K161971 · Bone Index Finland , Ltd. · MUA · Jan 9, 2017 · Radiology
Device Facts
Record ID
K161971
Device Name
Bindex BI-2
Applicant
Bone Index Finland , Ltd.
Product Code
MUA · Radiology
Decision Date
Jan 9, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1180
Device Class
Class 2
Indications for Use
Bindex measures apparent cortical bone thickness at the proximal tibia and can be used in conjunction with other clinical risk factors or patient characteristics as an aid to the physician in the diagnosis and other medical conditions leading to reduced bone strength and in the determination of fracture risk.
Device Story
Bindex BI-2 is a handheld ultrasound bone sonometer used in clinical settings by healthcare providers. The system consists of a handheld ultrasound transducer connected to a computer via USB. The operator applies gel to the patient's skin and manually places the transducer perpendicularly on the proximal tibia (1/3 length). The device measures apparent cortical bone thickness and calculates a Density Index (DI), which estimates hip bone density as measured by DXA. The output is displayed via computer software to assist physicians in diagnosing reduced bone strength and assessing fracture risk. The device requires manual transducer orientation and five repeated measurements per location. The transducer is disinfected between uses with isopropyl alcohol.
Clinical Evidence
No new clinical data provided. The device relies on the clinical and pre-clinical data previously established for the predicate Bindex BI-100. Performance testing included bench-level biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-1), electrical safety, EMC, and mechanical/acoustic testing (drop test, ball pressure test, moulding stress relief).
Technological Characteristics
Handheld ultrasound transducer; 3.0 MHz nominal center frequency. Powered via PC USB port. Operates in temperature range +10°C to +40°C. Complies with IEC 60601-1, IEC 60601-2-37, and IEC 60601-1-2. Biocompatibility per ISO 10993-1. Software-controlled measurement of apparent cortical bone thickness.
Indications for Use
Indicated for patients requiring assessment of cortical bone thickness at the proximal tibia to aid physicians in diagnosing conditions leading to reduced bone strength and determining fracture risk.
Regulatory Classification
Identification
A bone sonometer is a device that transmits ultrasound energy into the human body to measure acoustic properties of bone that indicate overall bone health and fracture risk. The primary components of the device are a voltage generator, a transmitting transducer, a receiving transducer, and hardware and software for reception and processing of the received ultrasonic signal.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Bone Sonometers.” See § 892.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9, 2017
Bone Index Finland, Ltd. % Janne Karjalainen Chief Technology Officer (CTO) Microkatu 1 Kuopio 70211 FINLAND
Re: K161971 Trade/Device Name: Bindex BI-2 Regulation Number: 21 CFR 892.1180 Regulation Name: Bone sonometer Regulatory Class: II Product Code: MUA Dated: July 12, 2016 Received: December 15, 2016
Dear Janne Karjalainen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161971
Device Name Bindex BI-2
Indications for Use (Describe)
Bindex measures apparent cortical bone thickness at the proximal tibia and can be used in conjunction with other clinical risk factors or patient characteristics as an aid to the physician in the diagnosis and other medical conditions leading to reduced bone strength and in the determination of fracture risk.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### SUBMITTER l.
Bone Index Finland, Ltd. Microkatu 1 Kuopio, Finland 70211 Phone: +358 45 896 2650 Contact Person: Janne Karjalainen Date Prepared: 2 Dec, 2016
# II. DEVICE
Name of Device: Bindex BI-2 Common or Usual Name: Bone sonometer Classification Name: Bone sonometer (21 CFR 892.1180) Regulatory Class: II Product Code: MUA
## III. PREDICATE DEVICE
Bindex BI-100 bone sonometer, K152020.
This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
# IV. DEVICE DESCRIPTION
The Bindex Bl-2 system consists of handheld ultrasound transducer and software. Bindex Bl-2 is connected to the USB port of a computer and controlled with computer software. Bindex Bl-2 is used for measurement of cortical bone thickness and it provides Density Index (DI), a parameter which estimates bensity at the hip as measured with DXA. For measurements, gel is applied on skin and ultrasound transducer is manually placed on the measurement location. Standardized measurement location is at proximal tibia (1/3 length of tibia). Transducer is manually oriented perpendicularly to the cortical bone to achieve accepted measurement. Measurement is repeated five times at each measurement location. Finally, transducer is disinfected by wiping gel off with isopropyl alcohol moistened cloth.
The associated accessories include:
- Measurement stick
- Ultrasound gel (optional)
#### V. INDICATIONS FOR USE
Bindex measures apparent cortical bone thickness at the proximal tibia and can be used in conjunction with other clinical risk factors or patient characteristics as an aid to the physician in the diagnosis and other medical conditions leading to reduced bone strength and in the determination of fracture risk.
### Indications for use discussion
The Indications for Use statement for the Bindex Bl-2 is identical to the predicate device; thus raises no additional questions on safety and effectiveness.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE
The predicate and subject devices are based similar ultrasound technology and are based on the same measurement principle. The main differences between the devices are structural in terms mechanics and electronics. Importantly, the transducer (responsible for conversion of electrical energy to kinetic mechanical energy), is the same for both devices.
The predicate and subject devices are compared in table 1.
Table 1. Device comparison table.
| # | Item | Bindex BI-100 (Predicate) | Bindex BI-2 (subject) |
|---|---------------------|------------------------------------------------------------------------|-----------------------|
| 1 | Indications for use | Bindex measures apparent cortical bone thickness at the proximal tibia | Same |
{5}------------------------------------------------
| | | and can be used in conjunction with<br>other clinical risk factors or patient<br>characteristics as an aid to the<br>physician in the diagnosis of<br>osteoporosis and other medical<br>conditions leading to reduced bone<br>strength and in the determination<br>of fracture risk. | |
|---|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| 2 | Measurement mode | Apparent Cortical Thickness<br>(Ct.Th.). | Same |
| 3 | Probe compatibility | One transducer, centre frequency =<br>3.0 MHz (nominal) | Same |
| 4 | Electrical safety | IEC 60601-2-37:2001 including<br>Amendments 1 and 2; and IEC<br>60601-1:2005 | Same |
| 5 | Electromagnetic<br>compatibility | Complies with IEC 60601-1-2:2007 | Same |
| 6 | Power supply | PC USB port powered. | Same |
| 7 | Operating<br>Environment | Temperature: +10°C to +40°C<br>Humidity: 5% to 85% RH, non-<br>condensing<br>Atmospheric Pressure: 600 to 1060<br>Pc | Same |
| 8 | Storage Environment | Temperature: +10°C to +40°C<br>Humidity: 5% to 85% RH, non-<br>condensing | Same |
| 9 | Calibration | Calibration before each use. | Same |
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the Bindex Bl-2 device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing along the ISO 10993-1 would recommend the following tests:
- Cytotoxicity
- Sensitization
- Irritation
{6}------------------------------------------------
The Bindex Bl-2 is in skin contact for duration of less than 24 hours. Since the use of Bindex involves very short skin contact (typically less than 10 minutes) on a healthy skin and therefore poses a very low risk. Safety of the manufacturing and used materials has been further discussed in Biocompatibility report (Section 13).
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the measurement system, consisting of the Bindex Bl-2 and a laptop computer (on battery use or connected to power supply with isolation transformer). The system complies with the IEC 60601-1, and IEC 60601-2-37 standards for safety and the IEC 60601-1-2 standard for EMC.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
The software for this device was considered as a "moderate" level of concern.
## Mechanical and acoustic testing
The acoustic output and mechanical properties have been tested along the principles presented in harmonized standards IEC 62359 and IEC 60601-1:
- . Acoustic Testing has been conducted on the predicate device and test documentation in this submission shows equivalence of the devices on behalf of acoustic output.
- Drop test (hand piece dropped from 1m on hard surface).
- . Ball pressure test (pressed with steel ball 5mm in diameter with force of 20N one hour at a temperature of 75±2°C)
- Moulding stress relief (device placed in circulating air oven at 70 °C for 7 hours and let cool down)
## Clinical Studies
The same clinical and pre-clinical data supports the safe and effective use of Bindex Bl-100, to which substantial equivalence is shown.
The safety and effectiveness of Bindex Bl-2 device has been discussed in clinical evaluation document Section 16.
#### CONCLUSIONS VIII.
{7}------------------------------------------------
The related safety standards are the same for the predicate and subject device. The same clinical evidence applies to both devices. In the present submission, the performance testing comparing the two models shows that the subject device is substantially equivalent to the predicate device and does not raise additional questions on safety or effectiveness.
Date: 12/02/2016
mm / dd / year
signature
Janne Karjalainen, CTO, Bone Index Finland Ltd.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.