The CR25.0 is indicated for use to provide diagnostic quality images to aid in physician diagnosis. The CR25.0 is intended to be used mainly in chest, skeletal, and gastro-intestinal x-ray imaging applications.
Device Story
CR25.0 is a computed radiography imaging system. It utilizes an imaging plate coated with photo-stimulatable storage phosphors to capture X-ray latent images. A laser scanner reads the plate, converting the latent image into a digital bit stream. Data is stored locally or transmitted to a PACS network via DICOM. The device is a compact, lower-cost version of the ADC Compact Plus, featuring reorganized, smaller electronics with lower power requirements. It is intended for use by healthcare professionals in clinical settings to provide diagnostic quality images for physician review and diagnosis. The system maintains the resolution and image quality of its predicate.
Clinical Evidence
No clinical data. Bench testing only; performance demonstrated through in-house reliability and imaging performance testing, and compliance with EN 60601-1-1 and EN 60601-1-2 standards.
Technological Characteristics
Computed radiography system using photo-stimulatable storage phosphor imaging plates. Laser scanning mechanism for digital conversion. DICOM connectivity. Complies with EN 60601-1-1 and EN 60601-1-2 safety standards. Compact form factor with reorganized electronics.
Indications for Use
Indicated for patients requiring chest, skeletal, and gastro-intestinal x-ray imaging to aid in physician diagnosis.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K053634 — COMPUTED RADIOGRAPHY SYSTEM WITH NX 1.0 WORKSTATION · Agfa Corporation · Jan 17, 2006
K032210 — ALARA CRYSTALVIEW COMPUTED RADIOGRAPHY SYSTEM · Alara, Inc. · Oct 2, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 2 2 2004
041701
# 510(k) Summary
## CR25.0
Common/Classification Name: Computed Radiography, 21 CFR 892.1650
Agfa Corporation 10 South Academy Street Greenville, SC 29602-9048
Contact: Jeffery A. Jedlicka, Prepared: June 17, 2004
#### LEGALLY MARKETED PREDICATE DEVICES A.
This is a Special 510(k) for Device Modification. The predicate device is the device that was modified to produce the CR25.0, namely the ADC Compact Plus, which was cleared by FDA on. 28 September 2001 as K013138.
#### B. DEVICE DESCRIPTION
The ADC Compact Plus, the predicate device, is a computed radiology imaging system. Instead of screens and photographic film for producing the diagnostic image, the ADC Compact Plus system utilizes an "imaging plate." a plate coated with photo-stimulatable storage phosphors that are sensitive to X-rays and capable of retaining a latent image. This imaging plate is inserted into a device that scans it with a laser and releases the latent image in the form of light which is converted into a digital bit stream. The bit stream of image data is stored locally and can also be stored in a PACS network in DICOM format.
The CR25.0 is very similar to the ADC Compact Plus and the ADC Solo. The electronics are being reorganized and made smaller, which will result in lower power requirements. However, the basic principles of operation are unchanged. Instead of upgrading the currently marketed economy system called the ADC Solo, components of the high-end ADC Compact Plus were reintegrated into a compact lower-cost system, resulting is no loss of resolution or other measures of image quality.
#### C. INTENDED USE
The CR25.0 is indicated for use to provide diagnostic quality images to aid in physician diagnosis. The CR25.0 is intended to be used mainly in
{1}------------------------------------------------
chest, skeletal, and gastro-intestinal x-ray imaging applications.
#### SUBSTANTIAL EQUIVALENCE SUMMARY D.
The CR25.0 has the same indications for use as the legally marketed predicate device, so the first decision point in the 510(k) Decision Algorithm is straight-forward. The CR25.0 has the same technological characteristics as the predicate device. This premarket notification has described most of the characteristics of the CR25.0 in sufficient detail to assure substantial equivalence. For the few characteristics that may not be precise enough to ensure equivalence, performance data was collected, and this data demonstrates substantial equivalence. In keeping with the format of a Special 510(k) for Device Modification, performance data were not included in the submission, but the declarations in Section IV provide certification that the data demonstrate equivalence.
#### TECHNOLOGICAL CHARACTERISTICS E.
The technological characteristics are the same in the proposed and predicate devices.
#### TESTING F.
The CR25.0 has been tested for proper performance to specifications through various in-house reliability and imaging performance demonstration the significal device also meets the requirements of EN 60601-1-1 and EN 60601-1-2.
#### CONCLUSIONS G.
This Special 510(k) for Device Modification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three tail feathers and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. The eagle is depicted in a simple, abstract style, with bold lines forming its wings and body.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Jeffery A. Jedlicka Manager of Regulatory Affairs AGFA Corporation Healthcare 10 South Academy Street GREENVILLE SC 29601
Re: K041701 Trade/Device Name: CR25.0 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: June 22, 2004 Received: June 22, 2004
AUG 23 2013
Dear Mr. Jedlicka:
This letter corrects our substantially equivalent letter of July 22, 2004.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name:___ CR25.0
Indications For Use:
The CR25.0 is indicated for use to provide diagnostic quality images to aid in physician diagnosis. The CR25.0 is intended to be used mainly in chest, skeletal, and gastrointestinal x-ray imaging applications.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel B. Simmons
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number KOH701
Page 1 of 1
000046
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.