K132444 · Monteris Medical, · MOS · Jan 27, 2014 · Radiology
Device Facts
Record ID
K132444
Device Name
ATAMA SYSTEM
Applicant
Monteris Medical,
Product Code
MOS · Radiology
Decision Date
Jan 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of the AtamA™ System (comprised of two major subsystems: a MR head coil apparatus and a patient stabilization apparatus), in conjunction with a magnetic resonance (MR) imaging system, is the collection of MR data and images of the human brain before, during, and at the end of brain surgery, in a standard operating room, diagnostic MR rooms, or in a MR intra-operative room, while stabilizing the patient's head during neurosurgical procedures and imaging. The patient head-stabilizing apparatus can also be used for other neurosurgical procedures requiring the patient's head to be stabilized or fixed. The head coil can also be used as a standard diagnostic head coil for diagnostic MR imaging. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Device Story
AtamA System consists of MR head coil and patient stabilization apparatus; facilitates neurosurgical workflow by maintaining head fixation during transport between OR table, gurney, and MRI couch; eliminates need to unpin/re-pin patient. Head coil functions as receive-only 8-channel split array for MR data collection. Used by neurosurgeons and radiology staff in OR or diagnostic imaging suites. Provides MR images for physician interpretation to assist in diagnosis and surgical guidance. Benefits include streamlined workflow, reduced patient movement, and consistent head stabilization during imaging and surgery. This submission expands compatibility to 3.0 Tesla MRI systems.
Clinical Evidence
Bench testing only. NEMA standard testing performed to validate 3.0 Tesla compatibility for the head coil. Material analysis conducted to confirm 3.0 Tesla compatibility for the patient stabilization subsystem.
Technological Characteristics
Receive-only, 8-channel split array head coil; patient stabilization apparatus. Materials: thermoplastics, titanium, brass. Compatible with 1.5T and 3.0T MRI systems. No software/algorithm components described.
Indications for Use
Indicated for patients undergoing neurosurgical procedures or diagnostic MR imaging requiring head stabilization and/or brain imaging. Used in standard OR, diagnostic MR rooms, or intra-operative MR suites.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Monteris Medical AtamA Head Coil and Stabilization System (K110411)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Monteris Medical. The logo consists of the word "MONTERIS" in large, bold letters, with the word "MEDICAL" in smaller letters underneath. Above the word "MONTERIS" is a graphic of several small circles arranged in a semi-circular shape.
Premarket Notif cation K132444
# 510(k) Summary
JAN 2 7 2014
న
# Device Information
| Category | Comments |
|---------------------------------------|--------------------------------------------|
| Date Summary Prepared | 22 January 2014 |
| Sponsor | Monteris Medical Corporation |
| | 16305 36th Avenue North, Suite 200 |
| | Plymouth, Minnesota 55446 USA |
| | Tel: 763-253-4710 |
| | Fax: 763-746-0084 |
| | Contact: Brooke Ren, Ph.D. |
| | Sr. VP of Operations |
| | BRen@Monteris.com |
| Correspondent Contact Information | Craig J. Coombs |
| | President, Coombs Medical Device |
| | Consulting, Inc. |
| | 1193 Sherman St |
| | Alameda, CA 94501 |
| | Tel: 510-337-0140 |
| | Fax: 510-337-0416 |
| | CraigJCoombs@gmail.com |
| Device Trade Name | AtamA™ System |
| Device Common Name | Head Coil and Patient Stabilization System |
| Device Classification & Product Code | Class II, MOS |
| Device Classification Name & Citation | Magnetic Resonance Diagnostic Device |
| | 21CFR 892.1000 |
| Device Proprietary Name | Monteris Medical AtamA™ System |
# Predicate Device Information
| Predicate Device Trade Name | Monteris Medical AtamA™ Head Coil and<br>Stabilization System |
|---------------------------------------|---------------------------------------------------------------|
| Manufacturer | Monteris Medical |
| 510(k) Number | K110411 |
| Device Common Name | Head Coil and Patient Stabilization System |
| Device Classification Name & Citation | Magnetic Resonance Diagnostic Device<br>21CFR 892.1000 |
| Device Classification & Product Code | Class II, MOS |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Monteris Medical. The logo consists of the word "MONTERIS" in large, bold, sans-serif font, with the word "MEDICAL" in a smaller font below it. Above the word "MONTERIS" is a series of small circles arranged in a semi-circular pattern.
Premarket Notification K132444
## Description of Device
The AtamA"" System is comprised of two subsystems:
- (i) MR Head Coil
- (ii) Patient Stabilization System
These two subsystems have been designed to work independently and in unison to meet the varied needs of neurosurgical procedures utilizing MRI. In addition, the patient head fixation ring subsystem and patient board can also be used for other neurosurgical procedures requiring the patient's head to be stabilized or fixed. Also, the head coil can also be used as a standard diagnostic head coil for diagnostic MR imaging.
Specifically, the AtamA System allows a streamlined workflow for procedures requiring the patient to undergo MRI examination before, during, or after neurosurgical procedures. The AtamA System maintains patient head fixation while providing a patient board for convenient, safe and effective method of moving the patient from an OR table to transport gurney to MRI couch without the need to unpin and re-pin the patient.
This submission clears the use of the additional 3.0 Telsa AtamA Head Coil for use with specific Siemens and Imris 3.0 Telsa MRI Systems, and the expanded labeling of the components of the Patient Stabilization System for their compatibility with both 1.5 Tesla and 3.0 Telsa systems.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Monteris Medical. The logo consists of the word "MONTERIS" in large, bold, sans-serif font, with the word "MEDICAL" in a smaller font below it. Above the word "MONTERIS" are several small circles arranged in a semi-circular pattern.
Premarket Notification K132444
## Indications for Use
The intended use of the AtamA™ System (comprised of two major subsystems: a MRhead coil apparatus and a patient stabilization apparatus), in conjunction with a magnetic resonance (MR) imaging system, is the collection of MR data and images of the human brain before, during, and at the end of brain surgery, in a standard operating room, diagnostic MR rooms, or in a MR intra-operative room, while stabilizing the patient's head during neurosurgical procedures and imaging.
The patient head-stabilizing apparatus can also be used for other neurosurgical procedures requiring the patient's head to be stabilized or fixed.
The head coil can also be used as a standard diagnostic head coil for diagnostic MR imaging.
When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
### Comparison to Predicate Device
The proposed Monteris Medical AtamA System is substantially equivalent in intended use, technology, design, materials, and physician use to the current Monteris Medical AtamA Head Coil and Patient Stabilization System, cleared under 510(k) Number K110411 for use in 1.5 Tesla MR Systems.
The testing described below demonstrates that the proposed addition of a 3.0 Tesla MR head coil and associated labeling changes on the AtamA System for 3.0 Tesla MR conditional compatibility do not raise any issues of safety or efficacy.
All patient contacting materials are identical in composition, source, and use with respect to the predicate device. The technical modes of action and technical principles are materially the same the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Monteris Medical. The logo consists of the word "MONTERIS" in large, bold, sans-serif font, with the word "MEDICAL" in a smaller font size underneath. Above the word "MONTERIS" are several small circles arranged in a semi-circular shape.
Premarket Notification K132444
| Parameter or<br>Characteristic | Predicate AtamA™<br>System | New AtamA<br>System |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| 510(k) Number | K110411 | K132444 |
| FDA Product<br>Classification / Code | 21 CFR 892.1000 / MOS | Same |
| Indications for Use | The intended use of the AtamA™ System<br>(comprised of two major subsystems: a MR<br>head coil apparatus and a patient stabilization<br>apparatus), in conjunction with a magnetic<br>resonance (MR) imaging system, is the<br>collection of MR data and images of the human<br>brain before, during, and at the end of brain<br>surgery, in a standard operating room, diagnostic<br>MR rooms, or in a MR intra-operative room,<br>while stabilizing the patient's head during<br>neurosurgical procedures and imaging.<br>The patient head-stabilizing apparatus can also<br>be used for other neurosurgical procedures<br>requiring the patient's head to be stabilized or<br>fixed.<br>The head coil can also be used as a standard<br>diagnostic head coil for diagnostic MR imaging.<br>When interpreted by a trained physician, these<br>images provide information that can be useful in<br>determining diagnosis. | Same |
| Operating Principle /<br>Technology | Receive-only, 8-channel split array head coil<br>accompanied by head fixation apparatus that<br>interfaces with MRI systems. | Same |
| Design | The AtamA System is comprised of two<br>subsystems:<br>• MR Head Coil<br>• Patient Stabilization system | Same |
| Materials | Thermoplastics / Titanium / Brass | Same |
| Use Location | Operating Room / MRI Diagnostic Suite | Same |
| MRI Compatibility of<br>Head Coil | 1.5T Conditional | 1.5T / 3.0T<br>Conditional |
| MRI Compatibility of<br>Stabilization Subsystem | 1.5T Conditional | 1.5T / 3.0T<br>Conditional |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Monteris Medical. The logo consists of the word "MONTERIS" in large, bold, sans-serif font, with the word "MEDICAL" in a smaller font size underneath. Above the word "MONTERIS" is a semi-circular arrangement of small, circular shapes, resembling cells or bubbles.
Page 5 of 5
Premarket Notification
K132444
# Summary of Supporting Data
The proposed AtamA 3.0 Tesla MR Head Coil is made by the same manufacturer, and is of the same basic design. as the currently marketed AtamA 1.5 Telsa MR Head Coil. Bench testing in accordance with NEMA Standards has demonstrated that the 3.0 Tesla MR Head Coil is compatible with 3.0 Telsa MRI systems.
An analysis of materials has shown that the AtamA Patient Stabilization Subsystem is also compatible with 3.0 Tesla MR environments and can be labeled with 1.5T and 3.0 T MR Conditional Compatibility.
### Conclusions
It can be concluded from the results that the Monteris Medical AtamA System is safe and effective for use in both the 1.5 and 3.0 Tesla MRI environments and can be labeled with both 1.5 T and 3.0 T MR Conditional Compatibility.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circle around the eagle. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27, 2014
### MONTERIS MEDICAL BROOKE REN 1193 SHERMAN STREET ALAMEDA CA 94501
Re: K132444
Trade/Device Name: AtamA 10 System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: December 9, 2013 Received: December 11, 2013
Dear Dr. Ren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general contrôls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subiect to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll
{6}------------------------------------------------
Page 2-Dr. Ren
free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.lda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Hrehar D. O'Hara
For
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
#### 510(k) Number (if known) K132444
#### Device Name
#### Monteris Medical AtamATM System
#### Indications for Use (Describe)
The intended use of the AtamA™ System (comprised of two major subsystems: a MR head coil apparatus and a patient stabilization apparatus), in conjunction with a magnetic resonance (MR) imaging system, is the collection of MR data and images of the human brain before, during, and at the end of brain surgery, in a standard operating room, diagnostic MR rooms, or in a MR intra-operative room, while stabilizing the patient's head during neurosurgical procedures and imaging.
The patient head-stabilizing apparatus can also be used for other nequiring the paient's head to be stabilized or fixed.
The head coil can also be used as a standard diagnostic head coil for diagnostic MR imaging.
When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | Michael D. O'Hara |
|------------------------------------------------------------------------------|-------------------|
|------------------------------------------------------------------------------|-------------------|
FORM FDA 3881 (1/14)
PSC Publishing Services ( 10) ) 441-67-41 EF
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.