CAREVIEW 1800R

K141488 · Careray Digital Medical System Co., Ltd. · MQB · Sep 8, 2015 · Radiology

Device Facts

Record IDK141488
Device NameCAREVIEW 1800R
ApplicantCareray Digital Medical System Co., Ltd.
Product CodeMQB · Radiology
Decision DateSep 8, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2

Indications for Use

CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the detector enables the digital X-ray imaging intended for medical application. This product is not intended for mammography applications and dental X-ray applications.

Device Story

CareView 1800R is a stationary X-ray flat panel detector (FPD) component for digital radiography systems. It uses a TFT/PD image sensor with a direct-deposited CsI scintillator to convert X-ray photons into visible light, then into electron signals via a diode capacitor array. The device captures single-frame images, which are processed and transmitted to a PC via a user interface. Used in clinical settings by healthcare professionals, the detector integrates into a completed X-ray system to enable digital imaging. It supports external and manual synchronization modes with X-ray generators. The output is a digital image used by clinicians for diagnostic purposes, facilitating medical decision-making through high-resolution radiographic visualization.

Clinical Evidence

Clinical evidence included a concurrence study of 30 clinical images comparing the CareView 1800R to the predicate device. Results showed no significant difference between the images produced by the subject and predicate devices. Primary evidence for equivalence was based on laboratory performance data (MTF, DQE, NPS, sensitivity, lag, uniformity, etc.) and established scintillator material usage.

Technological Characteristics

Active area 43cm x 43cm; CsI scintillator; TFT readout mechanism; 16-bit ADC; 154µm pixel pitch; 2816x2816 pixel array. Stationary installation. Power consumption ~36W. Connectivity via wired communication. Tested to IEC/ES 60601-1 (electrical/mechanical/environmental safety) and IEC/EN 60601-1-2 (EMC).

Indications for Use

Indicated for digital imaging in general purpose diagnostic radiographic procedures. Not for mammography or dental applications.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 8, 2015 CareRay Digital Medical System Co., Ltd. % Jamie Ku Official Correspondent Compass Innovations, Inc. 3001 Winchester Blvd, Suite 3 CAMPBELL CA 95008 Re: K141488 Trade/Device Name: CareView 1800R X-ray Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: August 13, 2015 Received: August 17, 2015 Dear Jamie Ku: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. For Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K141488 Device Name CareView 1800R X-ray Flat Panel Detectors Indications for Use (Describe) CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the digital X-ray imaging intended for medical application. This product is not intended for mammography applications and dental X-ray applications. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input type="checkbox"/> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary [As required by 21 CFR 807.92] ## 1. Date Prepared [21 CFR807.92 (a) (1)] August 13, 2015 #### 2. Submitter's Information [21 CFR807.92 (a) (1)] | Company Name: | CareRay Digital Medical System Co., Ltd. | |------------------|---------------------------------------------------------------------------------------------------------| | Company Address: | Suite 201, Building A2, BIOBAY, 218 Xing Hu Street,<br>SuZhou Industrial Park, SuZhou 215123, P.R.China | | Contact Person: | Ms. Ding | | Phone Number: | (86) 512-86860288 | | Fax Number: | (86) 512-86860388 | | E-mail: | yl.ding@careray.com | ### 3. Trade Name, Common Name, Classification [21 CFR807.92(a)(2)] | Trade Name: | X-ray Flat Panel Detectors | |----------------------|----------------------------| | Model Name: | CareView 1800R | | Classification Name: | Stationary X-ray system | | Regulation Number: | 21 CFR 892.1680 | | Regulatory Class: | Class II | | Product Code: | MQB | #### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[ The identified predicates within this submission are as follows: | Manufacturer: | Canon Inc. | |----------------------|-------------------------| | Trade Name: | Canon | | Model Name: | CXDI-40EG | | Classification Name: | Stationary X-ray system | | Regulation Number: | 21 CFR 892.1680 | | Regulatory Class: | Class II | {4}------------------------------------------------ | Product Code: | MQB | |---------------|---------| | FDA 510(k) #: | K050987 | ### 5. Description of the Device [21 CFR 807.92(a)(4)] CareView 1800R X-ray Flat Panel Detectors supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 43cm×43cm. The sensor plate of CareView 1800R X-ray Flat Panel Detectors is direct-deposited with Csl scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface. The major function of the CareView 1800R X-ray Flat Panel Detectors detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. This detector is the key component of DR system, enables to complete the digitalization of the medical X-ray imaging with the DR system software. # 6. Intended Use [21 CFR 807.92(a)(5)] CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the detector enables the digital X-ray imaging intended for medical application. This product is not intended for mammography applications and dental X-ray applications. | Item | Proposed Device:<br>CareView 1800R X-ray Flat Panel<br>Detectors | Predicate Device: CXDI-40EG | |---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(K)<br>Number | To be assigned | K050987 | | Intended Use | The CareView 1800R X-ray Flat<br>Panel Detectors is indicated for<br>digital imaging solution designed<br>for providing general radiographic<br>system in all general purpose<br>diagnostic procedures. CareView<br>1800R X-ray Flat Panel Detectors<br>is a component of a digital imaging<br>system. Properly integrated into a | DIGITAL RADIOGRAPHY<br>CXDI-40EG provides digital<br>image capture for conventional<br>film/screen radiographic<br>examinations. The device is<br>intended to replace radiographic<br>film/screen systems in all general<br>purpose diagnostic procedures. | | | completed X-ray system, the<br>detector enables the digital X-ray<br>imaging intended for medical<br>application.<br>This product is not intended for<br>mammography applications and<br>dental X-ray applications. | | | Classification<br>Name | Stationary X-ray system | Stationary X-ray system | | Product<br>Code | MQB | MQB | | Regulation<br>Number | 21 CFR 892.1680 | 21 CFR 892.1680 | | Panel | Radiology | Radiology | | Class | II | II | | X-ray<br>Absorber | CsI Scintillator | GOS | | Installation<br>Type | Stationary, permanently installed | Stationary, permanently installed | | Readout<br>Mechanism | Thin Film Transistor | Thin Film Transistor | | Detector<br>Size | 492 x 492 x 33.5 mm | 550 x 550 x 67.5 mm | | Detector<br>Weight | 13 kg | 20 kg | | Pixel Array | 2816x2816 | 2688 x2688 | | Active Area | 43cm×43cm | 43cm×43cm | | Pixel Pitch | 154µm | 160µm | | ADC<br>Digitization | 16 bit | 14 bit | | Power<br>Consumption | ~36 W | Max.200 VA | | Modulation<br>Transfer<br>Function<br>(MTF) | 1 lp/mm (70±3)%<br>2 lp/mm (42±3)%<br>3 lp/mm (23±3)% | 1 lp/mm 68%<br>2 lp/mm 38%<br>3 lp/mm 18% | | Detective<br>Quantum<br>Efficiency<br>(DQE) | DQE at 5 µGy<br>0 lp/mm (63±3)%<br>1 lp/mm (45±3)%<br>2 lp/mm (33±3)%<br>3 lp/mm (20±3)% | DQE at 5 mR<br>0 lp/mm 37%<br>I lp/mm 34%<br>2 lp/mm 24%<br>3 lp/mm 12% | | Power<br>Supply | Voltage: 100-250 VAC<br>Frequency: 50/60 Hz | Voltage: 100V, 120V, 230/240 V<br>Frequency: 50/60 Hz | | Operating<br>Condition | Temperature: 5 ~ 35°C<br>Relative humidity: 30~75% RH<br>Air pressure: 700hPa~1060hPa | Temperature: 5~35°C<br>Relative humidity: 30~75%<br>Air pressure: 700hPa ~1060hPa | | Storage<br>Condition | Temperature: -20~55 °C<br>Relative humidity: 10~90%RH<br>Air pressure: 700hPa~1060hPa | Temperature: -30~60 °C<br>Relative humidity: 10~60%<br>Air pressure: 700hPa ~1060hPa | # 7. Technological Characteristics [21 CFR 807.92(a)(6)] {5}------------------------------------------------ {6}------------------------------------------------ # 8. System requirements to operate with other radiographic system components - 1) Recommended Generator Specification: Energy range: 40~150kVp mA range: 10~1000mA (depending on the generator power) ms range: 10~6300ms to produce 0.1~1000mAs (depending on the generator power) Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators, please contact your distributor or CareRay. - 2) Application Program Interface (API) for system integration manufacturer Peripheral hardware: CareView detector connected via wired communication. CPU: Intel (R) Core (TM) 2 Duo, 2.93GHz or above RAM: 2 GB or higher Hard disk: 160 GB or higher Monitor: 1280 x 1024 or higher OS: Windows XP or Windows 7 Development environment: MS Visual Studio 2005 - 3) X-ray exposure mode The synchronous connection mode is the signal transfer mode between the X-ray generator which sends the X-ray and the detector which receives the X-ray. CareView 1800R supported typical sync mode contains external sync, manual sync mode. The detector can't provide feedback to the generator to terminate the x-ray exposure. ## 9. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92(b)(2)] - > Electrical safety and EMC testing Electrical, mechanical, environmental safety and performance testing according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC/EN 60601-1-2. All test results are satisfactory. Section 201.7.2.7, 201.7.9.1, 201.7.9.2.1.103 and 203.4.1 of IEC 60601-2-54 were also conducted. - A Nonclinical considerations The following non-clinical studies have been performed and the results have shown that the CareView 1800R X-ray flat panel detector is substantially equivalent to the predicate devices on the Market (CXDI-40EG). {7}------------------------------------------------ -Detective quantum efficiency (DQE), Quantum limited performance, Modulation transfer function (MTF), Effects of aliasing, Sensitivity linearity, Lag(Erasure thoroughness), Change in detection sensitivity, Dose requirement and reciprocity changes, Stability of device characteristics with time, Uniformity of device characteristic, Noise power spectrum(NPS), Spatial resolution, Image Acquisition time, & Black level - A Clinical considerations A concurrence study of 30 clinical images was conducted to compare the performance of the CareView 1800R to that of the predicate device (K050987). Clinical images were provided; these images were not necessary to establish substantial equivalence based on the superior performance in comparison to the predicate based on laboratory testing and based on the use of a detector scintillator material that is already on the marketplace; however, they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended. There was no significant difference between the images of the CareView 1800R and those of the predicate device. #### 10. Conclusion [21 CFR 807.92(b) (3)] In accordance with the Federal Food, Druq and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, the CareView 1800R X-ray flat panel detector is substantially equivalent to predicate device CXDI-40EG (K050987). Both propose and predicate devices are same or very similar in the intended use, the design principle, the applicable standards and specification. Some characteristics, for example, the detector size, pixel pitch, weight and ADC digitization, X-ray absorber, storage condition are different. However the test reports in this submission documents provide demonstration that these differences do not raise any new questions of safety and effectiveness. Therefore, CareRay Digital Medical System Co., Ltd. concludes the CareView 1800R X-ray flat panel detector is substantially equivalent with the predicate device CXDI-40EG (K050987) of Canon Inc.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...