dS TorsoCardiac 1.5T, dS MSK S 1.5T, dS MSK M 1.5T
K212864 · Philips Healthcare (Suzhou) Co., Ltd. · MOS · Dec 1, 2021 · Radiology
Device Facts
Record ID
K212864
Device Name
dS TorsoCardiac 1.5T, dS MSK S 1.5T, dS MSK M 1.5T
Applicant
Philips Healthcare (Suzhou) Co., Ltd.
Product Code
MOS · Radiology
Decision Date
Dec 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Magnetic Resonance (MR) coil is used with a Philips 1.5T MR scanner. A trained physician interprets the diagnostic images (of the anatomy of interest) produced.
Device Story
Phased-array receive-only MR coils; used with Philips 1.5T scanners. Input: RF signals from patient anatomy (torso, cardiac, neck, wrist, foot, ankle, elbow, shoulder, hip). Operation: Coils capture MR signals; transmit to scanner via interface; no internal energy source. Construction: PU foil, flexible PCBA, EVA foam; bendable design for patient fit. Output: Diagnostic images for physician interpretation. Used in clinical imaging environments; operated by trained technicians/physicians. Benefits: High-resolution imaging; improved flexibility and serviceability compared to rigid predecessors.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including SNR (NEMA MS 1), image uniformity (NEMA MS 3), phased array characterization (NEMA MS 9), and RF coil heating (NEMA MS 14).
Technological Characteristics
Phased-array receive-only coils. Materials: PU foil, EVA30 foam, flexible PCBA. Energy: Derived from MR scanner. Connectivity: DCI connector to MR system. Standards: AAMI/ANSI ES60601-1, IEC 60601-2-33, IEC 60601-1-2, ISO 14971, IEC 62366.
Indications for Use
Indicated for use with Philips 1.5T MR scanners to produce diagnostic images of anatomy of interest for interpretation by a trained physician.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
1.5T 8-CHANNEL MEDIUM GENERAL PURPOSE FLEX COIL (K111673)
Submission Summary (Full Text)
{0}------------------------------------------------
December 1, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Philips Healthcare (Suzhou) Co., Ltd. % Seaman Shao Associate Regulatory Affairs Manager No.258, ZhongYuan Road, Suzhou Industrial Park Suzhou, Jiangsu 215024 CHINA
Re: K212864
Trade/Device Name: dS TorsoCardiac 1.5T, dS MSK S 1.5T, dS MSK M 1.5T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: August 26, 2021 Received: September 8, 2021
Dear Seaman Shao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212864
Device Name
dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T
Indications for Use (Describe)
The Magnetic Resonance (MR) coil is used with a Philips 1.5T MR scanner. A trained physician interprets the diagnostic images (of the anatomy of interest) produced.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | August 26, 2021 |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, ZhongYuan Road, Suzhou Industrial Park,<br>Suzhou Jiangsu, CHINA, 215024<br>Establishment Registration Number: 3009529630 |
| Primary<br>Contact Person: | Seaman Shao<br>Associate Regulatory Affairs Manager<br>Phone: +86-13913116320<br>E-mail: seaman.shao@philips.com |
| Secondary<br>Contact Person | Erhong Wang<br>Senior Regulatory Affairs Manager<br>Phone: +86-13021019589<br>E-mail: erhong.wang@philips.com |
| Device Name: | dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T |
| Classification: | Classification Name: Magnetic resonance diagnostic device |
| | Classification<br>Regulation: 21CFR §892.1000 |
| | Classification Panel: Radiology |
| | Device Class: Class II |
| | Primary Product code: MOS |
| Primary<br>Predicate<br>Device: | Trade Name: (1) SENSE Torso 16<br>(2) SENSE MSK S 8CH 1.5T<br>(3) 1.5T 8-CHANNEL MEDIUM GENERAL<br>PURPOSE FLEX COIL |
| | Manufacturer: Invivo Corporation |
| | 510(k) Clearance: (1) K122897, 21-12-2012<br>(2) K120122, 10-02-2012<br>(3) K111673, 23-12-2011 |
{4}------------------------------------------------
| Classification Regulation: | 21CFR §892.1000 |
|----------------------------|--------------------------------------|
| Classification Name: | Magnetic resonance diagnostic device |
| Classification Panel: | Radiology |
| Device Class: | Class II |
| Product Code: | MOS |
#### Device The proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T description: are designed to be used in conjunction with MR Scanner to produce diagnostic images of the anatomy of interest that can be interpreted by a trained physician.
The proposed dS TorsoCardiac 1.5T is a phased array receive-only coil for high resolution diagnostic imaging of torso, cardiac, and neck. The coil foam is composed of PU foil, flexible PCB and EVA to provide sufficient flexibility along Left-Right direction for patient body scan. The layers from outside to patient side are: PU foil (outer surface), EVA30 foam, PCBA of the coil and PU foil (inner surface). The foam looks flat at the top surface. A few parts, two feed-board boxes, cable housing and a small connector placed across the central Head-Feet axis are also at the top surface. Inner surface is naturally flat and is bendable along slots to fit well to the patient body.
The proposed dS MSK S 1.5T is a phased array receive-only coil for high resolution diagnostic imaging of wrist, foot-ankle, and elbow. The coil foam is composed of PU foil, flexible PCB and EVA to provide sufficient flexibility along Left-Right direction for patient body scan. The layers from outside to patient side are: PU foil (outer surface), EVA30 foam, PCBA of the coil, EVA 30 foam, and PU foil (inner surface). A few parts, feed-board boxes, cable housing and a small connector placed across the central Head-Feet axis are also at the top surface. Inner surface is naturally flat and is bendable along slots to fit well to the patient body.
The proposed dS MSK M 1.5T is a phased array receive-only coil for high resolution diagnostic imaging of shoulder, knee, hip, foot, and elbow. The coil foam is composed of PU foil, flexible PCB and EVA to provide sufficient flexibility along Left-Right direction for patient body scan. The layers from outside to patient side are: PU foil (outer surface), EVA30 foam, PCBA of the coil, EVA30 foam, and PU foil (inner surface). A few parts, feed-board boxes, cable housing and a small connector placed across
{5}------------------------------------------------
| | the central Head-Feet axis are also at the top surface. Inner surface is naturally flat and is bendable along slots to fit well to the patient body. |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T are designed to be used with the Philips 1.5T MRI Scanners. |
| Indications for Use: | The indications for use for the proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T are as follows: |
| | The Magnetic Resonance (MR) coil is used with a Philips 1.5T MR scanner. A trained physician interprets the diagnostic images (of the anatomy of interest) produced. |
| Fundamental Scientific Technology: | The 510(k) summary contains a summary of the technological characteristics of the proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T compared to the predicate devices |
| Attribute | Predicate Device,<br>SENSE Torso 16 | Proposed dS TorsoCardiac 1.5T |
|----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fundamental Scientific Technology | | |
| Type of coil | Phased-array Receive<br>only coil | Identical |
| Magnetic<br>Field<br>Orientation<br>(B0) | Head-Feet oriented | Identical |
| Frequency<br>range | 63.87MHz+/-0.75MHz | Identical |
| Housing<br>Material | PC and PU | Identical |
| Decoupling<br>method | Overlap, pre-amp<br>decouple, Active/Passive<br>PIN diode decoupling | Identical |
| Energy<br>source | Derived from MR<br>scanner, no internal<br>energy source | Identical |
| Workflow to<br>support the<br>connection<br>and signal<br>transmission<br>between the<br>coil and MR<br>system | The coil cable connects<br>to MR system tabletop<br>via coaxial and DC pins<br>mixed analogue<br>connector to fit any<br>intended coil position on<br>tabletop according to the<br>needs of MR scan. | The coil cable connects to<br>dS Interface S 1.5T or dS<br>Interface L 1.5T, which is<br>connected into DCI of MR<br>system tabletop, via a DCI<br>connector with coaxial<br>pins to fit any intended<br>coil position on tabletop<br>according to the needs of<br>MR scan.<br>No impact to device safety<br>and effectiveness since it<br>only provides a better<br>workflow. |
| Flexibility of<br>coil to ensure<br>well fit to<br>patient body | The coil foam is<br>composed of PU foil,<br>flexible PCBA and EVA<br>to provide sufficient<br>flexibility along the Left-<br>Right direction for<br>patient body scan,<br>whereas it is rigid in<br>Head-Feet direction due<br>to the existing of long<br>rigid PCBAs and baluns<br>along this direction. | The coil foam is composed<br>of PU foil, flexible PCBA<br>and EVA to provide<br>sufficient flexibility along<br>the Left-Right direction<br>for patient body scan, and<br>the coil offers a degree of<br>flexibility along the Head-<br>Feet direction since the<br>baluns are moved to the<br>coil cable and the size of<br>rigid PCBAs are<br>decreased.<br>No impact to device safety<br>and effectiveness since it |
| | | only provides a better<br>flexibility and workflow. |
| | | |
| Easily<br>removeable<br>structure to<br>support the<br>serviceability<br>of coil | The baluns and one of<br>rigid PCBA are<br>centralized in the middle<br>of coil, and a removeable<br>plastic cover is designed<br>to cover them, whereas<br>the other rigid PCBAs<br>can be repaired or<br>replaced only after the<br>entire coil foam is<br>disassembled. | All rigid PCBAs are<br>placed at the left and right<br>side of coil, and two<br>removeable plastic covers<br>are designed to cover<br>them, meanwhile the<br>baluns are covered by<br>plastic housings in the coil<br>cable, and they can be also<br>repaired or replaced by<br>removing housings.<br>No impact to device safety<br>and effectiveness since it<br>only provides a better<br>serviceability. |
{6}------------------------------------------------
{7}------------------------------------------------
| Attribute | Predicate Device,<br>SENSE MSK S 8CH 1.5T | Proposed dS MSK S 1.5T |
|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fundamental Scientific Technology | | |
| Type of coil | Phased-array Receive<br>only coil | Identical |
| Magnetic<br>Field<br>Orientation<br>(B0) | Head-Feet oriented | Identical |
| Frequency<br>range | 63.87MHz+/-0.75MHz | Identical |
| Housing<br>Material | PC and PU | Identical |
| Decoupling<br>method | Overlap, pre-amp<br>decouple, Active/Passive<br>PIN diode decoupling | Identical |
| Energy<br>source | Derived from MR<br>scanner, no internal<br>energy source | Identical |
| Workflow to<br>support the<br>connection<br>and signal<br>transmission<br>between the<br>coil and MR<br>system | The coil cable connects<br>to MR system tabletop<br>via coaxial and DC pins<br>mixed analogue<br>connector to fit any<br>intended coil position on<br>tabletop according to the<br>needs of MR scan. | The coil cable connects to<br>dS Interface S 1.5T or dS<br>Interface L 1.5T, which is<br>connected into DCI of MR<br>system tabletop, via a DCI<br>connector with coaxial<br>pins to fit any intended<br>coil position on tabletop<br>according to the needs of<br>MR scan.<br>No impact to device safety<br>and effectiveness since it<br>only provides a better<br>workflow. |
| Flexibility of<br>coil to ensure<br>well fit to<br>patient body | The coil foam is<br>composed of PU foil,<br>flexible PCBA and EVA<br>to provide sufficient<br>flexibility along the Left-<br>Right direction for<br>patient body scan,<br>whereas it is rigid in<br>Head-Feet direction due<br>to the existing of long<br>rigid PCBAs and baluns<br>along this direction. | The coil foam is composed<br>of PU foil, flexible PCBA<br>and EVA to provide<br>sufficient flexibility along<br>the Left-Right direction<br>for patient body scan, and<br>the coil offers a degree of<br>flexibility along the Head-<br>Feet direction since the<br>baluns are moved to the<br>coil cable and the size of<br>rigid PCBAs are decreased<br>significantly. |
| | | |
| No impact to device safety<br>and effectiveness since it<br>only provides a better<br>flexibility and workflow. | | |
{8}------------------------------------------------
{9}------------------------------------------------
| Attribute | Predicate Device,<br>1.5T 8-CHANNEL MEDIUM<br>GENERAL PURPOSE FLEX COIL | Proposed dS MSK M 1.5T |
|----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fundamental Scientific Technology | | |
| Type of coil | Phased-array Receive<br>only coil | Identical |
| Magnetic<br>Field<br>Orientation<br>(B0) | Head-Feet oriented | Identical |
| Frequency<br>range | 63.87MHz+/-0.75MHz | Identical |
| Housing<br>Material | PC and PU | Identical |
| Decoupling<br>method | Overlap, pre-amp<br>decouple, Active/Passive<br>PIN diode decoupling | Identical |
| Energy<br>source | Derived from MR<br>scanner, no internal<br>energy source | Identical |
| Workflow to<br>support the<br>connection<br>and signal<br>transmission<br>between the<br>coil and MR<br>system | The coil cable connects<br>to MR system tabletop<br>via coaxial and DC pins<br>mixed analogue<br>connector to fit any<br>intended coil position on<br>tabletop according to the<br>needs of MR scan. | The coil cable connects to<br>dS Interface S 1.5T or dS<br>Interface L 1.5T, which is<br>connected into DCI of MR<br>system tabletop, via a DCI<br>connector with coaxial<br>pins to fit any intended<br>coil position on tabletop<br>according to the needs of<br>MR scan.<br>No impact to device safety<br>and effectiveness since it<br>only provides a better<br>workflow. |
| Flexibility of<br>coil to ensure<br>well fit to<br>patient body | The coil foam is<br>composed of PU foil,<br>flexible PCBA and EVA<br>to provide sufficient<br>flexibility along the Left-<br>Right direction for<br>patient body scan,<br>whereas it is rigid in<br>Head-Feet direction due<br>to the existing of long<br>rigid PCBAs and baluns<br>along this direction. | The coil foam is composed<br>of PU foil, flexible PCBA<br>and EVA to provide<br>sufficient flexibility along<br>the Left-Right direction<br>for patient body scan, and<br>the coil offers a degree of<br>flexibility along the Head-<br>Feet direction since the<br>baluns are moved to the<br>coil cable and the size of<br>rigid PCBAs are decreased<br>significantly.<br>No impact to device safety<br>and effectiveness since it |
{10}------------------------------------------------
{11}------------------------------------------------
| | | | only provides a better<br>flexibility and workflow. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------|
| Based on the information provided above, dS TorsoCardiac 1.5T is<br>considered substantially equivalent to the primary currently marketed and<br>predicate device SENSE Torso 16 (K122897, 21-12-2012) in terms of<br>fundamental scientific technology<br><br>Based on the information provided above, dS MSK S 1.5T is considered<br>substantially equivalent to the primary currently marketed and predicate<br>device SENSE MSK S 8CH 1.5T (K120122, 10-02-2012) in terms of<br>fundamental scientific technology<br><br>Based on the information provided above, dS MSK M 1.5T is considered<br>substantially equivalent to the primary currently marketed and predicate<br>device 1.5T 8-CHANNEL MEDIUM GENERAL PURPOSE FLEX COIL<br>(K111673, 23-12-2011) in terms of fundamental scientific technology | | | |
| Summary of<br>Non-Clinical<br>Performance<br>Data: | The proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T<br>complies with the following international and FDA-recognized consensus<br>standards:<br>AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012,<br>C1:2009/(R) 2012 and A2:2010/(R) 2012 (Consolidated Text)<br>Medical Electrical Equipment - Part 1: General Requirements. For<br>Basic Safety And Essential Performance (IEC 60601-1:2012,<br>MOD)<br>FDA/CDRH recognition number 19-4 IEC60601-2-33 Ed. 3.2:2015 Medical electrical equipment - Part 2-<br>33: Particular requirements for the basic safety and essential<br>performance of magnetic resonance equipment for medical<br>diagnosis<br>FDA/CDRH recognition number 12-295 IEC60601-1-2 Ed. 4.0:2014 Medical electrical equipment - Part 1-<br>2: General requirements for basic safety and essential performance<br>- Collateral standard: Electromagnetic disturbances - Requirements<br>and tests<br>FDA/CDRH recognition number 19-8 ISO 14971 Ed. 2.0:2007 Medical devices – Application of risk<br>management to medical devices.<br>FDA/CDRH recognition number 5-40. | | |
{12}------------------------------------------------
- Magnetic Resonance (MR) Receive only Coil Performance ● Criteria for Safety and Performance Based Pathway - issued on December 11, 2020
- . Guidance for Industry and FDA Staff - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process" (issued June 16, 2016 - document number 1811)
- IEC 62366 Edition 1.1: 2014-01 Medical devices Application of . usability engineering to medical devices FDA/CDRH recognition number 5-87
- NEMA MS 1-2008(R2020) ● Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging
- NEMA MS 3-2008 (R2020) ● Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- . NEMA MS 9-2008 (R2020) Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images
- NEMA MS 14-2019 Characterization of Radiofrequency (RF) Coil . Heating in Magnetic Resonance Imaging Systems
There are no risks identified in risk management documentation that require clinical data for the purpose of clinical evaluation: Risk Management Plan as Appendix 001, Risk Management Report as Appendix 002, and Risk Management Matrix as Appendix 003 of dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T.
Sufficient evidence is available to demonstrate the ability of dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T achieve the intended performances during normal condition of use;
Full consistency exists between the state-of-the-art, the evaluated data, the risk management documentation and the information materials supplied.
Therefore, the proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T are substantially equivalent to the primary currently marketed and predicate device SENSE Torso 16 (K122897, 21-12-2012), SENSE MSK S 8CH 1.5T (K120122, 10-02-2012) and 1.5T 8-CHANNEL MEDIUM GENERAL PURPOSE FLEX COIL (K111673, 23-12-2011) in terms of safety and effectiveness.
{13}------------------------------------------------
The proposed dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T Summary of Clinical Data: do not require clinical study since substantial equivalence to the predicate device was demonstrated with the following attributes:
- . Design features.
- . Fundamental scientific technology
- Indications for use; ●
- Safetv and effectiveness ●
- Non-clinical performance testing; .
### Substantial The dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T are substantially equivalent to the primary currently marketed and predicate Equivalence device SENSE Torso 16, SENSE MSK S 8CH 1.5T, 1.5T 8-CHANNEL Conclusion: MEDIUM GENERAL PURPOSE FLEX COIL in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, substantial equivalence was demonstrated with non-clinical performance tests, which complied with the requirements specified in the international and FDA-recognized consensus standards. The results of these tests demonstrate that dS TorsoCardiac 1.5T, dS MSK S 1.5T and dS MSK M 1.5T met the acceptance criteria and are adequate for this intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.