K193173 · Careray Digital Medical System Co., Ltd. · MQB · Dec 13, 2019 · Radiology
Device Facts
Record ID
K193173
Device Name
CareView 1800Le
Applicant
Careray Digital Medical System Co., Ltd.
Product Code
MQB · Radiology
Decision Date
Dec 13, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The CareView 1800Le detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This product is not intended for mammography applications.
Device Story
CareView 1800Le is a digital X-ray flat panel detector (430mm x 430mm) used to replace film/screen systems. Device intercepts X-ray photons via CsI scintillator; converts to visible light; illuminates a-Si thin-film transistor array to generate electrical signals; signals digitized for monitor display. Used in clinical settings; operated by radiology staff. Connected via Gigabit Ethernet to external computer and X-ray generator. Provides digital images for diagnostic review by physicians. Benefits include improved imaging resolution (140μm pixel pitch) compared to predicate, facilitating general radiographic diagnosis.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including electrical safety, EMC, and performance metrics (MTF, DQE, spatial resolution) compared to the predicate device.
Technological Characteristics
CsI scintillator; a-Si TFT readout; 140μm pixel pitch; 3072 x 3072 image matrix; 16-bit grayscale; Gigabit Ethernet connectivity; carbon fiber plate; air-cooled; DC +24V power; wired cassette form factor (460x460x15mm). Complies with IEC/ES 60601-1 and IEC/EN 60601-1-2.
Indications for Use
Indicated for general radiographic diagnosis of human anatomy in patients requiring general-purpose diagnostic procedures. Not intended for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
December 13, 2019
CareRay Digital Medical Technology Co., Ltd. % Ms. Leilei Li RA Manager A2-201/B3-501, Biobay, 218 Xinghu Street, SuZhou Industrial Park Suzhou, Jiangsu 215123 P. R. CHINA
Re: K193173
Trade/Device Name: X-ray Flat Panel Detectors (CareView 1800Le) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: November 19, 2019 Received: November 22, 2019
Dear Ms. Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K193173
Device Name
X-ray Flat Panel Detectors (CareView 1800Le)
Indications for Use (Describe)
The Care View 1800Le detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This product is not intended for mammography applications.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 006_510(k) Summary
K193173
Page 1 of 6
{4}------------------------------------------------
# 510(k) Summary
K193173
[As required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR807.92 (a) (1)]
Nov. 04, 2019
#### 2. Submitter's Information [21 CFR807.92 (a) (1)]
| Company Name: | CareRay Digital Medical Technology Co., Ltd. |
|------------------|-------------------------------------------------------------------------------------------------|
| Company Address: | A2-201/B3-501, Biobay, 218 Xinghu Street, Suzhou<br>Industrial Park, Suzhou 215123, P. R. China |
| Contact Person: | Ms. Li |
| Phone Number: | (86) 512-86860288 |
| Fax Number: | (86) 512-86860388 |
| E-mail: | ll.li@careray.com |
#### 3. Trade Name, Common Name, Classification [21 CFR807.92(a)(2)]
| Trade Name: | X-ray Flat Panel Detectors |
|----------------------|----------------------------|
| Model Name: | CareView 1800Le |
| Classification Name: | Stationary X-ray system |
| Regulation Number: | 21 CFR 892.1680 |
| Regulatory Class: | Class II |
| Product Code: | MQB |
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicates within this submission are as follows:
| Manufacturer: | CareRay Digital Medical Technology Co., Ltd. |
|----------------------|----------------------------------------------|
| Trade Name: | X-ray Flat Panel Detectors |
| Model Name: | CareView 1800L |
| Classification Name: | Stationary X-ray system |
| Regulation Number: | 21 CFR 892.1680 |
{5}------------------------------------------------
| Regulatory Class: | Class II |
|-------------------|----------|
| Product Code: | MQB |
| FDA 510(k) #: | K153492 |
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
The CareView 1800Le detector is a class of digital X-ray flat panel detector that has an imaging area of 430mm×430mm. The detector communicates by a wired connection (Giga-bit Ethernet communication mode via connecting the power box).
The detector functions by intercepting X-ray photons. Then the scintillator emits visible spectrum photons that illuminate an array of photo detectors (a-Si) that create electrical signals. The electrical signals are then digitally converted to display an image on the monitor.
The detector should be connected to a computer and X-ray generator to digitize X-ray images and transfer radiography diagnostics. The x-ray generator, an essential part of a full x-ray system, is not part of the subject medical device.
The software that supports the functions of the digital detector CareView 1800Le is unchanged from the predicate.
Generally, CareView 1800Le is the same as the cleared product, CareView 1800L except the
pixel size. The pixel size of CareView 1800Le is 140 μm, while the CareView 1800L is 154 μm.
## 6. Intended Use [21 CFR 807.92(a)(5)]
The CareView 1800Le detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This product is not intended for mammography applications.
#### 7. Technological Characteristics [21 CFR 807.92(a)(6)]
| Item | Proposed Device:<br>X-ray Flat Panel Detectors | Predicate Device:<br>X-ray Flat Panel Detectors |
|------------------------|------------------------------------------------|-------------------------------------------------|
| 510(K) Number | To be assigned | K153492 |
| Model | CareView 1800Le | CareView 1800L |
| Classification<br>Name | Stationary X-ray system | Stationary X-ray system |
| Product Code | MQB | MQB |
| Regulation Number | 892.1680 | 892.1680 |
| Panel | Radiology | Radiology |
| Class | II | II |
| X-ray Absorber | CsI Scintillator | CsI Scintillator |
| Installation Type | Wired, Cassette | Wired, Cassette |
{6}------------------------------------------------
| Readout<br>Mechanism | Thin Film Transistor | Thin Film Transistor |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Image Matrix Size | 3072 × 3072 pixels | 2816 × 2816 pixels |
| Pixel Pitch | 140μm | 154μm |
| Effective Imaging<br>Area | 430 mm × 430 mm | 434 mm × 434 mm |
| Grayscale | 16 bit, 65536 grayscale | 16 bit, 65536 grayscale |
| Spatial Resolution | 3.57 line pair/mm | 3.3 line pair/mm |
| MTF | ≥63%..........(@ 1lp/mm)<br>≥35%..........(@ 2lp/mm)<br>≥17%..........(@ 3lp/mm)<br>≥10%..........(@ 3.5lp/mm) | ~70%..........(@ 1lp/mm)<br>~40%..........(@ 2lp/mm)<br>~22%..........(@ 3lp/mm) |
| DQE | (@RQA5, 30μGy)<br>≥62%..........(@ 0lp/mm)<br>≥45%..........(@ 1lp/mm)<br>≥17%..........(@ 3lp/mm)<br>≥10%..........(@ 3.5lp/mm) | (@RQA5, 30μGy)<br>~65%..........(@ 0lp/mm)<br>~20%..........(@ 3lp/mm) |
| Rated<br>Power<br>Supply | DC +24 V, Max.1 A<br>Powered by the power box using<br>interface cable | DC +24 V, Max.1 A<br>Powered by the power box<br>using interface cable |
| Communications | Gigabit Ethernet | Gigabit Ethernet |
| Imaging Plate | Carbon Fiber Plate | Carbon Fiber Plate |
| Cooling | Air cooling | Air cooling |
| Dimensions | 460 mm × 460 mm × 15 mm | 460 mm × 460 mm × 15 mm |
| Operation | Temperature: +5 ~ +35°C | Temperature: +5 ~ +35°C |
| | Humidity: 30 ~ 75% | Humidity: 30 ~ 75% |
| | (Non-Condensing) | (Non-Condensing) |
| | Atmospheric pressure: 700 ~ 1060 | Atmospheric pressure: 700 ~ |
| | hPa | 1060 hPa |
| Storage and<br>transportation | Temperature: -20 ~ +55°C | Temperature: -20 ~ +55°C |
| | Humidity: 10 ~ 90% | Humidity: 10 ~ 90% |
| | (Non-Condensing) | (Non-Condensing) |
| | Atmospheric pressure: 700 ~ 1060 | Atmospheric pressure: 700 ~ |
| | hPa | 1060 hPa |
# 8. System requirements to operate with other radiographic system components
The submitted medical device is the CareView 1800Le digital detector. The other x-ray system components referred below are for information purpose only.
- 1) Recommended Generator Specification:
- Energy range: 40~150kVp
mA range: 10~1000mA (depending on the generator power)
ms range: 10~6300ms to produce 0.1~1000mAs (depending on the generator power)
{7}------------------------------------------------
Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators, please contact your distributor or CareRay.
- 2) Application Program Interface (API) for system integration manufacturer
Peripheral hardware: CareView detector connected via wired communication.
CPU: Intel (R) Core (TM) 2 Duo, 2.93GHz or above
RAM: 6 GB or higher
Hard disk: 80 GB or higher
Monitor: 1280 x 1024 or higher
OS: Windows XP or Windows 7 or Windows 10
- Development environment: MS Visual Studio 2005
- 3) X-ray exposure mode
The synchronous connection mode is the signal transfer mode between the X-ray generator which sends the X-ray and the detector which receives the X-ray.
CareView 1800Le supported typical sync mode contains external sync, soft sync, manual sync and auto sync containing FFAED mode.
The detector can't provide feedback to the generator to terminate the x-ray exposure.
# 9. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92(b)(2)]
- A Electrical safety and EMC testing Electrical, mechanical, environmental safety and performance testing according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC/EN 60601-1-2. All test results are satisfactory.
A Nonclinical and clinical considerations
The proposed device (CareView 1800Le) and predicate device (CareView 1800L) share most of primary product specifications including intended use, technology, material, and imaging principle, power supply method etc. The only difference is the pixel size. While the predicate device pixel size is 154 um, the proposed device pixel size is 140 um. The difference of pixel size doesn't affect the technological parameters and clinical images.
## 10. Conclusion [21 CFR 807.92(b) (3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, the CareView 1800Le X-ray flat panel detector is substantially equivalent to predicate device CareView 1800L(K153492).
{8}------------------------------------------------
Both propose and predicate devices are same in the intended use, the design principle, the applicable standards and specification. Some characteristics, for example, pixel size is different. However the test reports in this submission documents provide demonstration that this difference doesn't raise any new questions of safety and effectiveness. Therefore, CareRay Digital Medical Technology Co., Ltd. concludes the CareView 1800Le X-ray flat panel detector is substantially equivalent with the predicate device CareView 1800L (K153492).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.