← Product Code [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS) · K223239

# 8ch Wrist Coil (K223239)

_Shenzhen RF Tech Co., Ltd. · MOS · Dec 8, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K223239

## Device Facts

- **Applicant:** Shenzhen RF Tech Co., Ltd.
- **Product Code:** [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS.md)
- **Decision Date:** Dec 8, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

The 8ch Wrist Coil by Shenzhen RF Tech Co., Ltd. is receive-only coil and is designed for use as general purpose coil. The 8ch Wrist Coil is designed to be use with GE 1.5T MRI systems to produce diagnostic images of the wrists and fingers that can be interpreted by a trained physician.

## Device Story

8ch Wrist Coil is a receive-only phased array MRI coil; designed for use with GE 1.5T MRI systems. Device captures RF signals corresponding to proton precession (63.86MHz) in 1.5T magnetic field; utilizes integrated preamplifiers to process signals from 8 channels. Used in clinical imaging environments; operated by trained medical professionals (radiologists/technicians). Output consists of diagnostic images of wrists and fingers; interpreted by physicians to aid in clinical diagnosis. Benefits include high-quality diagnostic imaging of extremities. Device connects via P-Port to compatible GE MRI scanners.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included SNR (NEMA MS-1-2008), image uniformity (NEMA MS 3-2008), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10).

## Technological Characteristics

Receive-only phased array coil; 8 channels; 1.5T field strength; 63.86MHz proton frequency. P-Port connector. Integrated preamplifiers (0.5dB noise). Active and passive decoupling. Biocompatible materials per ISO 10993-5/10. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA MS-1, NEMA MS-3.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- HRW-63-8 WRIST ARRAY COIL ([K050622](/device/K050622.md))
- HRW-127-8 WRIST ARRAY COIL ([K050622](/device/K050622.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 8, 2022

Shenzhen RF Tech Co., Ltd. % Gary Wang O&R Consultant Bonnier Quality Supervision Consulting (JM) Center Hailunxinyuan No.3203, Jianghai District Jiangmen City, Guangdong 529000 CHINA

Re: K223239

Trade/Device Name: 8ch Wrist Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: October 19, 2022 Received: October 20, 2022

Dear Gary Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K223239

Device Name 8ch Wrist Coil

Indications for Use (Describe)

The 8ch Wrist Coil by Shenzhen RF Tech Co.,Ltd is receive-only coil and is designed for use as general purpose coil. The 8ch Wrist Coil is designed to be use with GE 1.5T MRI systems to produce diagnostic images of the wrists and fingers that can be interpreted by a trained physician.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

--- Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the letters "RFT" in blue, with a red diamond in place of the dot on the "i". The letters are inside of a blue oval. The letters are large and bold, and the diamond is small and centered above the "T". The oval is thin and surrounds the letters closely.

# K223239 510(k) Summary

Date:2022-12-08(YY-MM-DD)

In accordance with 21 CFR 807.92 the following summary of information is provided:

| I. Applicant/Manufacturer:        | Shenzhen RF Tech Co., Ltd.<br>2-F,BLD4 Juhui Industrial Park,<br>Tianliao,Guangming,Shenzhen,<br>P.R.China 518132<br>Phone: (+86) 755-2664 1989-113<br>Fax: (+86)755-2664 2989 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/Correspondent:          | Bonnier Quality Supervision Consulting (JM) Center<br>Hailunxinyuan No.3203, Jianghai District,<br>Jiangmen City, China 529000<br>Phone: (+86) 13600366215                     |
| Contact Person:                   | Mr. Gary Wang<br>Q&R Consultant<br>Email: gary.wang@bonnier.net.cn                                                                                                             |
| II. Device Regulation Information |                                                                                                                                                                                |
| Common Name of Device:            | 8ch Wrist Coil                                                                                                                                                                 |
| Classification panel:             | Radiology                                                                                                                                                                      |
| Classification Names:             | Coil, Magnetic Resonance, Specialty                                                                                                                                            |
| Regulation Number:                | 21 CFR 892.1000                                                                                                                                                                |
| Regulation Class:                 | II                                                                                                                                                                             |
| Product Code:                     | MOS                                                                                                                                                                            |
| Type of 510(k) submission:        | Traditional 510(k)                                                                                                                                                             |
| III. Device Information           |                                                                                                                                                                                |
| Product Number:                   | 10-F32119                                                                                                                                                                      |
| Device Trade Name:                | 8ch Wrist Coil                                                                                                                                                                 |
| IV. Predicate Device Information  |                                                                                                                                                                                |
| Sponsor:                          | MRI DEVICES CORPORATION                                                                                                                                                        |
| Device:                           | HRW-63-8 WRIST ARRAY COIL;<br>HRW-127-8 WRIST ARRAY COIL                                                                                                                       |
| 510(K) Number:                    | K050622                                                                                                                                                                        |
| V. Device Description             |                                                                                                                                                                                |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a blue oval with the letters "RFT" in blue inside. The letter "F" has a red diamond in the middle. The letters are in a bold, sans-serif font. The oval appears to be a logo or emblem.

The 8ch Wrist Coil is receive only phased array coil to produce diagnostic images of the wrists and fingers in adult population.

The 8ch Wrist Coil is tuned to receive RF frequency corresponding to the proton precession in a 1.5 Tesla magnetic field, which is governed by the Larmor equation.

The 8ch Wrist Coil is based on phased array technique to produce the images from 8 different channels with intergraded preamplifiers. The 8ch Wrist Coil is tuned to the proton frequency of 63.86MHz.

### VI. Indications for Use

The 8ch Wrist Coil by Shenzhen RF Tech Co., Ltd. is receive-only coil and is designed for use as general purpose coil. The 8ch Wrist Coil is designed to be use with GE 1.5T MRI systems to produce diagnostic images of the wrists and fingers that can be interpreted by a trained physician.

### VII. Compatibility

The connector of 8ch Wrist Coil is P-Port type. The 8ch Wrist Coil is compatible with GE1.5T MRI systems where coil ID allows.

| Specification         | Predicate Device                                                                                                                                                              | Subject                                                                                                                                                                                                                                                                                                                                        | Result     |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| K number              | K050622                                                                                                                                                                       | K223239                                                                                                                                                                                                                                                                                                                                        |            |
| Manufacturer          | MRI DEVICES<br>CORPORATION                                                                                                                                                    | Shenzhen RF Tech Co.,Ltd                                                                                                                                                                                                                                                                                                                       |            |
| Indications for Use   | To be used in conjunction with<br>a magnetic resonance<br>scanner to produce diagnostic<br>images of the wrist and hand<br>that can be interpreted by a<br>trained physician. | The 8ch Wrist Coil<br>manufactured by Shenzhen<br>RF Tech Co.,Ltd is<br>receive-only coil and is<br>designed for use as general<br>purpose coil. The 8ch Wrist<br>Coil is designed to be use<br>with GE 1.5T MRI systems<br>to produce diagnostic<br>images of the Wrists and<br>fingers that can be<br>interpreted by a trained<br>physician. | Equivalent |
| Anatomical site       | Wrist/Hand                                                                                                                                                                    | Wrist/Fingers                                                                                                                                                                                                                                                                                                                                  | Same       |
| Transmit/Receive      | RF coil (receive only)                                                                                                                                                        | RF coil (receive only)                                                                                                                                                                                                                                                                                                                         | Same       |
| Number of<br>channels | 8CH                                                                                                                                                                           | 8CH                                                                                                                                                                                                                                                                                                                                            | Same       |
| Field strength        | 1.5 T                                                                                                                                                                         | 1.5 T                                                                                                                                                                                                                                                                                                                                          | Same       |
| Preamplifier noise    | 0.5dB                                                                                                                                                                         | 0.5dB                                                                                                                                                                                                                                                                                                                                          | Same       |
| Energy Source         | Scanner/DC 10V                                                                                                                                                                | Scanner/DC 10V                                                                                                                                                                                                                                                                                                                                 | Same       |
| Compatible<br>systems | GE 1.5T MRI systems                                                                                                                                                           | GE 1.5T MRI systems                                                                                                                                                                                                                                                                                                                            | Same       |

### VIII Technological Comparison

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the letters "RFT" in blue, with a red diamond shape in the middle. The letters are inside a blue oval shape. The letters are bold and easy to read. The logo is simple and clean.

| Coil design                                                                                      | phased array      | phased array          | Same |
|--------------------------------------------------------------------------------------------------|-------------------|-----------------------|------|
| Tuning                                                                                           | Hydrogen (~64MHz) | Hydrogen (~64MHz)     | Same |
| Decoupling<br>method                                                                             | Yes               | Yes, active + passive | Same |
| Patient-contacting<br>materials assessed<br>for compliance with<br>biocompatibility<br>standards | Yes               | Yes                   | Same |

The proposed device has the same technological characteristics on design, energy source and using environment as the predicate device. The biocompatibility of material used is both compliance with ISO 10993-5 and ISO 10993-10.

IX. Summary of verification Tests:

All verification tests have been performed according to below standard, the testing results are passed 1. ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl.

AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance;

2.IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for safety and

essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests;

3.ISO 10993-5:2009: Biological evaluation of medical devices. Part 5-Tests for in vitro cytotoxicity.

4.ISO 10993-10:2010 Biological evaluation of medical devices, Part 10-Tests for irritation and skin sensitization.

5.NEMA MS-1-2008 Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging.

6.NEMA MS 3-2008 Determination of Image Uniformity in Diagnostic Magnetic Resonance Images.

7.IEC 60601-2-33:2010+A1:2013+A2:2015 Medical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis.

Bench testing was conducted to demonstrate that image uniformity, SNR and coil surface heating met specified requirements. The test results also show that the 8ch Wrist Coil achieved the expected results and satisfied the standards listed above.

X. Conclusion:

Shenzhen RF Tech Co., Ltd. considers the 8ch Wrist Coil does not raise any new issues of safety or effectiveness, and performs as well as the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K223239](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K223239)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
