SATURN 9000-810

K080589 · New Medical Co., Ltd. · MQB · Jun 11, 2008 · Radiology

Device Facts

Record IDK080589
Device NameSATURN 9000-810
ApplicantNew Medical Co., Ltd.
Product CodeMQB · Radiology
Decision DateJun 11, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1680
Device ClassClass 2

Indications for Use

The Saturn 9000-810 system generates digital X-ray images that can be used for general X-ray system not for Mammography. The Saturn 9000-810 system can interface to traditional X-ray generator and get digital X-ray image. The Saturn 9000-810 is intended to be used in same clinical application as traditional film-screen based general radiography system.

Device Story

Saturn 9000-810 is a digital X-ray imaging system; utilizes a-Se (amorphous selenium) flat panel detector to convert X-ray photons into electrical charges; matrix scan array reads charges into modulated electrical signals. System includes X-ray generator interface, acquisition workstation (AWS) with monitor, keyboard, mouse, and computer. Operated by clinical staff; AWS manages image acquisition, processing, display, database management, and transmission to remote viewing sites or archives. Replaces traditional film-screen radiography; provides digital images for diagnostic review; facilitates electronic storage and workflow efficiency.

Clinical Evidence

Bench testing only. Performance evaluated against IEC 62220-1 standards to confirm compatibility with predicate device. No clinical data provided.

Technological Characteristics

Digital X-ray system; a-Se (amorphous selenium) flat panel detector; X-ray generator interface; acquisition workstation (AWS). Connectivity: supports remote viewing and electronic archiving. Standards: IEC 62220-1. Regulatory Class II, Product Code MOB.

Indications for Use

Indicated for general radiographic imaging applications, excluding mammography. Intended for use in clinical settings where traditional film-screen radiography is utilized.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Premarket Notification New Medical Co.,LTD. - Saturn 9000-810 January 3d, 2008 Koross.9 JUN 1 1 2008 ## 510 (K) Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92 9 (c) Image /page/0/Picture/5 description: The image shows a stylized letter N in bold, sans-serif font. The letter is predominantly black, with a subtle gradient effect that gives it a three-dimensional appearance. The right side of the N has a textured pattern, resembling vertical lines or ridges, adding depth and visual interest to the design. 1. Submitter: New Medical Co.,LTD. 4F, No.42, SingZhong Rd., NeiHu District Taipei 114, Taiwan Contact Person: Young-Hoon Shin, Ph.D. R&D Manager yhshin@newmedical.com.tw Telephone: +886-2-8791-6268 January 30 2008 Date Prepared: 2. Device Name: Saturn 9000-810 - 3. Market Device: new submission - 4. Device Description: The Saturn 9000-810 is equipped with a digital detector. X-ray generator interface, and image acquisition workstation. The digital detector is a-Se deposited plat panel detector. The input X-ray photons are absorbed in a-Se layer that creates an electrical charge which is representation of the X-ray input. The charge is read-out by a matrix scan of the array that converts the charges into a modulated electrical signal. X-ray exposure is controlled by switch box, which also controls the acquisition timing of the detector. Saturn 9000-810 includes acquisition workstation ("AWS") monitor, keyboard and mouse, computer, electronics, and accessory storage. The resultant output signal can be transmitted to remote viewing sites, and/or it can be stored electronically for later viewing. The AWS is used for image acquisition, processing, and display. The AWS can also be used for database management and can send images to archive, review or filming. 5. Indications for Use: The Saturn 9000-810 system generates digital X-ray images that can be used for general X-ray system not for Mammography. The Saturn 9000-810 system can {1}------------------------------------------------ interface to traditional X-ray generator and get digital X-ray image. The Saturn 9000-810 is intended to be used in same clinical application as traditional filmscreen based general radiography system. - 6. Comparison with Predicate Device: The Saturn 9000-810 Is substantially equivalent to the Saturn 9000. It has the same technological characteristic and is comparable in key safety and effectiveness features. It utilizes similar design, construction, and interface scheme. The only difference is the pixel size and imaging area of the detector. It has same intended uses and basic operation modes as the predicate device. - 7. Non-cilnical Tests: The performance of the device has been evaluated and has been founded to confirm with applicable medical device performance and safety standards. Based on industrial standard, IEC 62220-1, many technical characteristic values were measured and computed and it is conformed that the device performance is compatible with predicate device. - 8. Clinical Tests: Not applicable for special 510(k) procedure. - 9. Conclusion: Intended uses and other key features are consistent with previously cleared solid state X-ray imaging devices. The device conforms to applicable medical device safety standards and compliance was verified. Saturn 9000-810 is substantially equivalent to currently marketed devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the Department of Health and Human Services. The logo consists of a symbol on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The symbol is a stylized representation of a human figure, and the text is in a sans-serif font. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 AUG 2 3 2013 Mr. Young-Hoon Shin Director of R&D New Medical Company, Ltd. 4F. No. 42. SingZhong Road NEIHU DISTRICT TAIPEI 114 TAIWAN Re: K080589 Trade/Device Name: Saturn 9000-810 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: January 15, 2008 Received: May 16, 2008 Dear Mr. Shin: This letter corrects our substantially equivalent letter of June 11, 2008. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {3}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice mourements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirents as set forth in the quality 2) vice as described in your Section 510(k) premarket witi anow you to ocgin finding of substantial equivalence of your device to a legally marketed notheation. The I Dri imanig of Succion for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and If you desire specific advice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-007), product on the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to ulder the MDA regulation (27 °CF RT at 005); produced/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if Known): To be assigned Device Name: Satum 9000-810 Indications For Use: The Saturn 9000-810 system generates digital X-ray images that can be used for general X-ray system not for Mammography. The Saturn 9000-810 system can interface to traditional X-ray generator and get digital X-ray image. The Saturn 9000-810 is intended to be used in same clinical application as traditional film-screen based general radiography system. Prescription Use V AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CIONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) tonth Whang (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...