K223239 · Shenzhen RF Tech Co., Ltd. · MOS · Dec 8, 2022 · Radiology
Device Facts
Record ID
K223239
Device Name
8ch Wrist Coil
Applicant
Shenzhen RF Tech Co., Ltd.
Product Code
MOS · Radiology
Decision Date
Dec 8, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The 8ch Wrist Coil by Shenzhen RF Tech Co., Ltd. is receive-only coil and is designed for use as general purpose coil. The 8ch Wrist Coil is designed to be use with GE 1.5T MRI systems to produce diagnostic images of the wrists and fingers that can be interpreted by a trained physician.
Device Story
8ch Wrist Coil is a receive-only phased array MRI coil; designed for use with GE 1.5T MRI systems. Device captures RF signals corresponding to proton precession (63.86MHz) in 1.5T magnetic field; utilizes integrated preamplifiers to process signals from 8 channels. Used in clinical imaging environments; operated by trained medical professionals (radiologists/technicians). Output consists of diagnostic images of wrists and fingers; interpreted by physicians to aid in clinical diagnosis. Benefits include high-quality diagnostic imaging of extremities. Device connects via P-Port to compatible GE MRI scanners.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included SNR (NEMA MS-1-2008), image uniformity (NEMA MS 3-2008), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Receive-only phased array coil; 8 channels; 1.5T field strength; 63.86MHz proton frequency. P-Port connector. Integrated preamplifiers (0.5dB noise). Active and passive decoupling. Biocompatible materials per ISO 10993-5/10. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA MS-1, NEMA MS-3.
Indications for Use
Indicated for use in adult population to produce diagnostic images of wrists and fingers using GE 1.5T MRI systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 8, 2022
Shenzhen RF Tech Co., Ltd. % Gary Wang O&R Consultant Bonnier Quality Supervision Consulting (JM) Center Hailunxinyuan No.3203, Jianghai District Jiangmen City, Guangdong 529000 CHINA
Re: K223239
Trade/Device Name: 8ch Wrist Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: October 19, 2022 Received: October 20, 2022
Dear Gary Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K223239
Device Name 8ch Wrist Coil
Indications for Use (Describe)
The 8ch Wrist Coil by Shenzhen RF Tech Co.,Ltd is receive-only coil and is designed for use as general purpose coil. The 8ch Wrist Coil is designed to be use with GE 1.5T MRI systems to produce diagnostic images of the wrists and fingers that can be interpreted by a trained physician.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
--- Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the letters "RFT" in blue, with a red diamond in place of the dot on the "i". The letters are inside of a blue oval. The letters are large and bold, and the diamond is small and centered above the "T". The oval is thin and surrounds the letters closely.
# K223239 510(k) Summary
Date:2022-12-08(YY-MM-DD)
In accordance with 21 CFR 807.92 the following summary of information is provided:
| I. Applicant/Manufacturer: | Shenzhen RF Tech Co., Ltd.<br>2-F,BLD4 Juhui Industrial Park,<br>Tianliao,Guangming,Shenzhen,<br>P.R.China 518132<br>Phone: (+86) 755-2664 1989-113<br>Fax: (+86)755-2664 2989 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/Correspondent: | Bonnier Quality Supervision Consulting (JM) Center<br>Hailunxinyuan No.3203, Jianghai District,<br>Jiangmen City, China 529000<br>Phone: (+86) 13600366215 |
| Contact Person: | Mr. Gary Wang<br>Q&R Consultant<br>Email: gary.wang@bonnier.net.cn |
| II. Device Regulation Information | |
| Common Name of Device: | 8ch Wrist Coil |
| Classification panel: | Radiology |
| Classification Names: | Coil, Magnetic Resonance, Specialty |
| Regulation Number: | 21 CFR 892.1000 |
| Regulation Class: | II |
| Product Code: | MOS |
| Type of 510(k) submission: | Traditional 510(k) |
| III. Device Information | |
| Product Number: | 10-F32119 |
| Device Trade Name: | 8ch Wrist Coil |
| IV. Predicate Device Information | |
| Sponsor: | MRI DEVICES CORPORATION |
| Device: | HRW-63-8 WRIST ARRAY COIL;<br>HRW-127-8 WRIST ARRAY COIL |
| 510(K) Number: | K050622 |
| V. Device Description | |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a blue oval with the letters "RFT" in blue inside. The letter "F" has a red diamond in the middle. The letters are in a bold, sans-serif font. The oval appears to be a logo or emblem.
The 8ch Wrist Coil is receive only phased array coil to produce diagnostic images of the wrists and fingers in adult population.
The 8ch Wrist Coil is tuned to receive RF frequency corresponding to the proton precession in a 1.5 Tesla magnetic field, which is governed by the Larmor equation.
The 8ch Wrist Coil is based on phased array technique to produce the images from 8 different channels with intergraded preamplifiers. The 8ch Wrist Coil is tuned to the proton frequency of 63.86MHz.
### VI. Indications for Use
The 8ch Wrist Coil by Shenzhen RF Tech Co., Ltd. is receive-only coil and is designed for use as general purpose coil. The 8ch Wrist Coil is designed to be use with GE 1.5T MRI systems to produce diagnostic images of the wrists and fingers that can be interpreted by a trained physician.
### VII. Compatibility
The connector of 8ch Wrist Coil is P-Port type. The 8ch Wrist Coil is compatible with GE1.5T MRI systems where coil ID allows.
| Specification | Predicate Device | Subject | Result |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| K number | K050622 | K223239 | |
| Manufacturer | MRI DEVICES<br>CORPORATION | Shenzhen RF Tech Co.,Ltd | |
| Indications for Use | To be used in conjunction with<br>a magnetic resonance<br>scanner to produce diagnostic<br>images of the wrist and hand<br>that can be interpreted by a<br>trained physician. | The 8ch Wrist Coil<br>manufactured by Shenzhen<br>RF Tech Co.,Ltd is<br>receive-only coil and is<br>designed for use as general<br>purpose coil. The 8ch Wrist<br>Coil is designed to be use<br>with GE 1.5T MRI systems<br>to produce diagnostic<br>images of the Wrists and<br>fingers that can be<br>interpreted by a trained<br>physician. | Equivalent |
| Anatomical site | Wrist/Hand | Wrist/Fingers | Same |
| Transmit/Receive | RF coil (receive only) | RF coil (receive only) | Same |
| Number of<br>channels | 8CH | 8CH | Same |
| Field strength | 1.5 T | 1.5 T | Same |
| Preamplifier noise | 0.5dB | 0.5dB | Same |
| Energy Source | Scanner/DC 10V | Scanner/DC 10V | Same |
| Compatible<br>systems | GE 1.5T MRI systems | GE 1.5T MRI systems | Same |
### VIII Technological Comparison
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters "RFT" in blue, with a red diamond shape in the middle. The letters are inside a blue oval shape. The letters are bold and easy to read. The logo is simple and clean.
| Coil design | phased array | phased array | Same |
|--------------------------------------------------------------------------------------------------|-------------------|-----------------------|------|
| Tuning | Hydrogen (~64MHz) | Hydrogen (~64MHz) | Same |
| Decoupling<br>method | Yes | Yes, active + passive | Same |
| Patient-contacting<br>materials assessed<br>for compliance with<br>biocompatibility<br>standards | Yes | Yes | Same |
The proposed device has the same technological characteristics on design, energy source and using environment as the predicate device. The biocompatibility of material used is both compliance with ISO 10993-5 and ISO 10993-10.
IX. Summary of verification Tests:
All verification tests have been performed according to below standard, the testing results are passed 1. ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl.
AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance;
2.IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for safety and
essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests;
3.ISO 10993-5:2009: Biological evaluation of medical devices. Part 5-Tests for in vitro cytotoxicity.
4.ISO 10993-10:2010 Biological evaluation of medical devices, Part 10-Tests for irritation and skin sensitization.
5.NEMA MS-1-2008 Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging.
6.NEMA MS 3-2008 Determination of Image Uniformity in Diagnostic Magnetic Resonance Images.
7.IEC 60601-2-33:2010+A1:2013+A2:2015 Medical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis.
Bench testing was conducted to demonstrate that image uniformity, SNR and coil surface heating met specified requirements. The test results also show that the 8ch Wrist Coil achieved the expected results and satisfied the standards listed above.
X. Conclusion:
Shenzhen RF Tech Co., Ltd. considers the 8ch Wrist Coil does not raise any new issues of safety or effectiveness, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.