The FLAATZ 600 Series (FLAATZ 600/601) is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures (excluding fluoroscopic, angiographic, and mammographic applications).
Device Story
FLAATZ 600 Series is a digital radiographic image acquisition system. It uses a direct conversion detector panel with an amorphous selenium (a-Se) layer to absorb X-ray photons and convert them into digital images. The system includes a detector panel, case, grid, power box, switch box, and interconnecting cables. Operated by technicians in radiology environments, the system routes image data to an operating PC via Ethernet or wireless connection. The API provides image capture and defect correction functions. Healthcare providers use the resulting images for diagnostic purposes, replacing traditional film/screen systems. The device benefits patients by providing digital diagnostic imaging for various body parts, including chest, spine, and extremities.
Clinical Evidence
Clinical study compared the FLAATZ 600 Series to the predicate FLAATZ 560 across eight body parts: chest, shoulder, L-Spine (lateral and AP), hand, forearm, foot, and knee. Results demonstrated that the FLAATZ 600 Series produces diagnostic images of equivalent quality to the predicate device.
Technological Characteristics
Direct conversion digital X-ray detector using amorphous selenium (a-Se) layer. Connectivity via Ethernet and wireless. Complies with IEC 62220-1:2003 (DQE/MTF), IEC 60601-1:2005 (Safety), IEC 60601-1-2:2007 (EMC), and NEMA PS 3.1-3.20:2011 (DICOM).
Indications for Use
Indicated for generating radiographic images of human anatomy for general-purpose diagnostic procedures. Excludes fluoroscopic, angiographic, and mammographic applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
K132842
Page 1 of 3
# 510(k) Summary
FEB 2 1 2014
| | 1. Submitter | DRTECH Corporation<br>Suite No. 2, 3 Floor, 29, Dunchon-daero541beon-gil,<br>Jungwon-gu, Seongnam-si, Gyeonggi-do, Republic of Korea<br>Post code: 462-807<br>www.drtech.co.kr |
|--|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 2. Contact Person | Name: Choul-Woo Shin<br>Position: Vice President<br>E-mail: cwshin@drtech.co.kr<br>Telephone: + 82-31-784-8856 |
| | 3. Date Prepared | August 16, 2013 |
| | 4. Device Classification Name | Stationary x-ray system |
| | 5. Device Name | Digital X-Ray Detector<br>FLAATZ 600 Series (FLAATZ 600/601) |
| | 6. Reason for Submission | New Device |
| | 7. Classification | 21 CFR 892.1680 |
| | 8. Product Code | MQB |
| | 9. Predicate Device | FLAATZ 560<br>DRTECH Corporation<br>510(k) No. K111583 |
### 10. Device Description
The FLAATZ 600 Series (FLAATZ 600/601) is a radiographic image acquisition device. It is a fully integrated image capture and routing system under human operator control. This system may be usable by a technician in a typical radiology environment.
The FLAATZ 600 Series (FLAATZ 600/601) system includes a Detector Panel, Case, Grid, Power Box, Switch Box, Interconnecting Cables, and API. The Detector Panel is a direct conversion device in the form of a rectangular plate in which the input x-ray photons are absorbed in an a-Se layer. The Power Box functions as a buffer between the Detector
DRTECH
CONFIDENTIAL Premarket Notification: FLAATZ 600 Series
{2}------------------------------------------------
Panel and Operating PC while also supplying power to the Detector Panel. The Switch Box transfers signals between the Power Box and X-ray Generator and also indicates the status of the panel using LED lights. . Finally, the API contains functions for image data capture and correction of defects on the image data.
- 11. Intended Use
The FLAATZ 600 Series (FLAATZ 600/601) is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures (excluding fluoroscopic, angiographic, and mammographic applications).
- 12. Summary of the Technological Characteristics to the Predicate Device
The modification to the predicate device is the change of data communication methods as Ethernet & Wireless type. There are no changes to the product performance specifications, device indications for use/intended use.
- 13. Summary of Non-Clinical Testing
The performance of Detector can be evaluated by DQE and MTF according to IEC 62220-1:2003 Standard.DQE & MTF have been tested by acquiring the X-ray image with designated devices. Moreover, FLAATZ 600 Series is in accordance with Safety, EMC and DICOM standards (IEC 606001-1:2005, IEC 60601-1- 2:2007 and NEMA PS 3.1-3.20:2011) The test results conclude that FLAATZ 600 Series (FLAATZ 600/601) is substantially equivalent compared to the predicate device (FLAATZ 560)
- 14. Summary of Clinical Testing
Clinical study is to investigate the diagnostic equivalency of detector panels with. the same a-Se technology and same pixel sizes. We've tested for eight body parts (Chest, Shoulder, L-Spine L, L-Spine_AP, Hand, Forearm, Foot and Knee) compared to the predicate device. The result of tests, FLAATZ 600 Series (FLAATZ 600/601) produces diagnostic images of equivalent quality as the predicate device (FLAATZ 560)
- 15. Substantial Equivalence
The proposed FLAATZ 600 Series (FLAATZ 600/601) has identical indication for use / intended use, the principal of operation and user environment. The proposed FLAATZ 600 Series (FLAATZ 600/601) has similar performance in terms of DQE, MTF and is the same physical characteristics as using Amorphous Selenium (a-Se) material. Moreover, FLAATZ Series (FLAATZ 600/601) is the upgraded model of its
DRTECH
CONFIDENTIAL Premarket Notification: FLAATZ 600 Series
{3}------------------------------------------------
K1327842
Page 3 of 3
previous model FLAATZ 560. TFT panel and main components are almost same.
- 16. Functional and Safety Testing
The FLAATZ 600 Series (FLAATZ 600/601) has been evaluated as per FDA's "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices" and has shown good performance, substantially equivalent to the predicate device.
The FLAATZ 600 Series (FLAATZ 600/601) has also met applicable Electro Magnetic Compatibility (EMC) requirements.
- 17. Conclusion
The FLAATZ 600 Series (FLAATZ 600/601) is substantially equivalent to the Predicate Device in design and performance.
- 18. Manufacturing Facility
- DRTECH Corporation Suite No. 2, 3 Floor, 29, Dunchon-daero541beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Republic of Korea Post code: 462-807
Establishment Registration Number: 3005172103
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with text around the perimeter. Inside the circle is a stylized design featuring three overlapping shapes that resemble human figures or abstract forms. The text around the circle reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 21, 2014
DRTECH Corporation % Choul-Woo Shin 333-1, Sangdaewon 1-Dong Jungwon-Gu Seongnam-Shi, Gyeonggi-Do, 462-807 SOUTH KOREA
Re: K132842
Trade/Device Name: FLAATZ 600 Series (FLAATZ 600/601) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: January 13, 2014 Received: January 24, 2014
Dear Choul-Woo Shin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2- Choul-Woo Shin
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director. Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K132842
Device Name
FLAATZ 600 Series (FLAATZ 600/601)
#### Indications for Use (Describe)
The FLAATZ 600 Series (FLAATZ 60060) is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures (excluding fluoroscopic, and mammographic applications).
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.f)
FORM FDA 3881 (1/14)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.