← Product Code [MQB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB) · K210619

# SKR 3000 (K210619)

_Konica Minolta, Inc. · MQB · Aug 24, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K210619

## Device Facts

- **Applicant:** Konica Minolta, Inc.
- **Product Code:** [MQB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB.md)
- **Decision Date:** Aug 24, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures. This device is not indicated for use in mammography, fluoroscopy, and angiography applications.

## Device Story

Digital radiography system (SKR 3000/AeroDR 3) captures X-ray images of human anatomy; utilizes planar X-ray detectors (FPDs P-65, P-75) to convert radiation into digital signals; signals processed by Console CS-7 software; output transmitted to filing systems, printers, or displays for diagnostic review by clinicians. System supports serial radiography up to 40 seconds. Used in clinical settings; operated by trained personnel. CS-7 performs image enhancements including gradation, frequency processing, dynamic range compression, and Intelligent-Grid scatter correction. Antibacterial surface material on detectors. Wired/wireless connectivity via Ethernet or IEEE 802.11a/n (WPA2-PSK encryption). Benefits include replacement of film/screen systems with digital workflow, improved image processing, and extended serial acquisition for dynamic object observation.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation testing performed per FDA guidance for Solid State X-ray Imaging Devices and internal risk analysis; results met all predetermined acceptance criteria.

## Technological Characteristics

Indirect conversion digital radiography system. Detectors: CsI scintillator, 100/200/400 µm pixel size. Connectivity: Wired Ethernet, Wireless LAN (IEEE 802.11a/n, WPA2-PSK). Materials: Surface infused with silver ions (antibacterial). Dimensions: P-65 (384x460x15mm), P-75 (460x460x15mm). IPX6 rating. Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, ISO 10993-1.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- SKR 3000 ([K182688](/device/K182688.md))

## Reference Devices

- Intelligent-Grid ([K151465](/device/K151465.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 24, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Konica Minolta, Inc. % Mr. Jan Maniscalco QA Manager Konica Minolta Healthcare Americas, Inc. 2217 US Highway 70 E GARNER NC 27529

### Re: K210619

Trade/Device Name: SKR 3000 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB, LLZ Dated: July 16, 2021 Received: July 20, 2021

### Dear Mr. Maniscalco:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

. for

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210619

Device Name SKR 3000

#### Indications for Use (Describe)

This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures.

This device is not indicated for use in mammography, fluoroscopy, and angiography applications.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the Konica Minolta logo. The logo features a blue circle with several horizontal white lines across the middle. Below the circle is the text "KONICA MINOLTA" in a bold, sans-serif font.

# 510(k) Summary K210619

| Company:           | KONICA MINOLTA, INC.<br>1 Sakura-machi, Hino-shi, 191-8511 Japan                                                                                                                     |  |  |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Contact:           | Tsutomu Fukui<br>Senior Manager of Regulatory & QMS Division<br>1 Sakura-machi, Hino-shi, 191-8511 Japan<br>Telephone: +81 42 589 8429<br>Email: tsutomu.fukui1@konicaminolta.com    |  |  |
| Date Prepared:     | August 23, 2021                                                                                                                                                                      |  |  |
| Device Name:       | SKR 3000                                                                                                                                                                             |  |  |
| Common Name:       | Digital Radiography                                                                                                                                                                  |  |  |
| Regulation Number: | 21 CFR 892.1680                                                                                                                                                                      |  |  |
| Regulation Name:   | Stationary x-ray system                                                                                                                                                              |  |  |
| Regulatory Class:  | Class II                                                                                                                                                                             |  |  |
| Product Code(s):   | MQB, LLZ                                                                                                                                                                             |  |  |
| Predicate Device:  | K182688 - SKR 3000 (KONICA MINOLTA, INC.)<br>Regulation Number: 21 CFR 892.1680<br>Regulation Name: Stationary x-ray system<br>Regulatory Class: Class II<br>Product Codes: MQB, LLZ |  |  |

# Device Description

The digital radiography SKR 3000 performs X-ray imaging of the human body using an Xray planar detector that outputs a digital signal, which is then input into an image processing device, and the acquired image is then transmitted to a filing system, printer, and image display device as diagnostic image data.

- The subject device SKR3000 is not intended for use in mammography
- This device is also used for carrying out exposures on children.

- Page 1 of 5 -

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the Konica Minolta logo. The logo features a blue circle with several horizontal white lines running across it. Below the circle, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

The Console CS-7, which controls the receiving, processing, and output of image data, is required for operation. CS-7 implements the following image processing; gradation processing, frequency processing, dynamic range compression, smoothing, rotation, reversing, zooming, and grid removal process/scattered radiation - correction (Intelligent-Grid). The Intelligent-Grid is cleared in K151465.

The proposed SKR 3000 is modified to consist of new FPD P-65 and P-75 in addition to previously cleared P-61, P-71, and P-81, Console CS-7 and other peripherals. The DR Detector uses the exposure signal or exposure from the X-ray device to generate X-ray digital image data for diagnosis, including serial exposure images, and send to the image processing controller.

The operator console software, Console CS-7, is a software program for installation on a OTC PC. Software module modifications have been made to use new FPDs (P-65 and P-75) (Cassette Type Detection Software (CTDS)) and to support 40 seconds serial radiography (SIC).

The FPDs used in SKR 3000 can communicate with the image processing device through the wired Ethernet and/or the Wireless LAN (IEEE802.11a/n and FCC compliant). The WPA2-PSK (AES) encryption is adopted for a security of wireless connection.

The new DR panels, P-65 and P-75, employ the surface material containing antibacterial agent in both radiation and irradiation sides. In the serial radiography settings, acquisition time has been changed from up to 20 seconds to 40 seconds to observe a variety of dynamic objects. Other control parameters of serial radiography are not changed from the predicate device.

The SKR 3000 is distributed under a commercial name AeroDR 3.

## Indications for Use

This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures.

This device is not indicated for use in mammography, fluoroscopy, and angiography applications.

- Page 2 of 5 -

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue circle at the top with several horizontal white lines running across it. Below the circle, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

# Comparison Table

The comparison to the predicate device was summarized in the table blow.

|                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                             |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                                                   | SKR 3000                                                                                                                                                                                                                                                                                                                                                                                                                     | SKR 3000                                                                                                                                                                                                                                                                                                                                                                                                                     |
| 510(K) Number                                                                 | K210619                                                                                                                                                                                                                                                                                                                                                                                                                      | K182688                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Indications for Use                                                           | This device is indicated for use in<br>generating radiographic images of<br>human anatomy. It is intended to replace<br>radiographic film/screen<br>system in general-purpose diagnostic<br>procedures.<br>This device is not indicated for use in<br>mammography, fluoroscopy, and<br>angiography applications.                                                                                                             | The SKR 3000 is indicated for use in<br>generating radiographic images of<br>human anatomy.     It is intended to<br>replace radiographic film/screen<br>system in general-purpose diagnostic<br>procedures.<br>The SKR 3000 is not indicated for use<br>in mammography, fluoroscopy and<br>angiography applications.                                                                                                        |
| Specification<br>Detection method<br>Scintillator<br>Image area size          | Indirect conversion method<br>Csl (Cesium Iodide)<br>P-65: 348.8×425.6mm<br>(3,488×4,256 pixels)<br>P-75: 424.8×424.8mm<br>(4,248×4,248 pixels)                                                                                                                                                                                                                                                                              | Indirect conversion method<br>Csl (Cesium Iodide)<br>P-61: 348.8×425.6mm<br>(3,488×4,256 pixels)<br>P-71: 424.8×424.8mm<br>(4,248×4,248 pixels)<br>P-81: 245.6×296.8mm<br>(2,456×2,968 pixels)                                                                                                                                                                                                                               |
| Pixel size                                                                    | 100 µm / 200 µm / 400 µm                                                                                                                                                                                                                                                                                                                                                                                                     | 100 µm / 200 µm / 400 µm                                                                                                                                                                                                                                                                                                                                                                                                     |
| A/D conversion<br>MTF(1.0 cycle/mm)<br>DQE (1.0 cycle/mm)<br>DQE (0 cycle/mm) | 16 bit (65,536 gradients)<br>(Non-binning) 0.62<br>(2x2 binning) 0.58<br>56% @ 1mR<br>65% @ 0.02mR                                                                                                                                                                                                                                                                                                                           | 16 bit (65,536 gradients)<br>(Non-binning) 0.62<br>(2x2 binning) 0.58<br>56% @ 1mR<br>65% @ 0.02mR                                                                                                                                                                                                                                                                                                                           |
| Mechanical<br>External dimensions<br>IPX                                      | P-65: 384(W)×460(D)×15(H)mm<br>P-75: 460(W)×460(D)×15(H)mm<br>IPX6                                                                                                                                                                                                                                                                                                                                                           | P-61: 384(W)×460(D)×15(H)mm<br>P-71: 460(W)×460(D)×15(H)mm<br>P-81: 282(W)×333(D)×15(H)mm<br>IPX6                                                                                                                                                                                                                                                                                                                            |
| Battery<br>(Lithium-ion capacitor)<br>Battery duration in<br>standby status   | P-65: Approx. 13.2 hours<br>P-75: Approx. 12.2 hours                                                                                                                                                                                                                                                                                                                                                                         | P-61: Approx. 13.2 hours<br>P-71: Approx. 12.2 hours<br>P-81: Approx. 6.3 hours                                                                                                                                                                                                                                                                                                                                              |
| Surface Material                                                              | Surface infused with Silver ions<br>(antibacterial properties)                                                                                                                                                                                                                                                                                                                                                               | No antibacterial agent                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                             |
| Communication I/F                                                             | Wired and Wireless                                                                                                                                                                                                                                                                                                                                                                                                           | Wired and Wireless                                                                                                                                                                                                                                                                                                                                                                                                           |
| Peripherals,<br>Cables/minor<br>components                                    | AeroDR Interface Units, Detector<br>Interface Units, Power Supply Unit,<br>Generator Interface Units, Battery<br>Charger, etc.                                                                                                                                                                                                                                                                                               | AeroDR Interface Units, Detector<br>Interface Units, Power Supply Unit,<br>Generator Interface Units, Battery<br>Charger, etc.                                                                                                                                                                                                                                                                                               |
| Operator console<br>(Software)                                                | CS-7<br>- Software module modification to use<br>new FPDs (Cassette Type Detection<br>Software (CTDS))<br>- Software module modification to<br>support 40 seconds serial radiography<br>(SIC)                                                                                                                                                                                                                                | CS-7                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Image Processing                                                              | Auto-gradation process<br>Frequency processing (F process)<br>Equalization processing (E process)<br>Hybrid processing (HF process - HE<br>process)<br>Hybrid smoothing process (HS process)<br>REALISM processing (RE process - RF<br>process)<br>Realism smoothing process (RS<br>process)<br>Grid removal process/Scattered<br>Radiation Correction (Intelligent-Grid)<br>Automatic exposure field recognition<br>process | Auto-gradation process<br>Frequency processing (F process)<br>Equalization processing (E process)<br>Hybrid processing (HF process - HE<br>process)<br>Hybrid smoothing process (HS process)<br>REALISM processing (RE process - RF<br>process)<br>Realism smoothing process (RS<br>process)<br>Grid removal process/Scattered<br>Radiation Correction (Intelligent-Grid)<br>Automatic exposure field recognition<br>process |
| Serial radiography                                                            | Max. acquisition time: 40 seconds                                                                                                                                                                                                                                                                                                                                                                                            | Max. acquisition time: 20 seconds                                                                                                                                                                                                                                                                                                                                                                                            |

— Page 3 of 5 ————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue globe-like shape with horizontal white lines running across it. Below the graphic is the text "KONICA MINOLTA" in a bold, sans-serif font.

# Performance Data

The SKR 3000 is designed to comply with the following standards; AAMI/ANSI ES 60601-1 (Ed.3.1), IEC 60601-1-2 (Ed.4.0), and ISO 10993-1 (2018). The performance tests according to the "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices" and the other verification and validation including the items required by the risk analysis for the SKR3000 were performed and the results demonstrated that the predetermined acceptance criteria were met. The results of risk management did not require clinical studies to demonstrate the substantial equivalency of the proposed device.

Page 4 of 5

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the Konica Minolta logo. The logo features a blue globe-like shape with several horizontal white lines running across it. Below the globe, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

## Conclusion

The SKR 3000 has the same intended use and indications for use, technological characteristics, and principal operations. The technological differences raised no new issues of safety or effectiveness as compared to its predicate device (K182688). Performance tests demonstrate that the SKR 3000 performs according to specifications and functions as intended. All the information to demonstrate assurance of our evaluation is attached to relevant sections of this submission.

Therefore, as for our conclusion, the SKR 3000 is substantially equivalent to the predicate devices and presents no new questions of safety or effectiveness.

Page 5 of 5 -

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K210619](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K210619)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
