GE SIGNA PROFILE 6 INCH GP COIL

K041921 · Ge Medical Systems, LLC · MOS · Jul 26, 2004 · Radiology

Device Facts

Record IDK041921
Device NameGE SIGNA PROFILE 6 INCH GP COIL
ApplicantGe Medical Systems, LLC
Product CodeMOS · Radiology
Decision DateJul 26, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The GE 0.2T 6-inch GP Coil is a 2-turn solenoid receive only coil designed for MR imaging. The coil has a soft pad, which helps to place the position to be examined in the center of the coil and provides comfort as well. The coil is a general-purpose coil that can be used to image the Cervical Spine, the neck, shoulder, and other extremity anatomy. The coil connects to the Signa MRI system by way of the interface connector.

Device Story

The GE 0.2T 6-inch GP Coil is a receive-only, 2-turn solenoid MR coil designed for use with GE Signa MRI systems. It features a soft pad for patient positioning and comfort. The device functions by receiving magnetic resonance signals from the anatomy placed within the coil; it includes blocking networks to decouple the coil during body coil transmit conditions. It is used in clinical settings by trained MRI technologists to image the cervical spine, neck, shoulder, and extremities. The output is processed by the host MRI system to generate diagnostic images, which are reviewed by physicians to inform clinical decision-making. The device benefits patients by providing high signal-to-noise ratio imaging for diagnostic purposes.

Clinical Evidence

Bench testing only. The device was evaluated against NEMA performance standards and IEC 60601-1 (Medical Equipment Safety) and IEC 60601-2-33 (Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis) to demonstrate that the design met predetermined acceptance criteria.

Technological Characteristics

2-turn solenoid receive-only MR coil; includes blocking networks for decoupling during transmit; soft pad for patient positioning; connects to Signa MRI system via interface connector. Complies with IEC 60601-1 and IEC 60601-2-33 standards.

Indications for Use

Indicated for MR imaging of the cervical spine, neck, shoulder, and other extremity anatomy in patients requiring magnetic resonance diagnostic imaging. Contraindications not specified.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and appear to be hand-drawn. The logo is black and white. K041921 GE Healthcare Technologies P.O. Box 414, Milwaukee, WI 53201 ## 510(K) SUMMARY This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c). #### Applicant and Contact Information | Applicant & Initial Distributor: | GE Healthcare Technologies<br>3200 N. Grandview Blvd.<br>Waukesha, WI 53188 | |----------------------------------|-----------------------------------------------------------------------------| | Est. Registration No: | 2183553- [per 21 CFR 807.87(b)] | | Corresponding Official: | Larry A. Kroger, Ph.D. | | Title: | Senior Regulatory Programs Manager | | Mailing Address: | GE Healthcare Technologies W-400<br>P.O. Box 414<br>Milwaukee, WI 53201 | | Telephone: | 262-544-3894 | | Fax: | 262-548-4768 | | E-mail: | Larry.Kroger@med.ge.com | #### Device Name 0.2T 6-inch GP Coil Magnetic Resonance Coil, 21 CFR 892.1000, 90-MOS #### Marketed Device The 0.2T 6-inch receive only GP coil is substantially equivalent to the currently marketed GE Signa Profile 9 inch coil (K972296), which is also a coil for 0.2T MR imaging. #### Device Description The GE 0.2T 6-inch GP Coil is designed for MR imaging of general purpose of the human anatomy. The coil is a 2-turn solenoid coil in series that provides high signal to noise ratio. The coil has blocking networks to decouple the coil during body coil transmit condition. #### Indications for Use The GE 0.2T 6-inch GP Coil is a 2-turn solenoid receive only coil designed for MR imaging. The coil has a soft pad, which helps to place the position to be examined in the center of the coil and provides comfort as well. The coil is a general-purpose coil that can be used to image the Cervical Spine, the neck, shoulder, and other extremity anatomy. The coil connects to the Signa MRI system by way of the interface connector. {2}------------------------------------------------ P.O. Box 414, Milwaukee, WI 53201 Image /page/1/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are black, creating a bold and recognizable emblem. #### Comparison with Predicate The predicate device is the GE Signa Profile 9 inch coil (K972296). The coil is a similar design and technology. The main difference is the diameter of the coil. #### Summary of Studies The 0.2T 6-inch GP coil is evaluated to the appropriate NEMA performance standards as well as the IEC 60601-1 International Medical Equipment Safety standard and labeling requirement of IEC 60601-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis. Testing was performed to demonstrate that the design met predetermined acceptance criteria. ## Conclusion: It is the opinion of GE that the 0.2T 6-inch GP Coil is substantially equivalent to the GE Signa Profile 9 inch coil (K972296). The GE 0.2T 6 inch GP Coil includes all the indications for use of that of the predicate device. The use of this device will not result in any new potential hazards. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 2 6 2004 Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems, LLC GE Healthcare Technologies P.O. Box 414, W-400 MILWAUKEE WI 53201 Re: K041921 Trade/Device Name: GE 0.2T 6 inch GP coil Regulatory Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: July 15, 2004 Received: July 16, 2004 Dear Dr. Kroger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours, Nancy C. Brygdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The circle is surrounded by a pattern of swirling lines, creating a dynamic and recognizable emblem for the company. GE Healthcare Technologies P.O. Box 414, Milwaukee, WI 53201 # STATEMENT OF INTENDED USE 510(k) Number (if known): _Ko4 /92 / Device Name: 0.2T 6-inch GP Coil ## Indications for Use: The GE 0.2T 6-inch GP Coil is a 2-turn solenoid receive only coil designed for MR imaging. The coil has a The GL 0.2 r o-fich of Or our S a 2 tarn oclonial room of the center of the coil and provides comfort son pag, which helps to plood the posted that can be used to image the Cervical Spine, the neck, as well. The coll is a general purpood our than be soss of the Signa Profile MRI system by way of the interface connector. (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801-109) OR E-2 Over-The-Counter Use Nancy Brogdon (Division Sign-Off) Division of Reproductive, Abdominal,and Radiological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...