K201101 · Neocoil, LLC · MOS · Jun 2, 2020 · Radiology
Device Facts
Record ID
K201101
Device Name
1.5T 16ch Shoulder Coil, 3.0T 16ch Shoulder Coil
Applicant
Neocoil, LLC
Product Code
MOS · Radiology
Decision Date
Jun 2, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
To be used in conjunction with GEHC 1.5T Magnetic Resonance Scanners to produce diagnostic images of the shoulder that can be interpreted by a trained physician. To be used in conjunction with GEHC 3.0T Magnetic Resonance Scanners to produce diagnostic images of the shoulder that can be interpreted by a trained physician.
Device Story
Receive-only phased array RF coils; designed for 1.5T or 3.0T GE Healthcare MRI scanners. Inputs: magnetic resonance signals from hydrogen nuclei in shoulder anatomy. Operation: coils receive RF signals while blocking high-frequency magnetic fields from scanner; tuned to proton precession frequency per Larmor equation. Output: diagnostic images (axial, sagittal, coronal oblique slices). Used in clinical MRI settings by trained physicians/technicians. Output enables visualization of shoulder structures (humerus, rotator cuff, etc.) to assist clinical diagnosis of shoulder pathology. Benefits: improved signal-to-noise ratio (SNR) and uniform coverage compared to previous 8-channel designs.
Clinical Evidence
Clinical performance data provided via sample shoulder MRI images (Oblique Axial PD, Oblique Coronal T1/T2, Oblique Sagittal T2). No adverse events reported. Bench testing included biocompatibility, electrical safety (IEC 60601-1, 60601-2-33, 60601-1-2), usability, surface temperature, and NEMA MS 6-2008 SNR/uniformity testing. Results confirm performance consistent with indications.
Technological Characteristics
Receive-only phased array RF coils; 16-channel configuration. Materials: biocompatible components. Energy source: passive (powered by MRI scanner). Form factor: rigid posterior array with flexible anterior array. Connectivity: proprietary system cable connectors for GEHC MRI scanners. Standards: IEC 60601-1, 60601-2-33, 60601-1-2, NEMA MS 6-2008.
Indications for Use
Indicated for patients requiring diagnostic MRI imaging of the shoulder anatomy, including humerus, humeral head, labrum, glenoid, scapula, clavicle, and rotator cuff regions, for interpretation by a trained physician.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
June 2, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
NeoCoil, LLC % Ms. Katie Gonzalez Regulatory & Quality Engineering Manager N27 W23910A Paul Road PEWAUKEE WI 53072
Re: K201101
Trade/Device Name: 1.5T 16ch Shoulder Coil. 3.0T 16ch Shoulder Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: April 22, 2020 Received: April 24, 2020
Dear Ms. Gonzalez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K201101
Device Name 1.5T 16ch Shoulder Coil
### Indications for Use (Describe)
To be used in conjunction with GEHC 1.5T Magnetic Resonance Scanners to produce diagnostic images of the shoulder that can be interpreted by a trained physician.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name 3.0T 16ch Shoulder Coil
Indications for Use (Describe)
To be used in conjunction with GEHC 3.0T Magnetic Resonance Scanners to produce diagnostic images of the shoulder that can be interpreted by a trained physician.
Type of Use (Select one or both, as applicable)
| <div> <div> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.25 4.75L6.75 11.25L2.75 7.25" stroke="#000000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </div> </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <div> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </div> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for NeoCoil. The logo consists of a black circular shape with a white crescent moon shape inside, and a black arrow pointing to the right. To the right of the logo is the word "NeoCoil" in blue and black lettering.
# K201101
## 5. Traditional 510(k) Summary
- 5.1. Applicant
NeoCoil, LLC N27 W23910A Paul Rd Pewaukee, WI 53072 USA
- 5.2. Contact
Katie Gonzalez Regulatory & Quality Engineering Manager 262-522-6124 (office) 262-347-1251 (fax) Katie.Gonzalez@neocoil.com
- 5.3. Preparation Date
April 22, 2020
- 5.4. Name of Device
- Trade/Proprietary name(s): 1.5T 16ch Shoulder Coil . 3.0T 16ch Shoulder Coil . Common name: Magnetic Resonance Specialty Coil . Classification name: 21 CFR 892.1000, Magnetic resonance diagnostic device, Product Code MOS
## 5.5. Predicate Device
NeoCoil 3.0T 8-Channel Shoulder Array Coil, K071611 cleared on 06/28/2007.
## 5.6. Device Description
The 1.5T 16ch Shoulder Coil and 3.0T 16ch Shoulder Coil are receive-only phased array RF coils designed for optimum signal-to-noise ratio (SNR) and uniform coverage of the shoulder anatomy for use with GE Healthcare Maqnetic Resonance Imaging (MRI) scanners. The coils receive magnetic resonance signals generated in hydrogen nuclei (protons) in the shoulder while blocking the high-frequency magnetic field applied by the MRI scanner at specified timings.
Images are typically generated as axial, sagittal, coronal oblique slices and include coverage of the Humerus, Humeral Head, Labrum, Labral Tear, Glenoid, Scapula, Clavicle, and Rotator Cuff regions of the shoulder anatomy.
The 1.5T 16ch Shoulder Coil and 3.0T 16ch Shoulder Coil are tuned to receive RF frequency corresponding to the proton precession in a 1.5 tesla and 3.0 tesla magnetic field (respectively), which is governed by the Larmor equation.
The 1.5T 16ch Shoulder Coil and 3.0T 16ch Shoulder Coil are intended for use in a manner that is identical to the predicate device described in this submission.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black coil-like symbol with an arrow on the left side, followed by the word "NeoCoil" in blue and black. The "Neo" part of the word is in blue, while the "Coil" part is in black.
Proposed labeling is documented in the Instructions for Use manuals for the 1.5T 16ch Shoulder Coil (NC143IFUF-EN) and 3.0T 16ch Shoulder Coil (NC143IFU-EN).
- 5.7. Intended Use
The intended use for the 1.5T 16ch Shoulder Coil and the 3.0T 16ch Shoulder Coil are identical to that of the predicate device, the 3.0T 8-Channel Shoulder Array Coil, K071611 cleared on 06/28/2007.
The intended use is:
ldentical to that of routine imaging; specifically to produce diagnostic images of the shoulder.
Use of the device in conjunction with an MRI scanner is unchanged.
- 5.8. Indications for Use
### 1.5T 16ch Shoulder Coil:
To be used in coniunction with GEHC 1.5T Magnetic Resonance Scanners to produce diagnostic images of the shoulder that can be interpreted by a trained physician.
### 3.0T 16ch Shoulder Coil:
To be used in conjunction with GEHC 3.0T Magnetic Resonance Scanners to produce diagnostic images of the shoulder that can be interpreted by a trained physician.
### 5.9. Technological Characteristics
Both the 1.5T 16ch Shoulder Coil and the 3.0T 16ch Shoulder Coil are similar in design, material, chemical composition and energy source to the legally marketed device, the 3.0T 8-Channel Shoulder Array Coil, K071611 cleared on 06/28/2007.
At a high level, the 1.5T 16ch Shoulder Coil and the 3.0T 16ch Shoulder Coil included as part of this submission and the predicate device are based on the following same technological elements:
- Prescription use;
- Coil designs are receive-only phased array coils; ●
- Decoupling methodology; ●
- Patient contacting materials and chemical composition are known materials that . have been assessed for compliance with recognized biocompatibility standards;
- . Energy source for the coils is the MRI scanner;
- . No energy is supplied by the coils;
- Coils designs are targeted for imaging the shoulder anatomy; .
- Mechanical designs are contoured for the shoulder anatomy: ●
- Manufactured for use with the same MRI scanner manufacturer.
The following technological differences exist between the subject and predicate devices:
- The exterior of the predicate device is a flexible foam whereas the 1.5T 16ch ● Shoulder Coil and 3.0T 16ch Shoulder Coil exterior is comprised of a rigid posterior array and a flexible anterior array;
- Predicate device incorporates 8 channels whereas the1.5T 16ch Shoulder Coil and . the 3.0T 16ch Shoulder Coil incorporate 16 channels;
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with a white crescent shape inside, and a black arrow pointing to the right. To the right of the circular shape is the word "NeoCoil" in blue and black letters. The "Neo" part of the word is in blue, and the "Coil" part is in black.
- . Different system cable connectors for the 1.5T 16ch Shoulder Coil and the 3.0T 16ch Shoulder Coil compared to the predicate to interface with the newer MRI scanners.
- . Field strength for the 1.5T 16ch Shoulder Coil
The Indications for Use for the 1.5T 16ch Shoulder Coil is similar to the predicate device, the 3.0T 8-Channel Shoulder Array Coil, K071611 cleared on 06/28/2007 differing in compatible scanner(s) and field strength.
With the exception of compatible scanner(s), the Indications for Use for the 3.0T 16ch Shoulder Coil are identical to the 3.0T 8-Channel Shoulder Array Coil, K071611 cleared on 06/28/2007.
Clinical testing demonstrates that the differences in the compatible scanners of the devices do not affect the safety and/or the effectiveness of the device when used as labeled.
Clinical and non-clinical testing demonstrates that the safety and effectiveness of the 1.5T 16ch Shoulder Coil compared to the predicate device is not adversely affected as a result of the differences.
- 5.10. Testing
A combination of clinical and non-clinical performance data is included, referenced or relied on to demonstrate that the 1.5T 16ch Shoulder Coil and 3.0T 16ch Shoulder Coil are safe and effective and perform in a manner that demonstrates substantial equivalence to the predicate device, the 3.0T 8-Channel Shoulder Array Coil, K071611 cleared on 06/28/207.
## Performance testing - Bench:
A Test Report Summary for non-clinical Bench Testing performed, including testing to FDArecognized consensus standards identified as relevant in FDA guidance document Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016 is outlined below:
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions /<br>Conclusions |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>Assessment | Assess potential<br>biological risks | Evaluation of<br>data; historical<br>use, biologic<br>testing where<br>warranted | Acceptable<br>level of risk | Pass | No identified significant<br>risks. |
| Electrical<br>Safety | Basic electrical<br>safety/essential<br>performance, 60601-1 | Test Lab | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements for basic<br>electrical safety and<br>essential performance<br>met. |
| Electrical<br>Safety | Particular electrical<br>requirements; MR<br>equipment, 60601-2-<br>33 | Test Lab | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements of the<br>particular standard<br>were met. |
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions /<br>Conclusions |
| Electrical<br>Safety | Collateral electrical<br>safety/essential<br>performance, 60601-<br>1-2 | Test Lab /<br>Bench Testing | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements of the<br>collateral standard were<br>met. |
| Usability<br>Assessment | Devices meet<br>customer, end user<br>and patient needs | Actual,<br>simulated or<br>retrospective<br>evaluation of<br>the device<br>and/or data | Pre-defined<br>requirements | Pass | The devices met the<br>needs of the customer,<br>end user and patient. |
| Entrapment,<br>Trapping Zone<br>and Cable<br>Looping<br>(assessment w/<br>scanner) | Assess the device for<br>pinch points,<br>entrapment, cable<br>looping - interfacing<br>with MRI scanner | Evaluation of<br>coil-to-scanner<br>entrapment,<br>trapping and<br>cable looping<br>not covered by<br>test lab<br>assessments. | Requirements<br>based on pre-<br>defined<br>requirements<br>in 60601-1<br>and customer<br>requirements | Pass | Requirements were<br>met. |
| Surface<br>Temperature | Surface temperatures<br>do not exceed limits | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | Surface temperatures<br>were within IEC limits. |
| Unplugged<br>Surface<br>Temperature | Devices remain safe in<br>first fault condition | MRI scanner<br>test | Acceptable<br>level of risk | Pass | Surface temperatures<br>were within IEC limits<br>when the coil is left<br>unplugged in the MRI<br>scanner. |
| Blocking<br>Network<br>Analysis | Ensures devices are<br>designed with<br>adequate active and<br>passive transmit<br>decoupling | Theoretical<br>calculations | Adequate<br>transmit<br>decoupling | Pass | Blocking network<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| Maximum B1<br>Peak | Demonstrate the<br>devices can withstand<br>the maximum B1 peak<br>without obvious signs<br>of arcing, burning,<br>voltage breakdown | MRI scanner<br>test and visual<br>inspection | Pre-defined<br>performance<br>standards | Pass | Coils were able to<br>withstand maximum B1<br>peak without obvious<br>signs of arcing, burning<br>or voltage breakdown. |
| B1 Field<br>Distortion | Measure amount of<br>distortion produced<br>due to presence of an<br>RF coil in the scanner | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | B1 field inhomogeneity<br>meets performance<br>requirements and<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| B0 Filed<br>Distortion | Measure amount of<br>distortion produced<br>due to presence of an<br>RF coil in the scanner | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | B0 field inhomogeneity<br>meets performance<br>requirements and<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions /<br>Conclusions |
| NEMA MS 6-<br>2008 | Evaluate single-<br>channel non-volume<br>special purpose radio-<br>frequency (RF) coils<br>for use with magnetic<br>resonance (MR)<br>imaging (MRI)<br>systems | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | SNR and Image<br>Uniformity are<br>consistent with the<br>requirements for<br>indications for use. |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue. The word "Neo" is in a lighter shade of blue than the word "Coil".
T
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for NeoCoil. The logo consists of a black circular design with an arrow pointing to the right, followed by the word "NeoCoil" in blue and black. The word "Neo" is in blue, while the word "Coil" is in black.
## Performance testing - Clinical:
Clinical data exhibits a mix of technical factors and anatomy as recommended in the FDA guidance, Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices issued November 18, 2016.
No adverse events were reported during clinical performance testing.
| 1.5T 16ch Shoulder Coil | 3.0T 16ch Shoulder Coil |
|----------------------------------|---------------------------------|
| Oblique Axial PD | Oblique Axial PD |
| Oblique Coronal T1 (no fat sat) | Oblique Coronal T1 (no fat sat) |
| Oblique Sagittal T2 (no fat sat) | Oblique Coronal T2 (fat sat) |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black coil-like shape with an arrow at the end, followed by the word "NeoCoil" in blue. The coil shape is on the left side of the logo, and the word "NeoCoil" is on the right side.
Image /page/9/Picture/2 description: This image shows two MRI scans of a shoulder, comparing the results from a 1.5T 16ch Shoulder Coil and a 3.0T 16ch Shoulder Coil. The scan on the left, labeled "Oblique Sagittal T2 (fat sat)," was taken using the 1.5T coil, while the scan on the right, labeled "Oblique Sagittal T2 (no fat sat)," was taken using the 3.0T coil. The image shows the difference in image quality and fat saturation between the two coils.
## 5.11. Conclusion
This submission demonstrates by means of nonclinical and clinical testing that the 1.5T 16ch Shoulder Coil and 3.0T 16ch Shoulder Coil are safe and effective and perform as well as or better than the predicate device, the 3.0T 8-Channel Shoulder Array Coil, K071611 cleared on 06/28/2007 for the indications claimed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.