← Product Code [MQB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB) · K203849

# General Radiography CXDI-CS01 (K203849)

_Canon, Inc. · MQB · Jan 26, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K203849

## Device Facts

- **Applicant:** Canon, Inc.
- **Product Code:** [MQB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB.md)
- **Decision Date:** Jan 26, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The DIGITAL RADIOGRAPHY CXDI-CS01 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

## Device Story

Solid-state X-ray detector series; intercepts X-ray photons; scintillator (CsI(Tl)) emits visible light; photodetector array converts light to electrical signals. Signals transmitted via wired/wireless connection to PC; CXDI Control Software (CCS) converts signals to images for display. Multi Box (standard component) synchronizes X-ray exposure, supplies power, manages communication between FPD, image capture computer, and WLAN access point. Used in clinical settings by healthcare professionals. Output enables diagnostic visualization of human anatomy; assists clinical decision-making by replacing traditional film/screen systems.

## Clinical Evidence

Bench testing only. Verification and validation activities demonstrated that modifications to the Multi Box, X-ray I/F unit, and software did not impact compliance with FDA guidance for Solid State X-ray Imaging Devices or relevant IEC safety/performance standards (60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 62366, 60601-2-54, 62220-1, 62304).

## Technological Characteristics

Solid-state flat panel X-ray detector; CsI(Tl) scintillator; 125μm pixel pitch; 35% MTF@2lp/mm. Case materials: Magnesium alloy or Carbon Fiber. Connectivity: Wired/Wireless. Software: CXDI Control Software V2.19. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 62366, 60601-2-54, 62220-1, 62304.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- DIGITAL RADIOGRAPHY CXDI-702C Wireless ([K192632](/device/K192632.md))
- CXDI-402C Wireless ([K192632](/device/K192632.md))

## Reference Devices

- DIGITAL RADIOGRAPHY CXDI-410C Wireless ([K171270](/device/K171270.md))
- DIGITAL RADIOGRAPHY CXDI-810C Wireless ([K170332](/device/K170332.md))
- DIGITAL RADIOGRAPHY CXDI-710C Wireless ([K170332](/device/K170332.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 26, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Canon Inc. % Mr. Gregory Woodard Biomedical Engineer Ken Block Consulting, LLC 800 East Campbell Road, Suite 202 RICHARDSON TX 75081

Re: K203849

Trade/Device Name: Digital Radiography CXDI-CS01 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: December 30, 2020 Received: December 31, 2020

Dear Mr. Woodard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K203849

Device Name DIGITAL RADIOGRAPHY CXDI-CS01

Indications for Use (Describe)

The DIGITAL RADIOGRAPHY CXDI-CS01 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

Type of Use (Select one or both, as applicable):

| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Canon" in red font. The font is bold and slightly italicized. The word is easily readable and stands out against the white background.

## 510(k) SUMMARY

K203849

|                    | K203849                                                                                                                  |                                                                                                                                                                                                                        |
|--------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:         | Canon Inc.<br>30-2 Shimomaruko, 3-chrome<br>Ohta-ku, Tokyo 146-8501 Japan                                                |                                                                                                                                                                                                                        |
| Contact Person:    | Mr. Akira Hirai<br>General Manager<br>TEL: 81-3-3758-2111;<br>FAX: 044-739-6695<br>hirai.akira@mail.canon                |                                                                                                                                                                                                                        |
| Date Prepared:     | 12/30/2020                                                                                                               |                                                                                                                                                                                                                        |
| Submission Type:   | Special 510(k) Submission                                                                                                |                                                                                                                                                                                                                        |
| Proposed Device:   | Manufacturer:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Classification:<br>Product Code:                 | Canon Inc.<br>DIGITAL RADIOGRAPHY CXDI-CS01<br>Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Stationary X-ray System<br>892.1680, Stationary X-ray System<br>MQB                                             |
| Predicate Device:  | 510k Number:<br>Manufacturer:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Classification:<br>Product Code: | K192632<br>Canon, Inc.<br>DIGITAL RADIOGRAPHY CXDI-702C Wireless,<br>CXDI-402C Wireless<br>Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Stationary X-ray System<br>892.1680, Stationary X-ray System<br>MQB |
| Reference Devices: | 510k Number:<br>Manufacturer:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Classification:<br>Product Code: | K171270<br>Canon, Inc.<br>DIGITAL RADIOGRAPHY CXDI-410C Wireless<br>Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Stationary X-ray System<br>892.1680, Stationary X-ray System<br>MQB                        |
|                    | 510k Number:<br>Manufacturer:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Classification:                  | K170332<br>Canon, Inc.<br>DIGITAL RADIOGRAPHY CXDI-810C Wireless,<br>CXDI-710C Wireless<br>Solid State X-ray Imager<br>Stationary X-ray System<br>892.1680, Stationary X-ray System                                    |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Canon" in red font. The font is bold and slightly italicized. The background is white. 

Product Code: MQB

Device Description: The DIGITAL RADIOGRAPHY CXDI-CS01 is a series of solid-state x-ray detectors. The detector intercepts x-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photodetectors that create electrical signals. After the electrical signals are generated, the digital values are sent to the PC via wired or wireless connection, converted to images with the CXDI Control Software (CCS), then displayed on monitors. The digital value can be communicated to the operator console via wired or wireless connection.

> The subject of this Special 510(k) submission is a change to the DIGITAL RADIOGRAPHY CXDI-702C Wireless and CXDI-402C Wireless to make a series of detectors including the DIGITAL RADIOGRAPHY CXDI-702C Wireless (K192632), CXDI-402C Wireless (K192632), CXDI-710C Wireless (K170332), CXDI-810C Wireless (K170332), and CXDI-410C Wireless (K171270), under the proposed device the DIGITAL RADIOGRAPHY CXDI-CS01. This change will change the Multi Box to be a Standard Component from an optional component for the detectors. The Multi Box has the functions of connecting to the X-ray generator to synchronize the X-ray exposure and photographing, supplying power to the FPD (front panel detector), communicating between the FPD and the image capture computer, connecting the status indicator to the Multi Box to turn the FPD on and off and switch to a ready state, connecting to the WLAN access point to communicate to the FPD wirelessly, and connecting to the docking station to charge the FPD and communicate between the FPD and image capture computer. These functions of the Multi Box have not changed since its clearance under the predicate device (K192632). The X-ray I/F (Interface) Unit has been removed as an optional component for the DIGITAL RADIOGRAPHY CXDI-702C Wireless and CXDI-402C Wireless as the Multi Box performs the function of the X-ray I/F Unit (connecting to the X-ray generator to synchronize the x-ray exposure and photographing). In addition, the detectors will all use the version software cleared under the DIGITAL RADIOGRAPHY CXDI-702C Wireless and CXDI-402C Wireless (K192632). Together, these changes make up the DIGITAL RADIOGRAPHY CXDI-CS01.

Indication for Use: The Indication for Use statement is identical to the predicate device. The intended use of the modified device, as described in the labeling, has not changed as a result of the modification(s).

> The DIGITAL RADIOGRAPHY CXDI-CS01 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This device is not intended for mammography applications.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the word "Canon" in a bold, red font. The word is slightly stylized, with a unique design for each letter. The background is plain white, which makes the red text stand out.

Summary of Technological Characteristics: Comparisons with the predicate devices show the characteristics of the proposed modifications (making the Multi Box a standard component, removing the X-ray I/F Interface as an option, and updating the CXDI Control Software to version 2.19) to the DIGITAL RADIOGRAPHY CXDI-CS01 to be substantially equivalent to the predicate device.

|                                 | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Reference Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |           |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Trade Name                      | CXDI-CS01                                                                                                                                                                                                                                                                                                                                                                                                          | CXDI-702C Wireless /<br>CXDI-402C Wireless                                                                                                                                                                                                                                                                                                                                                                                                                                      | CXDI-710C Wireless / CXDI-<br>810C Wireless / CXDI-410C<br>Wireless                                                                                                                                                                                                                                                                                                                                                                                                                                                     |           |
| 510(k)<br>Submitter<br>[Number] | Canon, Inc.<br>[TBD]                                                                                                                                                                                                                                                                                                                                                                                               | Canon, Inc.<br>[K192632]                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Canon, Inc.<br>[K170332]<br>[K171270]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | IDENTICAL |
| Indication for<br>Use           | The DIGITAL<br>RADIOGRAPHY CXDI-<br>CS01 provides digital image<br>capture for conventional<br>film/screen radiographic<br>examinations. This device is<br>intended to capture, for<br>display, radiographic images<br>of human anatomy, and to<br>replace radiographic<br>film/screen systems in all<br>general purpose diagnostic<br>procedures. This device is<br>not intended for<br>mammography applications. | The DIGITAL<br>RADIOGRAPHY CXDI-<br>402C Wireless and<br>DIGITAL RADIOGRAPHY<br>CXDI-702C Wireless<br>provides digital image<br>capture for conventional<br>film/screen radiographic<br>examinations. This device is<br>intended to capture, for<br>display, radiographic images<br>of human anatomy, and to<br>replace radiographic<br>film/screen systems in all<br>general purpose diagnostic<br>procedures. This device is<br>not intended for<br>mammography applications. | The DIGITAL<br>RADIOGRAPHY CXDI-710C<br>Wireless, DIGITAL<br>RADIOGRAPHY CXDI-810C<br>Wireless, and DIGITAL<br>RADIOGRAPHY CXDI-410C<br>Wireless provides digital image<br>capture for conventional<br>film/screen radiographic<br>examinations. This device is<br>intended to capture, for display,<br>radiographic images of human<br>anatomy, and to replace<br>radiographic film/screen<br>systems in all general purpose<br>diagnostic procedures. This<br>device is not intended for<br>mammography applications. | IDENTICAL |
| Application                     | General Radiography                                                                                                                                                                                                                                                                                                                                                                                                | General Radiography                                                                                                                                                                                                                                                                                                                                                                                                                                                             | General Radiography                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | IDENTICAL |
| Case Material                   | [Detector of CXDI-702C<br>Wireless and CXDI-402C<br>Wireless]<br>Magnesium alloy<br>[Detector of CXDI-710C<br>Wireless. CXDI-810C<br>Wireless and CXDI-410C<br>Wireless]<br>Carbon Fiber                                                                                                                                                                                                                           | Magnesium alloy                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Carbon Fiber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | IDENTICAL |
| Scintillator                    | CsI(Tl)<br>[Cesium Iodide doped with<br>Thallium]                                                                                                                                                                                                                                                                                                                                                                  | CsI(TI)<br>[Cesium Iodide doped with<br>Thallium]                                                                                                                                                                                                                                                                                                                                                                                                                               | CsI(Tl)<br>[Cesium Iodide doped with<br>Thallium]                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | IDENTICAL |
| Pixel Pitch                     | 125μm                                                                                                                                                                                                                                                                                                                                                                                                              | 125μm                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 125μm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | IDENTICAL |
| Spatial<br>Resolution           | 35% [MTF@2lp/mm]                                                                                                                                                                                                                                                                                                                                                                                                   | 35% [MTF@2lp/mm]                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 35% [MTF@2lp/mm]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | IDENTICAL |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Canon" in red font. The font is bold and slightly blurred. The background is white.

Performance:

|                        | Proposed Device                                                                                                                                                                                                                                                                                                       | Predicate Device                                                                                                                                               | Reference Devices                                                                                                                           |           |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Trade Name             | CXDI-CS01                                                                                                                                                                                                                                                                                                             | CXDI-702C Wireless /<br>CXDI-402C Wireless                                                                                                                     | CXDI-710C Wireless / CXDI-<br>810C Wireless / CXDI-410C<br>Wireless                                                                         |           |
| IP Level               | [Detector of CXDI-702C<br>Wireless and CXDI-402C<br>Wireless]<br>IP54<br>[Detector of CXDI-710C<br>Wireless, CXDI-810C<br>Wireless and CXDI-410C<br>Wireless]<br>IPX7                                                                                                                                                 | IP54                                                                                                                                                           | IPX7                                                                                                                                        | IDENTICAL |
| Standard<br>Components | Detectors (Sensors), Battery<br>Pack,<br>Documentation (Manuals),<br>Multi Box                                                                                                                                                                                                                                        | Detector (Sensor), Battery<br>Pack,<br>Documentation (Manuals)                                                                                                 | Detector (Sensor), Battery Pack,<br>Documentation (Manuals)                                                                                 | MODIFIED  |
| Options                | Docking Station, Battery<br>Charger, Battery Pack,<br>Status Indicator, Wiring<br>Cable, PC Connection<br>Cable, Ready Indicator                                                                                                                                                                                      | Docking Station, Battery<br>Charger, Battery Pack,<br>Multi Box, Status Indicator,<br>Wiring Cable, PC<br>Connection Cable, Ready<br>Indicator, X-ray I/F unit | Docking Station, Battery<br>Charger, Battery Pack, Multi<br>Box, Status Indicator, Wiring<br>Cable, PC Connection Cable,<br>Ready Indicator | MODIFIED  |
| Software               | CXDI Control Software<br>V2.19                                                                                                                                                                                                                                                                                        | CXDI Control Software<br>V2.19                                                                                                                                 | CXDI Control Software V2.16                                                                                                                 | MODIFIED  |
| Photographing<br>Mode  | [Detector of CXDI-702C<br>Wireless and CXDI-402C<br>Wireless]<br>Standard Synchronization<br>Mode, Non Generator<br>Connection Mode<br>[Detector of CXDI-710C<br>Wireless, CXDI-810C<br>Wireless and CXDI-410C<br>Wireless]<br>Standard Synchronization<br>Mode, Non Generator<br>Connection Mode,<br>Standalone Mode | Standard Synchronization<br>Mode, Non Generator<br>Connection Mode                                                                                             | Standard Synchronization Mode,<br>Non Generator Connection<br>Mode, Standalone Mode                                                         | IDENTICAL |

The User's and Installation Manuals provide detailed instructions and information for safe and effective use of the device and users are expected to adhere to the instructions and other information. The User's Manual explains how to use the detector and other equipment. Connected medical equipment, such as X-ray generators, must comply with IEC 60601-1. Before using the product, be sure to read the manual thoroughly in order to utilize it more effectively. The fundamental scientific technology of the detectors under DIGITAL RADIOGRAPHY CXDI-CS01 have not been modified. The detector units of the DIGITAL RADIOGRAPHY CXDI-CS01 has not been modified, and the change is to update the software in the DIGITAL RADIOGRAPHY CXDI-710C Wireless, CXDI-810C Wireless, and CXDI-410C Wireless; make the Multi Box a standard component; and remove the X-ray I/F Unit as an optional component.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Canon" in a bold, red font. The letters are slightly slanted, giving the word a dynamic appearance. The logo is simple and recognizable, representing the well-known electronics and imaging company.

| Documentation was provided demonstrating that the changes to DIGITAL<br>RADIOGRAPHY CXDI-702C Wireless and CXDI-402C Wireless do not impact<br>the device's compliance with FDA requirements stated in “Guidance for the<br>Submission of 510(k)'s for Solid State X-ray Imaging Devices”.                                                                                                                                                                                  |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Evaluation of the changes to the DIGITAL RADIOGRAPHY CXDI-CS01 confirmed<br>that the changes did not impact DIGITAL RADIOGRAPHY CXDI-CS01<br>conformance with the U.S. Performance Standard for radiographic equipment and<br>with relevant voluntary safety standards for Electrical safety and Electromagnetic<br>Compatibility testing, specifically IEC standards 60601-1, 60601-1-2, 60601-1-3,<br>60601-1-6, 62366, 60601-2-54, 62220-1, and 62304.                   |
| These verification/validation activities successfully demonstrated that the device<br>continues to meet the standards for the areas impacted by the device modifications<br>to the predicate device and raises no new questions regarding either safety or<br>effectiveness when compared to the predicate device. Therefore, the<br>verification/validation conducted supports a determination of substantial<br>equivalence for the DIGITAL RADIOGRAPHY CXDI-CS01 device. |
| Conclusion:                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Canon Inc. considers the DIGITAL RADIOGRAPHY CXDI-CS01 to be substantially<br>equivalent to the predicate device listed above. This conclusion is based on the<br>similarities in primary intended use, principles of operation, functional design, and<br>established medical use.                                                                                                                                                                                         |

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K203849](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K203849)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
