K080820 · Mr Instruments, Inc. · MOS · Apr 2, 2008 · Radiology
Device Facts
Record ID
K080820
Device Name
TEM 3002G- HN DUAL TUNE HEAD COIL
Applicant
Mr Instruments, Inc.
Product Code
MOS · Radiology
Decision Date
Apr 2, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The MR Instruments TEM 3002G-HN Dual Tune headcoil is to be used in conjunction with a 3T Magnetic Resonance Imaging system to produce proton ('H) images and spectra of the internal structures of the head. In addition, sodium (20Na) images and spectra may be obtained during the same exam without changing the coil. These images and/or spectral information of the head, when interpreted by a trained physician, may yield information that may assist in diagnosis.
Device Story
TEM 3002G-HN Dual Tune headcoil is a quadrature transmit/receive RF coil for 3T MRI systems; designed for head imaging. Device captures proton (1H) and sodium (23Na) signals simultaneously without coil changes. Rigid plastic housing protects patient/environment. Operated by MRI technicians/radiologists in clinical settings. Output consists of MR images and spectra; interpreted by physicians to assist in diagnosis. Benefits include dual-nuclei imaging capability within a single exam session, improving workflow compared to single-tuned or non-dual-resonant coils.
Clinical Evidence
Bench testing only. Verification and validation included electrical bench measurements, electrical/mechanical safety tests, and MR system safety and performance tests using phantoms and volunteer scans on a GE Healthcare Signa 3T MR system.
Technological Characteristics
Quadrature transmit/receive RF coil; dual-tuned for proton (1H) and sodium (23Na) frequencies at 3T. Rigid plastic housing. Resonant elements with matching and quadrature splitting circuitry. Standalone hardware component for MRI systems.
Indications for Use
Indicated for patients undergoing 3T MRI of the head to obtain proton (1H) and sodium (23Na) images and spectra for diagnostic assistance. Prescription use only.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
GE Healthcare Split Head Coil Assembly for G3 (K040444)
Siemens Medical Solutions 31P/1H headcoil for MAGNETOM Allegra MR system (K042617)
Siemens Medical Solutions 13C/1H headcoil for MAGNETOM Allegra MR system (K042718)
Submission Summary (Full Text)
{1}------------------------------------------------
Koro820
Image /page/0/Picture/1 description: The image shows the logo for MR Instruments INC. The logo consists of the letters "mrii" in a stylized font. The "m" and the two "i"s are solid black, while the "r" is a textured gray. Below the letters is the text "MR INSTRUMENTS INC" in a smaller, sans-serif font.
#### TEM 3002G-HN Dual Tune Headcoil
### Section 2 - 510(k) Summary and Certification
[As required by 21 CFR 807.92(c)]
APR - 2 2008
### 1. Contact Person
| Mark Watson | Tel: (952) 229-8813 |
|-----------------|---------------------|
| VP, Engineering | Fax: (952) 746-1437 |
MR instruments, Inc. 5610 Rowland Road, Suite 145 Minnetonka, MN 55343
### 2. General Information
| Name: | TEM 3002G-HN Dual Tune headcoil |
|----------------------|--------------------------------------------------------------------------------------|
| Trade Name: | TEM 3002G-HN Dual Tune headcoil |
| Common Name: | Headcoil |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Classification: | This device is classified by the Radiology<br>Panel into Class II, (21 CFR 892.1000) |
### 3. Device Description
The TEM 3002G-HN Dual Tune headcoil is a quadrature transmit/receive RF coil tuned to proton ('H) and sodium (22Na) frequencies 3T. The TEM resonant elements and associated matching and quadrature splitting circuitry are enclosed in a rigid plastic housing to prevent any exposure to patient or environment.
The predicate devices are also quadrature transmit/receive RF headcoils. All have similar construction, size and shape. All are tuned to proton ('H) resonant frequency at 3T and two are dual resonant, one at phosphorus (31P) and one at carbon (18C).
#### 4. Intended Use
The MR Instruments TEM 3002G-HN Dual Tune headcoil is to be used in conjunction with a 3T Magnetic Resonance Imaging system to produce proton ('H) images and spectra of the internal structures of the head. In addition, sodium (20 a) images and spectra may be obtained during the same exam without changing the coil. These images and/or spectral information of the head, when interpreted by a trained physician, may yield information that may assist in diagnosis.
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for MR Instruments Inc. The logo consists of the letters "mrii" in a stylized font, with the letters stacked on top of each other. The letters are black, and the background is white. Below the letters, the words "MR INSTRUMENTS INC" are written in a smaller font.
### TEM 3002G-HN Dual Tune Headcoil
## 5. Substantial Equivalence Comparison
The TEM 3002G-HN Dual Tune headcoil is substantially equivalent to the following devices with respect to safety, intended use and design:
| Coil Name | Premarket<br>Notification | Clearance Date |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-------------------|
| GE Healthcare<br>Split Head Coil Assembly for G3, for use<br>with a Signa 3T MR system for producing<br>proton (1H) MR images and spectra | K040444 | April 14, 2004 |
| Siemens Medical Solutions<br>31P/1H headcoil for MAGNETOM Allegra MR<br>system for producing phosphorus (31P) and<br>proton (1H) MR images and spectra | K042617 | November 5, 2004 |
| Siemens Medical Solutions<br>13C/1H headcoil for MAGNETOM Allegra MR<br>system for producing carbon (13C) and proton<br>(1H) MR images and spectra | K042718 | November 12, 2004 |
Similarities between the devices are devices in that they are all transmit and receive RF coils designed to work in conjunction with 3T MR System for acquiring proton ('H) diagnostic images and spectra of the internal structures of the head. The double resonance coils are similar in frequency as noted below. In addition, all coils are similar in size, shape and construction.
Note: The resonance frequency of 13C = 32.4 MHz, 22 Na = 33.8 MHz, and 31P = 51.6 MHz at 3T.
The primary differences between the devices are that the TEM 3002G-HN coil adds sodium (22) imaging capability instead of phosphorus (31P) or carbon (130). The GE Healthcare Split Top Coil is not a double resonant coil and requires switching coils to obtain images or spectra at other frequencies.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo for MR Instruments INC. The logo is black and white and features a stylized "m" followed by what appears to be a textured circle and two vertical lines. Below the symbol is the text "MR INSTRUMENTS INC" in a simple sans-serif font.
# 6. Summary of Studies
Verification and validation testing was performed to ensure that the coil design requirements specification and customer requirements specification were met. Testing activities included electrical bench measurements, electrical/mechanical safety tests, MR system safety and performance tests with phantoms and volunteer scans. MR system tests were conducted using the GE Healthcare Signa 3T MR system running standard clinical applications.
### 7. Conclusion (statement of equivalence)
The data and information provided in this submission supports a substantial equivalence determination, and, therefore, 510(k) premarket notification clearance of the TEM 3002G-HN Dual Tune headcoil.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
# APR - 2 2008
MR Instruments, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K080820
Trade/Device Name: TEM 3002G-HN Dual Tune headcoil Regulation Number: 21 CFR 872.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: March 22, 2008 Received: March 24, 2008
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a circular logo for the FDA Centennial, celebrating the years 1906-2006. The logo features the letters "FDA" prominently in the center, with the word "Centennial" written below. The logo is surrounded by a circular border with stars and other decorative elements.
Protecting and Promoting Public Health
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Page 2
Enclosure
{6}------------------------------------------------
#### TEM 3002G-HN Dual Tune Headcoil
Image /page/5/Picture/1 description: The image shows a logo for MR Instruments Inc. The logo consists of a stylized "m" followed by "r" and two "i"s. Below the logo, the words "MR INSTRUMENTS INC" are printed in a simple sans-serif font.
### Indications for Use Statement
Page 1 of 1
The MR Instruments TEM 3002G-HN Dual Tune headcoil is to be used in conjunction with a 3T Magnetic Resonance Imaging system to produce proton ('H) images and spectra of the internal structures of the head. In addition, sodium (20Na) images and spectra may be obtained during the same exam without changing the coil. These images and/or spectral information of the head, when interpreted by a trained physician, may yield information that may assist in diagnosis.
This device is for prescription use.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office, of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
ices
208082
MR Instruments, Inc.
March 6, 2008
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.