← Product Code [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS) · K172695

# AIR Anterior Array, AIR Posterior Array (K172695)

_Ge Healthcare Coils (Usa Instruments, Inc.) · MOS · Nov 22, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K172695

## Device Facts

- **Applicant:** Ge Healthcare Coils (Usa Instruments, Inc.)
- **Product Code:** [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS.md)
- **Decision Date:** Nov 22, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

The AIR Anterior Array is a receive-only RF coil designed for use with GE 3.0T MRI systems. When used with other coils, the indications for use include chest, cardiac, abdomen, torso, pelvis, prostate, hips, peripheral vascular, and long bone imaging. The nucleus detected is hydrogen. The AIR Posterior Array is a receive-only RF coil designed for use with GE 3.0T MRI Systems. The indications for use include spine. When used with other coils, the indications for use also include abdomen, torso, pelvis, prostate, cardiac, hips, and long bone imaging. The nucleus detected is hydrogen.

## Device Story

AIR Anterior and Posterior Arrays are receive-only RF coils for GE 3.0T MRI systems; designed to capture hydrogen nuclei signals for diagnostic imaging. Anterior Array features 30 elements in a soft, pliable, padded housing for patient comfort and anatomical conformity. Posterior Array features 60 elements embedded within the MR system patient table. Operated by MRI technicians/radiologists in clinical settings. Coils transform RF signals into image data processed by the MRI system. Output is diagnostic MR images used by radiologists for clinical decision-making. Benefits include improved signal-to-noise ratio and uniform coverage compared to previous semi-rigid designs.

## Clinical Evidence

No clinical studies were required. Substantial equivalence supported by non-clinical bench testing, including risk analysis, design verification, safety testing (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, MS 6-2008, MS 9-2008), and biocompatibility testing. Sample clinical images were provided.

## Technological Characteristics

Receive-only RF coils for 3.0T MRI. Anterior Array: 30 elements, soft/pliable housing. Posterior Array: 60 elements, embedded in table. Materials: Flame retardant, biocompatible (ISO 10993). Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA MS 6-2008, NEMA MS 9-2008.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- 3.0T AA ([K143345](/device/K143345.md))
- TDI Posterior Array ([K143345](/device/K143345.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

December 8, 2017

GE Healthcare Coils (USA Instruments, Inc.) Mary Mayka Regulatory Affairs Manager 1515 Danner Drive Aurora, Ohio 44202

Re: K172695

Trade/Device Name: AIR Anterior Array, AIR Posterior Array Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: November 3, 2017 Received: November 6, 2017

Dear Mary Mayka:

This letter corrects our substantially equivalent letter of November 22, 2017.

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K172695

Device Name AIR Anterior Array

Indications for Use (Describe)

The AIR Anterior Array is a receive-only RF coil designed for use with GE 3.0T MRI Systems. When used with other coils, the indications for use include chest, cardiac, abdomen, torso, pelvis, prostate, hips, peripheral vascular and long bone imaging. The nucleus detected is hydrogen.

Type of Use (Select one or both, as applicable):

| <span> <input checked="true" type="checkbox"/>    Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
|----------------------------------------------------------------------------------------------------------|
| <span> <input type="checkbox"/>    Over-The-Counter Use (21 CFR 801 Subpart C)   </span>                 |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known)

Device Name AIR Posterior Array

Indications for Use (Describe)

The AIR Posterior Array is a receive-only RF coil designed for use with GE 3.0T MRI Systems. The indications for use include spine. When used with other coils, the indications for use also include abdomen, torso, pelvis, prostate, cardiac, hips, and long bone imaging. The nucleus detected is hydrogen.

Type of Use (Select one or both, as applicable)

| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has a decorative border with wave-like patterns around the edge.

# GE Healthcare 510(k) Premarket Notification Submission

## 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

| Date:                     | September 05, 2017                                                                                              |
|---------------------------|-----------------------------------------------------------------------------------------------------------------|
| Submitter:                | GE Healthcare Coils (USA Instruments, Inc.)<br>1515 Danner Drive<br>Aurora, OH 44202<br>USA                     |
| Primary Contact Person:   | Mary A. Mayka, Ph.D.<br>Regulatory Affairs Manager<br>GE Healthcare<br>Phone: 262-527-3148<br>Fax: 262-364-2785 |
| Secondary Contact Person: | James D. McMahon<br>Regulatory Affairs Director<br>GE Healthcare<br>Phone: 508-382-2858<br>Fax: 262-364-2785    |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. The logo is simple, recognizable, and represents the multinational conglomerate corporation.

GE Healthcare
 510(k) Premarket Notification Submission

|                       | AIR Anterior Array                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | AIR Posterior Array                                                                                                                                                                                                                                                                                                              |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name:    | AIR Anterior Array                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | AIR Posterior Array                                                                                                                                                                                                                                                                                                              |
| Common/Usual Name:    | Coil, Magnetic Resonance, Specialty                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Coil, Magnetic Resonance, Specialty                                                                                                                                                                                                                                                                                              |
| Classification Names: | Magnetic Resonance Diagnostic Device per<br>21 CFR 892.1000                                                                                                                                                                                                                                                                                                                                                                                                                                       | Magnetic Resonance Diagnostic Device per<br>21 CFR 892.1000                                                                                                                                                                                                                                                                      |
| Product Code:         | MOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | MOS                                                                                                                                                                                                                                                                                                                              |
| Predicate Device(s):  | 3.0T AA (K143345)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | TDI Posterior Array (K143345)                                                                                                                                                                                                                                                                                                    |
| Device Description:   | The AIR Anterior Array is a receive-only coil<br>designed to provide optimum signal-to noise<br>ratio and uniform coverage of the chest, cardiac,<br>abdomen, torso, pelvis, prostate, hips, peripheral<br>vascular and long bone. The AIR Anterior Array<br>contains 30 elements and the housing of the AIR<br>Anterior Array is made of a soft, pliable material<br>that conforms to the patient's anatomy. The coil<br>is padded with a soft material designed to<br>maximize patient comfort. | The AIR Posterior Array is a receive-only coil<br>designed to provide optimum signal-to noise<br>ratio and uniform coverage of the spine,<br>abdomen, torso, pelvis, prostate, cardiac, hips<br>and long bone. The AIR Posterior Array contains<br>60 elements and is embedded within the patient<br>table of the MR System.     |
| Intended Use:         | The AIR Anterior Array is a receive-only RF<br>coil designed for use with GE 3.0T MRI<br>systems. When used with other coils, the<br>indications for use include chest, cardiac,<br>abdomen, torso, pelvis, prostate, hips, peripheral<br>vascular, and long bone imaging. The nucleus<br>detected is hydrogen.                                                                                                                                                                                   | The AIR Posterior Array is a receive-only RF<br>coil designed for use with GE 3.0T MRI<br>Systems. The indications for use include spine.<br>When used with other coils, the indications for<br>use also include abdomen, torso, pelvis, prostate,<br>cardiac, hips, and long bone imaging. The<br>nucleus detected is hydrogen. |

Page 2 of 6

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are enclosed within a circular border. The color of the letters and the border is a light blue.

GE Healthcare
 510(k) Premarket Notification Submission

|                                                    | AIR Anterior Array                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | AIR Posterior Array                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of<br>Intended Use:                     | Both the AIR Anterior Array and the predicate<br>device are classified as coils for magnetic<br>resonance imaging devices and are intended for<br>diagnostic use. Both indications for use<br>statements are functional in nature, and do not<br>list specific diseases or conditions. The AIR<br>Anterior Array and the predicate device are<br>indicated for the same patient population, and<br>for the same clinical setting. The predicate<br>device, the 3.0T AA, is one coil as part of the<br>TDI Coil Suite and covers brachial-plexus, pelvis,<br>hips, prostate, abdominal, cardiac, lower extremities,<br>blood, vessels, or long bone anatomies. Therefore,<br>GE Healthcare believes that the AIR Anterior<br>Array has the same intended use as the predicate<br>device in accordance with the FDA's guidance<br>document "The 510(k) Program: Evaluating<br>Substantial Equivalence in Premarket<br>Notifications [510(k)]", dated 28 July 2014. | Both the AIR Posterior Array and the predicate<br>device are classified as coils for magnetic<br>resonance imaging devices and are intended for<br>diagnostic use. Both indications for use<br>statements are functional in nature, and do not<br>list specific diseases or conditions. The AIR<br>Posterior Array and the predicate device are<br>indicated for the same patient population, and for<br>the same clinical setting. The predicate device,<br>the 3.0T AA, is one coil as part of the TDI Coil<br>Suite and covers pelvis, hips, prostate, abdominal,<br>cardiac, lower extremities, or long bone anatomies.<br>Therefore, GE Healthcare believes that the AIR<br>Posterior Array has the same intended use as the<br>predicate device in accordance with the FDA's<br>guidance document "The 510(k) Program:<br>Evaluating Substantial Equivalence in Premarket<br>Notifications [510(k)]", dated 28 July 2014. |
| Comparison of<br>Technological<br>Characteristics: | Overall, the AIR Anterior Array employs the<br>same fundamental scientific technology as its<br>predicate device.<br><br>Coil Design: The most notable technological<br>differences between the AIR Anterior Array and<br>the predicate 3.0T AA is that                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Overall, the AIR Posterior Array employs the<br>same fundamental scientific technology as its<br>predicate device.<br><br>Coil Design: The most notable technological<br>difference between the AIR Posterior Array and<br>the predicate TDI Posterior Array is that the AIR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

Page 3 of 6

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and the space between the letters and the border is filled with a light blue color. The overall design is simple, recognizable, and conveys a sense of tradition and innovation.

GE Healthcare
 510(k) Premarket Notification Submission

| AIR Anterior Array                                                                                                                                                                                                        | AIR Posterior Array                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1) the AIR Anterior Array contains 30<br>elements as compared to the predicate's<br>16 elements and                                                                                                                       | Anterior Array contains 60 elements as<br>compared to the predicate's 32 elements.                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 2) The housing of the AIR Anterior Array<br>is made of a soft, pliable material that<br>conforms to the patient's anatomy as<br>compared to the predicate's semi-rigid<br>construction.                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Operating Principles: The AIR Anterior Array and AIR Posterior Array both operate on the same<br>principles as the predicate devices.                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Materials: The AIR Anterior Array and AIR Posterior Array and the predicate devices use flame<br>retardant materials.                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Safety and Performance Testing: The AIR Anterior Array and AIR Posterior Array and the<br>predicate devices comply with the same safety and performance testing (see Determination of<br>Substantial Equivalence, below). |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| These technological differences do not raise any different questions of safety and effectiveness.                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Determination of<br>Substantial Equivalence:                                                                                                                                                                              | Summary of Non-Clinical Tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                                                                                                                                                           | The predicate and modified devices have been subject to<br>similar risk management testing to demonstrate substantial<br>equivalence of safety and performance. Testing included:                                                                                                                                                                                                                                                                                                                  |
|                                                                                                                                                                                                                           | AAMI/ANSI ES60601-1 IEC 60601-1-2 IEC 60601-2-33 MS 6-2008 MS 9-2008 Maximum B1 peak Network blocking analysis Heat Testing                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                           | Additionally, both predicate and modified devices have a<br>successful biocompatibility track record, as demonstrated<br>by ISO 10993 testing and by their history of use in<br>previously cleared devices.                                                                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                           | The following quality assurance measures were applied to<br>the development of the devices:  Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation)                                                                                                                                                           |
|                                                                                                                                                                                                                           | Summary of Clinical Tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                           | The subjects of this premarket submission, the AIR<br>Anterior Array and AIR Posterior Array, do not require<br>clinical studies to support substantial equivalence. Sample<br>clinical images have been included in this submission.                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                           | Substantial Equivalence Conclusion:                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                                                                                                                                                                           | The indications for use of the proposed devices are<br>comparable to the claimed predicate devices. The AIR<br>Anterior Array and AIR Posterior Array employs<br>equivalent technology to the claimed predicate devices.<br>Additionally, the results from the above non-clinical tests<br>demonstrate that the devices perform as intended. Thus,<br>the AIR Anterior Array and AIR Posterior Array are<br>substantially equivalent to the predicate devices to which<br>they have been compared. |
| Conclusion:                                                                                                                                                                                                               | GE Healthcare considers the AIR Anterior Array and AIR<br>Posterior Array to be as safe, as effective, and performance<br>is substantially equivalent to the predicate devices.                                                                                                                                                                                                                                                                                                                    |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized, cursive font. The letters are enclosed within a circular border that has a swirling, decorative pattern around the edge. The logo is presented in a blue color.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle is surrounded by decorative swirls, also in blue, which add a sense of motion and elegance to the design.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K172695](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K172695)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
