Head Coil - 12 is a 12 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis. Head Coil - 32 is a 32 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis. Carotid Coil - 8 is an 8 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for obtaining diagnostic images of carotid. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis. Temporomandibular Joint Coil - 4 is a 4 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for diagnostic images of the mandible region. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis. Infant Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for imaging of the infant's brain, spine, heart, torso and extremities. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis. Cardiac Coil - 24 is a 24 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The indications for use include imaging of the heart, and mediastinum regions. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis. Foot & Ankle Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for foot and ankle imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Device Story
Receive-only multichannel phased array RF coils; designed for UIH 3.0T MRI systems. Input: RF signals from hydrogen protons (1H) in patient anatomy. Operation: linear loop coils resonant at system frequency; independent low noise amplifiers; active/passive decoupling circuits protect from transmit field. Output: diagnostic quality MR images. Used in clinical settings by trained physicians to assist in diagnosis. Benefits: optimized image quality for specific anatomical regions; improved patient comfort via rigid/flexible enclosure designs.
Clinical Evidence
Bench testing only. Performance validated via SNR (MS 1-2008, MS 6-2008), image uniformity (MS 3-2008), and phased array characterization (MS 9-2008). Biocompatibility confirmed per ISO 10993-5 and 10993-10. Sample clinical images provided to demonstrate diagnostic quality.
Technological Characteristics
Receive-only phased array RF coils; 4 to 32 channels. Materials: flame/impact resistant, biocompatible. Sensing: linear loop coils resonant at 3T frequency. Decoupling: active (PIN diode) and passive. Connectivity: UIH 3.0T MRI system interface. Standards: IEC 60601-1, IEC 60601-2-33, ISO 10993.
Indications for Use
Indicated for diagnostic MRI imaging of specific anatomical regions: head (12/32 channel), carotid (8 channel), mandible (4 channel), infant brain/spine/heart/torso/extremities (24 channel), heart/mediastinum (24 channel), and foot/ankle (24 channel). For use by trained physicians on UIH 3.0T MRI systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
Shanghai United Imaging Healthcare Co., Ltd. Xin Gao Regulatory Affairs Specialist No. 2258 Chengbei Rd., Jiading Industrial District Shanghai, 201807 Cn
February 8, 2019
#### Re: K183186
Trade/Device Name: Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil -4. Infant Coil - 24. Cardiac Coil - 24. Foot & Ankle Coil - 24 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: November 15, 2018 Received: November 19, 2018
#### Dear Xin Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183186
#### Device Name
Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24
#### Indications for Use (Describe)
Head Coil - 12 is a 12 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Head Coil - 32 is a 32 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Carotid Coil - 8 is an 8 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for obtaining diagnostic images of carotid. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Temporomandibular Joint Coil - 4 is a 4 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for diagnostic images of the mandible region. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Infant Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for imaging of the infant's brain, spine, heart, torso and extremities. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Cardiac Coil - 24 is a 24 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The indications for use include imaging of the heart, and mediastinum regions. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Foot & Ankle Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for foot and ankle imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
Type of Use (Select one or both, as applicable)
| <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------|
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" symbol, which is also in a bold, sans-serif font. The color of the text and the "U" symbol is a dark gray or blue-gray.
# 510 (k) SUMMARY
## 1. Date Prepared [21 CFR 807.92(a)(1)] November 15, 2018
## 2. General Information [21 CFR 807.92(a)(1)]
| Manufacturer: | Shanghai United Imaging Healthcare Co., Ltd<br>2258 Chengbei Rd., Jiading District, Shanghai, 201807 |
|-----------------|------------------------------------------------------------------------------------------------------|
| Contact Person: | Xin GAO<br>Regulatory Affairs Specialist |
| | Tel: +86 (21) 67076888-5386 |
| | Fax: +86-021-67076889 |
| | Email: xin.gao@united-imaging.com |
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | Head Coil - 12, Head Coil - 32, Carotid Coil - 8,<br>Temporomandibular Joint Coil - 4, Infant Coil - 24,<br>Cardiac Coil - 24, Foot & Ankle Coil - 24 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Magnetic Resonance Diagnostic Device |
| Model: | Head Coil - 12, Head Coil - 32, Carotid Coil - 8,<br>Temporomandibular Joint Coil - 4, Infant Coil - 24,<br>Cardiac Coil - 24, Foot & Ankle Coil - 24 |
| Product Code: | MOS |
| Regulation Number: | 21 CFR 892.1000 |
| Device Class: | II |
#### 4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[
The identification of predicate devices within this submission are as follow:
| Predicate Device 1 | |
|--------------------|--------------------------------------------------|
| Manufacturer: | GE Healthcare Coils (USA Instruments, Inc.) |
| Device Name: | 48CH Head Coil |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K163205 |
| Predicate Device 2 | |
| Manufacturer: | Quality Electrodynamics, LLC |
| Device Name: | 16ch Foot/ Ankle SPEEDER |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K181697 |
| Predicate Device 3 | |
| Manufacturer: | INVIVO |
| Device Name: | CARDIOVASCULAR ARRAY COILS |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K061952 |
| Predicate Device 4 | |
| Manufacturer: | Resonance Innovations LLC |
| Device Name: | Orbit and Mandible Array Family |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K133986 |
| Predicate Device 5 | |
| Manufacturer: | Shanghai Chenguang Medical Technologies Co., Ltd |
| Device Name: | Carotid Coil |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K112002 |
| Predicate Device 6 | |
| Manufacturer: | Quality Electrodynamics, LLC |
| Device Name: | Pediatric 16, A 3T Tim Coil |
| Product Code: | MOS |
| Device Class: | II |
| Regulation Number: | 21 CFR 892.1000 |
| FDA 510 (k) #: | K140998 |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, sans-serif font. To the right of the text is a stylized icon that resembles a shield with a horizontal line through the middle. The logo is simple and modern in design.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the use of bold fonts gives it a strong and confident look.
## 5. Device Description [21 CFR 807.92(a)(4)]
Head Coil - 12, Head Coil - 32, Carotid Coil - 8, Temporomandibular Joint Coil - 4, Infant Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24 are receive-only multichannel phased array coils designed for use with UIH 3T MRI systems. The suffix number in coil name stands for the receive channel of the resonant nucleus is hydrogen protons (1H). All coil elements are protected from transmit field by both active and passive decoupling circuits. Each coil is designed to provide optimum image quality for its dedicated human anatomy region as defined in intended use.
# 6. Intended Use [21 CFR 807.92(a)(5)]
- Head Coil - 12 is a 12 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Head Coil - 32 is a 32 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for head imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Carotid Coil - 8 is an 8 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for obtaining diagnostic images of carotid. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Temporomandibular Joint Coil - 4 is a 4 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for diagnostic images of the mandible region. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Infant Coil - 24 is a 24 channel receive-only RF coil designed for use with UIH 3.0T MRI systems. The coil is indicated for use for imaging of the infant's brain, spine, heart, torso and extremities. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
- Cardiac Coil - 24 is a 24 channel receive-only RF Coil designed for use with UIH 3.0T MRI systems. The indications for use include imaging of the heart, and mediastinum regions. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, uppercase letters, stacked on top of each other. To the right of the text is a stylized graphic that resembles the letter "U" with a horizontal line through the middle, creating a cross-like shape within the "U".
- Foot & Ankle Coil 24 is a 24 channel receive-only RF coil designed for use ● with UIH 3.0T MRI systems. The coil is indicated for use for foot and ankle imaging. When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
# 7. Technological Characteristic [21 CFR 807.92(a)(6)]
Head Coil - 12. Head Coil - 32. Carotid Coil - 8. Temporomandibular Joint Coil - 4. Infant Coil - 24, Cardiac Coil - 24, Foot & Ankle Coil - 24 are built with wellestablished technology and design principles in the industry.
All coils are formed with linear loop coils resonant at the MR system frequency into a phased array. Each coil element is connected to an independent low noise amplifier with a low noise figure to maintain optimum SNR.
Two means of protection from RF transmit field are used for each coil element. An active decoupling circuit is controlled by a PIN diode driven by external control signal provided by the MR system. A passive decoupling circuit is also used in case the coil is left unplugged in the system bore.
The rigid coil enclosures are designed with shape matching the anatomy. Flexible flaps are also used in some coils to enhance patient comfort and adjustability. Coil materials are test to be flame and impact resistant, and biocompatible.
## 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]
## Summary of Non-Clinical Tests:
The following testing was conducted on the proposed devices as the predicate devices:
- A ES60601-1:2005/(R)2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- A IEC 60601-2-33 Ed. 3.1:2013, Medical Electrical Equipment - Part 2-33: Particular Requirements For The Basic Safety And Essential Performance Of Magnetic Resonance Equipment For Medical Diagnostic
- A ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
- A ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- A MS 1-2008(R2014), Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Images
- A MS 3-2008(R2014), Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- A MS 6-2008(R2014), Determination of Signal-to-Noise Ratio and Image Uniformity for Single-Channel Non-Volume Coils in Diagnostic MR Imaging
- > MS 9-2008(R2014), Standards Publication Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized graphic element that resembles the letter "U" or a shield, with a vertical white line running through the center, creating a negative space "I" shape. The color scheme is a muted teal or gray-blue.
The test results demonstrated that the devices performs as expected and thus, it is substantially equivalent to the predicate devices to which it has been compared.
## Summary of Clinical Tests:
- Sample clinical images were provided to support the ability of the proposed A devices to generate diagnostic quality images in accordance with the MR guidance on premarket notification submissions.
## 9. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, we conclude that the proposed devices substantially equivalent to the predicate devices. They do not introduce new indications for use, and have the same technological characteristics and do not introduce new potential hazards or safety risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.