The Fuji Computed Radiography (FCR) Velocity Image Reader with Image Plate (IP) reading is intended to be used for the identification, capture, digitization, and processing of diagnostic x-ray images, and associating palient and exam identification with the images.
Device Story
FCR Velocity is a computed radiography image reader; captures x-ray images using photostimulable phosphor imaging plates (IPs) composed of europium-activated barium fluorohalide compounds. Device features built-in IP in an upright exam stand configuration. Process: x-ray irradiated IP moved to reading position; latent image read via laser emission (658 nm laser diodes); photostimulated luminescence collected, sampled, and digitized. Image data associated with patient/exam ID; digitally processed per user parameters; output to CRT monitor, hard copy printer, workstation, or optical disk. Used in clinical settings by radiology staff. Benefits include efficient digital capture of diagnostic x-ray images, replacing traditional film/intensifying screens.
Indicated for identification, capture, digitization, and processing of diagnostic x-ray images and patient/exam identification association. No specific patient population, age, or gender restrictions provided.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
NOV 26 2003
K033561
# 1&1-Imaging & Information
# FUJIFILM MEDICAL SYSTEMS USA, INC.
419 WEST AVENUE STAMFORD, CT 06902 Telephone: 203/324-2000 Fax: 203/353-0926
# 510(K) Summary
In accordance with the requirements of the Safe Medical Device Act, FUJIFILM Medical Systems, USA, Inc. herewith submits a 510(K) summary of safety and effectiveness for the following device.
| SUBMITTER NAME / ADDRESS: | FUJIFILM Medical Systems, USA, Inc.<br>419 West Avenue<br>Stamford, CT 06902 |
|---------------------------|------------------------------------------------------------------------------|
| CONTACT PERSON / TEL NO: | Frank Gianelli<br>Regulatory Coordinator<br>Tel No: (203) 602-3774 |
| DATE SUMMARY PREPARED: | August 21, 2003 |
| ESTABLISHMENT NO.: | 2443168 |
| TRADE/PROPRIETARY NAME: | Fuji Computed Radiography (FCR) Velocity |
| COMMON/USUAL NAME: | Computed Radiography Image Reader |
| CLASSIFICATION NAME: | Solid State X-Ray Imager |
| CLASS/PANEL: | Class II, 90-MQB, 21CFR 892.1650 |
| PREDICATE DEVICE(S): | Fuji FCR 9501-HQ |
#### DEVICE DESCRIPTION:
A Fuji Computed Radiography (FCR) system typically consists of an image reader (IR), palient ID terminal, imaging plates (IPs), IP cassettes, interface board, positioning monitor, laser printer for hard copy oulput, and optionally an image workstation, optical disk file, and network interface. This notification is for the image reader and associated imaging plates. IPs are used as twodimensional radiation detectors in place of radiographic film and intensifying screens to capture a portion of the projected x-ray patient image in space. In the image reader, the captured image data is associated electronically with patient and exam identification data and the latent image is read by laser emission by the phenomenon of photostimulable luminescence. The photostimulated fuminescence is collected, detected, sampled, and digitized. The image data is then digitally processed according to exam and user-specified parameters and may be displayed on a CRT monitor to confirm patient positioning, printed by a hard copy device (such as laser printer, or dry printer), or output to a workstation, optical disk file, or other destination. The device performs no lossy compression of image data.
FCR Velocity consists of an Image Reader and an Imaging Plate (described below). The Image Reader is cassetteless because the Image Plate is built into the upright exam stand configuration of the FCR Velocity Image Reader. Imaging plates are exposed via conventional X-ray devices. The Xray irradiated IP is then moved from the exposure position to the reading position, and images are read. After reading, the IP is erased, and moved to the exposure position again.
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the Fujifilm symbol on the left and the word "FUJIFILM" in bold, black letters on the right. Below the word "FUJIFILM" is the text "I&I-Imaging & Information" in a smaller font size.
### FUJIFILM MEDICAL SYSTEMS USA, INC.
141-Imaging & Informatio
419 WEST AVENUE STAMFORD. CT 06902 Telephone: 203/324-2000 Fax: 203/353-0926
As with other FCR image readers, the FCR Velocity will feature a photostimulable phosphor imaging plate (IP) composed of europium activated barium fluorohalaid compounds in a crystal form held in an organic binder. The IP used with the subject device will have a rigid substrate, which enables it to be held in a constant plane. IPs are used as two-dimensional radiation detectors in place of radiographic film and intensitying screens to capture a portion of the projected x-ray palient image in space. In the image reader, the captured image data is associated electronically with patient and exam identification data and the latent image is read by laser emission by the phenomenon of photostimulable luminescence. The photostimulated luminescence is collected, sampled, and digitized. The image data is then digitally processed according to exam and user-specified parameters and may be displayed on a CRT monitor to confirm patient positioning, printed by a hard copy device (such as laser printer), or dry printer), or output to a workstation, optical disk file, or other destination. The device performs no lossy compression of image data.
#### INTENDED USE:
The Fuji Computed Radiography (FCR) Velocity Image Reader with Image Plate (IP) reading is intended to be used for the identification, capture, digitization, and processing of diagnostic x-rav images, and associating palient and exam identification with the images.
#### PREDICATE DEVICE AND SUBSTANTIAL EQUIVALENCE INFORMATION:
FCR Velocity is considered comparable and substantially equivalent to the FCR 9501-HQ Image Reader manufactured by Fuji. FCR 9501-HQ has been granted a 510(k) clearance. Refer to accession number K951373.
| PARAMETER | FUJI FCR VELOCITY U | FUJI FCR 9501HQ | | |
|----------------------------|------------------------------|------------------------------|-----------|----|
| Image Recording | | | | |
| Patient Identification | Digital Data (from Console) | Digital Data (from Console) | | |
| Recording Method | Photostimulable Luminescence | Photostimulable Luminescence | | |
| No. of Imaging Plates Used | One Built-in Imaging Plate | Four Built-in Imaging Plates | | |
| Imaging Plate Size | 497x456 mm | 495x377 mm | | |
| Image Reading | | | | |
| Reading Laser | Laser Diodes (658 nm) | He-Ne Laser (675 nm) | | |
| Gray Scale | 10 bits (1024 gray levels) | 10 bits (1024 gray levels) | | |
| Sampling Raster | IP Reading Area | IP Reading Area | | |
| | Pixels/mm | Pixels/mm | | |
| | 17x17 in. | 10 | N/A | |
| | 17x14 in. | 10 | N/A | |
| | 14x17 in. | 10 | 14x17 in. | 10 |
| | 14x14 in. | 10 | 14x14 in. | 10 |
| | 10x12 in. | 10 | 10x12 in. | 10 |
| | 12x10 in. | 10 | | |
| | 8x10 in. | 10 | 8x10 in. | 10 |
| | 10x8in. | 10 | | |
| | 18x43 cm | 10 | N/A | |
| Physical | | | | |
| WxHxD (mm) | 645x1835x450 (Image Reader) | 1163x1733x935 (Image Reader) | | |
| Weight (kg) | 220 | 580 | | |
| Throughput (Approximate) | 240 14x17 in. IP's/hr. | 80 14x14 in. IP's/hr | | |
| Processing Time | 80 seconds | 132 seconds | | |
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in bold, sans-serif font, with a stylized "F" symbol to the left. Below the main logo, the text "I&I-Imaging & Information" is printed in a smaller font size.
# FUJIFILM MEDICAL SYSTEMS USA, INC.
419 WEST AVENUE STAMFORD, CT 06902 Telephone: 203/324-2000 Fax: 203/353-0926
#### SAFETY INFORMATION:
FCR Velocity introduces no new safety and efficacy issues other than those already identified with the predicale device. The results of a hazard analysis, combined with the appropriate preventive measures taken indicate that the device is of minor level of concern as per the May 29, 1998 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
FCR Velocity complies with the following mandatory and voluntary standards:
- 21 CFR 1040.10 Performance Standards for Light Emitting Products (Laser Products) .
- 21 CFR 1020.30 Performance Standards for lonizing Radiation Emitting Products . (Diagnostic X-ray System and their Major Components)
- Medical Electrical Equipment Part 1: General Requirements for Safety UL Standard 60601-1 . (IEC 60601-1-1 included)
- . Medical Electrical Equipment - Part 1: General Requirements for Safety; Electromagnetic Compatibility - Requirements and Tests: IEC 60601-1-2
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper half of the circle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
FUJIFILM Medical Systems USA, Inc. % Mr. William J. Sammons Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Drive, P.O. Box 13995 Research Triangle Park, NC 27709-3995
AUG 23 2013
Re: K033561
Trade/Device Name: Fuji Computed Radiography (FCR) Velocity Image Reader (CR-IR364) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: November 5, 2003 Received: November 12, 2003
Dear Mr. Sammons:
This letter corrects our substantially equivalent letter of November 26, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice modion as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirements as be form ar an are ag your device as described in your Section 510(k) premarket wifi anow you to began finding of substantial equivalence of your device to a legally marketed predication. The results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you desire specific darres to: your use in Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket o 150. Thise, promotion and 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name : FCR Velocity Image Reader (CR-IR364)
Indications For Use :
The indications for use of the Fuji Computed Radiography (FCR) Velocity Image Reader with Image Plate (IP) reading is the identification, capture, digitization, and processing of diagnostic x-ray images, and associating palient and exam identification with the images.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive Abdominal
Division of Reproductive, Abdo and Radiological Devices 510(k) Number
Prescription Use ( Per 21 CFR 801.109
Over-The-Counter Use
(Optional Formal 1-2-96)
Page A3-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.