← Product Code [MQB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB) · K153401

# GR40CW (K153401)

_Samsung Electronics Co., Ltd. · MQB · Dec 21, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K153401

## Device Facts

- **Applicant:** Samsung Electronics Co., Ltd.
- **Product Code:** [MQB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB.md)
- **Decision Date:** Dec 21, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.

## Device Story

GR40CW is a digital X-ray retrofit kit designed to upgrade existing analog X-ray systems. Components include a Gd2O2S flat-panel detector, power supply, battery, access point, Control Interface Box (CIB), workstation, and barcode scanner. The detector captures X-rays passing through a patient, converting them into electrical signals; these are amplified, digitized, and processed by the S-Station software into DICOM files for PACS storage. Integration occurs via CIB mode (passive signal detection from the hand switch) or AED mode (automatic radiation sensing). The system is operated by healthcare professionals in clinical settings. It does not control X-ray radiation parameters, which remain managed by the existing analog generator. Output allows physicians to view and interpret radiographic images for diagnostic purposes, facilitating treatment planning.

## Clinical Evidence

Clinical images were obtained and evaluated by a professional radiologist. Results were determined to be equivalent or better than the predicate device. The manufacturer noted that clinical data were not strictly necessary to establish substantial equivalence, as non-clinical bench testing (MTF, DQE per IEC 62220-1) was sufficient.

## Technological Characteristics

Detector: Gd2O2S scintillator; 14"x17" active area; 140 µm pixel pitch. Connectivity: Wireless/wired, DICOM compliant. Standards: ES 60601-1 (electrical/mechanical safety), IEC 60601-1-2 (EMC), IEC 62220-1 (MTF/DQE), ISO 14971, ISO 13485, 21 CFR 1020.30/31. Operation: CIB or AED mode for exposure detection.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- GR40CW ([K152094](/device/K152094.md))
- LLX240AB01 ([K102587](/device/K102587.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 21, 2015

SAMSUNG ELECTRONICS CO., LTD. % Ms. Chulsin Kim Regulatory Affairs Manager 129. Samsung-ro. Yeongtong-gu Suwon-si, Gyeonggi-do 443-742 REPUBLIC OF KOREA

Re: K153401

Trade/Device Name: GR40CW Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: November 25, 2015 Received: November 27, 2015

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ochs

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K153401

Device Name GR40CW

Indications for Use (Describe)

The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.

| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

{3}------------------------------------------------

SAMSUNG ELECTRONICS Co., Ltd.

510(k) Premarket Notification - Traditional

Image /page/3/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed on a blue, rounded rectangle.

## Section 5: 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92

- 1. Date: November 20, 2015
### 2. Submitter

- A. Company Name: SAMSUNG ELECTRONICS Co., Ltd.
- B. Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-742, Republic of Korea

#### 3. Primary Contact Person

- A. Name: KIM, CHULSIN
- B. Title: Regulatory Affairs Manager
- C. Phone Number: +82-31-200-7661
- D. FAX Number: +82-31-200-1199
- E. E-Mail: chulsin.kim@samsung.com

#### 4. Secondary Contact Person

- A. Name: Ninad Gujar
- B. Title: Requlatory Affairs Manager
- C. Phone Number: 978-564-8503
- D. FAX Number: 978-750-6677
- E. E-Mail: ngujar@samsungneurologica.com

## 5. Proposed Device

- A. Trade Name: GR40CW
- B. Device Name: GR40CW
- C. Common Name: Retrofit Kit
- D. Classification Name: Solid State X-ray Imager (Flat panel/Digital imager)
- E. Product Code: MQB
- F. Regulation: 21 CFR 892.1680

#### 6. Predicate Devices

|                         | The 1st Predicate Device                                | The 2nd Predicate Device                                |
|-------------------------|---------------------------------------------------------|---------------------------------------------------------|
| Manufacturer            | SAMSUNG ELECTRONICS Co.,<br>Ltd.                        | SAMSUNG MOBILE DISPLAY Co.,<br>Ltd.                     |
| Device Name             | GR40CW                                                  | LLX240AB01                                              |
| Common Name             | Retrofit Kit                                            | Digital Flat Panel X-Ray Detector                       |
| Classification<br>Name  | Solid State X-ray Imager (Flat<br>panel/Digital imager) | Solid State X-ray Imager (Flat<br>panel/Digital imager) |
| Product Code            | MQB                                                     | MQB                                                     |
| Regulation              | 21 CFR 892.1680                                         | 21 CFR 892.1680                                         |
| 510(k) Number           | K152094                                                 | K102587                                                 |
| 510(k) Decision<br>Date | Oct. 16, 2015                                           | Dec. 01, 2010                                           |

### 7. Device Description

{4}------------------------------------------------

## SAMSUNG ELECTRONICS Co., Ltd.

510(k) Premarket Notification - Traditional

Image /page/4/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue oval shape.

The GR40CW digital X-ray imaging system consists of Detector, Power supply box, Battery pack, Battery charger, Access point, CIB(Control Interface Box), Workstation, Barcode scanner and Main cable. This system is used to capture images by transmitting X-ray to a patient's body.

The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device being sent to the S-Station (Operation Software) and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images.

#### Integration Information

This is a retrofit system consisting of Detectors, Power supply box, Battery charger, Access point, CIB (Control Interface Box), Workstation, Barcode scanner and Main cable. This retrofit system is designed to generate a digital image while using the current analoque X-ray system by upgrading only the part of an analogue cassette film to the digital panel(detector), and does not get involved in controlling X-ray radiation related parameters, which is still controlled by the existing X-ray system.

The GR40CW retrofit system can be applied to the existing analogue X-ray system by two ways (CIB and AED modes). 1) In CIB mode, CIB is only connected to a signal line of a hand switch for passively detecting the signal, as On or Off, coming out from the hand switch to the X-ray Generator Console, to make the digital detector ready to active or inactive to receive X-ray radiation. 2) In AED (Automatic Exposure Detection) mode, without CIB, the detector is sensing of radiation exposure. Once it recognizes the exposure, it become active right away to receive X-ray radiation. This whole process is independently operated from the existing analogue X-ray system.

Image /page/4/Figure/8 description: The image shows an integration diagram of the GR40CW Retrofit Kit and an existing X-ray system. The GR40CW Retrofit Kit includes a detector, workstation, and CIB, while the existing X-ray system includes an X-ray generator console and a hand switch. The diagram illustrates the connections and signals between the components, including the 'READY' signal from the first step press and the 'EXPOSURE' signal from the second step press. The system is operating in CIB mode.

#### [Integration Diagram]

{5}------------------------------------------------

Image /page/5/Picture/10 description: The image contains the text "SAMSUNG ELECTRONICS Co., Ltd." in bold, black font. To the right of the text is the Samsung logo, which is a blue oval with the word "SAMSUNG" written in white inside. The logo is partially cut off on the right side of the image.

510(k) Premarket Notification - Traditional

Image /page/5/Figure/2 description: The image shows a diagram of the GR40CW Retrofit Kit and the existing X-ray system. The GR40CW Retrofit Kit includes a detector, workstation, and CIB. The existing X-ray system includes an X-ray generator console and a hand switch. The hand switch has two steps: the first step sends a READY signal to the X-ray generator console, and the second step sends an EXPOSURE signal.

In conclusion, this retrofit kit does not either require a modification / alteration or control in the part of X-ray radiation control of the current X-ray system in anyway.

## 8. Intended Use

The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems may be used. This device is not intended for mammographic applications.

## 9. Summary of Technological characteristic of the proposed device compared with the predicate device

The proposed GR40CW has a digital X-ray detector (S4335-WV), which type is Gd2O2S as same as the 200 predicate device (LLX240AB01) cleared with 510(k) K102587, to be added into a list of detectors of the 1st predicate device, GR40CW (K152094). The proposed GR40CW has same software feature, workstation and CIB with the 1st predicate device. There is no significant difference in materials, energy source or technological characteristics compared to the predicate devices. The following technological characteristics are compared and the results demonstrate the substantial equivalent to the predicate devices.

| Specification | 1stPredicate Device                                                                                           | 2ndPredicate Device                                                                                                                  | Proposed Device                                                                                               | Discussion                         |
|---------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|------------------------------------|
| Device Name   | GR40CW                                                                                                        | LLX240AB01                                                                                                                           | GR40CW                                                                                                        | -                                  |
| Manufacturer  | SAMSUNG<br>ELECTRONICS                                                                                        | SAMSUNG MOBILE<br>DISPLAY                                                                                                            | SAMSUNG<br>ELECTRONICS                                                                                        | -                                  |
| 510(k) Number | K152094                                                                                                       | K102587                                                                                                                              | N/A                                                                                                           | -                                  |
| Appearance    | Image: First predicate device appearance                                                                      | Image: Second predicate device appearance                                                                                            | Image: Proposed device appearance                                                                             | Same as 1st<br>predicate<br>device |
| Intended Use  | The GR40CW Digital<br>X-ray Imaging<br>System is intended<br>for use in general<br>projection<br>radiographic | LLX240AB01 Digital<br>Flat Panel X-Ray<br>Detector is indicated<br>for digital imaging<br>solution designed for<br>providing general | The GR40CW Digital<br>X-ray Imaging<br>System is intended<br>for use in general<br>projection<br>radiographic | Same as 1st<br>predicate<br>device |

{6}------------------------------------------------

# AMSUNG ELECTRONICS Co., Ltd.

Image /page/6/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded rectangle.

510(k) Premarket Notification - Traditional

| applications<br>wherever<br>conventional screen-<br>film systems or CR<br>systems may be<br>used. This device is<br>not intended for<br>mammographic<br>applications. | radiographic<br>diagnosis of human<br>anatomy targeting<br>both adult and<br>children. It is<br>intended to replace<br>film based<br>radiographic<br>diagnostic systems<br>and provide a case<br>diagnosis and<br>treatment planning<br>for physicians and<br>other health care<br>professionals. Not to<br>be used for<br>mammography. | applications<br>wherever<br>conventional screen-<br>film systems or CR<br>systems may be<br>used. This device is<br>not intended for<br>mammographic<br>applications. |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| Manufacturer<br>Contents                           | GR40CW<br>(K152094)                         | LLX240AB01<br>(K102587)  | GR40CW                   | Discussion                   |
|----------------------------------------------------|---------------------------------------------|--------------------------|--------------------------|------------------------------|
| Detector                                           | Note: S4335-WV has not been cleared before. |                          |                          |                              |
| Name                                               | S4335-W                                     | LLX240AB01               | S4335-WV                 | -                            |
| Detector Type                                      | CsI                                         | Gd2O2S                   | Gd2O2S                   | Same as 2nd predicate device |
| Detector Area                                      | 14"X17"<br>(345mmX425mm)                    | 17"X17"<br>(439mmX439mm) | 14"X17"<br>(345mmX425mm) | Same as 1st predicate device |
| Pixel Pitch (µm)                                   | 140                                         | 143                      | 140                      | Same as 1st predicate device |
| High Contrast<br>Limiting<br>Resolution<br>(lp/mm) | 3.5                                         | 3.6                      | 3.5                      | Same as 1st predicate device |

Since imaging Workstation and CIB(Control Interface Box) are same as the 18t predicate device has, there is no difference.

In non-clinical data, the proposed device's detector shows curves and measurements of MTF and DQE that demonstrate a substantially equivalent or improvement to the 200 predicate device.

### 10.Safety, EMC and Performance Data

Electrical, mechanical, environmental safety and performance testing were conducted according to standard ES 60601-1(2012), and EMC testing was conducted according to IEC 60601-1-2(2007). Wireless function was tested and verified followed by the guidance for Radio frequency Wireless Technology in Medical Devices. All test results were satisfying with the standards

#### 11.Non-clinical data

Non-clinical testing data was provided in conformance to the FDA "Guidance for the Submission

{7}------------------------------------------------

# SAMSUNG ELECTRONICS Co., Ltd.

510(k) Premarket Notification - Traditional

Image /page/7/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The wordmark is set against a blue, oval-shaped background, which is tilted slightly.

of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1. The proposed device shows no difference in non-clinical testing data such as MTF and DQE measurements from the predicate device.

It conforms to the followings: ISO14971, ISO13485, 21 CFR Subchapter J 1020.30 and 1020.31

### 12.Clinical data

In clinical data, clinical images were obtained in accordance with FDA guidance for the submission of 510(k)'s for Solid State X-Ray Imaging Devices. They were evaluated by a professional radiologist, and shown equivalent and better than the 20d predicate device. The provided clinical data were determined to be not necessary to establish substantial equivalence with the predicate devices.

## 13.Conclusions

The results of the non-clinical data & clinical data demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed devices.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K153401](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MQB/K153401)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
