← Product Code [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS) · K213397

# InkSpace Imaging Pediatric Body Array (K213397)

_Inkspace Imaging, Inc. · MOS · Dec 17, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K213397

## Device Facts

- **Applicant:** Inkspace Imaging, Inc.
- **Product Code:** [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS.md)
- **Decision Date:** Dec 17, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

The InkSpace Imaging Pediatric Body Array is a receive-only coil, used for obtaining diagnostic images of pediatric cardiac and body in GE 3.0T magnetic resonance imaging systems. These images, when interpreted by a trained physician, yield information that may assist in diagnosis.

## Device Story

The InkSpace Imaging Pediatric Body Array is a 24-channel, receive-only phased array coil designed for use with GE 3.0T MRI scanners. It consists of two flexible pads (anterior and posterior) containing 12 elements each, designed to conform to pediatric anatomy. The device detects hydrogen nuclei signals to generate high-resolution diagnostic images. It is operated by trained clinical staff in a radiology setting. The resulting images are interpreted by a physician to assist in clinical diagnosis. The flexible design improves signal-to-noise ratio and uniformity compared to rigid coils, potentially enhancing diagnostic quality for pediatric patients.

## Clinical Evidence

No clinical data. Bench testing only, including SNR, image uniformity, surface heating, decoupling, EMC, and electrical/mechanical safety. Biocompatibility (ISO 10993), usability, and transit testing were performed. Sample clinical images were reviewed by a board-certified radiologist to support image quality.

## Technological Characteristics

24-channel receive-only phased array coil; flexible, non-sterile, reusable; operates at 127.73 MHz (hydrogen resonance) for 3.0T MRI systems. Materials are soft and pliable. Biocompatibility per ISO 10993. No active software/algorithm components.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Pediatric Body-Cardiac Coil ([K101949](/device/K101949.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 17, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

InkSpace Imaging, Inc. % Taras Bouzakine Director, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 SAN JOSE CA 95110

Re: K213397

Trade/Device Name: InkSpace Imaging Pediatric Body Array Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: October 15, 2021 Received: October 18, 2021

Dear Taras Bouzakine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name

InkSpace Imaging Pediatric Body Array

Indications for Use (Describe)

The InkSpace Imaging Pediatric Body Array is a receive-only coil, used for obtaining diagnostic images of pediatric cardiac and body in GE 3.0T magnetic resonance imaging systems. These images, when interpreted by a trained physician, yield information that may assist in diagnosis.

| Type of Use ( <i>Select one or both, as applicable</i> )                                                                                  |                                                                                                                                          |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Notification K213397

### GENERAL INFORMATION [807.92(a)(1)]

#### Applicant:

InkSpace Imaging, Inc. 5635 West Las Positas Blvd, Suite 403/404 Pleasanton, CA 94588 USA Phone: 925-425-7410

### Contact Person:

Taras Bouzakine Director, Regulatory Affairs, Experien Group, LLC Regulatory Consultant for InkSpace Imaging, Inc. 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA

### Date Prepared: October 15, 2021

### DEVICE INFORMATION [807.92(a)(2)]

#### Trade Name:

InkSpace Imaging Pediatric Body Array

### Generic/Common Name:

Coil, Magnetic Resonance, Specialty

### Classification:

Class II per 21 CFR§892.1000

# Product Code:

MOS

# PREDICATE DEVICE(S) [807.92(a)(3)]

Pediatric Body-Cardiac Coil (K101949)

{4}------------------------------------------------

# DEVICE DESCRIPTION [807.92(a)(4)]

The InkSpace Imaging Pediatric Body Array is a phased array, receive-only coil intended to work with GE 3.0T MRI scanners to provide optimal signal-to-noise ratio and uniformity for high resolution pediatric body MR imaging examinations. The size and layout of individual channels of the InkSpace Imaging Pediatric Body Array were designed specifically to fit the pediatric population for body imaging. The coil is comprised of 24 total channels, with two individual flexible pads, intended to be placed anterior and posterior to the patient, containing 12 individual elements. The elements are optimized to be flexible to conform to a pediatric patient's anatomy.

## INDICATIONS FOR USE [807.92(a)(5)]

The InkSpace Imaging Pediatric Body Array is a receive-only coil, used for obtaining diagnostic images of pediatric cardiac and body in GE 3.0T magnetic resonance imaging systems. These images, when interpreted by a trained physician, yield information that may assist in diagnosis.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]

The InkSpace Imaging Pediatric Body Array and its predicate, the Pediatric Body-Cardiac Coil (K101949), are classified as receive-only coils for 3.0 Tesla magnetic resonance imaging systems and are intended for use in obtaining diagnostic images of the body and cardiac regions of pediatric populations. Both devices detect hydrogen as the resonant nuclei and use phased array coil configurations to enhance overall signal to noise ratio. The proposed device and its predicate device have similar technological characteristics and employ the same operating principles. The primary technological difference between the two devices is that the predicate device utilizes an 8-element phased array RF structure, as opposed to the 24-elements of the proposed device. These minor differences in technological characteristics between the InkSpace Imaging Pediatric Body Array and the Pediatric Body-Cardiac Coil have been evaluated and determined to not raise any new or different questions of safety or effectiveness and thus is substantially equivalent to the predicate device.

### SUBSTANTIAL EQUIVALENCE

A comparison table summarizing the specifications and features of the InkSpace Imaging Pediatric Body Array and the predicate device, the Pediatric Body-Cardiac Coil, is included in the Substantial Equivalence Table.

{5}------------------------------------------------

## PERFORMANCE DATA [807.92(b)]

Performance testing was performed in accordance with the FDA Guidance Document entitled, "Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance Based Pathway," issued December 11, 2020.

All necessary nonclinical testing was conducted on the InkSpace Imaging Pediatric Body Array to support a determination of substantial equivalence to the predicate device.

## Nonclinical Testing Summary [807.92(b)(1)]

The nonclinical testing included:

- Image Signal to Noise (SNR)
- Image Uniformity ●
- . Surface Heating
- Acquired Image Quality ●
- Decoupling Circuit
- EMC Immunity, Electrostatic Discharge ●
- General Electrical/Mechanical Safety ●

In addition to the above nonclinical testing, the InkSpace Imaging Pediatric Body Array also underwent the following testing:

- Biocompatibility Testing (Cytotoxicity, Sensitization, Irritation)
- Usability Testing
- Transit Testing .

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and the design of the InkSpace Imaging Pediatric Body Array meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the InkSpace Imaging Pediatric Body Array does not raise new or different questions of safety or effectiveness for pediatric body MR imaging examinations when compared to the predicate device.

### Clinical Testing Summary [807.92(b)(2)]

Not applicable. Clinical testing was not required to support substantial equivalence of the InkSpace Imaging Pediatric Body Array to the predicate device. Acquired image quality was analyzed based on sample clinical images. These images, as well as a review of their clinical quality by a board-certified radiologist, have been included as part of this submission.

### CONCLUSIONS [807.92(b)(3)]

Based on the results from the nonclinical tests performed in support of the InkSpace Imaging Pediatric Body Array, it is concluded that the proposed device is safe, effective, and performs at least as safely and effectively as the legally marketed predicate device.

### SUMMARY

The InkSpace Imaging Pediatric Body Array is substantially equivalent to the predicate device.

{6}------------------------------------------------

# Substantial Equivalence Table

| Device Name                                       | InkSpace Imaging Pediatric Body Array<br>(Subject Device)                                                                                                                                                                                                                                                       | Pediatric Body-Cardiac Coil<br>(Predicate)                                                                                                                                                                                                                                                                        | Comparison                                                                                                                                                                                           |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification                                    | Magnetic Resonance Diagnostic Device<br>per 21 CFR§892.1000                                                                                                                                                                                                                                                     | Magnetic Resonance Diagnostic Device<br>per 21 CFR§892.1000                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                 |
| Product Code                                      | MOS (Coil, Magnetic Resonance,<br>Specialty)                                                                                                                                                                                                                                                                    | MOS (Coil, Magnetic Resonance,<br>Specialty)                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                 |
| Indications for<br>Use                            | The InkSpace Imaging Pediatric Body<br>Array is a receive-only coil, used for<br>obtaining diagnostic images of pediatric<br>cardiac and body in GE 3.0T magnetic<br>resonance imaging systems. These images,<br>when interpreted by a trained physician,<br>yield information that may assist in<br>diagnosis. | The Pediatric Body-Cardiac Coil is a<br>receive-only coil, used for obtaining<br>diagnostic images of pediatric cardiac and<br>body in Philips Achieva 3.0T magnetic<br>resonance imaging systems. These images<br>when interpreted by a trained physician,<br>yield information that may assist in<br>diagnosis. | Similar. Only difference is the type of<br>MRI system. The proposed device is used<br>with the GE 3.0T MRI system while the<br>predicate device is used with the Philips<br>Achieva 3.0T MRI system. |
| Patient Anatomy<br>and Population                 | Imaging of general body part regions<br>including cardiac regions; designed for<br>pediatric patients.                                                                                                                                                                                                          | Imaging of general body part regions<br>including cardiac regions; designed for<br>pediatric patients.                                                                                                                                                                                                            | Same                                                                                                                                                                                                 |
| Comparison of<br>Technological<br>Characteristics | 24 channel, receive-only phased array,<br>designed for 3.0T MR systems (127.73<br>MHz, hydrogen).<br>Device is reusable, non-sterile, and by<br>prescription only.<br>Device is made of soft, pliable materials,<br>and is composed of an anterior and a<br>posterior pad.                                      | 8 channel, receive-only phased array,<br>designed for 3.0T MR systems (127.73<br>MHz, hydrogen).<br>Device is reusable, non-sterile, and by<br>prescription only.<br>Device is made of rigid materials and is<br>composed of a base plate coil and an<br>anterior coil.                                           | Similar. Differences do not raise any new<br>or different questions of safety and<br>effectiveness.                                                                                                  |
|                                                   | Compatible MRI<br>Systems                                                                                                                                                                                                                                                                                       | GE 3.0T MRI Systems                                                                                                                                                                                                                                                                                               | Philips 3.0T MRI Systems                                                                                                                                                                             |
| Biocompatibility                                  | ISO 10993 testing                                                                                                                                                                                                                                                                                               | ISO 10993 testing                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                 |

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K213397](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K213397)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
