Embrace Neonatal MRI System

K254277 · Aspect Imaging, Ltd. · LNH · Mar 13, 2026 · Radiology

Device Facts

Record IDK254277
Device NameEmbrace Neonatal MRI System
ApplicantAspect Imaging, Ltd.
Product CodeLNH · Radiology
Decision DateMar 13, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesPediatric

Indications for Use

The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Device Story

Embrace Neonatal MRI System is a 1 Tesla permanent magnet MRI; captures MR images of neonatal head. Device utilizes static magnetic field, RF pulses, and gradient magnetic fields to induce precession of magnetization; signals sampled and processed into image data. Used in hospital settings; operated by trained clinicians. System includes integrated patient bed with temperature control (20.5°C–36.5°C). Clinicians view images on 24-inch LED display via workstation to assist in diagnostic decision-making. Benefits include non-invasive visualization of neonatal internal head structures. Revision E introduces software enhancements for usability and cybersecurity, plus updated imaging sequences (e.g., Propeller FSE, TOF).

Clinical Evidence

Bench testing only. Performance validated via NEMA MS-1, MS-2, MS-3, MS-5, MS-12, MS-14, MS-4, and MS-8 standards. Software verification and validation, biocompatibility testing (ISO 10993-1), and simulated use validation performed. No clinical data presented.

Technological Characteristics

1 Tesla permanent magnet MRI; 150 mT/m gradient strength; 500 T/m/sec slew rate. Includes integrated patient bed with air-based temperature control. Connectivity via Windows PC workstation. Software life-cycle per IEC 62304. Biocompatibility per ISO 10993-1. Standards: IEC 60601-1, 60601-1-2, 60601-2-33, 60601-2-20, 60601-1-6.

Indications for Use

Indicated for neonates with head circumference up to 38 cm and weight between 1 kg and 4.5 kg requiring MR imaging of the head.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 13, 2026 Aspect Imaging, Ltd. % Alex Cadotte Vice President, Digital Health, AI, and Radiology Mcra 803 7th St. NW Washington, District of Columbia 20001 Re: K254277 Trade/Device Name: Embrace Neonatal MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, MOS Dated: December 9, 2025 Received: December 30, 2025 Dear Alex Cadotte: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K254277 - Alex Cadotte Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K254277 - Alex Cadotte Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K254277 | | | Device Name Embrace Neonatal MRI System | | | Indications for Use (Describe) The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K254277 # 510(k) Summary Device Trade Name: Embrace Neonatal MRI System Manufacturer: Aspect Imaging LTD 27 Shaked St. Industrial Area Hevel Modi'in Shoham 60850, Israel Contact Person: Yaron Eshel VP, Quality Assurance and Regulatory Affairs Yaron.Eshel@aspectimaging.com Prepared by: MCRA, LLC 803 7th Street, NW, 3rd Floor Washington, DC 20001 Office: 202.552.5800 Date Prepared: March 10, 2026 Classifications: 21 CFR 892.1000 Class: II Product Codes: LNH, MOS Primary Predicate: Embrace Neonatal MRI System (Aspect Imaging, K170978) Page 1 of 8 {5} Page 2 of 8 # Indications For Use: The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. # Device Description The Embrace Neonatal MRI System is a 1 Tesla Permanent magnet MRI system producing MR images of the neonatal head. During MRI scan, body parts to be imaged are held within a uniform static magnetic field, and are subject to sequences of RF pulses and gradient magnetic fields. The signal from the precession of the magnetization created by these fields is sampled and processed to produce image data. # Predicate Device: The purpose of this Traditional 510(k) is to demonstrate that the Embrace Neonatal MRI System (Revision E) is substantially equivalent to predicate device for the purpose of introducing the system to interstate commerce. It has been determined that the Embrace Neonatal MRI System (Revision E) are substantially equivalent to the following predicate: - Primary Predicate: Embrace Neonatal MRI System (Aspect Imaging, K170978) | Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences | | --- | --- | --- | --- | | Product Code | LNH, MOS | LNH, MOS | Same | | Classification / Regulation | Magnetic Resonance Diagnostic Device (21 CFR.892.1000) | Magnetic Resonance Diagnostic Device (21 CFR 892.1000) | Same | {6} | Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences | | --- | --- | --- | --- | | Intended Use/Indication for Use | The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that display the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | Same | | Patient Population | Patients requiring MR images of the Neonatal Head | Patients requiring MR images of the Neonatal Head | Same | | Anatomical Sites | Neonatal Head | Neonatal Head | Same | | Environment of Use | Hospital setting | Hospital setting | Same | | Energy Used and/or delivered | Magnetic Resonance | Magnetic Resonance | Same | | Human Factors | The Embrace Neonatal MRI System is designed similar to other commercially available MRI systems and therefore is familiar and easy for use for the user. Furthermore, the device contains a user-friendly software interface through which the user may easily access all device functions. | The Embrace Neonatal MRI System is designed similar to other commercially available MRI systems and therefore is familiar and easy for use for the user. Furthermore, the device contains a user-friendly software interface through which the user may easily access all device functions. | Same | | Hardware Specifications: | | | | | Magnet -Physical | | | Same | Page 3 of 8 {7} | Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences | | --- | --- | --- | --- | | Diminutions | 171×145×229 cm | 171×145×229 cm | | | -Bore Opening | 184×260mm | 184×260mm | | | -Weight | 5500 (5680 with patient bed) Kg | 5500 (5680 with patient bed) Kg | | | -Field Strength | 1 Tesla Permanent Magnet | 1 Tesla Permanent Magnet | | | Gradient | | | | | -Strength | 150 mT/m | 150 mT/m | Same | | -Rise Time | 0.300mSec | 0.300mSec | | | -Slew Rate | 500 T/m/Sec | 500 T/m/Sec | | | IT infrastructure | Windows PC on workstation table | Windows PC on workstation table | Revision E contains software enhancements for usability and strengthened cybersecurity | | Computer Display | 24” LED Display | 24” LED Display | Same | | RF Coils | 1 Head Coils | 1 Head Coils | | | -Coil Type | TX/RX | TX/RX | | | -Coil Geometry | Cylindrical | Cylindrical | Same | | -Inner dimensions (mm) | 143 Diameter | 143 Diameter | | | -Coil Design | Linear Volume | Linear Volume | | | Patient Bad | Integrated patient bed | Integrated patient bed | Improvements to the bed for usability and serviceability are included in revision E | | Target population | Neonates with head circumference of up to 38 cm and weight | Neonates with head circumference of up to 38 cm and weight | Same | Page 4 of 8 {8} | Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences | | --- | --- | --- | --- | | | between 1Kg and 4.5Kg | between 1Kg and 4.5Kg | | | -Dimensions | 60.6cm W x 120cm H x 140cm L | 60.6cm W x 120cm H x 140cm L | | | -Patient weight capacity | 4.5 Kg Max | 4.5 Kg Max | Same | | -Compartment length | 58 cm | 58 cm | | | -Set Temperature | 20.5°C - 36.5°C | 20.5°C - 36.5°C | | | -Warm Up time | 45 minutes | 45 minutes | | | -Temperature control | Air | Air | Same | | -Humidity control | No | No | | | -MRI compatibility | Aspect Embrace 1T | Aspect Embrace 1T | | | Performance Testing | | | | | Imaging Performance: (Recognition no. in brackets) | • NEMA MS 1 - Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging (12-188) • NEMA MS 12 - Quantification and Mapping of Geometric Distortion for Special Applications (12-306) • NEMA MS 2 - Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images (12-196) • NEMA MS 3 - Determination of Image Uniformity in Diagnostic Magnetic Resonance Images (12-187) • NEMA MS 5 - Determination of Slice | • NEMA MS 1 - Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging (12-188) • NEMA MS 12 - Quantification and Mapping of Geometric Distortion for Special Applications (12-306) • NEMA MS 14 - Characterization of Radiofrequency (RF) Coil Heating in Magnetic Resonance Imaging Systems (12-331) • NEMA MS 2 - Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images (12-196) • NEMA MS 3 - | MS-14 was added to conform with FDA applicable standards | Page 5 of 8 {9} | Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences | | --- | --- | --- | --- | | | Thickness in Diagnostic Magnetic Resonance Imaging (12-322) | Determination of Image Uniformity in Diagnostic Magnetic Resonance Images (12-187) • NEMA MS 5 - Determination of Slice Thickness in Diagnostic Magnetic Resonance Imaging (12-322) | | | Safety Performance: (Recognition no. in brackets) | | • NEMA MS 4 - Acoustic Noise Measurement Procedure for Diagnostic Magnetic Resonance Imaging Devices (12-232) • NEMA MS 8 - Characterization of the Specific Absorption Rate (SAR) for MRI Systems (12-315) | MS-4 and MS -8 were added to conform with FDA safety standards | | Electromagnetic Compatibility, Electrical, Mechanical, and Thermal Safety | | | | | Voluntary consensus standards (Recognition no. in brackets) | • IEC 60601-1 - Medical equipment/medical electrical equipment - Part 1: General requirements for basic safety and essential performance (19-37) • IEC 60601-1-2 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (19-36) • IEC 60601-2-33 - Medical electrical equipment - Part 2-33: Particular requirements for the basic | • IEC 60601-1 - Medical equipment/medical electrical equipment - Part 1: General requirements for basic safety and essential performance (19-37) • IEC 60601-1-2 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (19-36) • IEC 60601-2-33 - Medical electrical equipment - Part 2-33: Particular requirements for the | Voluntary consensus standards (Recognition no. in brackets) | Page 6 of 8 {10} | Specification | Predicate Device Embrace Neonatal MRI System (K170978) | Subject Device Embrace Neonatal MRI System | Discussion of Differences | | --- | --- | --- | --- | | | safety and essential performance of magnetic resonance equipment for medical diagnosis (12-347) • IEC 60601-2-20 - Medical Electrical Equipment Part 2-20: Particular requirements for the basic safety and essential performance of infant transport incubators (6-462) • IEC 60601-1-6 - Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability (5-132) • IEC 62304 - Software life-cycle processes (13-79) | basic safety and essential performance of magnetic resonance equipment for medical diagnosis (12-347) • IEC 60601-2-20 - Medical Electrical Equipment Part 2-20: Particular requirements for the basic safety and essential performance of infant transport incubators (6-462) • IEC 60601-1-6 - Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability (5-132) • IEC 62304 - Software life-cycle processes (13-79) | | | Sequence available to the User | • Gradient Echo • Gradient Echo 3D • Spin Echo • Fast Spin Echo • ADC Map Single Shot Splice Axial • ADC Map Spin Echo | • 2D Spin Echo • (SE) 2D Fast Spin Echo (FSE) • 3D Fast Spin Echo (FSE) • 3D Gradient Echo (GRE) • ADC map (Diffusion Weighted imaging with ADC mapping) • 2D Time of Flight (TOF) • 3D Time of Flight (TOF) 2D • Propeller FSE • 2D ADC Propeller | Modification of sequences for the user | Software: In addition, system and software verification and Validation testing were performed to demonstrate performance of the device as part of design controls activity. Biocompatibility: Patient-contacting materials were assessed per FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Patient contact surfaces were classified based on type of contact and duration. All surfaces are considered surface contacting for durations less than or equal to 24 Page 7 of 8 {11} hours. Where required, evidence of testing was provided for endpoints including cytotoxicity, sensitization, and irritation. All patient contact surfaces of the Embrace Neonatal MRI System (Revision E) meet biocompatibility requirements consistent with contact type and duration. The following design control activities were applied to the development of the system: - Risk Management - Requirements Management - Design Reviews - Unit level module verification - System Integration verification - Performance verification (see table above for specific testing performed) - Safety and EMC testing - Simulated use validation testing - Sample Phantom images ## Substantial Equivalence: The subject device was demonstrated to be substantially equivalent to the predicate cited above with respect to indications, design materials, function, manufacturing, and performance. The non-clinical tests performed by the company demonstrated that any differences in the Embrace Neonatal MRI System (Revision E) do not raise new questions of safety and effectiveness. ## Conclusion: The subject device and the predicate device have the same intended use, have similar technological characteristics, and are made of the same materials. The subject and predicate devices have similar software and EMC characteristics that were validated via new testing. The data included in this submission demonstrates substantial equivalence to the predicate device listed above. The Embrace Neonatal MRI System (Revision E) is as safe, as effective, and performs as well as, or better, than the predicate device. Page 8 of 8
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...