K160143 · Jpi Healthcare, Inc. · MQB · Oct 28, 2016 · Radiology
Device Facts
Record ID
K160143
Device Name
ClearVision ExamVue Flat Panel Detector
Applicant
Jpi Healthcare, Inc.
Product Code
MQB · Radiology
Decision Date
Oct 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The ClearVision ExamVue Flat Panel detector is indicated for use in general radiology, specialist radiology including podiatry, orthopedic, and other specialties, and in mobile x-ray systems. The ClearVision ExamVue Flat Panel detector is not indicated for use in mammography.
Device Story
ClearVision ExamVue Flat Panel Detector consists of three solid-state x-ray detector models (tethered and wireless cassette-sized panels, fixed panel); designed for radiologists and technicians. Device captures digital x-ray images by converting x-rays to electronic signals via a Cesium Iodide (CsI) scintillator and amorphous Silicon (aSi) TFT array; eliminates film/chemical processing. ExamVueDR software performs image processing, presentation, and storage in DICOM format. Used in clinical settings; interfaces with x-ray generators via hard-wired control signals or Automatic Exposure Detection (AED). Output is a digital image for clinical review; assists in diagnostic decision-making and treatment planning. Benefits include improved workflow efficiency and digital image availability.
Clinical Evidence
Clinical images provided to supplement laboratory performance data; not required to establish equivalence due to identical detector technology. Bench testing included safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility for patient-contacting surfaces.
Technological Characteristics
Solid-state x-ray imager using aSi TFTD technology with CsI scintillator. Models: 14x17in tethered, 14x17in wireless (with AED), 17x17in tethered. Connectivity: Gigabit Ethernet or wireless. DICOM compliant. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for general radiology, including podiatry, orthopedic, and other specialties, and mobile x-ray systems. Not indicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
DIGITAL FLAT PANEL X-RAY DETECTOR, MODEL LTX240AA01-A (K090742)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract design of three human profiles facing to the right, with a stylized caduceus-like symbol above them. The design is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 28, 2016
JPI Healthcare, Inc. % Mr. William Little Senior Product Manager JPI Healthcare Solutions, Inc. 52 Newtown Plaza PLAINVIEW NY 11803
Re: K160143
Trade/Device Name: ClearVision Exam Vue Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: September 12, 2016 Received: September 13, 2016
Dear Mr. Little:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K160143
Device Name
ClearVision ExamVue Flat Panel Detector
Indications for Use (Describe)
The ClearVision ExamVue Flat Panel detector is indicated for use in general radiology, specialist radiology including podiatry, orthopedic, and other specialties, and in mobile x-ray systems.
The ClearVision ExamVue Flat Panel detector is not indicated for use in mammography.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510 (k) Summary
August 10, 2016.
- 1. Company and Correspondant Making the Submission:
Name: JPI Healthcare Co., LTD Address: Woolim e-BIZ center #608 170-5, Kuro-3-dong, Kuro-gu Seoul, 152-769 Telephone: +82-2-2108 - 2580 (Ext. 500-504) Fax: +82-2-2108-1180 Contact: John Lim Website: http://www.jpi.co.kr/
- 2. Identification of Device
| Classification Name: | Solid State X-ray Imager (flat panel/digital imager) |
|-------------------------|------------------------------------------------------|
| Common Name: | Solid State X-ray Imager |
| Trade/Proprietary Name: | ClearVision ExamVue Flat Panel Detector |
#### 3. Predicate Device
| Manufacturer: | Atlaim |
|----------------------|------------------------------------------------------|
| Device: | ATAL-8 |
| 510(k) Number: | K113812 |
| Classification Name: | Solid State X-ray Imager (flat panel/digital imager) |
| Common Name: | Solid State X-ray Imager |
| Regulatory Number: | 21 CFR 892.1680 |
| Regulatory Class: | II |
| Product Code: | 90 MQB |
| Manufacturer: | Samsung Mobile Display |
|----------------------|----------------------------------------------------------|
| Device: | DIGITAL FLAT PANEL X-RAY DETECTOR, MODEL<br>LTX240AA01-A |
| 510(k) Number: | K090742 |
| Classification Name: | Solid State X-ray Imaging Device |
| Common Name: | Digital Flat Panel Xray Detector |
| Regulatory Number: | 21 CFR 892.1680 |
| Regulatory Class: | II |
| Product Code: | 90 MQB |
{4}------------------------------------------------
- Product Classification Names and Citations 4.
Regulatory Number: 21 CFR 892.1680 Regulatory Class: II Product Code: 90 MQB
#### 5. Description:
The ClearVision ExamVue Flat Panel Detector consists of a line of 3 different models of solid state x-ray detectors, of differing size and characteristics, designed for use by radiologists and radiology technicians for the acquisition of digital x-ray images. The ClearVision ExamVue Flat Panel Detector captures digital images of anatomy through the conversion of x-rays to electronic signals, eliminating the need for film or chemical processing to create a hard copy image. The ClearVision ExamVue Flat Panel Detector incorporates the ExamVueDR software, which performs the processing, presentation and storage of the image in DICOM format.
All models of the ClearVision ExamVue Flat Panel Detector use aSi TFTD for the collection of light generated by a CsI scintillator, for the purpose of creating a digital x-ray image. The three available models are:
- a. A 14x17in (35x43cm) tethered cassette sized panel
- b. A 14x17in (35x43cm) wireless cassette sized panel with automatic exposure detection
- c. A 17x17in (43x43cm) tethered panel for fixed installations.
#### 6. Indication for use
The ClearVision ExamVue Flat Panel Detector is indicated for use in general radiology, specialist radiology including podiatry, orthopedic, and other specialties, and in mobile x-ray systems.
The ClearVision ExamVue Flat Panel Detector is not indicated for use in mammography.
### 7. Comparison with Predicate Device:
JPI Healthcare Co., Ltd, believes that the ClearVision ExamVue Flat Panel Detector is substantially equivalent to the ATAL-8 detector of Atlaim and the LTX240AA01-A of Samsung Mobile Display, Ltd.
The ClearVision ExamVue Flat Panel Detectors have similar or superior characteristics to the predicate devices, including:
- The same aSi TFTD technology, converting visible light from a a. scintillator into a digital image.
- b. Similar pixel pitch (140um-143um for the Clear Vision Exam Vue Flat
{5}------------------------------------------------
Panels vs 139um-143um for the predicate devices)
- Exclusively CsI scintillators, compared to the option of CsI or the lower C. performance Gadolinium Oxide scintillator in the predicate devices.
- d. Similar indications for use, focused on general radiography and excluding mammography.
- Have been integrated to function with the same image acquisition e. software, ExamVueDR, for final processing and presentation.
Clinical images were provided; these images were not necessary to establish equivlance based on the modifications to the device (x-ray detector technology identical to predicate devices) but they provide further evidence in addition to the laboratory performance data to show that the subject device works as intended.
The image acquisition control interface testing was performed as part of laboratory and clinical testing. This control interface consists of hard wired control signal to the x-rav generator (all models) or AED (FDX3543RPW). The software control interface for exposure settings on selected generators had been previously tested in the notification of the ExamVueDR software as a stand-alone product (K142930).
- 8. Safety, EMC and Performance Data
Bench tests reports have been provided, detailing the safety and EMC testing, as well as non-clinical performance data for the ClearVision ExamVue Flat Panel Detector.
Biocompatibility data has been provided for patient-contacting surfaces of the device, demonstrating that there are no known adverse reactions to skin contact with the device.
We have also provided clinical testing of the hardware, showing comparison images using identical software, as recommended by the FDA guidance document "Guidance for Industry and/or for FDA Reviewers/Staff and/or Compliance: Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices"
- 9. Conclusions:
In accordance with the Federal Food. Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification JPI Healthcare Co., Ltd. concludes that The ClearVision ExamVue Flat Panel Detector is safe and effective and substantially equivalent to predicate devices as described herein.
A detailed comparison supporting this conclusion can be found in Exhibit 1, Substantial Equivalence Chart.
{6}------------------------------------------------
# EXHIBIT 1
## SUBSTANTIAL EQUIVALENCE CHART
| Device Name | ExamVue<br>FPD<br>FDX3543RP | Examvue FPD<br>FDX3543RPW | Examvue<br>FPD<br>FDX4343R | ATAL-8 | LTX240AA01-<br>A |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Toshiba | Toshiba | Toshiba | ATLAIM | Samsung<br>Mobile Display |
| FDA 510(k)<br>K# | K160143 | K160143 | K160143 | K113812 | K090742 |
| Intended use. | The ClearVision ExamVue Flat Panel detector<br>is indicated for use in general radiology,<br>specialist radiology including podiatry,<br>orthopedic, and other specialties, and in mobile<br>x-ray systems.<br>The ClearVision ExamVue Flat Panel detector<br>is not indicated for use in mammography. | The ClearVision ExamVue Flat Panel detector<br>is indicated for use in general radiology,<br>specialist radiology including podiatry,<br>orthopedic, and other specialties, and in mobile<br>x-ray systems.<br>The ClearVision ExamVue Flat Panel detector<br>is not indicated for use in mammography. | The ClearVision ExamVue Flat Panel detector<br>is indicated for use in general radiology,<br>specialist radiology including podiatry,<br>orthopedic, and other specialties, and in mobile<br>x-ray systems.<br>The ClearVision ExamVue Flat Panel detector<br>is not indicated for use in mammography. | The ATAL 8<br>and ATAL 8C<br>are indicated<br>for use in<br>general<br>radiographic<br>images of<br>human<br>anatomy. It is<br>intended to<br>replace<br>radiographic<br>film/screen<br>systems in all<br>general-<br>purpose<br>diagnostic<br>procedures,<br>excluding<br>fluoroscopic,<br>angiographic,<br>and<br>mammographic<br>applications. | LTX240AA01-<br>A digital flat<br>panel X-ray<br>detector is<br>indicated for<br>digital imaging<br>solution<br>designed for<br>providing<br>general<br>radiographic<br>diagnosis of<br>human<br>anatomy<br>targeting both<br>adult and<br>children. It is<br>intended to<br>replace film<br>based<br>radiographic<br>diagnostic<br>systems and<br>provide a case<br>diagnosis and<br>treatment<br>palnning for<br>physicians and<br>other health<br>care<br>professionals.<br>Not to be used<br>for<br>mammography. |
| Configuration | | This submission is for the digital panel and | | This | This |
{7}------------------------------------------------
| | software only, no generator or stand provided. | | | submission is<br>for the digital<br>panel only, no<br>generator or<br>stand provided. | submission is<br>for the digital<br>panel only, no<br>generator or<br>stand provided. |
|-----------------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Detector Type | Amorphous Silicon (a-Si) Photodiode | | | SAME | SAME |
| Pixel Pitch | 143um | 140um | 143um | 139um | 143um |
| Limiting<br>Resolution | 3.7lp/mm | 3.7lp/mm | 3.7lp/mm | Over 3lp/mm | 3lp/mm |
| A/D<br>Conversion | 16 bit | 14 bit | 14 bit | 14 bit | 14 bit |
| Active Area | 14x17 inch | 14x17 inch | 16.9x17.3 inch | 17x17 inch | 17x17 inch |
| Dimensions<br>(mm) | 384(w) x<br>460(D) x<br>15(H) | 384(w) x<br>460(d) x 15(h) | 512(w) x<br>495(d) x<br>43(h) | 500(w) x<br>500(1) x 25(h) | 500(w) x<br>497(1) x 44mm<br>(H) |
| Weight | 3kg | 3kg | 9kg | 7.8kg | 13.4kg |
| Pixels | 2448x2984<br>(7.3Mpx) | 2466x3040<br>(7.5Mpx) | 3008x3072<br>(9.2Mpx) | 3072x3072<br>(9.4Mpx) | 3072x3072<br>(9.4Mpx) |
| DQE @ 1<br>lp/mm | 57% | 60% | 58% | 33% Gadox /<br>46% CsI | 45% Gadox /<br>65% CsI |
| MTF @ 1<br>lp/mm | 63% | 68% | 65% | 63% Gadox /<br>72% CsI | 57% Gadox /<br>59% CsI |
| DICOM | Yes | Yes | Yes | Yes | Yes |
| Scintillator | CsI | CsI | CsI | CsI/Gadox | CsI/Gadox |
| Interface | Gigabit<br>Ethernet | Gigabit<br>Ethernet or<br>Wireless | Gigabit<br>Ethernet | Gigabit<br>Ethernet | Gigabit<br>Ethernet |
| Automatic<br>Exposure<br>Detection | No | Yes | No | Yes | No |
| Generator<br>Exposure<br>Trigger or<br>Sensor | Yes | Yes | Yes | Yes | Yes |
| Power Source | AC Line | AC Line | AC Line | AC Line | AC Line |
| Standards | Electrical<br>Safety per<br>IEC 60601-1<br>and EMC per<br>IEC 60601-<br>1-2. | Electrical<br>Safety per IEC<br>60601-1 and<br>EMC per IEC<br>60601-1-2. | Electrical<br>Safety per<br>IEC 60601-1<br>and EMC per<br>IEC 60601-<br>1-2. | Electrical<br>Safety per IEC<br>60601-1 and<br>EMC per IEC<br>60601-1-2. | Electrical<br>Safety per IEC<br>60601-1 and<br>EMC per IEC<br>60601-1-2. |
| Acquisition<br>and Control<br>Software | Exclusively<br>ExamVueDR<br>(K142930) | Exclusively<br>ExamVueDR<br>(K142930) | Exclusively<br>ExamVueDR<br>(K142930) | *Option for<br>ExamVueDR<br>(K142930) | *Option for<br>ExamVueDR<br>(K142930) |
{8}------------------------------------------------
*: The ExamVueDR software 510(k) notification (K149230) includes the indicated detector as one of the detectors the software can be integrated with to form a retrofit solution. The ExamVueDR software is not referenced in the 510(k) notification for the detector itself.
The intended use and environment of the device is the same as the predicate devices, with only minor differences in features that are not integral to the function of the device. The image acquisition and user interface software in the device has been previously 510(k) notified for use with the predicate devices. For this reason, we believe it is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.