K994378 · Hitachi Medical Systems America, Inc. · MOS · Mar 7, 2000 · Radiology
Device Facts
Record ID
K994378
Device Name
AIRIS II
Applicant
Hitachi Medical Systems America, Inc.
Product Code
MOS · Radiology
Decision Date
Mar 7, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The AIRIS II Magnetic Resonance Diagnostic Device with the CTL Phased Array Coil P/N MR-CTL-51 provides for continuous imaging of the cervical, thoracic, and lumbar spine and adjacent anatomies. The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Device Story
The CTL Phased Array Coil (P/N MR-CTL-51) is an accessory for the Hitachi AIRIS II MRI system. It utilizes an 8-element multiple receiver coil configuration to enable continuous imaging of the cervical, thoracic, and lumbar spine. The coil operates in three selectable modes (C, T/L, L) using a multi-position switch to route current through PIN diode-activated high-impedance tanks. The device captures RF emissions from hydrogen nuclei excited by magnetic resonance, which are spatially encoded via magnetic field gradients. The system processes these signals into cross-sectional images (transverse, coronal, sagittal, oblique, curved). Used in clinical settings by physicians to visualize internal structures, the output aids in diagnosis and facilitates real-time MR guidance for minimally invasive procedures. The coil design includes a removable nosepiece bridge for the C-spine section and removable loop sections for T/L and L sections.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
8-element phased array RF coil; PIN diode-activated high-impedance tanks for mode switching; quadrature and linear element combinations; compatible with AIRIS II MRI system; ACR/NEMA/DICOM 3 compliant; non-invasive; no ionizing radiation.
Indications for Use
Indicated for continuous imaging of cervical, thoracic, and lumbar spine and adjacent anatomies, as well as head, body, and extremities. Used for non-invasive diagnostic imaging to provide physiological and clinical information via proton density, T1, T2, and flow measurements. Contraindications include use of non-MRI-compatible instrumentation.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K994378
## Attachment 1
# 510(k) Summary of Safety and Effectiveness
{1}------------------------------------------------
#### SUBMITTER INFORMATION: 1.0
- Hitachi Medical Systems America 1.1 Submitter: 1959 Summit Commerce Park Twinsburg, OH 44087 PH: 330 425-1313 FX: 330 425-1410
- James Jochen Rogers 1.2 Contact:
- 1.3 Date: December 21, 1999
#### DEVICE NAME: 2.0
- Radiology 2.1 Classification Panel:
- 2.2 Classification Number: 892.1000 Magnetic Resonance Diagnostic Device
- Product Nomenclature: 2.3 System, Nuclear Magnetic Resonance Imaging
- 2.4 Product Code(s): 90LNH 90MOS (Magnetic Resonance Specialty Coil)
- 2.5 Trade/Proprietary Name: AIRIS II
- 2.6 PREDICATE DEVICE(s):
Hitachi AIRIS II
#### 3.0 DEVICE DESCRIPTION:
- 3.1 FUNCTION
Identical to the AIRIS II 510(k) K974212.
The AIRIS II MRI Magnetic Resonance Diagnostic Device is being enhanced by one additional RF Coil (CTL Phased Array Coil) to increase the clinical utility of the AIRIS II in the stationary configuration.
The CTL Phased Array Coil P/N MR-CTL-51 provides for continuous imaging of the cervical, thoracic, and lumbar spine and adjacent anatomies, and utilizes a proprietary multiple receiver coil configuration (8 elements) operating in the phased array mode. The coil operates in three modes:
| Imaging Mode | Coil Size |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| C | One quadrature combination, saddle and<br>solenoid, identical to current QD -spine |
| T/L | Two quadrature combinations, one routed<br>through the solenoid and saddle inputs of<br>channel one, the other consists of one element<br>combined in quadrature in the coil with another,<br>this signal subsequently is routed into the saddle<br>input of channel two |
| L | One quadrature combination, saddle and<br>solenoid, and one linear saddle element |
A multi-position switch is used to select between the modes, by routing a current through PIN diode-activated high impedance tanks.
{2}------------------------------------------------
The C-spine section of the coil is nearly identical in design to the USAI QD C-spine coil P/N MR-QCS-51, and features a removable nosepiece bridge. The T/L and L sections are combinations of flat elements in the coil's base and removable loop sections, connecting to mounting blocks at the edge of the coil.
AIRIS II software did not need to be revised in order to support full functionality of this CTL Phased Array Coil.
#### SCIENTIFIC CONCEPTS 3.2
Magnetic Resonance (MR) is based on the fact that certain atomic nuclei have electromagnetic properties which cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nucleus used in current imaging experiments in magnetic resonance. When placed in a magnetic field, there is a slight net orientation or alignment of these atomic nuclei with the magnetic field. The introduction of a short burst of radiofrequency (RF) excitation of wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a reorientation of the proton's magnetization vector. When the RF excitation is removed, the proton relaxes and returns to its original orientation. The rate of relaxation is exponential, and varies with the character of the proton and its adjacent molecular environment. This reorientation process is characterized by two exponential relaxation times called T1 and T2 which can be measured.
These relaxation events are accompanied by an RF emission or echo which can be measured and used to develop a representation of these emissions on a three dimensional matrix. Spatial localization is encoded into the echo by varying the RF excitation and by appropriately applying magnetic field gradients in x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of NMR characteristics of the nuclei under consideration can be constructed by using image processing techniques similar to those used in CT.
For magnetic fields up to 1.5T, the RF frequencies commonly used range up to 65MHz. The RF fields have pulse powers from several watts to greater than 10 kilowatts, and repeat at rates from once every few seconds to greater than fifty per second. The timevarying magnetic gradient fields have a typical duration of sub-millisecond to several milliseconds.
#### PHYSICAL AND PERFORMANCE CHARACTERISTICS 3.3
MR is currently of great interest because it is capable of producing high quality anatomical images without the associated risks of ionizing radiation. In addition, the biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In x-ray imaging, differences in x-ray attenuation, largely based on differences in electro density are responsible for the contrast observed in x-ray images. In MR imaging, differences in proton density, blood flow, and relaxation times T1 and T2 all may contribute to image contrast. In addition, by varying the duration and spacing of the RF pulses, images may be produced in which the contrast is primarily dependent on T1 relaxation, T2 relaxation, proton density, or a combination of all three.
#### DEVICE INTENDED USE: 4.0
The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or
{3}------------------------------------------------
extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
- Head, Body, Spine, Extremities Anatomical Region: ●
- Nucleus excited: Proton
- Diagnostic uses:
2D T1- / T2-weighted imaging T1, T2, proton density measurements
MR Angiography
image processing
Imaging capabilities:
- 2D Spin Echo (SE); also with rephase
- 2D Gradient Field Echo (GE); also with rephase (2D GR)
- 2D Steady state acquisition with rewinded GE (SARGE); also with RF spoiling, rephasing
- 2D Inversion Recovery (IR)
- 2D Fast Spin Echo (FSE); also with rephase
- 2D Fast Inversion Recovery (FIR)
- 2D Dual Slice acquisition (SEDS)
- 3D Spin Echo (SE)
- 3D Gradient Field Echo (GE)
- 3D Steady state acquisition with rewinded GE (SARGE); also with RF spoiling, rephasing
- MR Angiography (2D TOF, 3D TOF, half echo, high resolution/high definition, sloped slab profile, magnetization transfer contrast)
- MR Fluoroscopy
- RF Coil Uniformity
- Adaptive Image post-processing
- ACR/NEMA/DICOM 3 compliant
- Aids in the performance of minimally invasive, diagnostic, and therapeutic procedures of the head, body, and extremities which mav be facilitated by real-time MR guidance. Such procedures must be performed with MRI-compatible instrumentation, as selected and evaluated by the clinical user.
#### DEVICE TECHNOLOGICAL CHARACTERISTICS: 5.0
Identical to the Predicate Device.
{4}------------------------------------------------
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
## MAR - 7 2000
James Jochen Rogers Manager, Regulatory Affairs Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park Twinsburg, Ohio 44087
Re:
Rockville MD 20850 K994378 CTL Phased Array Coil for use with AIRIS II Magnetic Resonance Diagnostic Device Dated: December 21, 1999 Received: December 27, 1999 Regulatory class: II 21 CFR 892.1000/Procode: 90 MOS
Dear Mr. Rogers:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requiation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D. Captain, USPHS Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
1994278 510(k) Number (if known): __
### Device Name: Additional RF Coil for AIRIS II (CTL Phased Array)
Indications for Use:
The AIRIS II Magnetic Resonance Diagnostic Device with the CTL Phased Array Coil P/N MR-CTL-51 provides for continuous imaging of the cervical, thoracic, and lumbar spine and adjacent anatomies.
The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information. obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
- Anatomical Region: Head, Body, Spine, Extremities
- Nucleus excited:
- Diagnostic uses:
Proton 2D T1- / T2-weighted imaging T1, T2, proton density measurements MR Angiography image processing
- Imaging capabilities:
- 2D. 3D Spin Echo (SE)
- 2D, 3D Fast Spin Echo (FSE)
- 2D Inversion Recovery (IR)
- 2D, 3D Fast Inversion Recovery (FIR)
- 2D.3D Gradient Echo (GE)
- 2D, 3D Steady state acquisition with rewinded GE (SARGE)
- 2D. 3D RF-spoiled SARGE (RSSG)
- 2D, 3D Time-reversed SARGE (TRSG)
- MR Angiography (half echo, high resolution/high definition, sloped slab profile, magnetization transfer contrast) 2D, 3D TOF
- 2D, 3D TOF RSSG
- ECG, Peripheral Gating
- RF Coil Uniformity
Adaptive Image post-processing
ACR/NEMA/DICOM 3 compliant
Aids in the performance of minimally invasive, diagnostic, and therapeutic procedures of the head, body and extremities which may be facilitated by real-time MR guidance. Such procedures must be performed with MRI-compatible instrumentation, as selected and evaluated by the clinical user.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801-109
Vinid C. Seymour
OR
Division Sion-Off) ctive, Abdominal, ENT. 510(k) Numbe
Over-the-Counter Use _
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.