K023596 · Medrad, Inc. · MOS · Nov 8, 2002 · Radiology
Device Facts
Record ID
K023596
Device Name
MEDRAD 8-RECEIVER PHASED ARRAY NEUROVASCULAR COIL
Applicant
Medrad, Inc.
Product Code
MOS · Radiology
Decision Date
Nov 8, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Medrad Phased Array Neurovascular coil is a receive-only coil intended to be used with the General Electric Superconducting MRI scanners. The coil will facilitate complete MR imaging of the intracranial/extracranial Neurovascular, skull base and C-Spine without moving the patient or the coil, i.e., no scan room intervention.
Device Story
Receive-only phased array coil; used with GE 1.5T Superconducting MRI scanners; captures MR signals from vertex of skull to aortic arch; enables imaging of intracranial/extracranial neurovascular, skull base, and C-Spine regions; supports parallel imaging; eliminates need for patient/coil repositioning during scan; operated by MRI technicians/radiologists in clinical imaging environment; output is raw MR signal data processed by MRI system into diagnostic images; benefits patient by reducing scan time and improving workflow efficiency.
Clinical Evidence
Bench testing only. Design verification and validation testing performed to confirm compliance with design specifications and international standards. No clinical data provided.
Technological Characteristics
Phased array receive-only coil; housing made of Polyurethane (UL 94V-0); comfort pad made of cotton/urethane (CAL 117); connects via phased array quick disconnect port; compatible with GEMS 1.5T Signa Excite platforms; supports parallel imaging; no external tuning/matching required.
Indications for Use
Indicated for patients requiring MR imaging of the intracranial/extracranial neurovascular, skull base, and C-Spine regions using General Electric Superconducting MRI scanners.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
8 2002 NOV
## KO23596
## Special 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| Submitter's Name: | Medrad Inc. |
|----------------------|--------------------------------------------------------|
| Submitter's Address: | One Medrad Drive, Indianola, PA 15051 USA |
| Telephone Number: | (412) 767-2400, ext. 3326 |
| Fax Number: | (412) 767-2475 |
| Contact Person: | Jim Ferguson, Jr. |
| Date: | October 25, 2002 |
| Proprietary Name: | Medrad 8-Receiver Phased Array Neurovascular Coil |
| Common Name: | MR Imaging Surface Coil |
| Classification: | Class II, 892.1000 |
| Classification Name: | Magnetic Resonance Diagnostic device |
| Predicate Device: | Medrad 1.5T Phased Array Neurovascular Coil<br>K984257 |
Substantial Equivalence: The information provided in this premarket notification demonstrates that the proposed device is substantially equivalent to a legally marketed device. The proposed 8-Receiver Phased Array Neurovascular Coil is substantially equivalent to the 1.5T Phased Array Neurovascular Coil (K984257)
The Medrad 8-Receiver Phased Array Neurovascular Coil maintains the same intended use, similar operational parameters, similar labeling and is used in a manner similar to the predicate device.
Like the Medrad 4-Receiver Neurovascular Coil, the Medrad 8-Receiver Neurovascular Coil is a receive-only coil intended to be used with the General Electric Superconducting MRI scanners for MR imaging of the intracranial Neurovascular, skull base and C-Spine without moving the patient or the coil, i.e., no scan room intervention
Medrad has established, as part of its Quality System, design controls in compliance with FDA's Quality System Regulations (QSRs). These design controls are applied to all Medrad product development processes and product design changes. These design controls were applied to the development of the Stellant CT Injector and meet the requirements of the FDA's QSRs.
As part of the design control a risk analysis was performed, and design verification and validation testing was conducted to support the conclusion drawn by the risk analysis.
Test results demonstrate that the design specifications for the Medrad 8-Receiver Phased Array Neurovascular Coil were met and that the Medrad 8-Receiver Phased Array Neurovascular Coil meets the applicable requirements of the international standards cited. Therefore, it has been determined that the Medrad 8-Receiver Phased Array Neurovascular Coil is substantially equivalent to the predicate device, its predecessor, for its intended use when used as prescribed in the User Operation Manual.
A comparison of features and principles of operation between the proposed device and predicate device is provided in the table below.
{2}------------------------------------------------
| Feature | (Proposed) Medrad 1.5T 8-<br>Receiver Neurovascular | Medrad 1.5T Neurovascular<br>Array Coil (Predicate) |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Coil type | Phased Array Receive-Only Coil | Phased Array Receive-only<br>Quadrature coil. |
| Region of Interest | Vertex of the skull to the aortic<br>arch. | Vertex of the skull to the aortic<br>arch. |
| Compatibility | All phased array GEMS 1.5T<br>Signa Excite platforms with 8-<br>Receiver capability. All Signa<br>System pulse sequences and<br>appropriate imaging options. | All phased array GEMS 1.5T<br>Signa Horizon platforms All Signa<br>System pulse sequences and<br>appropriate imaging options<br>excluding parallel imaging. |
| Tuning | No external tuning, or matching,<br>is necessary since the coil is<br>matched to the recommended<br>anatomy of interest. | No external tuning, or matching, is<br>necessary since the coil is<br>matched to the recommended<br>anatomy of interest. |
| System connection | The coil plugs into the MRI<br>System by way of the Phased<br>Array quick disconnect port | The coil plugs into the MRI System<br>by way of the Phased Array quick<br>disconnect port |
| Imaging configurations | High resolution Head, Parallel<br>imaging Fast Brain,<br>Neurovascular, C-Spine (user<br>optional), Volume Neck (user<br>optional), High resolution Head<br>and C-Spine (user optional) | High resolution Head, Fast<br>Brain, Neurovascular, C-Spine<br>(user optional), Volume Neck<br>(user optional), High<br>resolution Head and C-Spine<br>(user optional) |
| Patient contacting materials comparison information | | |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| | Medrad 1.5T Neurovascular (Proposed) | Medrad 1.5T Neurovascular Array Coil (Predicate) |
| Housing | The housing material is made from Polyurethane; Fire Rated UL 94V-0 | The housing material is made from Glass Filled Polyurethane; Fire Rated UL 94V-0 and Kydex, Fire Rated UL 94V-0 |
| Comfort Pad | Comfort Pad material is made with a cotton material embedded with urethane and is fire rated to CAL 117 | Comfort Pad material is made with a cotton material embedded with urethane and is fire rated to CAL 117 |
:
、*
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a partial view of the Department of Health & Human Services logo. The logo features the department's emblem, which is a stylized representation of an eagle with three lines extending from its body. The text "DEPARTMENT OF HEALTH &" is partially visible, curved along the left side of the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
8 2002 NOV
Re: K023596
Trade/Device Name: Medrad 8 Receiver Phased Array Neurovascular Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II
Product Code: 90 MOS Dated: October 25, 2002 Received: October 28, 2002
Dear Mr. Ferguson:
Mr. Jim Ferguson, Jr.
One Medrad Drive
Medrad, Inc.
Regulatory Affairs Analyst
INDIANOLA PA 15051
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Intended Use
## Indications for Use Statement
510(k) Number:
KO23596
Device Name:
Medrad 8-Receiver Phased Array Neurovascular Coil
Indications for Use:
The Medrad Phased Array Neurovascular coil is a receive-only coil intended to be used with the General Electric Superconducting MRI scanners. The coil will facilitate complete MR imaging of the intracranial/extracranial Neurovascular, skull base and C-Spine without moving the patient or the coil, i.e., no scan room intervention.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David L. Simpson
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K023596
Prescription Use
/
OR
Over-The-Counter Use _ (Per 21 CFR 801.109)
· Medrad, Inc. · Special 510(k) Premarket Submission · 8-Receiver Neurovascular · Confidential ·
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.