← Product Code [MQB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB) · K170480

# Models 3543A, 4343A, 2929A and A843B (K170480)

_Shanghai PZ Medical Technolgoy Co., Ltd. · MQB · Jul 7, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K170480

## Device Facts

- **Applicant:** Shanghai PZ Medical Technolgoy Co., Ltd.
- **Product Code:** [MQB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB.md)
- **Decision Date:** Jul 7, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

Model#'s 3543A, 4343A, 2929A and A843B; Solid State X-ray Imager (Flat Panel/Digital Imager) are indicated for use in generating radiographic images of human anatomy. They are intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding dental, fluoroscopic, angiographic, and mammographic applications. As prescribed by a licensed physician.

## Device Story

Solid state X-ray imagers (flat panel detectors) used in medical X-ray imaging systems; consist of GOS scintillator screen and thin-film transistors (TFTs). Scintillator converts X-rays to visible light; TFTs convert light to electrical charge; system performs analog-to-digital conversion to produce digital images. Operated by trained personnel in medical facilities for single-image diagnosis. Output is a DICOM image viewed by healthcare providers to assist in clinical decision-making for general radiographic procedures. Benefits include replacement of traditional film/screen systems with digital imaging capabilities.

## Clinical Evidence

No clinical trials performed. Substantial equivalence supported by bench testing and comparative image quality analysis using phantom/clinical images against the Atlaim Atal 9 (K132151) predicate.

## Technological Characteristics

Flat panel digital X-ray detector; GOS scintillator; thin-film transistor (TFT) array; 16-bit A/D conversion; pixel pitch 140 µm. Connectivity: Gigabit Ethernet (wired) and IEEE 802.11ac (wireless). Power: AC line or rechargeable lithium battery. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Atlaim's model ATAL 9 ([K152151](/device/K152151.md))
- IRAY model 1417 ([K161730](/device/K161730.md))
- CareRay model Care view 1500C & 1500L ([K153058](/device/K153058.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a symbol above them that resembles a caduceus without the snake.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Shanghai PZ Medical Technology Co., Ltd. % Jeff Moeller Vice President OrganizeUS. Inc. 8317 Lakeside Drive Downers Grove, Illinois 60516

July 7, 2017

Re: K170480

Trade/Device Name: Models 3543A, 4343A, 2929A and A843B Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MOB Dated: April 30, 2017 Received: May 5, 2017

Dear Jeff Moeller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael D. O'Hara
For

Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K170480

Device Name Models 3543A, 4343A, 2929A and A843B

Indications for Use (Describe)

Model#'s 3543A, 4343A, 2929A and A843B; Solid State X-ray Imager (Flat Panel/Digital Imager) are indicated for use in generating radiographic images of human anatomy. They are intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding dental, fluoroscopic, and mammographic applications. As prescribed by a licensed physician

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| X  Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

�rganizeUS, Inc.

8317 Lakeside Drive Downers Grove, IL 60516

# 510(k) Summary

(Revised 06/30/17)

| Owners name:                            | Shanghai PZ Medical Technology Co., Ltd.<br>5F, Building 13 No.2 Suide Road<br>200331 Shanghai                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person / firm: OrganizeUS, Inc. | 8317 Lakeside Drive<br>Downers Grove, IL 60516<br>847-337-3004<br>Jeff Moeller, Vice President                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Date:                                   | March 17, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Trade name:                             | Models 3543A, 4343A, 2929A and A843B                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common name:                            | Solid State X-Ray Imager (Flat Panel/Digital Imager)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification name:                    | Stationary X-Ray system, 892.1680, code MQB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate devices:                      | Atlaim's model ATAL 9 - K152151<br>Stationary X-Ray system, 892.1680, code MQB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                         | IRAY model 1417 - K161730<br>Stationary X-Ray system, 892.1680, code MQB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                         | CareRay model Care view 1500C & 1500L - K153058<br>Stationary X-Ray system, 892.1680, code MQB                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Description:                            | Our device is used in medical x-ray imaging systems. The product can only be<br>used by trained personnel of medical facilities. The product is only used for<br>single image diagnosis applications. The flat panel detector consists of GOS<br>scintillator screen and thin-film transistors. The scintillator screen converts the<br>x-rays into visible light. Thin-film transistors convert the visible light to an<br>electrical charge. The flat panel detector can then obtain a digital image by<br>analog to digital conversion and associated circuits. |

{4}------------------------------------------------

Intended use: Model#'s 3543A, 4343A, 2929A and A843B; Solid State X-ray Imager (Flat Panel/Digital Imager) are indicated for use in generating radiographic images of human anatomy. They are intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding dental, fluoroscopic, angiographic, and mammographic applications. As prescribed by a licensed physician.

Refer to the table below (2 pages) Technological:

| Characteristic               | Atlaim ATAL 9 K152151                                                                                                                                                                                                                                              | PZMedical 4343A,3543A,2929A,A843B                                                                                                                                                                                                                                                                                                                                                                                              |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                 | Indicated for use in general<br>radiographic images of human<br>anatomy. It is intended to replace<br>radiographic film/screen systems in all<br>general-purpose diagnostic procedures,<br>excluding fluoroscopic, angiographic,<br>and mammographic applications. | Model#'s 3543A, 4343A, 2929A and A843B;<br>Solid State X-ray Imager (Flat Panel/Digital<br>Imager) are indicated for use in generating<br>radiographic images of human anatomy. They are<br>intended to replace radiographic film/screen<br>systems in all general-purpose diagnostic<br>procedures, excluding dental, fluoroscopic,<br>angiographic, and mammographic applications.<br>As prescribed by a licensed physician. |
| Configuration                | This submission is for the Digital<br>Panel and Software only, no generator<br>or stand provided.                                                                                                                                                                  | This submission is for the Digital Panel and<br>Software only, no generator or stand provided.                                                                                                                                                                                                                                                                                                                                 |
| Digital Panel                | Atlaim ATAL 9                                                                                                                                                                                                                                                      | PZMedical 4343A,3543A,2929A,A843B                                                                                                                                                                                                                                                                                                                                                                                              |
| Scintillator                 | GOS/CsI                                                                                                                                                                                                                                                            | GOS/CsI                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Pixel Pitch                  | 139 µm                                                                                                                                                                                                                                                             | 140 µm                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Limiting<br>Resolution       | Over 3 lp/mm                                                                                                                                                                                                                                                       | 3.6 lp/mm                                                                                                                                                                                                                                                                                                                                                                                                                      |
| DOE                          | 26% at 2 lp/mm (CsI) Not specified for<br>GOS                                                                                                                                                                                                                      | 32% at 2 lp/mm (CsI) 15% at 2 lp/mm (GOS)                                                                                                                                                                                                                                                                                                                                                                                      |
| MTF                          | 42% at 2 lp/mm (CsI) Not specified for<br>GOS                                                                                                                                                                                                                      | 33% at 2 lp/mm (CsI) 24% at 2 lp/mm (GOS)                                                                                                                                                                                                                                                                                                                                                                                      |
| A/D Conversion               | 16 bit                                                                                                                                                                                                                                                             | 16 bit                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Active Area                  | 17 x 17 inch                                                                                                                                                                                                                                                       | 4343A: 430.08 x 430.08 mm or 16.9 x 16.9 inch<br>3543A: 350.00 x 427.28 mm or 13.8 x 16.8 inch<br>2929A: 286.72 x 286.72 mm or 11.3 x 11.3 inch<br>A843B: 1075.2 x 430.08 mm or 42.3 x 16.9 inch                                                                                                                                                                                                                               |
| Dimensions                   | 460(W)×461(L)×15(H)                                                                                                                                                                                                                                                | 4343A: 460 x 460 x 15 mm<br>3543A: 383 x 460 x 15 mm<br>2929A: 316 x 316 x 15 mm<br>A843B: 1120 x 465 x 20 mm                                                                                                                                                                                                                                                                                                                  |
| Weights                      | 2.9kg (2.4kg w/o Battery) (up to 10<br>hours of battery life)                                                                                                                                                                                                      | 4343A: 4.3 kg (wireless, w/ batteries) 4.3 kg<br>(wired, w/o batteries)<br>3543A: 3.3 kg (wireless, w/ batteries) 2.9 kg<br>(wired, w/o batteries)<br>2929A: 1.8 kg (wireless, w/ batteries) 1.5 kg<br>(wired, w/o batteries)<br>A843B: 11.6 kg (wireless, w/ batteries) 10.9 kg<br>(wired, w/o batteries)                                                                                                                     |
| Pixels                       | 3072 x 3072 (9.4 Million)                                                                                                                                                                                                                                          | 3543A: 2500 x 3052 (7.6 Million)<br>4343A: 3072 x 3072 (9.4 Million)<br>2929A: 2048 x 2048 (4.2 Million)<br>A843B: 7680 x 3072 (23.6 Million)                                                                                                                                                                                                                                                                                  |
| Interface                    | Wired: Gigabit Ethernet (1000Base-T)<br>Wireless: IEEE802.11ac, backward<br>compatible                                                                                                                                                                             | Wired: Gigabit Ethernet (1000Base-T)<br>Wireless: IEEE802.11ac, backward compatible                                                                                                                                                                                                                                                                                                                                            |
| Power Source                 | AC Line and/or Rechargeable Lithium<br>Battery (10 hr run time)                                                                                                                                                                                                    | AC Line and/or Rechargeable Lithium Battery                                                                                                                                                                                                                                                                                                                                                                                    |
| Electrical safety<br>and EMC | Electrical Safety per IEC 60601-1 and<br>EMC per<br>IEC 60601-1-2 as well as IEEE<br>802.11ac.<br>Meets FCC requirements.                                                                                                                                          | Electrical Safety per IEC 60601-1 and EMC per<br>IEC 60601-1-2 as well as IEEE 802.11ac.<br>Meets FCC requirements.                                                                                                                                                                                                                                                                                                            |

{5}------------------------------------------------

## SOFTWARE RELATED

| Characteristic                                                                                | Atlaim ATAL 9 K152151 | PZMedical 4343A,3543A,2929A,A843B |
|-----------------------------------------------------------------------------------------------|-----------------------|-----------------------------------|
| Software                                                                                      | Outputs a DICOM image | Outputs a DICOM image             |
| SW revision                                                                                   | K152151               | PZDR V2.0.1                       |
| DICOM                                                                                         | Yes                   | Yes                               |
| DICOM Image Transmission                                                                      | Yes                   | Yes                               |
| Image Acquisition Interface                                                                   | Yes                   | Yes                               |
| Patient Body Part Selection                                                                   | Yes                   | Yes                               |
| Image Processing                                                                              | Yes                   | Yes                               |
| Basic Image Editing and<br>Marking (pan, zooming,<br>window/level adjusting, text<br>marking) | Yes                   | Yes                               |
| Image Browsing                                                                                | Yes                   | Yes                               |
| Patient Registration and<br>Management                                                        | Yes                   | Yes                               |

Non-Clinical data:

Testing results have shown that our devices Models 3543A, 4343A, 2929A and A843B, Solid State X-Ray Imager (Flat Panel/Digital Imager) are substantially equivalent to cleared predicate devices.

Atlaim Corporation, Atal 9, K132151 Careray, Care View 1500C / Care View 1500L, K153058 IRay, Wireless Digital Flat Panel Detertor, K161730

Specifically in relation too; DQE, MTF, Quantum limited performance, Effects of aliasing, Sensitivity linearity, Lag, Change in detection sensitivity, Dose requirement and reciprocity changes, Stability of the device characteristics with time, Uniformity of device characteristics, NPS, spatial resolution, Image Acquisition time and Black level.

{6}------------------------------------------------

### Clinical data:

We have included comparison images using our devices Models 3543A, 4343A, 2929A and A843B and a Atlaim Atal 9 cleared device (K132151) as direct comparison of the image quality of our devices in order to support substantial equivalence. We believe our image quality is substantially equivalent to cleared predicate devices.

### Conclusions:

In accordance with 21 CFR, Part 807 and supported by the information provided in this premarket notification (K170480). Shanghai PZ Medical Technology Co., Ltd. 5F, Building 13 No.2 Suide Road 200331 Shanghai concludes that our devices Models 3543A, 2929A and A843B Solid State X-Ray Imager (Flat Panel/Digital Imager) is substantially equivalent to predicate devices in regards to safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K170480](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K170480)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
