The DIGITAL RADIOGRAPHY CXDI-401C Wireless provides digital image capture for conventional film/screen radiographic examinations. This device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Device Story
Solid state X-ray imager; captures X-ray photons via scintillator (CsI) and amorphous silicon (a-Si) array; converts to electrical signals then digital values. Communicates via wired or wireless connection to operator console. Features Non-Generator Connection Mode for X-ray detection without direct generator interface. Used in clinical settings by radiology staff. Output displayed on monitors for diagnostic review by physicians. Benefits include digital workflow efficiency and replacement of conventional film/screen systems.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including verification/validation to internal functional specifications, software requirements, and image quality comparisons against predicate devices. Compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-32, and FCC standards for SAR and EMI demonstrated.
Technological Characteristics
Flat panel detector; Scintillator (CsI) and amorphous silicon (a-Si); 125x125um pixel pitch; 3,320 x 3,408 pixels; 16-bit dynamic range; wireless and wired connectivity; AC/DC power; 3.8kg weight. Complies with IEC 60601 series and FCC Part 15/ICES-003.
Indications for Use
Indicated for general purpose diagnostic radiographic examinations in patients requiring conventional film/screen replacement. Not indicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
# Canon
## 5. 510(k) SUMMARY
**JUL 01 2014**
**K133693**
*Page 1 of 2*
| Submitter: | Canon, Inc. - Medical Equipment Group<br>30-2 Shimomaruko, 3-chrome<br>Ohta-ku, Tokyo 146-8501 Japan |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Applicant:<br>Mr. Shinji Mori<br>Manager<br>TEL: 81-3-3758-2111<br>FAX: 81-3-5482-3960<br>mori.shinji@canon.co.jp |
| Date Prepared: | November 22, 2013 (revised Jan 9, 2014) |
| Trade Name: | DIGITAL RADIOGRAPHY CXDI-401C Wireless |
| Common Name: | Flat Panel Digital Imager |
| Classification<br>Name: | MQB (Solid State X-Ray Imager, Flat Panel/Digital Imager)<br>892.1680 (Stationary X-Ray System) |
| Predicate Devices: | K131106 MQB DIGITAL RADIOGRAPHY CXDI-701C Wireless, Canon, Inc.<br>K103591 MQB DIGITAL RADIOGRAPHY CXDI-401C COMPACT, Canon, Inc. |
| Device Description: | The model of detector included in this submission is a solid state x-ray imager.<br>Model CXDI-401C Wireless has an approximate imaging area of 41.5 x 42.6 cm.<br>For this model, the detector intercepts x-ray photons and the scintillator emits<br>visible spectrum photons that illuminate an array of photo-detectors that create<br>electrical signals. After the electrical signals are generated, the signals are converted<br>to digital values and the images will be displayed on monitors. The digital value can<br>be communicated to the operator console via wiring connection or wireless.<br>The proposed model includes the Non-Generator Connection Mode, allowing this<br>model to detect x-ray irradiation without direct electrical connection to the x-ray<br>generator. |
| Statement of<br>Intended Use: | The DIGITAL RADIOGRAPHY CXDI-401C Wireless provides digital image<br>capture for conventional film/screen radiographic examinations. This device is<br>intended to replace radiographic film/screen systems in all general purpose<br>diagnostic procedures. This device is not intended for mammography applications. |
| Summary of<br>Technological<br>Characteristics: | Comparisons with the predicate devices show the technological characteristics of<br>the proposed DIGITAL RADIOGRAPHY CXDI-401C Wireless device to be<br>substantially equivalent to the predicate devices. The proposed device is<br>functionally identical to the predicate devices.<br>The CXDI-401C Wireless is a modification of the CXDI-401C COMPACT cleared<br>under K103591. The differences between the CXDI-401C Wireless and the CXDI-<br>401C COMPACT are dynamic range, wireless capability, power supply, and<br>external dimensions. However, for each of these features where the CXDI-401C<br>Wireless differs from the CXDI-401C COMPACT, the CXDI-401C Wireless is<br>identical to the CXDI-701C Wireless. |
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the word "Canon" in a bold, sans-serif font. The letters are closely spaced together, creating a compact and recognizable logo. The word is presented in a simple, monochromatic color scheme, with the letters appearing in black against a white background.
**K133693**
Page 2
Tests
included
#### 5. 510(k) SUMMARY (continued)
risk hazards.
Summary of Technological Characteristics: (continued)
Summary of
Test Data:
Non-Clinical /
The CXDI-401C Wireless utilizes a flat panel detector (Scintillator [Cs]] and amorphous silicon [a-Si]) [identical to both K131106 and K103591] with a pixel pitch of 125x125um [identical to both K131106 and K103591]. 3.320 x 3,408 pixels [identical to K103591], external dimensions of 460 x 15.4mm [equivalent to K103591]. a 16bit dynamic range [identical to K131106], wireless capability fidentical to K131106], an approximate weight of 3.8kg fequivalent to K131106], and has both AC and DC power supply [identical to K131106]. The CXDI-401C Wireless can only be used with systems running Canon CXDI Control NE software.
Tests were performed on the proposed model which demonstrated that the device is safe and effective, performs comparably to the predicate device(s), and is substantially equivalent to to the predicate device(s). verification/validation testing to internal functional specifications (including software) and non-clinical image comparisons involving flat panel display images taken with the new device and the predicate devices. Documentation was provided demonstrating compliance of the CXDI-401C Wireless to all FDA requirements stated in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, including results of verification/validation plus
Documentation was provided demonstrating that the CXDI-401C Wireless RF technology is in accordance with the FDA Guidance Radio Frequency Wireless Technology in Medical Devices.
traceability of verification/validation tests to software requirements and software
Documentation was provided demonstrating that the CXDI-401C Wireless complies with the FDA requirements stated in Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices. The evaluations of the Non-clinical and Clinical Considerations of the CXD1-701C Wireless compared to the CXD1-401C Wireless, including the image quality evaluation, show the CXDI-401C Wireless to be equivalent to the CXDI-701C Wireless.
Testing confirmed that the CXDI-401C Wireless complies with the U.S. Performance Standard for radiographic equipment and with relevant voluntary safety standards for Electrical safety and Electromagnetic Compatibility testing. specifically IEC standards 60601-1, 60601-1-2, 60601-1-3, and 60601-2-32,
Testing confirmed that the CXDI-401C Wireless complies with the FCC test standard for SAR. specifically 47CFR 2.1093 and for EMI test regulations FCC Part 15 Subpart B:2012 Class A and ICES-003 Issue 5:2012 Class A.
Together, these verification/validation activities successfully demonstrated that the CXDI-401C Wireless correctly performs as designed, has been validated for its intended use, and raises no new questions regarding either safety or effectiveness when compared to the predicate device(s). Therefore, the verification/validation testing conducted supports a determination of substantial equivalence for the CXDI-401C Wireless device.
Conclusion:
Canon. Inc. - Medical Equipment Group considers the DIGITAL RADIOGRAPHY CXDI-401C Wireless device to be substantially equivalent to the predicate devices CXDI-401C COMPACT and CXDI-701C Wireless. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract eagle or bird-like figure with three curved lines representing wings or feathers.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 1, 2014
Canon. Inc. % Ms. Diane Rutherford Submissions Manager Ken Block Consulting 1201 Richardson Drive, Suite 280 RICHARDSON TX 75080
Re: K133693
Trade/Device Name: Digital Radiography CXDI-401C Wireless Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: May 23, 2014 Received: May 29, 2014
Dear Ms. Rutherford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. "The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Diane Rutherford
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Sammy H. Menge
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K133693
#### Device Name
Digital Radiography CXDI-401C Wireless
#### Indications for Use (Describe)
The Digital Radiography CXD1-401C Wireless provides digital image capture for conventional film/screen radiographic examinations. This device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Stenngh. Menge
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the The burden time for this onloolion of maintains to uces, gather and maintain the data needed and complete time to review that dolled of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.