← Product Code [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS) · K021433

# MODEL 415PH-64/42: PHASED ARRAY SHOULDER COIL (K021433)

_Medical Advances, Inc. · MOS · Jun 13, 2002 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K021433

## Device Facts

- **Applicant:** Medical Advances, Inc.
- **Product Code:** [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS.md)
- **Decision Date:** Jun 13, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

Magnetic resonance imaging (MRI) of the shoulder.

## Device Story

Phased array shoulder coil designed for use with Philips 1.5T/1.0T MRI systems; utilizes Synergy MultiConnect interface. Device functions as radiofrequency (RF) receiver coil to capture magnetic resonance signals from shoulder anatomy. Input consists of RF signals generated by patient tissue within static magnetic field; coil transforms these into electrical signals for processing by MRI system console. Used in clinical radiology settings by trained MRI technologists/radiologists. Output provides 2D/3D images (proton density, T1/T2 weighted, time-of-flight, phase contrast) for diagnostic interpretation. Enhances signal-to-noise ratio and image quality for joint structure visualization, aiding clinical decision-making regarding shoulder pathology.

## Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing and performance parameter comparisons (SNR, image uniformity, geometric distortion, slice thickness) demonstrating no change in safety or imaging performance relative to predicate devices.

## Technological Characteristics

Phased array RF receiver coil; compatible with Philips 1.5T/1.0T MRI systems. Materials include plastic housing (UL 94 flammability rated). Complies with UL 2601-1 and IEC 601-1 safety standards. Connectivity via Synergy MultiConnect interface. Passive device; no internal software algorithms or active energy sources.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

- JUN 1 3 2002
# 510(k) Summary of Safety and Effectiveness

្នំ
ប្រទេស

K02/433

,

| Device Name                       | Model 415PH-64/42: Phased Array Shoulder Coil<br>With Model 855PH: Synergy MultiConnect                          |                                                                          |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Applicability                     | Compatible with a Philips 1.5T/1.0T MRI System                                                                   |                                                                          |
| Reason for 510(k)                 | Control mechanism change to an existing product                                                                  |                                                                          |
| Classification Name               | Magnetic Resonance Diagnostic Device                                                                             |                                                                          |
| Device Classification Panel       | Radiology                                                                                                        |                                                                          |
| Device Classification Number      | 892.1000                                                                                                         |                                                                          |
| Product Code                      | 90MOS                                                                                                            |                                                                          |
| Common Name                       | Magnetic Resonance Imaging Coil                                                                                  |                                                                          |
| Proprietary Name                  | Model 415PH-64/42: Phased Array Shoulder Coil                                                                    |                                                                          |
| Establishment Registration Number | 2183683                                                                                                          |                                                                          |
| Address of MFG Facility           | IGC-Medical Advances Inc.<br>10437 Innovation Drive<br>Milwaukee, WI 53226                                       |                                                                          |
| Point of Contact                  | Jerry Frohlich<br>Manager, Regulatory Affairs<br>414.258.3808 Ext. 206                                           |                                                                          |
| Classification                    | Class II                                                                                                         |                                                                          |
| Diagnostic Uses                   | 2D, 3D imaging, proton density, T1 and T2<br>weighted imaging. 2D, 3D time of flight, phase<br>contrast imaging. |                                                                          |
| Anatomic Regions                  | Shoulder and other related joint structures.                                                                     |                                                                          |
| Performance Standards             | None Established under Section 514                                                                               |                                                                          |
| Voluntary Safety Standards        | UL 2601-1                                                                                                        | Medical Electrical Equipment, Part<br>1: General Requirements for Safety |
|                                   | UL 94                                                                                                            | Tests for Flammability of Plastic<br>Materials                           |
|                                   | IEC 601-1                                                                                                        | General Safety Requirements for<br>Medical Electrical Equipment          |

{1}------------------------------------------------

#### Overview

The Radiology Devices Panel considered potential concerns regarding the safe and effective operation of Magnetic Resonance Diagnostic Devices when they recommended reclassification to Class II on July 27, 1987. After reclassification, the FDA's Center for Devices and Radiological Health (CDRH) released a draft guidance document for the content and review of Magnetic Resonance Diagnostic Device premarket notifications that offered clarification of these concerns. Due to considerable technological advances in MRDDs, CDRH issued an updated guidance document on November 14, 1998. The following is a summary of the information contained within this premarket notification that addresses these concerns:

The GE 1.5T Signa MRI system operated with the Medical Advances UltraImage "Temporal Lobe Array Coil is substantially equivalent to the same system operated with the legally marketed predicate devices listed in section 4.0, within the Class II definition of Magnetic Resonance Diagnostic Device with respect to the safety parameter action levels:

#### Safety Parameters

| Maximum Static Magnetic Field:          | No change |
|-----------------------------------------|-----------|
| Rate of Magnetic Field Strength Change: | No change |
| RF Power Deposition:                    | No change |
| Acoustic Noise Levels:                  | No change |
| Biocompatibility:                       | No change |

#### Imaging Performance Purameters

| Specification Volume:  | No change |
|------------------------|-----------|
| Signal-to-Noise Ratio: | No change |
| Image Uniformity:      | No change |

{2}------------------------------------------------

| Geometric Distortion:             | No change |
|-----------------------------------|-----------|
| Slice Thickness and Gap:          | No change |
| High Contrast Spatial Resolution: | No change |

### General Safety and Effectiveness Concerns

The device contains instructions for use. It includes indications for use, precautions, cautions, contraindications, warnings and quality assurance testing. This information assures sufe and effective use of the device.

## Substantial Equivalence Summary

The Philips 1.5T Intera MRI system operated with the Medical Advances Phased Array Shoulder Coil addressed in this PMN, has the same intended use and technological characteristics as the same system operated with the identified legally marketed predicate devices. The use of these coils does not affect the Philips Intera system safety parameter specifications.

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. R. Jerry Frohlich Manager, Ouality Assurance and Regulatory Affairs IGC-Medical Advances, Inc. 10437 Innovation Dr. MILWAUKEE WI 53226

Re: K021433

JUN 1 3 2002

Trade/Device Name: Model 415PH Series: Phased Array Coil with Synergy MultiConnect Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: May 3, 2002 Received: May 6, 2002

Dear Mr. Frohlich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page _ 1 of __________________________________________________________________________________________________________________________________________________________________

.

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Model 415PH Series: Medical Advances Phased Array Shoulder Coil with Synergy MultiConnect

Indications for Use:

Magnetic resonance imaging (MRI) of the shoulder.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96

David G. Tegenon

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 4021433 510(k) Number _

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K021433](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K021433)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
