iLoop Interventional Coil 0.55T

K241069 · Noras Mri Products GmbH · MOS · Jun 7, 2024 · Radiology

Device Facts

Record IDK241069
Device NameiLoop Interventional Coil 0.55T
ApplicantNoras Mri Products GmbH
Product CodeMOS · Radiology
Decision DateJun 7, 2024
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The intended use of the iLoop Interventional Coil 0.55T is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the human body. Used in a magnetic resonance tomograph, the iLoop Interventional Coil 0.55T is intended to produce transversal, sagittal, coronal and oblique images of the internal structures of the body. This device may be used for interventional procedures.

Device Story

iLoop Interventional Coil 0.55T is a 1-channel RF receive-only surface loop coil; used with Siemens 0.55T MRI systems for diagnostic imaging and MR-guided interventions. Device captures high-frequency signals from hydrogen nuclei induced by the MRI scanner's transmitter; produces transversal, sagittal, coronal, and oblique images of internal body structures. Used in clinical settings by physicians; coil placed on patient body regions; often used with sterile, self-adhesive OP-drape to facilitate interventional workflows. Output is raw MR signal data processed by the MRI system into diagnostic images; assists physicians in visualizing internal anatomy for diagnosis and procedural guidance. Benefits include localized imaging capability for interventional procedures.

Clinical Evidence

No clinical data. Bench testing only. Performance verified per FDA guidance 'Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance Based Pathway'. Tests included Image SNR, Image Uniformity, Surface heating, Acquired Image Quality, Decoupling circuit inspection, EMC immunity, and electrical/mechanical safety (IEC 60601-1, IEC 60601-2-33). Biocompatibility testing (ISO 10993) for cytotoxicity, irritation, and sensitization performed.

Technological Characteristics

1-channel RF receive-only surface loop coil; 19 cm inner conductor diameter; 1000 mm cable attached to plug. Compatible with 0.55T MRI systems. Materials tested for biocompatibility per ISO 10993. Safety standards: IEC 60601-1 (general electrical/mechanical), IEC 60601-2-33 (MR safety), IEC 60601-1-2 (EMC), and IEC 60601-1-6 (usability).

Indications for Use

Indicated for MR examination of the human body, including interventional procedures, to produce transversal, sagittal, coronal, and oblique images of internal structures when used with a 0.55T Magnetic Resonance Scanner.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 7, 2024 NORAS MRI products GmbH % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114 Re: K241069 Trade/Device Name: iLoop Interventional Coil 0.55T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: April 11, 2024 Received: April 22, 2024 Dear Daniel Kamm: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, FDA Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K241069 Device Name iLoop Interventional Coil 0.55T #### Indications for Use (Describe) The intended use of the iLoop Interventional Coil 0.55T is, in conjunction with a Magnetic Resonance Scamer, the MR examination of the human body. Used in a magnetic resonance tomograph, the iLoop Interventional Coil 0.55T is intended to produce transversal, sagittal, coronal and oblique images of the internal structures of the body. This device may be used for interventional procedures. Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size:10pt"> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" width="12" x="0" y="0"></rect> <path d="M2 2 L10 10 M10 2 L2 10" stroke="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:10pt"> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" width="12" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### 510(k) Summary K241069 Image /page/3/Picture/1 description: The image is a logo for NORAS MRI products. The logo is in blue and white, with the word "NORAS" in large, bold letters. To the right of "NORAS" is a line that resembles a heartbeat, followed by the words "MRI products" in smaller letters. The logo is set against a white background with a blue border. NORAS MRI Products GmbH Leibnizstrasse 4 Hoechberg Germany 97204 Telephone: 49-931-299270 E-Mail: mri@noras.de www.noras.de Date Prepared: May 30, 2024 - NORAS MRI Products GmbH, Inc A) SUBMITTED BY: - B) CONTACT: Manuel Noras, Chief Executive Officer - C) Device Identification Trade/Device Names: iLoop Interventional Coil 0.55T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: ll Product Code: MOS - D) Predicate Device: K203443 Manufacturer: Siemens Medical Solutions USA, Inc. Trade/Device Name: MAGNETOM (Including Flex Loop Large local coil) Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: ll Product Code: MOS - E) DEVICE DESCRIPTION: The iLoop Interventional Coil 0.55T is a 1-channel receiver coil that can be used for MR imaging before, during and after MR-guided interventions with a Siemens MRI System with field strength 0.55T. The iLoop Interventional Coil 0.55T can also be used as a standard diagnostic coil for diagnostic examination of the human body. The iLoop Interventional Coil 0.55T is a coil that can be used on many different regions of the body. Used in conjunction with a sterile, self-adhesive OP-drape, the coil enables the workflow for MR-guided interventions - F) INDICATIONS FOR USE The intended use of the iLoop Interventional Coil 0.55T is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the human body. Used in a magnetic resonance tomograph, the iLoop Interventional Coil 0.55T is intended to produce transversal, sagittal, coronal and oblique images of the internal structures of the body. This device may be used for interventional procedures. {5}------------------------------------------------ - G) SUBSTANTIAL EQUIVALENCE: The predicate Flex Loop Large Coil from Siemens has a very similar geometry to the device under evaluation. It is also a one-channel receiving coil and has the same diameter of the inner conductor. The absence of directly attached cable to the plug highlights the difference in design. Both devices are compatible with different MRI scanners and used on different field strengths but perform identical functions using similar designs and materials. | | H) SUBSTANTIAL EQUIVALENCE COMPARISON AND DISCUSSION<br>K203443 MAGNETOM Vida, MAGNETOM<br>Sola, MAGNETOM Lumina, MAGNETOM<br>Altea with syngo MR XA31A (Includes Flex<br>Loop Large local coil) | iLoop Interventional Coil 0.55T | Comparison | |-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------| | Indication for<br>Use | Your MAGNETOM system is indicated for<br>use as a magnetic resonance diagnostic<br>device (MRDD) that produces transverse,<br>sagittal, coronal and oblique cross sectional<br>images, spectroscopic images and/or<br>spectra, and that displays the internal<br>structure and/or function of the head,<br>body, or extremities. Other physical<br>parameters derived from the images<br>and/or spectra may also be produced.<br>Depending on the region of interest,<br>contrast agents may be used. These images<br>and/or spectra and the physical parameters<br>derived from the images and/or spectra<br>when interpreted by a trained physician<br>yield information that may assist in<br>diagnosis. | The intended use of the iLoop<br>Interventional Coil 0.55T is, in<br>conjunction with a Magnetic<br>Resonance Scanner, the MR<br>examination of the human body.<br>Used in a magnetic resonance<br>tomograph, the iLoop<br>Interventional Coil 0.55T is<br>intended to produce transversal,<br>sagittal, coronal and oblique images<br>of the internal structures of the<br>body. This device may be used for<br>interventional procedures. | Functionally<br>the same.<br>The new<br>iLoop<br>represents<br>a subset of<br>the<br>predicate<br>indications. | | Appearance | Image: Flex Loop Large Coil | Image: iLoop Coil | SIMILAR | | Construction/Use | • one-channel RF receive-only surface loop coil<br>• the diameter of the inner conductor is 19 cm<br>• it does not have a 1000 mm cable attached to<br>the plug<br>• used with a field strength of 1.5T and 3T<br>• used on the Magnetom Altea, Magnetom<br>Avanto and Magnetom Sola systems | • one-channel RF receive-only surface<br>loop coil<br>• the diameter of the inner conductor<br>is 19 cm<br>• a 1000 mm cable is directly attached<br>to the plug<br>• used with a field strength of 0.55T<br>• used on the MAGNETOM Free.Max<br>system | Similar<br>function<br>but for use<br>on different<br>systems. | | Dimensions | 19 cm opening | 19 cm opening | SAME | | Function | Employs a coil system which serves solely<br>as a receiving coil for the reception of high<br>frequency signals from the hydrogen -(1-H)<br>nuclei. The hydrogen nuclei are induced<br>into precession by the transmitting coil of<br>the MRI device. | Identical mode of operation | SAME | | | K203443 MAGNETOM Vida, MAGNETOM<br>Sola, MAGNETOM Lumina, MAGNETOM<br>Altea with syngo MR XA31A (Includes Flex<br>Loop Large local coil) | iLoop Interventional Coil 0.55T | Comparison | | Biocompatibility | Patient contact parts tested to<br>ISO 10993 | SAME | SAME | ## H) SUBSTANTIAL EQUIVALENCE COMPARISON AND DISCUSSION {6}------------------------------------------------ #### SUMMARY OF PERFORMANCE TESTING. = This device was tested in accordance with the FDA special controls guidance: Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance Based Pathway, Guidance for Industry and Food and Drug Administration Staff Document issued on December 11, 2020. We conducted these tests successfully as called for in the guidance: - Image Signal to Noise (SNR) ● - Image Uniformity - Surface heating - Acquired Image Quality ● - Decoupling circuit inspection - EMC Immunity, electrostatic discharge, Methodology: FDA currently recognized version of IEC ● 60601-1-2, FDA Recognition (Performed within the test report for IEC 60601-2-33:2022) - . General electrical/mechanical safety: IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 (FDA Recognition 19-49 Patient contact materials were subjected to ISO 10993 tests for: Cytoxicity, Irritation, and sensitization. Testing was done according to IEC60601-2-33 Edition 4.0 2022-08 Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis, FDA Recognition 12-347 Usability testing was performed in accordance with: IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability, FDA Recognition 5-132. In addition, testing was performed in accordance with IEC standard: IEC60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability, FDA Recognition 5-132. A detailed risk analysis was also performed. - J) CLINICAL TESTING: Not required. - K) CONCLUSION: In all material aspects, the iLoop interventional coil 0.55T model of MR coil performed equivalently to the predicate device, had similar construction methods, and has not raised any new issues of safety or effectiveness. The predicate MAGNETOM (Including Flex Loop Large local coil) has a similar intended use as the predicate, thus rendering the new model of MRI coil substantially equivalent to the predicate.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...