TxRx Knee 15 Flare MR Coil 3.0T

K163066 · Quality Electrodynamics, LLC · MOS · Nov 29, 2016 · Radiology

Device Facts

Record IDK163066
Device NameTxRx Knee 15 Flare MR Coil 3.0T
ApplicantQuality Electrodynamics, LLC
Product CodeMOS · Radiology
Decision DateNov 29, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The TxRx Knee 15 Flare MR Coil 3.0T is intended for use with Siemens 3.0T MR systems to produce diagnostic images of the knee that can be interpreted by a trained physician.

Device Story

The TxRx Knee 15 Flare MR Coil 3.0T is a 15-channel transmit/receive phased array RF coil designed for knee MRI. It functions as an accessory to Siemens 3.0T MR systems, capturing radiofrequency signals from the knee anatomy. The device features a split-top rigid polycarbonate housing shaped to fit the knee and leg. It is used in clinical imaging environments by trained technologists; the resulting images are interpreted by physicians to aid in clinical diagnosis. The device utilizes active PIN diode and passive blocking circuitry for signal management. It connects to the MR system via a single system cable. The coil provides diagnostic-quality images, enabling non-invasive visualization of knee structures.

Clinical Evidence

Bench testing included SNR and uniformity analysis per IEC 62464, electrical safety/EMC per AAMI/ANSI ES60601-1 and IEC 60601-2-33, and SAR measurement per NEMA MS-8. Surface heating remained below 41°C. Clinical evidence consisted of volunteer knee scans demonstrating diagnostic image quality.

Technological Characteristics

15-channel transmit/receive phased array RF coil. Materials: Polycarbonate housing. Connectivity: Single system cable interface for Siemens 3.0T MR systems. Sensing: Active PIN diode and passive blocking circuitry. Standards: AAMI/ANSI ES60601-1, IEC 60601-2-33, IEC 62464, NEMA MS-8.

Indications for Use

Indicated for patients requiring diagnostic MRI imaging of the knee using Siemens 3.0T MR systems.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle. Inside the circle is a stylized graphic of three human profiles facing right, stacked on top of each other, with flowing lines suggesting movement or connection. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 29, 2016 Quality Electrodynamics, LLC % Ms. Kathleen Aras Director, Regulatory and Quality Affairs 6655 Beta Drive. Suite 100 MAYFIELD VILLAGE OH 44143 Re: K163066 Trade/Device Name: TxRx Knee 15 Flare MR Coil 3.0T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: October 31, 2016 Received: November 2, 2016 Dear Ms. Aras: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K163066 Device Name TxRx Knee 15 Flare MR Coil 3.0T Indications for Use (Describe) The TxRx Knee 15 Flare MR Coil 3.0T is intended for use with Siemens 3.0T MR systems to produce diagnostic images of the knee that can be interpreted by a trained physician. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary #### 1. Applicant Quality Electrodynamics, LLC. (QED) 6655 Beta Drive, Suite 100 Mayfield Village, OH 44143 #### 2. Contact Kathleen Aras Director, Regulatory and Quality Affairs (440) 484-2964 kathleen.aras@qualedyn.com #### 3. Date Prepared 31 October 2016 #### 4. Tradenames TxRx Knee 15 Flare MR Coil 3.0T #### 5. Common name Coil, magnetic resonance, specialty #### Model Numbers 6. QED Model Number: Q7000168 #### 7. Classification Magnetic resonance diagnostic device (21 CFR 892.1000, Product Code MOS, Class II) #### 8. Unmodified Device 3.0T TxRx 15Ch Knee Coil, Quality Electrodynamics, LLC., K082636 #### 9. Device Description The TxRx Knee 15 Flare MR Coil 3.0T is a transmit/receive, 15-channel phased array coil designed for magnetic resonance imaging (MRI) using the Siemens 3.0T MR systems. The TxRx Knee 15 Flare MR Coil 3.0T is intended to be used for imaging the knee. {5}------------------------------------------------ The TxRx Knee 15 Flare MR Coil 3.0T is a reusable, non-invasive device with limited exposure with regard to duration of contact with the body. All coil elements are enclosed in a rigid plastic housing which is fire-rated, has impact and tensile strength, and has been tested for biocompatibility. #### 10. Indications for Use The TxRx Knee 15 Flare MR Coil 3.0T is intended for use with Siemens 3.0T MR systems to produce diagnostic images of the knee that can be interpreted by a trained physician. The Indications for Use statement for the modified TxRx Knee 15 Flare MR Coil 3.0T is not identical to the unmodified device (3.0T TxRx 15Ch Knee Coil); however, the differences do not alter the intended diagnostic use of the device nor do they affect the safety and effectiveness of the device relative to the unmodified device. Both Indications for Use statements for the modified TxRx Knee 15 Flare MR Coil 3.0T and unmodified 3.0T TxRx 15Ch Knee Coil indicate that the device is intended to be used in conjunction with a 3.0T MR system to produce images of the knee and that the images can be interpreted by a trained physician. ### 11. Summary of Technological Characteristics Compared to the Unmodified Device At a high level, the modified device and unmodified device are based on the following same technological elements: - . Transmit/receive phased array RF coils - Active PIN diode switching blocking circuitry. Passive blocking . circuitry. - Split-top mechanical design with an inner cross section shaped to fit . the knee and leg - . Polycarbonate housing material The following technological differences exist between the modified device and unmodified device: - Changes to ergonomics of patient-user interface such as modified . system cable connector and the use of one system cable (modified device) in comparison to two system cables (unmodified device) - Changes to dimensional specifications of the mechanical design of the . enclosure {6}------------------------------------------------ #### 12. Performance Data The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility Testing All surface materials on the TxRx Knee 15 Flare MR Coil 3.0T that are intended to come into direct or indirect contact with patient biological tissues, cells or body fluids have a history of safe use in previouslycleared devices. ## Electrical Safety and Electromagnetic Compatibility The TxRx Knee 15 Flare MR Coil 3.0T was tested to and found to be compliant with AAMI/ANSI ES60601-1 and IEC 60601-2-33. Surface heating was tested in accordance with AAMI/ANSI ES60601-1. The measured temperature of the surface of the coil never exceeded the maximum limit of 41°C. SAR measurement testing was performed per NEMA MS-8 to provide data supporting that the partial body limits for SAR are controlled within the limits described in IEC 60601-2-33. The testing showed that the system controls the SAR limits for the TxRx Knee 15 Flare MR Coil 3.0T such that they remain below the IEC 60601-2-33 partial body limits. ## Performance Testing - Bench The SNR and uniformity of the TxRx Knee 15 Flare MR Coil 3.0T was analyzed per IEC 62464 and was found to conform to predetermined acceptance criteria. ## Performance Testing - Clinical In accordance with the FDA Guidance for Industry: Guidance for the Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, clinical images from volunteer scanning of the knee were obtained from the TxRx Knee 15 Flare MR Coil 3.0T. These images were used to demonstrate that the TxRx Knee 15 Flare MR Coil 3.0T produces diagnostic quality images of the intended anatomy. #### 13. Conclusion The electrical safety and electromagnetic compatibility and biocompatibility data support the safety of the TxRx Knee 15 Flare MR {7}------------------------------------------------ Coil 3.0T and the bench testing per the IEC standards and diagnostic quality sample clinical images demonstrates the performance and effectiveness of the device under the specified use conditions. This testing demonstrates that the TxRx Knee 15 Flare MR Coil 3.0T performs as well as or better than the unmodified device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...