 |
|-------------|---------------------------------------------------|
| Device: | Full Field Digital, system, X-Ray<br>Mammographic |
| Regulation: | Medical Specialty Radiology |
## JUN 1 1 2013
{1}------------------------------------------------
## SIEMENS
Review Panel Radiology Product Code MUE Submission Type 510(k) Requlation Number 892.1715 Device Class 2
#### ട്. Device Description:
Mammomat Inspiration PRIME is a floor-mounted mammography system for screening, diagnostic and biopsy procedures on standing, seated or recumbent patients.
The Mammomat Inspiration PRIME provides optional gridless acquisition and progressive reconstruction. During progressive reconstruction a unique algorithm replicates the function of the grid. The grid slides back and no longer absorbs primary radiation, therefore less radiation dose is needed. The Mammomat Inspiration PRIME in gridless acquisition mode reduces radiation dose up to 30 percent.
The system consists of an examination stand with integrated, microprocessorcontrolled, high-frequency generator as well as a radiation shield with an optional height-adjustable control desk in which the Acquisition Workstation (AWS) can be integrated. A swivel arm contains the X-ray tube on the top end and the object table with the detector on the bottom end.
#### 6. Indication for Use:
The MAMMOMAT Inspiration PRIME system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals.
Mammographic images can be interpreted by either hard copy film or soft copy workstation.
#### 7. Substantial Equivalence:
The Siemens Mammomat Inspiration PRIME is substantially equivalent to the commercially available Siemens Mammomat Inspiration P030010/S006.
X-ray generation and control used with the Mammomat Inspiration PRIME is identical to the Mammomat Inspiration with VB30. Detector TFT specifications and general image processing algorithms remain unchanged. The Acquisition Workstation (AWS) is identical.
#### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The modified Mammomat Inspiration Prime can lower dose up to 30 percent compared to the predicate (P030010/S006) with equivalent image quality as evaluated with phantom testing and clinical image review. The dose can be reduced as the scatter radiation grid, which has been commonplace up to now, is replaced by a new algorithm for progressive image reconstruction. The conventional scatter radiation grid absorbs scattered radiation to ensure image quality. But as it also absorbs part of the important primary radiation, higher dose is needed to obtain images of the desired quality. The new algorithm can identify the scatter-causing
{2}------------------------------------------------
K123520
Page 3 of 3
# SIEMENS
structures and can calculate a corrected image. Thus, the primary radiation can be completely used and high-quality images can be achieved with less dose.
#### 9. Summary of Non-Clinical Tests:
The Siemens Mammomat Inspiration Prime complies with voluntary standards as detailed in Section 9 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- Risk Analvsis
- Requirements Reviews
- Design Reviews
- Integration testing (System verification) .
- Performance testing (Bench testing) .
#### 10 Summary of Clinical Tests:
For the subject of this premarket submission, Siemens Mammomat Inspiration Prime, included clinical testing to quantify the potential of dose savings. This clinical is described in Section 15 of this premarket submission.
#### 11. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Mammomat Inspiration is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed.
Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards. Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
#### 12. Conclusion as to Substantial Equivalence:
The Mammomat Inspiration PRIME is intended for the same indications for use as the predicate Mammomat Inspiration VB30. The imaging properties (TFT with pixel size and number and image processing algorithms) have not been modified. It is Siemens opinion, that the Mammomat Inspiration PRIME is substantially equivalent to the Mammomat Inspiration P030010/S006.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle-like symbol with three wing-like shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609
Silver Spring, MD 20993-0002
Public Health Service
June 11, 2013
Ms. Patricia D. Jones Siemens Medical Solution, Inc. Technical Specialist, Regulatory Submissions 51 Valley Stream Parkway, E-50 MALVERN PA 19355
Re: K123520
Trade/Device Name: Mammomat Inspiration PRIME Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: May 31, 2013 Received: June 4, 2013
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, practice, labeling, and prohibitions against.misbranding and_ adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Jones
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
5 10(k) Number (if known): __K 123520
Device Name: Mammomat Inspiration PRIME
Indications for Use:
The Mammomat Inspiration system is intended for mammography exams. screening, diagnosis, and stereotactic biopsies under supervision of medical professionals.
Mammographic images can be interpreted by either hardcopy film or softcopy workstation.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ ANDIOR (Part 21 CFR 301 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Sm.h.p.)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
K123520
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.