The OSTEOSPACE is a quantitative ultrasound bone sonometer device (QUS) to be used for the measurement of broadband ultrasound attenuation (BUA) of the calcaneus, as an aid, together with other clinical risk factors, to diagnose osteoporosis and other medical conditions leading to reduced bone strength and to estimate the risk of subsequent atraumatic fracture. The output is expressed in terms of BUA, T-score, and Z-score.
Device Story
OSTEOSPACE is a quantitative ultrasound (QUS) bone sonometer measuring calcaneal bone properties via high-frequency sound waves. System includes scanner, dedicated PC, and accessories. Operator positions patient's foot in scanner footwell; applies water-soluble ultrasound gel; uses low-power laser to align transducers with external malleolus. Transducers transmit/receive ultrasound pulses through heel; infrared beam ensures reproducible ROI positioning. Software calculates soft tissue thickness and broadband ultrasound attenuation (BUA) from 20 consecutive measurements. Output includes BUA (dB/MHz), T-score (compared to young normal Caucasian female), and Z-score (compared to age/gender/ethnicity-matched population). Used in clinical settings by trained operators. Results assist clinicians in diagnosing osteoporosis and assessing fracture risk by identifying reduced bone strength. Benefits include non-invasive, radiation-free bone status assessment comparable to DEXA.
Clinical Evidence
Clinical studies included a reference database study (n=410 healthy Caucasian females) and fracture risk studies (n=195 subjects across two centers). Reference study established normality curves and T-score/Z-score parameters. Fracture risk studies compared OSTEOSPACE to DEXA (QDR 4500) and other QUS systems (ACHILLES+, SAHARA). Results showed OSTEOSPACE BUA discriminated between fractured and control subjects with AUC 0.72-0.77, comparable to DEXA and other QUS devices. Precision study (n=56) reported RMS CV of 1.84%. No adverse events reported.
Technological Characteristics
Electrically operated QUS bone sonometer. Components: scanner with two ultrasound transducers, dedicated PC (Pentium 3+, 800MHz, Windows OS). Sensing: high-frequency sound wave transmission/reception. Connectivity: PC-based, parallel port (LTP1). Materials: standard medical-grade contact materials. Software: Minor Level of Concern. Compliance: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC). Acoustic output: Track 1 FDA guidance compliant.
Indications for Use
Indicated for use as an aid in diagnosing osteoporosis and estimating risk of atraumatic fracture in patients by measuring calcaneal broadband ultrasound attenuation (BUA).
Regulatory Classification
Identification
A bone sonometer is a device that transmits ultrasound energy into the human body to measure acoustic properties of bone that indicate overall bone health and fracture risk. The primary components of the device are a voltage generator, a transmitting transducer, a receiving transducer, and hardware and software for reception and processing of the received ultrasonic signal.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Bone Sonometers.” See § 892.1(e) for the availability of this guidance document.
Reference Devices
Hologic QDR 4500 (DEXA)
Lunar ACHILLES+ (QUS)
Hologic SAHARA (QUS)
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
I. GENERAL INFORMATION
Generic Name: Ultrasound Bone Sonometer
Device / Trade Name: OSTEOSPACE
Applicants name and address: MEDILINK
80, rue de l’Hortus
34280 CARNON
France
Applicant’s U.S. Representative: Michael Manolagas
494 Washington Street
Dedham, MA 02026
Dates of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P010058
Date of Good Manufacturing Practice Inspection: February 28, 2002
Date of notice of approval to the Applicant:
II. INDICATIONS FOR USE
The OSTEOSPACE is a quantitative ultrasound bone sonometer device (QUS) to be used for the measurement of broadband ultrasound attenuation (BUA) of the calcaneus, as an aid, together with other clinical risk factors, to diagnose osteoporosis and other medical conditions leading to reduced bone strength and to estimate the risk of subsequent atraumatic fracture. The output is expressed in terms of BUA, T-score, and Z-score.
III. CONTRAINDICATIONS
None.
IV. WARNINGS and PRECAUTIONS
See attached device labeling.
{1}
2
8
# V. DEVICE DESCRIPTION
OSTEOSPACE (Ultrasound Bone Sonometer) is an electrically operated quantitative ultrasound (QUS) bone sonometer which measures bone properties at the calcaneus using non-audible high-frequency sound waves. The device consists of the scanner, dedicated PC and accessories. The scanner consists of a footwell to position the foot and two ultrasound transducers that contact the heel so that ultrasound beam is passed through it. The minimum specifications for the computer are: Pentium 3 or better processor with at least 800MHz operating speed; Windows 98, 2000, or XP operating system; 128 Megabytes memory; 20 Gigabytes disk memory; and a main board with parallel output (LTP1). The accessories: keyboard, mouse, compatible printer and monitor (15'' or bigger) can be supplied by the customer, distributor or MEDILINK.
The OSTEOSPACE is controlled directly by a dedicated computer. All instructions and examination results are maintained in the computer, along with a software version of the User Manual. OSTEOSPACE allows the possibility to back-up the results onto disk. Printed reports of the examination results are produced using an external printer.
The results are given as Broadband Ultrasound Attenuation (BUA) in dB/MHz. This ultrasound parameter is based on the frequency dependent attenuation, with the higher BUA values corresponding to lower risk of fracture, and vice versa.
Before the BUA measurement can be used for a diagnosis it needs to be compared to the average value of the young normal Caucasian female. This comparison is done using an index called T-score, which represents the BUA value on a normalized scale. T-score above (below) 0 corresponds to a bone stronger (weaker) than that of the average young normal Caucasian woman. The T-score is the recommended parameter for assessing the risk of fracture.
Comparing the actual BUA value to the average value in a healthy population of the same gender, ethnic origin, and age, when expressed in terms of standard deviations (SDs) of that population, is called Z-score, which can be used as an aid in the detection of conditions associated with non-age-related bone loss.
An OSTEOSPACE examination includes the following steps: probe positioning, soft tissue thickness calculation, examination and processing operation. The OSTEOSPACE measurements are made with the patient seated in a chair with his/her foot placed in the footwell of the scanner. The heel is smeared with standard water-soluble ultrasound gel; the gel is the medium for the transmission of ultrasound. The transducers are positioned with the aid of a low power laser onto the external part of the malleolus. The transducers are then brought into contact with the heel. A transducer on one side of the heel converts an electrical signal into a sound wave which passes through the patient's heel. The second transducer on the opposite side of the patients' heel receives the sound wave and converts it into an electrical signal that is analyzed by the OSTEOSPACE software. An infrared beam positioner allows reproducible positioning to the same Region of Interest (ROI) for repeat examinations. The probe positioning occurs when the patient has placed their foot into the footwell, and the measurement length of the foot size is entered into the computer software. The low powered laser light is moved into position onto the external malleolus by the operator. The computer thus ensures that the probes are correctly
{2}
positioned to obtain the ROI of the calcaneus. This can be further verified by the operator. The soft tissue thickness calculation of the patient’s heel is made by the transmission and detection of the reflected wave.
Examination of the patient lasts several seconds. During this period, 20 consecutive ultrasound measurements are made on the foot, followed by the computer calculation of the broadband ultrasound attenuation (BUA) of the Region of Interest (ROI).
The quality of each examination can be verified by the operator in comparing the ultrasound signal displayed at the end of the examination on the computer monitor. The patient details, including laser / probes positioning are recorded on the computer to ensure continued follow-up of patient examinations and greater precision in repeat measurements.
## VI. ALTERNATIVE PRACTICES & PROCEDURES
Alternative methods for assessing bone status include single energy x-ray absorptiometry (SXA), dual energy x-ray absorptiometry (DEXA), quantitative computed tomography (QCT), single photon absorptiometry (SPA), and dual photon absorptiometry (DPS). Of these techniques, SXA, DEXA and SPA have been used specifically for the estimation of Bone Mineral Density (BMD) of the calcaneus. These established techniques estimate BMD at a variety of anatomical sites, including the heel, by measuring the attenuation of x-rays due to passage through the bone. In addition, there are several bone sonometers that are currently being marketed for assessment of a patient’s skeletal status (fracture risk).
## VII. MARKETING HISTORY
OSTEOSPACE has been commercially available in markets outside the United States since 1997. Two hundred eighty (280) units have been sold in over 35 countries world-wide, including Europe, Latin America, Asia and the Middle East.
No OSTEOSPACE has ever been withdrawn from any market due reasons related to safety or effectiveness.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
None known.
{3}
# IX. PRE-CLINICAL STUDIES
A. Electrical Safety
MEDILINK verified compliance of the OSTEOSPACE with the general safety requirements of IEC 60601-1: 1996 and the electrical safety requirements of IEC 60601-1-1:1996.
B. Electromagnetic Compatibility
MEDILINK verified compliance of the OSTEOSPACE with the electromagnetic compatibility requirements of IEC 60601-1-2: 1996 (EMC Directive 89/336/CEE).
C. Ultrasound Acoustic Evaluation
MEDILINK tested the physics of the 4 ultrasound transmitting/receiving probes in accordance with the Track 1 of the Food and Drug Administration’s (FDA) 510k guidance, entitled “Measuring and Reporting Acoustic Output of Diagnostic Ultrasound Medical Devices”, 1985 (Table 1). Acoustic levels were well below acceptable pre-amendment values.
| | Probe 1 | Probe 2 | Probe 3 | Probe 4 | Typical uncertainty (+/-) |
| --- | --- | --- | --- | --- | --- |
| MI | 0.068 | 0.070 | 0.069 | 0.064 | 12% |
| p_{1/2} (MPa) | 86 x10^{-3} | 88 x10^{-3} | 87 x10^{-3} | 81 x10^{-3} | 10% |
| I_{SPPA} (W/cm^{2}) | 0.40 | 0.45 | 0.47 | 0.43 | 27% |
| ISPTA (mW/cm^{2}) | 4.2 x10^{-3} | 4.4 x10^{-3} | 4.3 x10^{-3} | 4.1 x10^{-3} | 25% |
| Beam diameter (cm) | 7.0 | 6.5 | 6.1 | 7.4 | 6% |
Table 1-Acoustic Output Values
D. Software
OSTEOSPACE is classified as Minor Level of Concern device (Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 29, 1998). Software hazard risk analyses demonstrated that all hardware, software and user concerns were adequately addressed. Verification, validation and unit testing demonstrated that the device operates in a manner as described in the specifications.
E. Biological / Sterility
The patient contact materials used in the OSTEOSPACE are ones which have been used in the medical field without any known adverse effects or reactions. No additional testing was required, since the safety of the contact materials was well established.
The labeling instructs the user to clean the footwell with a disinfectant between patients. The labeling provides specific instructions on how to ensure the effective use of the disinfectant.
FDA-cleared water soluble ultrasound gel contains chemicals known in the medical field and provides no risk to the patient. No additional testing was deemed necessary.
4
10
{4}
# X. SUMMARY OF CLINICAL STUDIES
Clinical studies were conducted to assess the safety and effectiveness of the OSTEOSPACE, a quantitative ultrasound bone sonometer device, as an aid to establish the diagnosis of osteoporosis and to identify patients with high risk of osteoporotic fracture. Clinical studies were carried out in two U.S. centers, the University of Massachusetts (UMASS) and the University of California (UCSF), San Francisco, and in one European center, the Geneva University Hospital (HUG), Switzerland. The same protocol was followed in all the centers.
## A. Reference Database Study
**Objective:** This study was to establish a U.S. Reference Database (Normality curve) for the BUA of OSTEOSPACE on healthy or non-fractured Caucasian U.S. women aged 20 to 79 (Reference Database Study).
**Methods:** Four hundred ten (410) healthy Caucasian U.S. females, ranging in age from 20 to 79 years, were measured using the OSTEOSPAC5 to establish the normality curve.
**Results:** BUA was found statistically independent of age for 235 females ranging from 20 to 47. Thus, over this period, the reference curve could be represented as a constant equal to the average BUA over the group (BUA$_{20-47}$ = 66.16 dB/MHz). Over 47 years old, a 3rd order polynomial regression was found to fit the best.
**Conclusions:** The Normality Curve of OSTEOSPACE BUA for Caucasian U.S. Women displayed in Figure 1 shows that between the age of 48 and 60 years (post menopause), the BUA significantly declined by 3.5 dB/MHz (approximately 83% of the total range). Then, from 60 and 79 years old, the BUA further declined by 0.7 dB/MHz, i.e. approximately 17% of the range.
{5}

Figure 1 - Normality Curve of OSTEOSPACE BUA for Caucasian U.S. Women
The World Health Organization (WHO) criterion for T-score is the difference between the patient’s measurement and the mean of a healthy young female Caucasian reference population between the ages of 20 and 39 expressed as the number of standard deviations, for the reference database, between the two values. The reference population for this device shows that there was no difference between using 20-39 group and 20-47 group. However, the 20-39 age range was selected for the representative sample of the young normal Caucasian U.S. female reference population to maintain consistency with the WHO definition. This young reference population’s mean BUA, as well as its standard deviation (SD), were calculated for the purpose of generating T-scores (see Table 1).
| | Value (dB/MHz) | 95% Confidence Interval |
| --- | --- | --- |
| Mean BUA OSTEOSPACE | 66.16 | 65.5 - 66.8 |
| Standard Deviation | 4.6 | 3.8 - 5.4 |
Table 1 - Young Reference Value for OSTEOSPACE BUA (Data From 171 U.S. Caucasian Females, Ages 20 to 39)
Given the previous results, the T-score of the patient “j” is calculated as follows:
$$
T\text{-score}_j = \frac{BUA_j - 66.16}{4.6}
$$
where $BUA_j$ is the BUA measured on the patient “j”.
6
12
{6}
# B. Precision Study
Object: To estimate the in-vivo short-term precision of the BUA obtained by OSTEOSPACE (Precision Study) f
Methods: Fifty-six (56) subjects ranging in age from 20 to 79 were recruited by UMASS and UCSF the two U.S. centers and used to assess the measurement reproducibility. Each subject was examined three times with OSTEOSPACE, with foot repositioning before each examination.
Results: Precision was evaluated by calculating the RMS SD (Absolute Precision), the RMS CV(Relative Precision), the CV, the SCV(Standardized Coefficient of Variation) and the TSD(Standard Deviation of the T-score). (See section 17 of the User Manual for definitions). Results are displayed in Table 2.
| | BUA
OSTEOSPACE |
| --- | --- |
| RMS SD | 1.19dB/MHz |
| RMS CV | 1.84 % |
| CV | 1.31 % |
| SCV | 3.97 % |
| TSD | 0.26 |
Table 2- Results of the Evaluation of the OSTEOSPACE Precision (56 American Subjects Aged between 20 to 79)
Conclusions: The CVs for the OSTEOSPACE measurements show that the device can provide precise measurements of BUA.
# C. Fracture Risk Studies
Object: To establish the capability of OSTEOSPACE BUA, a) to assess the risk of fracture, b) to discriminate between patients who have suffered atraumatic fractures and age-matched control subjects who have never had an atraumatic fracture, and c) to compare the performance of the device with those of one DEXA (Hologic QDR 4500®) and two sonometer systems (Lunar ACHILLES+® and Hologic SAHARA®), in order to assess possible bias in selection of control patients ("Fracture Risk Studies"). The output of the QDR 4500 is bone density. The output of the ACHILLES+® is Stiffness and the SAHARA® is the Quantitative Ultrasound Index (QUI).
Methods: In order to assess the capacity of OSTEOSPACE to evaluate the risk of fracture and to discriminate the Osteoporotic patients, fractured subjects and age-matched controls were enrolled by the HUG and UCSF centers. UCSF measured 52 age-matched controls and 50 fractured patients. HUG measured 43 age-matched controls and 56 fractured patients. Subjects were measured using the OSTEOSPACE (both sites), Hologic QDR 4500® (UCSF only), Lunar ACHILLES+® (HUG only) and the Hologic SAHARA® (HUG only).
{7}
Results: Table 3 shows that the BUA results for the fractured group expressed in T-score or in Z-score are similar to neck or spine BMD.
| | Controls | Fractured | Z-score | T-Score |
| --- | --- | --- | --- | --- |
| BUA OSTEOSPACE | 62.6 ± 4.5 | 58.8 ± 4.9 | -0.9 | -1.6 |
| Neck BMD (QDR 4500®) | 0.695 ± 0.111 | 0.614 ± 0.111 | -0.7 | -2.1 |
| Spine BMD (QDR 4500®) | 0.960 ± 0.145 | 0.839 ± 0.141 | -0.8 | -1.9 |
Table 3- UCSF Center, OSTEOSPACE and DEXA Parameters of the Two Groups Expressed in Z-score and T-score
Table 4 shows that the OSTEOSPACE measurements for the fractured subjects, when expressed in T-score or in Z-score, are similar to the QDR 4500® neck or spine BMD, or to Hologic QUI and Lunar Stiffness results.
| | Controls | Fractured | Z-score | T-Score |
| --- | --- | --- | --- | --- |
| BUA OSTEOSPACE | 61.2 ± 5.0 | 55.8 ± 5.2 | -1.1 | -2.3 |
| QUI (SAHARA®) | 73.9 ± 15.3 | 59.8 ± 19.4 | -0.9 | -2.2 |
| Stiffness (ACHILLES+®) | 71.0 ± 11.3 | 58.6 ± 12.5 | -1.1 | -1.8 |
Table 4- HUG Center, OSTEOSPACE and QUS Parameters for the Two Groups Expressed in Z-score and in T-score
For each center, non-adjusted and adjusted Odds Ratios per standard deviation decrease were estimated, with their 95% confidence intervals, and the areas under the ROC curves were obtained (see Tables 5 and 6).
| | Non-Adjusted Odds Ratios (95% CI) | Adjusted Odds Ratios* (95% CI) | Area under the ROC Curve** (95% CI) |
| --- | --- | --- | --- |
| BUA OSTEOSPACE | 2.44 (1.49 - 3.98) | 1.80 (1.05 - 3.10) | 0.72 (0.62 - 0.82) |
| Neck BMD (QDR 4500®) | 2.30 (1.40 - 3.79) | 1.70 (1.02 - 2.99) | 0.71 (0.60 - 0.81) |
| Spine BMD(QDR 4500®) | 2.47 (1.53 - 3.98) | 2.32 (1.36 - 3.93) | 0.74 (0.64 - 0.84) |
*Adjusted by Age, Weight and Height.
**Not Adjusted by age.
Table 5- UCSF Center, Odds Ratios per Standard Deviation Decrease and Area under the ROC Curve for each Bone Parameters
14
{8}
| | Non-Adjusted Odds Ratios (95% CI) | Adjusted Odds Ratios* (95% CI) | Area under the ROC Curve** (95% CI) |
| --- | --- | --- | --- |
| BUA OSTEOSPACE | 3.04 (1.81 - 5.10) | 2.91 (1.57 - 5.37) | 0.77 (0.68 - 0.86) |
| QUI (SAHARA®) | 2.47 (1.46 - 4.17) | 1.88 (1.05 - 3.34) | 0.77 (0.68 - 0.86) |
| Stiffness (ACHILLES+®) | 3.17 (1.83 - 5.47) | 2.56 (1.39 - 4.73) | 0.77 (0.68 - 0.86) |
*Adjusted by Age, Weight and BMI.
**Not Adjusted by age.
Table 6- HUG Center, Odds Ratios per Standard Deviation Decrease and Area under the ROC Curve for each Bone Parameters
Conclusions: ROC curves as well as Odds Ratios analysis showed no statistical difference between OSTEOSPACE, DEXA, and QUS measurements, thus demonstrating the absence of any significant bias in selection of control patients, and also demonstrating the ability of the OSTEOSPACE to discriminate between fractured subjects and controls.
## XI. CONCLUSIONS DRAWN FROM THE STUDIES
### A. Safety
The safety of OSTEOSPACE has been demonstrated during the clinical evaluation, with no reports of adverse events or side effects. This clinical experience is consistent with the worldwide experience with OSTEOSPACE.
### B. Effectiveness
The results of the clinical studies demonstrate that both the capacity of OSTEOSPACE BUA to discriminate between osteoporotic and non-osteoporotic subjects and its ability to assess risk of fracture are comparable with those of BMD obtained by DXA absorptiometry technique. BUA measured by OSTEOSPACE has a comparable precision to other QUS systems. Therefore, OSTEOSPACE can be used as an aid to diagnosing osteoporosis and to determining the risk of subsequent atraumatic fracture.
## XII. PANEL RECOMMENDATION
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Radiologic Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
15
{9}
10
16
# XIII. CDRH DECISION
FDA issued an approval order on **MAR 15 2004**
The applicant’s manufacturing facility was inspected on February 28, 2002, and was found to be in compliance with the Quality Systems regulations.
# XIV. APPROVAL SPECIFICATIONS
Directions for use: See attached labeling.
The sale, distribution, and use of this device are restricted to prescription use in accordance with 21 CFR 801.109 within the meaning of section 520(e) of the Federal Food, Drug, and Cosmetic Act (the act) under the authority of section 515(d)(1)(B)(ii) of the act. FDA has also determined that, to ensure the safe and effective use of the device, the device is further restricted within the meaning of section 520(e) under the authority of section 515(d)(1)(B)(ii) insofar as the sale, distribution, and use must not violate sections 502(q) and (r) of the act.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, and Precautions in the attached labelling.
# XV. REFERENCES
General References:
[1] Glüer CC, Blake G, Lu Y, Blunt BA, Jergas M, Genant HK. Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int 1995;5(4):262-70.
[2] CG Miller, RJ Herd, T Ramalingam, I Fogelman, GM Blake Ultrasonic Velocity Measurement Through the Calcaneus: which velocity should be measured, Osteoporosis Int, 1993. 3.31-35
[3] Glüer CC, How to characterize the ability of diagnostic technique to monitor the skeletal changes. Journal of Bone and Mineral Research 1997; 12: S378.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.