Smart Fit TorsoCardiac 1.5T and Smart Fit Shoulder 1.5T
Applicant
Philips Healthcare (Suzhou) Co., Ltd.
Product Code
MOS · Radiology
Decision Date
Sep 1, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Intended Use
The Smart Fit TorsoCardiac 1.5T coil is designed to be used in conjunction with a Philips 1.5T MR system to produce diagnostic images of the torso (including chest, abdomen, pelvis), head and neck and heart that can be interpreted by a trained physician. The Smart Fit Shoulder 1.5T coil is designed to be used in conjunction with a Philips 1.5T MR system to produce diagnostic images of the Shoulder that can be interpreted by a trained physician.
Device Story
Phased-array receive-only RF coils for 1.5T MRI systems. Inputs: magnetic resonance signals from hydrogen nuclei in patient anatomy. Operation: coils receive RF signals, which are amplified and transmitted to the MRI system for image reconstruction. Used in clinical settings by trained physicians/technicians. Output: diagnostic cross-sectional images of torso, head, neck, heart, or shoulder. Assists physicians in diagnosis and therapy planning. Benefits: high-resolution imaging of specific anatomical regions.
Clinical Evidence
No clinical study was required. Substantial equivalence was established through non-clinical bench testing and comparison to predicate devices. Clinical images obtained using the coils were evaluated by qualified radiologists, confirming sufficient image quality for diagnostic use.
Technological Characteristics
Phased-array receive-only RF coils. Materials: PU foil, flexible PCB, EVA foam. Sensing: RF signal reception via phased-array elements. Energy: derived from MR scanner (no internal source). Connectivity: analog connector to dS Interface boxes. Standards: IEC 60601-1, IEC 60601-2-33, ISO 10993-1, NEMA MS series. Biocompatibility: ISO 10993-1 compliant for limited contact (<24h) with intact skin.
Indications for Use
Indicated for patients requiring MR examination of the torso (chest, abdomen, pelvis), head, neck, heart, or shoulder, where the coil provides diagnostic benefit in terms of Field of View or Signal-to-Noise ratio. Contraindications include physical characteristics (circumference, mass) exceeding system/coil capabilities and standard MR safety contraindications.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K102889 — HITACHI ECHELON NV ATTACHMENT AND SMALL EXTREMITY COILS MODEL ECHELON · Hitachi Medical Systems America, Inc. · Dec 20, 2010
K244054 — 16ch Flex SPEEDER Medium 1.5T (Q7000246); 16ch Flex SPEEDER Large 1.5T (Q7000247) · Quality Electrodynamics · Feb 28, 2025
K031366 — PREMIER 9000 8 CHANNEL PHASED ARRAY CTL SPINE COIL · Usa Instruments, Inc. · Jul 11, 2003
K093842 — 1.5T AND 3.0T GENERAL PURPOSE FLEX COILS · Invivo Corporation · Mar 5, 2010
K984588 — PHASED ARRAY FLEXIBLE CARDIAC COIL · Philips Medical Systems (Cleveland), Inc. · Feb 8, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
September 1, 2023
Philips Healthcare (Suzhou) Co., Ltd. % Li Sherry Regulatory Affairs Specialist No. 258, Zhongyuan Road Suzhou Industrial Park Suzhou, Jiangsu 215024 CHINA
Re: K232021
Trade/Device Name: Smart Fit TorsoCardiac 1.5T and Smart Fit Shoulder 1.5T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: June 12, 2023 Received: July 7, 2023
Dear Li Sherry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K232021
Device Name
Smart Fit TorsoCardiac 1.5T and Smart Fit Shoulder 1.5T
#### Indications for Use (Describe)
The Smart Fit TorsoCardiac 1.5T coil is designed to be used in conjunction with a Philips 1.5T MR system to produce diagnostic images of the torso (including chest, abdomen, pelvis), head and neck and heart that can be interpreted by a trained physician.
The Smart Fit Shoulder 1.5T coil is designed to be used in conjunction with a Philips 1.5T MR system to produce diagnostic images of the Shoulder that can be interpreted by a trained physician.
Type of Use (Select one or both, as applicable)
< Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
K232021
510(k) Summary of Safety and Effectiveness
[As required by 21 CFR 807.92(c)]
| [As required by 21 CFR 807.92(c)] | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Date<br>Prepared: | July 04, 2023 | | | |
| Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, Zhongyuan Road, Suzhou Industrial Park, Suzhou Jiangsu, CHINA, 215024<br>Establishment Registration Number: 3009529630 | | | |
| Primary<br>Contact<br>Person: | Sherry Li<br>Regulatory Affairs Specialist<br>Phone: +86-17849535031<br>E-mail: sherry.li@philips.com | | | |
| Secondary<br>Contact<br>Person | Erhong Wang<br>Senior Manager Regulatory Affairs<br>Phone: +86-512-67336804<br>E-mail: erhong.wang@philips.com | | | |
| Device Name: | Smart Fit TorsoCardiac 1.5T and Smart Fit Shoulder 1.5T | | | |
| Classification: | Classification name: | Magnetic Resonance Diagnostic Device | | |
| | Classification Regulation: | 21CFR 892.1000 | | |
| | Classification Panel: | Magnetic Resonance Diagnostic Device | | |
| | Device Class: | Class II | | |
| | Primary Product Code: | MOS | | |
| Predicate<br>Device for<br>Smart Fit<br>TorsoCardiac<br>1.5T | Trade name: | dS TorsoCardiac 1.5T | | |
| | Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd. | | |
| | 510(k) Clearance: | K212864 | | |
| | Classification Regulation: | 21CFR 892.1000 | | |
| | Classification name: | Magnetic Resonance Diagnostic Device | | |
| | Classification Panel: | Radiology | | |
| | Device class<br>Product Code: | Class II<br>MOS | | |
| Predicate<br>Device for<br>Smart Fit<br>Shoulder 1.5T | Trade name: | 1.5T 16 CH GE Shoulder Coils | | |
| | Manufacturer: | InVivo Corporation | | |
| | 510(k) Clearance: | K162001 | | |
| | Classification Regulation: | 21CFR 892.1000 | | |
| | Classification name: | Magnetic Resonance Diagnostic Device | | |
| | Classification Panel: | Radiology | | |
| | Device class<br>Product Code: | Class II<br>MOS |…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.