K233444 · Inkspace Imaging, Inc. · MOS · Jan 8, 2024 · Radiology
Device Facts
Record ID
K233444
Device Name
InkSpace Imaging Small Body Array
Applicant
Inkspace Imaging, Inc.
Product Code
MOS · Radiology
Decision Date
Jan 8, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The InkSpace Imaging Small Body Array is a receive-only coil, used for obtaining diagnostic images of general human anatomy in adult and pediatric patients such as cardiac, spine, shoulder, elbow, knee, foot, and prostate in Siemens 3.0T magnetic resonance imaging systems. These images, when interpreted by a trained physician, yield information that may assist in diagnosis.
Device Story
The InkSpace Imaging Small Body Array is a 24-channel, receive-only phased array RF coil designed for use with Siemens 3.0T MRI scanners. It consists of two flexible 12-channel pads that conform to patient anatomy to improve signal-to-noise ratio and image uniformity. The device is used in clinical settings by MRI technicians or radiologists to acquire hydrogen-based MR images. The output consists of raw MR signal data processed by the host MRI system into diagnostic images. These images are interpreted by a trained physician to assist in clinical diagnosis. The device is reusable, non-sterile, and prescription-only.
Clinical Evidence
No clinical trials were conducted. Substantial equivalence was demonstrated via nonclinical bench testing, including SNR, image uniformity, surface heating, decoupling, EMC, and electrical/mechanical safety. Biocompatibility (cytotoxicity, sensitization, irritation), usability, and transit testing were leveraged from the predicate device. Clinical image quality was assessed by a board-certified radiologist using sample images.
Technological Characteristics
24-channel receive-only phased array coil; two flexible 12-channel pads; 123.2 MHz operating frequency (hydrogen); reusable; non-sterile; biocompatible materials per ISO 10993; designed for Siemens 3.0T MRI systems.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic MR imaging of general anatomy, including cardiac, spine, shoulder, elbow, knee, foot, and prostate regions, using Siemens 3.0T MRI systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
January 8, 2024
InkSpace Imaging, Inc. % Taras Bouzakine Director, Regulatory Affairs Veranex. Inc. 224 Airport Parkway, Suite 250 San Jose. California 95110
Re: K233444
Trade/Device Name: InkSpace Imaging Small Body Array Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: December 14, 2023 Received: December 15, 2023
Dear Taras Bouzakine:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.g.K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233444
Device Name
InkSpace Imaging Small Body Array
Indications for Use (Describe)
The InkSpace Imaging Small Body Array is a receive-only coil, used for obtaining diagnostic images of general human anatomy in adult and pediatric patients such as cardiac, spine, shoulder, elbow, knee, foot, and prostate in Siemens 3.0T magnetic resonance imaging systems. These images, when interpreted by a trained physician, yield information that may assist in diagnosis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) Notification K233444
### GENERAL INFORMATION [807.92(A)(1)]
## Applicant:
InkSpace Imaging, Inc. 5635 West Las Positas Blvd, Suite 403/404 Pleasanton, CA 94588 USA Phone: 925-425-7410
### Contact Person:
Taras Bouzakine Director, Regulatory Affairs, Veranex, Inc. 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA Phone: 603-205-4874 Email: taras.bouzakine@veranex.com
Date Prepared: November 3, 2023
## DEVICE INFORMATION [807.92(A)(2)]
### Classification:
Class II per 21 CFR§892.1000
### Product Code:
MOS
# Trade Name:
InkSpace Imaging Small Body Array
### Generic/Common Name:
Coil, Magnetic Resonance, Specialty
# PREDICATE DEVICE(S) [807.92(A)(3)]
InkSpace Imaging Body Array (K223487)
# DEVICE DESCRIPTION [807.92(A)(4)]
The InkSpace Imaging Small Body Array is a phased array, receive-only coil intended to work with Siemens 3.0T MRI scanners to provide optimal signal-to-noise ratio and uniformity for high resolution body MR imaging examinations. The coil is comprised of 24 total channels, with two individual flexible pads, each containing 12 individual elements. The elements are optimized to be flexible to conform to patients' anatomies.
{6}------------------------------------------------
# INDICATIONS FOR USE [807.92(A)(5)]
The InkSpace Imaging Small Body Array is a receive-only coil, used for obtaining diagnostic images of general human anatomy in adult and pediatric patients such as cardiac, spine, shoulder, elbow, knee, foot, and prostate in Siemens 3.0T magnetic resonance imaging systems. These images, when interpreted by a trained physician, yield information that may assist in diagnosis.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(A)(6)]
The technological characteristics of the modified InkSpace Imaging Small Body Array are substantially equivalent to the predicate device, InkSpace Imaging Body Array (K223487). Table 1 lists the technological characteristics of the predicate and modified devices and provides rationale to support a determination of substantial equivalence. Any differences between the devices do not raise any new or different questions of safety or effectiveness.
{7}------------------------------------------------
# 510(k) SUMMARY
| Characteristics | Predicate Device:<br>InkSpace Imaging Body Array (K223487) | Subject Device:<br>InkSpace Imaging Small Body Array | Substantial Equivalence<br>Rationale |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Magnetic Resonance Diagnostic Device per 21<br>CFR§892.1000 | Magnetic Resonance Diagnostic Device per 21<br>CFR§892.1000 | N/A (Same). |
| Product Code | MOS (Coil, Magnetic Resonance, Specialty) | MOS (Coil, Magnetic Resonance, Specialty) | N/A (Same). |
| Intended Use | Pediatric Population:<br>The InkSpace Imaging Body Array is a receive-<br>only RF coil designed for use with GE 3.0T<br>MRI systems. It is used for obtaining diagnostic<br>images of the body and cardiac regions of<br>pediatric patients in magnetic resonance imaging<br>systems. The nucleus detected is hydrogen.<br>Anatomic Regions: Cardiac/Body for pediatric<br>patients.<br>Adult Population:<br>The InkSpace Imaging Body Array is a receive-<br>only RF coil designed for use with GE 3.0T<br>MRI systems. It is used for obtaining diagnostic<br>images of general human anatomy such as spine,<br>shoulder, elbow, knee, foot, and prostate for<br>adult patients in magnetic resonance imaging<br>systems. The nucleus detected is hydrogen. | The InkSpace Imaging Small Body Array is a<br>receive-only RF coil designed for use with<br>Siemens 3.0T MRI systems. It is used for<br>obtaining diagnostic images of general human<br>anatomy in adult and pediatric patients such as<br>cardiac, spine, shoulder, elbow, knee, foot, and<br>prostate in magnetic resonance imaging systems.<br>The nucleus detected is hydrogen. | Intended use of the modified<br>device maintains the same general<br>purpose and function as the<br>predicate device, differing only<br>by target MRI system. |
| Indications for<br>Use | Pediatric Population:<br>The InkSpace Imaging Body Array is a receive-<br>only coil, used for obtaining diagnostic images<br>of pediatric cardiac and body in GE 3.0T<br>magnetic resonance imaging systems. These<br>images, when interpreted by a trained physician,<br>yield information that may assist in diagnosis.<br>Adult Population:<br>The InkSpace Imaging Body Array is a receive-<br>only coil, used for obtaining diagnostic images<br>of general human anatomy in adults, such as | The InkSpace Imaging Small Body Array is a<br>receive-only coil, used for obtaining diagnostic<br>images of general human anatomy in adult and<br>pediatric patients, such as cardiac, spine, shoulder,<br>elbow, knee, foot, and prostate in Siemens 3.0T<br>magnetic resonance imaging systems. These<br>images, when interpreted by a trained physician,<br>yield information that may assist in diagnosis. | Indications for Use of the<br>modified device maintains the<br>same general purpose and<br>function as the predicate device,<br>differing only by target MRI<br>system. |
| Characteristics | Predicate Device:<br>InkSpace Imaging Body Array (K223487) | Subject Device:<br>InkSpace Imaging Small Body Array | Substantial Equivalence<br>Rationale |
| | spine, shoulder, elbow, knee, foot, and prostate<br>in GE 3.0T magnetic resonance imaging<br>systems. These images, when interpreted by a<br>trained physician, yield information that may<br>assist in diagnosis. | | |
| Patient Anatomy<br>and Population | Imaging of general body part regions including<br>cardiac regions; intended for pediatric patients.<br>Imaging of general human anatomy, such as<br>spine, shoulder, elbow, knee, foot, and prostate;<br>intended for adult patients. | Imaging of general human anatomy in adult and<br>pediatric patients, such as cardiac, spine, shoulder,<br>elbow, knee, foot, and prostate. | N/A (Same). |
| Comparison of<br>Technological<br>Characteristics | 24 channel, receive-only phased array, designed<br>for 3.0T MR systems (127.73 MHz, hydrogen).<br>Device is reusable, non-sterile, and by<br>prescription only.<br>Device is made of soft, pliable materials, and is<br>composed of an anterior and a posterior pad. | 24 channel, receive-only phased array, designed<br>for 3.0T MR systems (123.2 MHz, hydrogen).<br>Device is reusable, non-sterile, and by prescription<br>only.<br>Device is made of soft, pliable materials, and is<br>composed of two identical 12ch pads. | N/A (Same). |
| Compatible MRI<br>Systems | GE 3.0T MRI Systems | Siemens 3.0T MRI Systems | Differed target MRI system<br>vendor. Same magnetic field<br>strength. |
| Biocompatibility | ISO 10993 testing | ISO 10993 testing | N/A (Same). |
# Table 1: Summary of Technological Characteristics
{8}------------------------------------------------
# 510(k) SUMMARY
{9}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE
The InkSpace Imaging Small Body Array is substantially equivalent to the predicate device, InkSpace Imaging Body Array with regard to function, intended use, and physical characteristics. There are no differences in the technological characteristics between the devices which would raise any different questions of safety or efficacy. Thus, the InkSpace Imaging Small Body Array is substantially equivalent to the predicate device.
## PERFORMANCE DATA [807.92(B)]
Performance testing was performed in accordance with the FDA Guidance Document entitled "Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance Based Pathway," issued December 11, 2020.
Some of the nonclinical testing conducted on the InkSpace Imaging Body Array (predicate device) is still applicable to the InkSpace Imaging Small Body Array.
Additional analysis was conducted based on variations between the devices and differences between MR system vendors. These results and images listed below have been included as part of this submission.
## [807.92(b)(1)] Nonclinical Testing Summary:
The nonclinical, bench tests, per the FDA Guidance Document entitled "Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance Based Pathway," were performed on the InkSpace Imaging Small Body Array:
- . Image Signal to Noise (SNR)
- Image Uniformity ●
- Surface Heating ●
- Decoupling Circuit ●
- EMC - Immunity, Electrostatic Discharge
- General Electrical/Mechanical Safety
- Acquired Image Quality ●
- Biocompatibility Testing (Cytotoxicity, Sensitization, Irritation)
In addition to the above nonclinical testing, a risk analysis was performed to validate that the following testing performed on the InkSpace Imaging Body Array (predicate device) could be applied to the InkSpace Imaging Small Body Array without retesting:
- . Biocompatibility Testing (Cytotoxicity, Sensitization, Irritation)
- Usability Testing .
- Transit Testing ●
{10}------------------------------------------------
### 510(K) SUMMARY
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the InkSpace Imaging Small Body Array meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the InkSpace Imaging Small Body Array does not raise different questions of safety or effectiveness for MR imaging examinations with Siemens 3.0T MRI Systems when compared to the predicate device.
### [807.92(b)(2)] Clinical Testing Summary:
Not applicable. Clinical testing was not required to demonstrate substantial equivalence of InkSpace Imaging Small Body Array to the predicate device. Acquired image quality was analyzed based on sample clinical images. These images, as well as a review of their clinical quality by a board-certified radiologist, have been included as part of this submission.
### CONCLUSIONS [807.92(B)(3)]
Based on the results from the nonclinical tests performed in support of the InkSpace Imaging Small Body Array, it is concluded that the modified device is safe, effective, and performs at least as safely and effectively as the legally marketed predicate device.
#### SUMMARY
The InkSpace Imaging Small Body Array is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.