The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.
Device Story
GR40CW is a digital X-ray retrofit kit designed to upgrade existing analog X-ray systems. Components include a Gd2O2S flat-panel detector, power supply, battery, access point, Control Interface Box (CIB), workstation, and barcode scanner. The detector captures X-rays passing through a patient, converting them into electrical signals; these are amplified, digitized, and processed by the S-Station software into DICOM files for PACS storage. Integration occurs via CIB mode (passive signal detection from the hand switch) or AED mode (automatic radiation sensing). The system is operated by healthcare professionals in clinical settings. It does not control X-ray radiation parameters, which remain managed by the existing analog generator. Output allows physicians to view and interpret radiographic images for diagnostic purposes, facilitating treatment planning.
Clinical Evidence
Clinical images were obtained and evaluated by a professional radiologist. Results were determined to be equivalent or better than the predicate device. The manufacturer noted that clinical data were not strictly necessary to establish substantial equivalence, as non-clinical bench testing (MTF, DQE per IEC 62220-1) was sufficient.
Technological Characteristics
Detector: Gd2O2S scintillator; 14"x17" active area; 140 µm pixel pitch. Connectivity: Wireless/wired, DICOM compliant. Standards: ES 60601-1 (electrical/mechanical safety), IEC 60601-1-2 (EMC), IEC 62220-1 (MTF/DQE), ISO 14971, ISO 13485, 21 CFR 1020.30/31. Operation: CIB or AED mode for exposure detection.
Indications for Use
Indicated for general projection radiographic applications in adults and children. Not for mammographic applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 21, 2015
SAMSUNG ELECTRONICS CO., LTD. % Ms. Chulsin Kim Regulatory Affairs Manager 129. Samsung-ro. Yeongtong-gu Suwon-si, Gyeonggi-do 443-742 REPUBLIC OF KOREA
Re: K153401
Trade/Device Name: GR40CW Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: November 25, 2015 Received: November 27, 2015
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K153401
Device Name GR40CW
Indications for Use (Describe)
The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
{3}------------------------------------------------
SAMSUNG ELECTRONICS Co., Ltd.
510(k) Premarket Notification - Traditional
Image /page/3/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed on a blue, rounded rectangle.
## Section 5: 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92
- 1. Date: November 20, 2015
### 2. Submitter
- A. Company Name: SAMSUNG ELECTRONICS Co., Ltd.
- B. Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-742, Republic of Korea
#### 3. Primary Contact Person
- A. Name: KIM, CHULSIN
- B. Title: Regulatory Affairs Manager
- C. Phone Number: +82-31-200-7661
- D. FAX Number: +82-31-200-1199
- E. E-Mail: chulsin.kim@samsung.com
#### 4. Secondary Contact Person
- A. Name: Ninad Gujar
- B. Title: Requlatory Affairs Manager
- C. Phone Number: 978-564-8503
- D. FAX Number: 978-750-6677
- E. E-Mail: ngujar@samsungneurologica.com
## 5. Proposed Device
- A. Trade Name: GR40CW
- B. Device Name: GR40CW
- C. Common Name: Retrofit Kit
- D. Classification Name: Solid State X-ray Imager (Flat panel/Digital imager)
- E. Product Code: MQB
- F. Regulation: 21 CFR 892.1680
#### 6. Predicate Devices
| | The 1st Predicate Device | The 2nd Predicate Device |
|-------------------------|---------------------------------------------------------|---------------------------------------------------------|
| Manufacturer | SAMSUNG ELECTRONICS Co.,<br>Ltd. | SAMSUNG MOBILE DISPLAY Co.,<br>Ltd. |
| Device Name | GR40CW | LLX240AB01 |
| Common Name | Retrofit Kit | Digital Flat Panel X-Ray Detector |
| Classification<br>Name | Solid State X-ray Imager (Flat<br>panel/Digital imager) | Solid State X-ray Imager (Flat<br>panel/Digital imager) |
| Product Code | MQB | MQB |
| Regulation | 21 CFR 892.1680 | 21 CFR 892.1680 |
| 510(k) Number | K152094 | K102587 |
| 510(k) Decision<br>Date | Oct. 16, 2015 | Dec. 01, 2010 |
### 7. Device Description
{4}------------------------------------------------
## SAMSUNG ELECTRONICS Co., Ltd.
510(k) Premarket Notification - Traditional
Image /page/4/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue oval shape.
The GR40CW digital X-ray imaging system consists of Detector, Power supply box, Battery pack, Battery charger, Access point, CIB(Control Interface Box), Workstation, Barcode scanner and Main cable. This system is used to capture images by transmitting X-ray to a patient's body.
The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device being sent to the S-Station (Operation Software) and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images.
#### Integration Information
This is a retrofit system consisting of Detectors, Power supply box, Battery charger, Access point, CIB (Control Interface Box), Workstation, Barcode scanner and Main cable. This retrofit system is designed to generate a digital image while using the current analoque X-ray system by upgrading only the part of an analogue cassette film to the digital panel(detector), and does not get involved in controlling X-ray radiation related parameters, which is still controlled by the existing X-ray system.
The GR40CW retrofit system can be applied to the existing analogue X-ray system by two ways (CIB and AED modes). 1) In CIB mode, CIB is only connected to a signal line of a hand switch for passively detecting the signal, as On or Off, coming out from the hand switch to the X-ray Generator Console, to make the digital detector ready to active or inactive to receive X-ray radiation. 2) In AED (Automatic Exposure Detection) mode, without CIB, the detector is sensing of radiation exposure. Once it recognizes the exposure, it become active right away to receive X-ray radiation. This whole process is independently operated from the existing analogue X-ray system.
Image /page/4/Figure/8 description: The image shows an integration diagram of the GR40CW Retrofit Kit and an existing X-ray system. The GR40CW Retrofit Kit includes a detector, workstation, and CIB, while the existing X-ray system includes an X-ray generator console and a hand switch. The diagram illustrates the connections and signals between the components, including the 'READY' signal from the first step press and the 'EXPOSURE' signal from the second step press. The system is operating in CIB mode.
#### [Integration Diagram]
{5}------------------------------------------------
Image /page/5/Picture/10 description: The image contains the text "SAMSUNG ELECTRONICS Co., Ltd." in bold, black font. To the right of the text is the Samsung logo, which is a blue oval with the word "SAMSUNG" written in white inside. The logo is partially cut off on the right side of the image.
510(k) Premarket Notification - Traditional
Image /page/5/Figure/2 description: The image shows a diagram of the GR40CW Retrofit Kit and the existing X-ray system. The GR40CW Retrofit Kit includes a detector, workstation, and CIB. The existing X-ray system includes an X-ray generator console and a hand switch. The hand switch has two steps: the first step sends a READY signal to the X-ray generator console, and the second step sends an EXPOSURE signal.
In conclusion, this retrofit kit does not either require a modification / alteration or control in the part of X-ray radiation control of the current X-ray system in anyway.
## 8. Intended Use
The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems may be used. This device is not intended for mammographic applications.
## 9. Summary of Technological characteristic of the proposed device compared with the predicate device
The proposed GR40CW has a digital X-ray detector (S4335-WV), which type is Gd2O2S as same as the 200 predicate device (LLX240AB01) cleared with 510(k) K102587, to be added into a list of detectors of the 1st predicate device, GR40CW (K152094). The proposed GR40CW has same software feature, workstation and CIB with the 1st predicate device. There is no significant difference in materials, energy source or technological characteristics compared to the predicate devices. The following technological characteristics are compared and the results demonstrate the substantial equivalent to the predicate devices.
| Specification | 1stPredicate Device | 2ndPredicate Device | Proposed Device | Discussion |
|---------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|------------------------------------|
| Device Name | GR40CW | LLX240AB01 | GR40CW | - |
| Manufacturer | SAMSUNG<br>ELECTRONICS | SAMSUNG MOBILE<br>DISPLAY | SAMSUNG<br>ELECTRONICS | - |
| 510(k) Number | K152094 | K102587 | N/A | - |
| Appearance | Image: First predicate device appearance | Image: Second predicate device appearance | Image: Proposed device appearance | Same as 1st<br>predicate<br>device |
| Intended Use | The GR40CW Digital<br>X-ray Imaging<br>System is intended<br>for use in general<br>projection<br>radiographic | LLX240AB01 Digital<br>Flat Panel X-Ray<br>Detector is indicated<br>for digital imaging<br>solution designed for<br>providing general | The GR40CW Digital<br>X-ray Imaging<br>System is intended<br>for use in general<br>projection<br>radiographic | Same as 1st<br>predicate<br>device |
{6}------------------------------------------------
# AMSUNG ELECTRONICS Co., Ltd.
Image /page/6/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded rectangle.
510(k) Premarket Notification - Traditional
| applications<br>wherever<br>conventional screen-<br>film systems or CR<br>systems may be<br>used. This device is<br>not intended for<br>mammographic<br>applications. | radiographic<br>diagnosis of human<br>anatomy targeting<br>both adult and<br>children. It is<br>intended to replace<br>film based<br>radiographic<br>diagnostic systems<br>and provide a case<br>diagnosis and<br>treatment planning<br>for physicians and<br>other health care<br>professionals. Not to<br>be used for<br>mammography. | applications<br>wherever<br>conventional screen-<br>film systems or CR<br>systems may be<br>used. This device is<br>not intended for<br>mammographic<br>applications. |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>Contents | GR40CW<br>(K152094) | LLX240AB01<br>(K102587) | GR40CW | Discussion |
|----------------------------------------------------|---------------------------------------------|--------------------------|--------------------------|------------------------------|
| Detector | Note: S4335-WV has not been cleared before. | | | |
| Name | S4335-W | LLX240AB01 | S4335-WV | - |
| Detector Type | CsI | Gd2O2S | Gd2O2S | Same as 2nd predicate device |
| Detector Area | 14"X17"<br>(345mmX425mm) | 17"X17"<br>(439mmX439mm) | 14"X17"<br>(345mmX425mm) | Same as 1st predicate device |
| Pixel Pitch (µm) | 140 | 143 | 140 | Same as 1st predicate device |
| High Contrast<br>Limiting<br>Resolution<br>(lp/mm) | 3.5 | 3.6 | 3.5 | Same as 1st predicate device |
Since imaging Workstation and CIB(Control Interface Box) are same as the 18t predicate device has, there is no difference.
In non-clinical data, the proposed device's detector shows curves and measurements of MTF and DQE that demonstrate a substantially equivalent or improvement to the 200 predicate device.
### 10.Safety, EMC and Performance Data
Electrical, mechanical, environmental safety and performance testing were conducted according to standard ES 60601-1(2012), and EMC testing was conducted according to IEC 60601-1-2(2007). Wireless function was tested and verified followed by the guidance for Radio frequency Wireless Technology in Medical Devices. All test results were satisfying with the standards
#### 11.Non-clinical data
Non-clinical testing data was provided in conformance to the FDA "Guidance for the Submission
{7}------------------------------------------------
# SAMSUNG ELECTRONICS Co., Ltd.
510(k) Premarket Notification - Traditional
Image /page/7/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The wordmark is set against a blue, oval-shaped background, which is tilted slightly.
of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1. The proposed device shows no difference in non-clinical testing data such as MTF and DQE measurements from the predicate device.
It conforms to the followings: ISO14971, ISO13485, 21 CFR Subchapter J 1020.30 and 1020.31
### 12.Clinical data
In clinical data, clinical images were obtained in accordance with FDA guidance for the submission of 510(k)'s for Solid State X-Ray Imaging Devices. They were evaluated by a professional radiologist, and shown equivalent and better than the 20d predicate device. The provided clinical data were determined to be not necessary to establish substantial equivalence with the predicate devices.
## 13.Conclusions
The results of the non-clinical data & clinical data demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.