MAGNETOM PROJECT 024 SYSTEM (NUMARIS 3.5 SOFTWARE)
K970852 · Siemens Medical Solutions USA, Inc. · LNH · Jun 5, 1997 · Radiology
Device Facts
Record ID
K970852
Device Name
MAGNETOM PROJECT 024 SYSTEM (NUMARIS 3.5 SOFTWARE)
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Jun 5, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Intended Use
The SIEMENS MAGNETOM Project 024 is a 1Tesla whole body magnetic resonance imaging (MRI) system intended for general diagnostic use. This MRI system will present images which reflect the spatial distribution and/or magnetic resonance spectra which are a function of the frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging.
Device Story
The MAGNETOM Project 024 is a 1.0T whole body MRI system. It acquires magnetic resonance signals from hydrogen-1 nuclei to generate diagnostic images and spectra reflecting spatial distribution and physical parameters. The system features a compact design, redesigned patient table, and RF coil system. It operates under Numaris 3.5 software. Used in clinical settings by trained medical professionals, the system provides images for physician interpretation to aid in clinical decision-making and patient diagnosis. It serves as both a new system and an upgrade to the MAGNETOM Impact system.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and performance parameters consistent with previously cleared predicate devices.
Technological Characteristics
1.0T high field strength whole body MRI system. Features compact design, redesigned patient table, and RF coil system. Operates with Numaris 3.5 software. Class II device.
Indications for Use
Indicated for general diagnostic whole body magnetic resonance imaging (MRI) in patients requiring hydrogen-1 (proton) imaging and magnetic resonance spectra analysis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
SIEMENS MAGNETOM Vision
SIEMENS MAGNETOM Impact
Related Devices
K971684 — MAGNETOM PROJECT 047 SYSTEM (NUMARIS 3.5 SOFTWARE) · Siemens Medical Solutions USA, Inc. · Aug 5, 1997
K110942 — PICA WHOLE-BODY MRI SYSTEM · Time Medical Limited · Jul 26, 2011
K130262 — MAGNETOM ESSENZA WITH SYNGO MR D14 · Siemens Medical Solutions USA, Inc. · Mar 1, 2013
K211980 — SIGNA Prime · Ge Healthcare (Tianjin) Company Limited · Jan 16, 2022
K093984 — PICA WHOLE BODY MRI SYSTEM MODEL TMS-MRI-3500WB-01 · Time Medical Limited · Mar 24, 2010
Submission Summary (Full Text)
{0}
SIEMENS
K970852
JUN - 5 1997
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## I. General Information.
### Establishment:
- Address: Siemens Medical Systems, Inc.
186 Wood Avenue South
Iselin, N.J. 08830
Registration Number: 2240869
Contact Person: Ms. Kathleen Rutherford
Manager, Regulatory Submissions
(908) 321-4779
(908) 321-4841
Date of Summary Preparation: 3/06/97
### Device Name:
- Trade Name: MAGNETOM Project 024
- Classification Name: Magnetic Resonance Diagnostic Device, CFR § 892.1000
- Classification: Class II
- Performance Standards: None established under Section 514 of the Food, Drug, and Cosmetic Act.
## II. Safety and Effectiveness Information Supporting Substantial Equivalence.
### Device Description:
The SIEMENS MAGNETOM Project 024 is a 1.0T whole body magnetic resonance imaging (MRI) system. The MAGNETOM Project 024 is the first member of a family of magnetic resonance imaging (MRI) systems. The Project 024 is being introduced as a new system and as an upgrade to the commercially available MAGNETOM Impact system.
MAGNETOM Project 024
/ Numans 3.5 Software
510(k)
3/06/97
Appendix 9: Page 1
{1}
# SIEMENS
Intended Use:
Whole body magnetic resonance imaging.
Technological Characteristics:
The MAGNETOM Project 024 system is a high field strength system. The new MAGNETOM system has a new compact design and a redesigned patient table and rf coil system.
General Safety and Effectiveness Concerns:
The safety and performance characteristics for this system are within the scope of parameters cleared for market for the MAGNETOM Vision and MAGNETOM Impact systems. No additional safety concerns have been raised.
The device labeling for the Project 024 system will contain instructions for use, indications for use, precautions, cautions, contraindications, and warnings to the user.
Substantial Equivalence:
The MAGNETOM Project 024 system is substantially equivalent to the SIEMENS MAGNETOM Vision and MAGNETOM Impact magnetic resonance imaging systems.
Kathleen M. Rutherford
Manager, Regulatory Submissions
Imaging Systems Group
Siemens Medical Systems, Inc.
3/6/97
Date
MAGNETOM Project 024
Numaris 3.5 Software
510(k)
3/06/97
Appendix 9: Page 2
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN - 5 1997
Kathleen Rutherford
Manager, Regulatory Submissions
Siemens Medical Systems, Inc.
186 Wood Avenue South
Iselin, NJ 08830
Re: K970852
MAGNETOM Project 024 (Numeris 3.5 Software)
Dated: March 6, 1997
Received: March 7, 1997
Regulatory class: II
21 CFR 892.1000/Procode: 90 LNH
Dear Ms. Rutherford:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591 for Radiology devices, or 594-4613 for Ear, Nose and Throat devices. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{3}
06/04/97 13:55 FAX 908 321 4841 SMS MRL AFFAIRS --> CDRH DRAHED 003
# SIEMENS
510(k) Number (if known) _______________
Device Name: **MAGNETOM Project 024**
## Indications for Use:
The SIEMENS MAGNETOM Project 024 is a 1Tesla whole body magnetic resonance imaging (MRI) system intended for general diagnostic use. This MRI system will present images which reflect the spatial distribution and/or magnetic resonance spectra which are a function of the frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging.
(please do not write below this line- continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation
Prescription Use ☑ OR Over-The-Counter Use ☐
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K970852
Siemens Medical Systems, Inc. - Page 1 of 1 -
Tel: (908) 321-4500
(908) 494-2250
Imaging Systems Group 186 Wood Avenue South Iselin, NJ 08830
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.