OR Head Holder LUCY with HeadCoilSet Siemens 1.5T, OR Head Holder LUCY with HeadCoilSet Siemens 3T, OR Head Holder LUCY with HeadCoilSet Philips 1.5T, OR Head Holder LUCY with HeadCoilSet Philips 3T
K211658 · Noras Mri Products GmbH · MOS · Mar 3, 2022 · Radiology
Device Facts
Record ID
K211658
Device Name
OR Head Holder LUCY with HeadCoilSet Siemens 1.5T, OR Head Holder LUCY with HeadCoilSet Siemens 3T, OR Head Holder LUCY with HeadCoilSet Philips 1.5T, OR Head Holder LUCY with HeadCoilSet Philips 3T
Applicant
Noras Mri Products GmbH
Product Code
MOS · Radiology
Decision Date
Mar 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended medical use of the OR Head Holder LUCY and Head Coil Set 1.57/3T is, in conjunction with a Magnetic Resonance Scanner, intra-operative MR imaging during an open-skull neurosurgery. The OR Head Holder LUCY is used for safe securing the patient's head during a surgical intervention at the head. It is possible to make an intraoperative assessment during the intervention to determine its progress. In the MR area the Head Coils enable the diagnostic imaging. Using Computed Tomography or C-arm Fluoroscopy, it is mandatory to remove the Head Coils and the device serves as a skull clamp only.
Device Story
OR Head Holder LUCY is a mechanical skull fixation device used during neurosurgery. It secures the patient's head via three skull pins; provides stable geometry; allows height adjustment and Z-direction movement for positioning in MRI, CT, and C-arm scanners. When used with integrated NORAS Head Coils (1.5T/3T), it enables intra-operative MR imaging. Operated by physicians or trained staff in clinical settings. The device facilitates surgical planning and resection by allowing intra-operative assessment of procedure progress. It supports prone, lateral, or supine patient positioning. Benefits include secure head fixation and compatibility with multiple imaging modalities to assist neurosurgical interventions.
Clinical Evidence
Bench testing only. Verification and validation conducted per FDA-recognized standards, including DIN EN 60601-2-33:2017 and DIN EN ISO 60601-1 series, to confirm mechanical and electrical safety. Risk management performed per ISO 14971:2013.
Technological Characteristics
Mechanical skull clamp with three-pin fixation. Height-adjustable; Z-axis movement. Compatible with 1.5T/3T MRI, CT, and C-arm. Materials differ from predicates. Complies with DIN EN 60601-2-33:2017 for MRI safety. Force indicator range 0-500N.
Indications for Use
Indicated for patients undergoing open-skull neurosurgery requiring head fixation and intra-operative MR imaging. Compatible with CT and C-arm fluoroscopy when head coils are removed.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
March 3, 2022
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
NORAS MRI products GmbH % Ms. Patricia Kirchner Quality & Regulatory Affairs Leibnizstr. 4 Hoechberg, Bavaria 97204 GERMANY
#### Re: K211658
Trade/Device Name: OR Head Holder LUCY with HeadCoilSet Siemens 1.5T, OR Head Holder LUCY with HeadCoilSet Siemens 3T, OR Head Holder LUCY with HeadCoilSet Philips 1.5T, OR Head Holder LUCY with HeadCoilSet Philips 3T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS, HBL Dated: January 10, 2022 Received: January 12, 2022
Dear Ms. Kirchner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211658
#### Device Name
OR Head Holder LUCY with HeadCoilSet Siemens 1.5 T; OR Head Holder LUCY with HeadCoilSet Siemens 3T; OR Head Holder LUCY with HeadCoilSet Philips 1.5T; OR Head Holder LUCY with HeadCoilSet Philips 3T;
#### Indications for Use (Describe)
The intended medical use of the OR Head Holder LUCY and Head Coil Set 1.57/3T is, in conjunction with a Magnetic Resonance Scanner, intra-operative MR imaging during an open-skull neurosurgery.
The OR Head Holder LUCY is used for safe securing the patient's head during a surgical intervention at the head. It is possible to make an intraoperative assessment during the intervention to determine its progress. In the MR area the Head Coils enable the diagnostic imaging. Using Computed Tomography or C-arm Fluoroscopy, it is mandatory to remove the Head Coils and the device serves as a skull clamp only.
**Type of Use (Select one or both, as applicable)**
| <span style="vertical-align:middle">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------------|
| <span style="vertical-align:middle">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for NORAS MRI products. The logo features the word "NORAS" in bold, blue letters, with a horizontal line underneath. To the right of "NORAS" is a stylized image of a heartbeat line in black, with the words "MRI products" in smaller, blue letters underneath the heartbeat line. The background is white.
# 510(k) Summary
K211658
As required by section 807.92(c)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.
## 1. General Information Date Prepared: March, 01 2021 Manufacturer: NORAS MRI products GmbH Leibnizstr. 4 97204 Hoechberg / Germany Registration Number: 3004929307 Owner/Operator Number: 9071737 Contact Person: Manuel Noras CEO NORAS MRI products GmbH Leibnizstr. 4 97204 Hoechberg / Germany Tel.: (+49) 931 29927 - 29 Fax: (+49) 931 29927 – 20 E-Mail: Manuel.Noras@noras.de Device Name: OR Head Holder LUCY with HeadCoilSet Siemens 1.5 T OR Head Holder LUCY with HeadCoilSet Siemens 3T OR Head Holder LUCY with HeadCoilSet Philips 1.5T OR Head Holder LUCY with HeadCoilSet Philips 3T
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for NORAS MRI products. The logo has the word "NORAS" in bold, blue letters on the left side of the image. To the right of the word "NORAS" is a thin line that resembles a heartbeat. To the right of the heartbeat line are the words "MRI products" in smaller, blue letters.
| Trade Name: | OR Head Holder LUCY with HeadCoilSet Siemens 1.5 T (Ref. 119734)<br>OR Head Holder LUCY with HeadCoilSet Siemens 3T (Ref. 119736)<br>OR Head Holder LUCY with HeadCoilSet Philips 1.5T (Ref. 120281)<br>OR Head Holder LUCY with HeadCoilSet Philips 3T (Ref. 120282) |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | OR Head Holder LUCY with HeadCoilSet Siemens 1.5 T<br>OR Head Holder LUCY with HeadCoilSet Siemens 3T<br>OR Head Holder LUCY with HeadCoilSet Philips 1.5T<br>OR Head Holder LUCY with HeadCoilSet Philips 3T |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR § 892.1000 |
| Device Class: | II |
| Product Code: | MOS, HBL |
## 2. Safety and Effectiveness information supporting substantial equivalence
#### Device Description
The OR Head Holder LUCY can be used for intra-operative MR Imaging, CT Imaging and C-arm Imaging. In use with the NORAS OR Head Coils 1.5T or 3T, the OR Head Holder LUCY can be applied for intra operative MR imaging during an open skull neurosurgery at human patients.
The NORAS OR Head Holder LUCY is used for safe fixation of the patient's head during neurosurgical procedures on the patient's skull with planning and resection aid for MRI, CT and C-arm Imaging.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for NORAS MRI products. The logo features the word "NORAS" in bold, blue letters, with a horizontal line extending from the "S". To the right of the line, there is a stylized image of a heartbeat, followed by the words "MRI products" in smaller, blue letters. The background is white.
Image /page/5/Picture/1 description: The image shows two different views of a medical device used for brain surgery. The device is white and gold and has a curved shape. The first view shows the device without a head in it, and the second view shows the device with a head in it. The device appears to be used to hold the head in place during surgery.
Image: Head Holder Set up with coils and fixed patient
One additional feature of the OR Head Holder LUCY is the ability of height adjustment providing an ideal patient positioning in 70 cm MRI (Siemens Magnetom Aera and Skyra and Philips Ingenia 1.5T/3T, Ingenia Ambition 1.5T and Ingenia Elition 3.0T) and in CT- Scanners. Furthermore, the whole set up can be moved parallel to the bore direction (in Z- direction).
The configuration allows comfortable positioning of the patient in prone, lateral or supine position on the transfer board and allows an exact adjustment of the Head Holder during the positioning of the patient´s head. The system supports awake craniotomy.
Mechanical, the model of the Head Holder allows a slight movement of both posts of the three-point fixation at the Arc with a stable geometry, which minimizes unwanted movements. The patients head will be fixed by three skull pins.
#### Indications for use
The intended medical use of the OR Head Holder LUCY and Head Coil Set 1.5T/3T is, in conjunction with a Magnetic Resonance Scanner, intra-operative MR imaging during an open-skull neurosurgery.
The OR Head Holder LUCY is used for safe securing the patient's head during a surgical intervention at the head. It is possible to make an intraoperative assessment during the intervention to determine its progress. In the MR area the Head Coils enable the diagnostic imaging. Using Computed Tomography or Carm Fluoroscopy, it is mandatory to remove the Head Coils and the device serves as a skull clamp only.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for NORAS MRI products. The logo features the word "NORAS" in a bold, blue font, with a horizontal line extending from underneath the word. To the right of the line, there is a stylized waveform graphic, also in blue, followed by the words "MRI products" in a smaller, black font. The overall design is clean and professional, suggesting a company that specializes in MRI-related products.
## Substantial Equivalence
NORAS MRI products GmbH believes that, within the meaning of the Safe Medical Device Act of 1990, the OR Head Holder LUCY with Head Coil Set 1.5T/3T is substantially equivalent to the following Devices:
| Predicate Device<br>Name and<br>Manufacturer | 510(k)<br>Number | Clearance<br>Date | Product<br>Code | Comparable<br>Properties |
|------------------------------------------------------------------------------|------------------|-------------------|-----------------|-----------------------------------------------|
| NORAS OR Head<br>Holder Flexibility<br>NORAS MRI<br>products GmbH | K133506 | August 6,<br>2014 | MOS; HBL | Fixation of the<br>patient's head |
| MAYFIELD@ Infinity<br>XR2 Skull Clamp<br>Integra LifeSciences<br>Corporation | K130389 | April 23, 2013 | HBL | Fixation of the<br>patient's head<br>and neck |
#### Summary of Technological Characteristics of the Principal Device as
#### Compared with the Predicate Device
The OR Head Holder LUCY bears many similarities to its predicates, the NORAS OR Head Holder Flexibility and the MAYFIELD@ Infinity XR2 Skull Clamp. The OR Head Holder LUCY has the same indication for use as the predicate devices. All of them, the subject and the two predicate devices, are used for the fixation of the patient's head.
The NORAS OR Head Coil Set 1.5T and 3T can be used with both NORAS Head Holders, Flexibility and LUCY. The Coils were already registered with the FDA under the Premarket Submission Number K060758/K091546 and Listing Number E646984/D093265 by NORAS MRI Products GmbH.
The subject device offers the following improvements and minor differences over the predicate device "NORAS OR Head Holder Flexibility".
- -The subject device can be used in Computed Tomography or C-arm Fluoroscopy
- -The Subject device has other dimensional specifications. It is 330 mm less wide and deeper than the predicate device.
- The subject device is made of different materials than the predicate device. -
- -The subject device have a fixation only by three skull pins.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for NORAS MRI products. The logo features the word "NORAS" in bold, blue letters, with the letters slightly overlapping each other. To the right of "NORAS" is a thin, blue line that resembles a heartbeat waveform. Below the waveform, the words "MRI products" are written in a smaller, sans-serif font, also in blue.
The subject device offers the following minor differences over the predicate device "Mayfield Infinity XR2 Skull Clamp"
- -The subject device is made of different materials than the predicate device.
- The subject device uses NORAS or Brainlab Pins, Mayfield Infinity XR2 Skull -Clamp uses only Mayfield Pins.
- -The range of the force indicator of the subject device is higher (0-500N) than of the predicate device (0-356N).
- -The subject device uses slightly different cleaning, disinfection and sterilization procedures.
#### General Safety and Effectiveness Concerns
The OR Head Holder LUCY combined with the NORAS OR Head Coil 1.5T/3T will conform to the FDA recognized NEMA Standards for the measurement of performance and safety parameters and the DIN EN standards for safety issues with the Magnetic Resonance Imaging Devices (DIN EN 60601-2-33:2017). This will assure that the performance of this device can be considered safe and effective when used with the currently available Siemens Magnetom Aera 1.5T & Skyra 3T and Philips Ingenia 1.5T/3T. Ingenia Ambition 1.5T and Ingenia Elition 3.0T systems.
Risk management is ensured via a risk analysis in compliance with DIN EN ISO 14971:2013 to identify and provide mitigation to potential hazards in a risk analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in development, testing and product labeling. To minimize risks, NORAS adhere to recognized and established industry practices and standards to minimize safety and performance risks. Furthermore, the operators and end users of the device are doctors or trained staff and responsible for other hospital procedures.
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for NORAS MRI products. The logo features the word "NORAS" in bold, blue letters, with a horizontal line extending from the right side of the word. To the right of the line, there is a stylized image of a heartbeat, with the words "MRI products" in smaller, blue letters below it. The background of the logo is white.
#### Non-clinical Tests
Testing and analysis have been conducted to show that the verification, validation, and safety requirements have been met per the FDA as established standards. Risk Management is ensured via a risk analysis in compliance with DIN EN ISO 14971:2013 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. NORAS adhere to recognized and established industry standards, such as the DIN EN ISO 60601-1 series, to minimize electrical and mechanical hazards.
Non-clinical bench testing was conducted to support the intended use and to confirm that technological changes do not raise any new issues of safety or effectiveness over the predicate.
## Conclusion as to Substantial Equivalence
The subject device, OR Head Holder LUCY, is optimized for MRI, CT and C-arm supported surgery, just like the Mayfield Infinity XR2 Skull Clamp. While the other predicate device, NORAS OR Head Holder Flexibility, is optimized for its usefulness in interventional surgery applications. All three devices are Head Holders and the fundamental attributes of the subject device and the predicate devices are the same.
The conclusions from the non-clinical data suggest that the subject device has the same fundamental technological characteristics with respect to the predicate device and exhibits an equivalent safety and performance profile as that of the predicate device.
Following, NORAS MRI products GmbH believes, that the OR Head Holder LUCY with HeadCoilSet Siemens 1.5T/3T and HeadCoilSet Philips 1.5T/3T does not raise new questions of safety or effectiveness and, therefore, is substantially equivalent to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.