SENSE HEAD 16 MODEL 800647, SENSE SPINE 16 MODEL 800646; SENSE TORSO 16 MODEL 800645
K122897 · Invivo Corporation · MOS · Dec 21, 2012 · Radiology
Device Facts
Record ID
K122897
Device Name
SENSE HEAD 16 MODEL 800647, SENSE SPINE 16 MODEL 800646; SENSE TORSO 16 MODEL 800645
Applicant
Invivo Corporation
Product Code
MOS · Radiology
Decision Date
Dec 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The coils are indicated for use on the order of a physician, in conjunction with Philips 1.5T MR scanners as accessories to produce images of the head, spine and torso.
Device Story
SENSE Head-Spine-Torso (HST) MR Coil System consists of 16-channel receive-only coils (Head 16, Spine 16, Torso 16) designed for Philips 1.5T MRI systems. Coils function as accessories to the system's body coil, which transmits RF signals; HST coils receive resultant RF signals from excited nuclei to produce high-resolution images of head, neck, spine, thorax (including heart), and abdomen. Used in clinical settings by trained MRI technicians/radiologists. Coils can be used individually or in combination (Head+Spine+Torso, Head+Spine, Spine+Torso, or Spine alone). Output is provided to the MRI system for image reconstruction, aiding clinicians in diagnostic decision-making. Benefits include high-resolution anatomical imaging for improved diagnostic accuracy.
Clinical Evidence
Bench testing only. Evaluated per NEMA MS 1-2008 (SNR), MS 3-2008 (Image Uniformity), and MS 9-2008 (Phased Array Coil Characterization). Safety evaluated per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-33. Radiologist review of DICOM files confirmed image quality suitable for clinical applications.
Technological Characteristics
16-channel receive-only phased array coils. Materials: Polycarbonate and Polyethylene (UL94-HF-1 flammability rating). Decoupling: LC tank circuit. Connectivity: Multi-pin connector for Philips 1.5T MRI systems. Electrical isolation: Plastic housing with foam insulation. Standards: NEMA MS 1-2008, MS 3-2008, MS 9-2008; IEC 60601-1, 60601-1-2, 60601-2-33.
Indications for Use
Indicated for patients requiring diagnostic imaging of the head, spine, and torso regions using Philips 1.5T MR scanners. Used on the order of a physician.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
1.5T 8-Channel Medium General Purpose Flex Coil (K093842)
Submission Summary (Full Text)
{0}------------------------------------------------
K122897.
## 510(k) Summary of Safety and Effectiveness SENSE Head-Spine-Torso (HST) MR Coils
| Submitted By: | Invivo Corporation<br>3545 SW 47th Avenue<br>Gainesville, FL 32608 | DEC 21 2012 |
|------------------------------------|---------------------------------------------------------------------------------------------|-------------|
| Date: | December 20, 2012 | |
| Contact Person: | Lance Aulabaugh, Quality Engineer<br>Tel: (352) 336-0010, ext. 156 Fax: (352) 336-1410 | |
| Proprietary Name: | SENSE Head-Spine-Torso MR Coil System:<br>SENSE Spine 16<br>SENSE Head 16<br>SENSE Torso 16 | |
| Common Name: | Coil, Magnetic Resonance, Specialty | |
| Classification Name and Reference: | 21 CFR 892 1000 | |
A magnetic resonance diagnostic device, for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance, class II.
Device Product Code and Panel Code: MOS / Radiology / 90
Indications for Use:
The coils are indicated for use on the order of a physician, in conjunction with Philips 1.5T MR scanners as accessories to produce images of the head, spine and torso.
### Device Description:
The SENSE Head-Spine-Torso (HST) MR Coil System includes 16 channel coil components that are designed for use with Philips 16-Channel 1.5T Magnetic Resonance Imaging (MRI) magnet systems. The coils work in unison with the Body Coil of the MRI system, which will transmit the radio frequency (RF) signals, so that the coils may receive the resultant RF signal from the excited nuclei. The coils are receive-only coils for high resolution imaging of thorax (including heart), abdomen, cervical, lumbar, and thoracic regions of the spine and the head, neck regions. The coils provide unilateral images of the anatomy of interest. The SENSE HST coils are designed to be used together and separately as follows:
| Coil Models | Comments |
|----------------------|-----------------------------------------------------------------------|
| Head + Spine + Torso | All three coils may be used together. |
| Head + Spine | Head Coil does not function unless it is connected to the Spine Coil. |
| Spine + Torso | Spine and Torso may be used together without the Head Coil |
| Spine | Spine Coil may be used alone. |
Section 005 Page 1 of 3
{1}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness SENSE Head-Spine-Torso (HST) MR Coils
### Substantial Equivalence Discussion:
When compared to the predicate devices, the SENSE Head-Spine-Torso system is substantially equivalent to the 1.5T 8-Channel Medium General Purpose Flex Coil (K093842, clearsd March 5, 2010). Substantial equivalence is based on design rationale, safety features, magnet system. compatibility and overall intended use. The indications for use statements differ with respect to the target anatomies; however, the predicate and new device system both fall within the same FDA class II product code, MOS and have the same fundamental intended use. I.ille the predicate, the proposed devices are receive-only MRI coils that are designed for use with Philips 1.5T magnet systems. The electrical isolation methods, decoupling method and housing material flammability ratings are the same for the predicate and new device.
| | PREDICATE | NEW DEVICE |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| Type of Device<br>(FDA Product Code) | 1.5T 8-Channel Medium<br>General Purpose Flex Coil<br>K093842<br>Accessory to MRI Scanner<br>(MOS) | SENSE<br>Head-Spine-Torso<br>System<br>Accessory to MRI Scanner<br>(MOS) |
| Indications for Use | The coil is indicated for use on<br>the order of a physician, in<br>conjunction with Philips 1.5T<br>and 3.0T MR scanners as an<br>accessory to produce images<br>of the hip, knee, ankle, chest,<br>and pelvic regions, as an aid to<br>diagnosis. | The coils are indicated for use<br>on the order of a physician, in<br>conjunction with Philips 1.5T<br>MR scanners as accessories to<br>produce images of the head,<br>spine and torso. |
| Target Anatomies | hip, knee, ankle, chest, and<br>pelvic regions | head, spine and torso regions |
| System Compatibility | Philips 1.5T MRI Systems | Philips 1.5T MRI Systems |
| Type of Coil | Receive-Only | Receive-Only |
| Number of Channels | 8 Channels | 16 Channels |
| Decoupling Method | LC Tank Circuit | LC Tank Circuit |
| Electrical Isolation | Plastic housing with foam<br>insulation | Plastic housing with foam<br>insulation |
| Housing Materials &<br>Flammability Rating | Polycarbonate<br>Polyethylene<br>Rating: UL94-HF-1 | Polycarbonate<br>Polyethylene<br>Rating: UL94-HF-1 |
| Type of MRI System | Standard | Standard |
| Connector | Multi-pin Connector | Multi pin Connector |
{2}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness SENSE Head-Spine-Torso (HST) MR Coils
# Summary of Non-Clinical Tests for Substantial Equivalency Determination:
Per Guidance for Industry: Guidance for the Submission of Premarket Notification for Magnetic Resonance Diagnostic Devices', the following preclinical testing was performed per the standards listed below and supports substantial equivalency to the predicate device: Signal to Noise Ratio (SNR)2, Image Uniformity3, Characterization of for Diagnostic MR Images and IEC 60601 safety evaluation . DICOM files from the 510(k) subject device were reviewed by a radiologist who confirmed the image quality is suitable for clinical applications.
#### Standards Referenced:
- 1. Guidance for the Submission Of Premarket Notifications for Magnetic Resonance Diagnostic Devices Issued On November 14, 1998
- 2. NEMA Standards Publications MS 1-2008: Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging
- 3. NEMA Standards Publications MS 3-2008: Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- 4. NEMA Standards Publications MS 9-2008: Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images
- 5. IEC 60601 Standards:
IEC 60601-1, 1988; Medical Electrical Equipment - Part 1: General Amendment 1, 1991-11, Requirements for Safety Amendment 2, 1995. IEC 60601-1-2: 2001 Medical Electrical Equipment - Part 2: General Edition 2:2001 with Requirements for Safety Collateral Standard: Electromagnetic Compatibility-Requirements and Amendment 1:2004; Edition 2.1 (Edition:2001 consolidated with Tests Amendment 1:2004) IEC 60601-2-33: 2002 Medical Electrical Equipment - Part 2: Particular Requirements for the Safety of Magnetic Resonance Amendment 1:2005 Equipment for Medical Diagnosis
#### Substantial Equivalency Conclusion:
Based on a comparison of technological characteristics, conformance to safety standards preclinical test results, the SENSE Head-Spine-Torso MR coil system is substantially equivalent to the 1.5T 8-Channel Medium General Purpose Flex Coil cleared via K093842.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three arms, representing health, services, and people. The figure is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. The text is written in a sans-serif font and is arranged to follow the curve of the circle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 21, 2012
INVIVO CORP % Mr. Lance Aulabaugh Quality Engineer 3545 SW 47th Avenue Gainesville, FL 32608
Re: K122897
Trade/Device Name:
SENSE Head-Spine-Torso MR Coil System: SENSE Head 16 SENSE Spine 16 SENSE Torso 16
Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: December 6, 2012 Received: December 7, 2012
Dear Mr. Aulabaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - [Contact Person's Name]
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
### Janine M. Morris -S
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
{5}------------------------------------------------
### 510(k) Number (if known): K122897
# Device Name: SENSE Head 16, SENSE Spine 16, SENSE Torso 16
Indications for Use: The coils are indicated for use on the order of a physician, in conjunction with Philips 1.5T MR scanners as accessories to produce images of the head, spine and torso.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Janine M. Morris -S 2012.12.21 16:26:13 -05'00'
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
K122897 510(k):
> Section 004 Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.