← Product Code [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS) · K982919

# MEDRAD QUAD T/L COIL (1.5T, 1.0T & 0.5T) MODELS M64TLC, M42TLC, M21TLC (K982919)

_Medrad, Inc. · MOS · Nov 2, 1998 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K982919

## Device Facts

- **Applicant:** Medrad, Inc.
- **Product Code:** [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS.md)
- **Decision Date:** Nov 2, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

The Medrad Quad T/L Spine Coil is a receive only coil intended to be used with the General Electric Superconducting MRI Scanners. This coil is intended to facilitate complete MR imaging of the Quad T/L Spine anatomy. The Medrad Quad T/L Spine Coil is a receive only coil intended to be used in MRI Scanner Systems for imaging of the thoracic and lumbar regions of the spine anatomy. The Medrad Quad T/L Spine Coil is intended for use only under the supervision of a physician who is trained in the field of Diagnostic Magnetic Resonance Imaging.

## Device Story

Medrad Quad T/L Spine Coil is a receive-only surface coil for MRI scanners; designed to enhance imaging of thoracic and lumbar spine regions. Device functions as passive hardware component within MRI system; receives hydrogen nuclei signals during 2D/3D imaging. Used in clinical settings under physician supervision. Output is raw signal data processed by MRI scanner to produce diagnostic images. Device is mechanically and electrically identical to predicate; primary change is manufacturer labeling from GE to Medrad. Benefits include facilitating high-quality spinal imaging for diagnostic purposes.

## Clinical Evidence

Clinical images provided for 1.5T model to demonstrate effectiveness. SNR studies conducted for .5T, 1.0T, and 1.5T models compared to predicate. NEMA standards used for image uniformity evaluation. No significant static magnetic field or RF field distortion observed.

## Technological Characteristics

Receive-only surface coil; hydrogen nuclei excitation. Contains slightly magnetic materials positioned to prevent static field distortion. Electrical design includes blocking network to prevent induced currents. Compatible with GE Superconducting MRI scanners.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- General Electric (GE) Quad T/L Spine Coil ([K923025](/device/K923025.md))

## Reference Devices

- GE Quad T/L Spine Coil ([K902663](/device/K902663.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the contribution of the fi

NOV 2 1998

### 510(k) SUMMARY MEDRAD .5T, 1.0T, 1.5T QUAD T/L SPINE COILS

| OFFICIAL CONTACT:    | Jim Ferguson, Jr.<br>Sr. Regulatory Affairs Associate<br>Medrad, Inc.<br>One Medrad Drive<br>Indianola, PA 15051<br>(412) 767-2400 Ext. 3326 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| CLASSIFICATION NAME: | Magnetic Resonance Diagnostic Accessory<br>[21 CFR 892.1000]                                                                                 |
| COMMON/USUAL NAME:   | MR Imaging Surface Coil                                                                                                                      |
| PROPRIETARY NAME:    | Medrad Quad T/L Spine Coil                                                                                                                   |
| PREDICATE DEVICES:   | General Electric (GE) Quad T/L Spine Coil                                                                                                    |

# DEVICE DESCRIPTION:

The Medrad Quad T/L Spine Coil is a receive only coil designed to enhance the MR Imaging of the thoracic and lumbar regions of the spine anatomy.

#### INTENDED USE:

The Medrad Quad T/L Spine Coil is a receive only coil intended to be used with the General Electric Superconducting MRI Scanners. This coil is intended to facilitate complete MR imaging of the Quad T/L Spine anatomy.

| Anatomical Region: | thoracic and lumbar regions of the spine<br>anatomy. |
|--------------------|------------------------------------------------------|
| Nuclei Excited:    | Hydrogen                                             |
| Diagnostic Uses:   | 2D and 3D Imaging                                    |

{1}------------------------------------------------

510(k) Summary Medrad Quad T/L Spine Coil Page 2 of 3

## PROPOSED MEDRAD QUAD T/L SPINE COJL TECHNICAL COMPARISON TO PREDICATE DEVICES:

The following table compares the predicate GE Quad T/L Spine Coil and the proposed Medrad Quad T/L Spine Coil.

| GE Quad T/L Spine Coil<br>(K923025) | Medrad Quad T/L Spine Coil |
|-------------------------------------|----------------------------|
| GE labeling                         | Medrad labeling            |

Medrad was responsible for the design and manufacturing of the predicate device, and will remain responsible for the design and manufacturing for the proposed device. The device will be labeled and marketed as "Medrad Product". This is the only change to the device.

# PROPOSED MEDRAD QUAD T/L SPINE TECHNICAL COMPARISON TO PREDICATE DEVICES(Cont.):

| Patient contacting materials comparison information                                                                                                                                      |                                                                                   |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| GE Quad T/L Spine Coil                                                                                                                                                                   | Medrad Quad T/L Spine Coil                                                        |
| Reference K 902663 for material information                                                                                                                                              | All materials used are the same as the GE Quad T/L Spine coil. (Predicate device) |
| Certification: Medrad, Inc. certifies that the patient contacting materials and formulations for the proposed Medrad Quad T/L Spine Coils are unchanged from currently marketed devices. |                                                                                   |

{2}------------------------------------------------

## PERFORMANCE TEST DATA:

## SIGNAL TO NOISE RATIO (SNR)

A Signal to Noise Ratio(SNR) study was conducted to generate a Signal -To-Noise ratio comparison between the proposed Medrad Quad T/L Spine coil and the predicate GE Quad T/L Spine Coil.

IMAGE UNIFORMITY - The Medrad Quad T/L Spine Coil was evaluated using NEMA Standards to characterize the non-uniformity of the proposed coil. Contours of the images obtained with the coil were constructed for the axial image, sagittal image.

GEOMETRIC DISTORTION: None. Static magnetic field distortion results from the use of magnetic materials in the construction of a surface coil. This device contains slightly magnetic materials or components. However, such components have been positioned within the surface coil so that no observable distortion of the static magnetic field is present.

TRANSMIT RF FIELD [B1] DISTORTION - Analysis of the electrical design of the coil and its blocking network demonstrates that no significant currents are induced. No artifacts of any type were observed during imaging.

RESOLUTION, SLICE THICKNESS, AND CONTRAST - These performance parameters are not affected by the use of a surface coil and were not separately tested in the performance evaluation of the proposed Medrad Quad T/L Spine Coil.

CLINICAL EVALUATION - Clinical images for the proposed 1.5T Quad T/L Spine Coil have been provided with this submission to demonstrate the clinical effectiveness of the Quad T/L Spine coils. Based on these results, and the SNR results for the proposed .5T and 1.0T Quad T/L Spine Coils, Medrad concludes that the proposed .5T and 1.0T Quad T/L Spine Coils will produce similar clinical image results.

CONCLUSION - Extensive safety, verification, durability and clinical testing was conducted on the predicate GE device. (K 902663) Mechanically and electrically the proposed device and the predicate device are identical. This labeling change is the only change to the device. The proposed device is deemed by Medrad to be substantially equivalent.

{3}------------------------------------------------

Re:

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing to the right, stacked on top of each other. The profiles are connected by a flowing line that forms the shape of a ribbon. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the design.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2 1998

Jim Ferguson, Jr. Sr. Regulatory Affairs Associate Medrad, Inc. One Medrad Drive Indianola, PA 15051

K982919 Medrad Quad T/L Spine Coil Dated: August 18, 1998 Received: August 19, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 MOS

Dear Mr. Ferguson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitne diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdr/dsmaldsmamain.html".

Sincerely yours,

Lillian Yim, Ph.D.

Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) NUMBER (IF KNOWN) : K 98 19 19

DEVICE NAME: Medrad Quad T/L Spine Coil

INDICATIONS FOR USE:

The Medrad Quad T/L Spine Coil is a receive only coil intended to be used in MRI Scanner Systems for imaging of the thoracic and lumbar regions of the spine anatomy.

The Medrad Quad T/L Spine Coil is intended for use only under the supervision of a physician who is trained in the field of Diagnostic Magnetic Resonance Imaging.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | V |
|----------------------|---|
| (Per 21 CFR 801.109) |   |

OR

Over-The-Counter-Use (Optional Format 1-2-96)

Vorind h. Algerson

(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological De 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K982919](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K982919)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
