K123005 · Swissray Medical AG · MQB · Dec 7, 2012 · Radiology
Device Facts
Record ID
K123005
Device Name
DDRVERSA MOTION
Applicant
Swissray Medical AG
Product Code
MQB · Radiology
Decision Date
Dec 7, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The ddR Versa™ Motion System is intended for use by a qualified/trained doctor or technician on subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)
Device Story
Digital diagnostic X-ray system; captures radiographic images of skull, spine, chest, abdomen, and extremities. System utilizes ceiling-mounted X-ray tube and two digital X-ray receptor panels (wall-mounted and table-mounted). Operates via Windows-based workstation with touch-screen monitor, keyboard, and mouse. Used in clinical settings by qualified doctors or technicians. Transforms X-ray energy into digital images using A-Si TFT flat panel detectors with Cesium iodide scintillators. Output provides diagnostic-quality images for clinical review and decision-making. Benefits include improved workflow through motorized table, automatic positioning, and dual-panel configuration.
Clinical Evidence
Clinical images were obtained per FDA Guidance Document on Solid State Imaging Devices. Images were compared to predicate device images and evaluated by professional radiologists, who confirmed good diagnostic quality. No quantitative metrics (e.g., sensitivity/specificity) were reported.
Technological Characteristics
Digital X-ray system; high-frequency generator (64kW or 80kW); Trixell 4343 RC or Pixium Portable 3543 EZ A-Si TFT flat panel detectors with Cesium iodide (CsI) scintillator; 16-bit A/D conversion; 148 μm pixel size; 2880 x 2880 pixel matrix; 40-150 kVp energy range. Motorized table; ceiling-mounted tube support. Connectivity via Windows workstation. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for patients requiring diagnostic radiographic imaging of the skull, spine, chest, abdomen, and extremities. Suitable for patients in sitting, standing, prone, or supine positions. Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
## 510(k) Summary: K123005
- 1. Submitter: Swissray Medical AG Turbistrasse 25 - 27 CH-6280 Hochdorf Switzerland Phone +41 41 914 12 12 Fax +41 41 914 12 13 Date Prepared: November 13, 2012 Contact: Markus Bütler, Quality Manager
- Identification of the Device: ddRVersa™ Motion (Digital Diagnostic X-Ray System); 2. Recommended classification regulation: 21 CFR 892.1650, 892.1680 Device class: II, Panel: Radiology, Product code: MQB and KPR
- Predicate Device: This is a MODIFICATION to ddRElement™ under document number K110828 and 3. Suinsa (Now owned by Sedecal) (K083109).
- 4. A description of the device: ddRVersa™ Motion uses the same core technology as our ddRElement but is modified in that: The supplier for the digital x-ray receptor panel has changed. The system can now accommodate two digital x-ray receptor panels instead of just one. A Wi-Fi digital x-ray receptor panel may be chosen as the second panel. The tubestand is changed from a C-arm configuration to an overhead plus wall stand configuration. The table is changed from non-motorized to motorized. The generator supplier has been changed. The collimator has been changed. (See comparison table below)
- 5. Intended for use by a qualified/trained doctor or technician on Intended use of the device: subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography) (SAME as predicate)
- The ddRVersa™ Motion has essentially the same technological characteristics (i.e., design, material, 6. chemical composition, energy source) as the predicate device ddRElement™. See the comparison table below. There are really only two main differences: The tubehead is now an overhead version and the system can now accommodate two digital panels. The panel resolution and technology remains the same.
### Comparison Table
| Characteristic | Suinsa (Now<br>owned by<br>Sedecal)<br>(K083109) | Swissray<br>ddRElement™<br>(K110828) | Swissray ddRVersa™<br>Motion<br>(K123005) (New<br>modified device) |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Overall Configuration | Image: Suinsa (Now owned by Sedecal) (K083109) | Image: Swissray ddRElement™ (K110828) | Image: Swissray ddRVersa™ Motion (K123005) (New modified device) |
| Characteristic | Suinsa (Now<br>owned by<br>Sedecal)<br>(K083109) | Swissray<br>ddRElement™<br>(K110828) | Swissray ddRVersa™<br>Motion<br>(K123005) (New<br>modified device) |
| Intended Use | General purpose<br>radiography except<br>not for<br>mammography | SAME | SAME |
| Control System:<br>Positioning | Software Driven with<br>Touch Panel LCD<br>(Sedecal product) | Manual | Software Driven via<br>Windows workstation with<br>touch-screen monitor,<br>keyboard, and mouse |
| Control System: Exposure | Software Driven with<br>Touch Panel LCD<br>(Sedecal product) | Software Driven via<br>Windows workstation with<br>touch-screen monitor,<br>keyboard, and mouse<br>(Samsung LTX240AA01-A,<br>K090742) | Software Driven via<br>Windows workstation with<br>touch-screen monitor,<br>keyboard, and mouse<br>(Pixium 4343 RC or Portable<br>3543 EZ) |
| Configuration: Tube | Ceiling-mounted<br>single tube; automatic<br>positioning (utilizing<br>Sedecal Nova FA X-<br>Ray tube overhead<br>support system) | Single tube column-<br>mounted on U-Arm with<br>detector; manual<br>positioning | SAME AS SEDECAL<br>Ceiling-mounted single<br>tube; automatic positioning<br>(utilizing Sedecal Nova FA X-<br>Ray tube overhead support<br>system) |
| Configuration:<br>Detector/Image Receptor | Not applicable | One detector system:<br>Column-mounted U-Arm<br>with tube with Samsung<br>LTX240AA01-A panel<br>(K090742) | Two detector system: Wall-<br>mounted (column) and<br>Table-mounted |
| Receptor Details | Not applicable | Technology Single A-Si TFT<br>+ photodiode plate<br>Cesium iodide scintillator<br>(Csl)<br>A/D conversion 14 bit<br>Active detector area 43 cm<br>x 43 cm<br>Spatial resolution 3.0<br>lp/mm<br>Active pixel matrix 3072 x<br>3072 pixels<br>Pixel size 143 μm<br>Energy range 40-150 kVp<br>Sensitivity > 120 LSB/μGy | Technology Single A-Si TFT +<br>photodiode plate Cesium<br>iodide scintillator (Csl)<br>A/D conversion 16 bit<br>Active detector area 43 cm<br>x 43 cm<br>Spatial resolution 3.5<br>lp/mm<br>Active pixel matrix Active<br>pixel matrix 2880 x 2880<br>pixels<br>Pixel size 148 μm<br>Energy range 40- 150 kVp<br>Sensitivity up to 850<br>LSB/μGy |
| Generator: Type | High frequency made<br>by Sedecal | CPI Canada CMP200 high<br>frequency generator | High Frequency SHF<br>635/645 or SHF 835/845<br>made by Sedecal |
| Generator: Capacity | 30 kW to 64 kW | 65 - 80 kW | 64kW or 80kW |
| Characteristic | Suinsa (Now<br>owned by<br>Sedecal)<br>(K083109) | Swissray<br>ddRElement™<br>(K110828) | Swissray ddRVersa™<br>Motion<br>(K123005) (New<br>modified device) |
| Detector/Image Receptor:<br>Resolution | Not applicable | Samsung LTX240AA01-A<br>(K090742) | Trixell 4343 RC a-si TFT flat<br>panel or Pixium Portable<br>3543 EZ : 148 microns pixel<br>size, with 2880 x 2880 pixels<br>(8.2 mil pixels) 16 bit |
| Tube: Type | Toshiba 7869x 0.6/1.2 | Dunlee DR1436CRSR | Toshiba 7869x 0.6/1.2 |
| Tube: Capacity | 150KV | 150KV | 150KV |
| Collimator + cu filter | Huestis | Siemens ML02 | Ralco 225 43x43cm<br>Motorized filters<br>1mm AL +0.1mm Cu<br>1mm AL +0.2mm Cu<br>2mm AL |
| Energy Source: | 120V, 230/240 V<br>50/60 Hz | SAME | SAME |
| Performance Standard | 21 CFR 1020.30 | SAME | SAME |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 |
DEC 7 2012
{1}------------------------------------------------
:
.
:
{2}------------------------------------------------
- 7. Description of non-clinical tests. The modified unit has undergone electrical safety and electromagnetic compatibility testing, as well as software validation and risk analysis. The technical characteristics of the new panel have been measured and included in the bench testing information.
- 8. Description of clinical tests. Clinical images were obtained in accordance with the FDA Guidance Document on Solid State Imaging Devices. They were compared to our predicate images and evaluated by professional radiologists and found to be of good diagnostic quality.
- 9. Conclusions drawn: The nonclinical and clinical tests that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph 3, above.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem is a graphic representation of an eagle or bird-like figure, with three curved lines forming its body and wings. The overall design is simple and recognizable, representing the department's role in health and human services.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 7, 2012
Swissray Medical AG % Mr. Daniel Kamm, P.E. Submission Correspondent Kamm & Associates 8870 Ravello Ct NAPLES FL 34113
Re: K123005
Trade/Device Name: ddRVersa™ Motion Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: MQB Dated: November 16, 2012 Received: November 20, 2012
Dear Mr. Kamm:
We have reviewed your Section.510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
Janine M. Morris, M.S. Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123005
Device Name: ddRVersa™ Motion
Indications for Use:
The ddR Versa™ Motion System is intended for use by a qualified/trained doctor or technician on subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) K123005
Page 1 of
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.