The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Device Story
MAGNETOM Avanto Fit, Skyra Fit, Sola Fit, and Viato.Mobile are MRI systems utilizing magnetic resonance to produce cross-sectional images and spectra of human anatomy. Systems operate in clinical environments, managed by trained physicians and technicians. New hardware includes myExam 3D Camera and BM Head/Neck 20 coil; software updates include AI-based features (Deep Resolve, AutoMate Cardiac) and advanced pulse sequences. Inputs include RF signals and gradient-encoded magnetic resonance data; processing involves image reconstruction, denoising, and automated workflow assistance. Outputs are diagnostic images/spectra displayed for physician interpretation to assist in clinical diagnosis. AI algorithms (Deep Resolve) enhance image quality (sharpness, SNR) via deep learning networks. Systems support interventional procedures using MR-safe accessories. Benefits include improved image quality, workflow efficiency, and diagnostic accuracy.
Clinical Evidence
No clinical trials conducted. Evidence consists of bench testing, software verification/validation, and sample clinical images. AI features (Deep Resolve) validated using large datasets (e.g., >25,000 slices for TSE, >1,000,000 for EPI Diffusion) with performance metrics including PSNR, SSIM, and visual sharpness ratings. Literature cited for specific imaging features (e.g., 3D Whole Heart, AutoMate Cardiac).
Technological Characteristics
MRI systems (1.5T/3T). Components: Magnet, RF system, gradient system, patient table. New hardware: myExam 3D Camera. Software: syngo MR XA70A. Standards: IEC 60601-1, IEC 60601-2-33, ISO 14971, IEC 62304, NEMA MS 6-2008. Connectivity: DICOM compliant. AI/ML: Deep learning-based image reconstruction/enhancement.
Indications for Use
Indicated for patients requiring diagnostic MR imaging of the head, body, or extremities to visualize internal structure and function. Used for diagnostic imaging and interventional procedures with MR-compatible devices. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 16, 2025
Siemens Healthcare GmbH
Alina Goodman
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, Pennsylvania 19355
Re: K250443
Trade/Device Name: MAGNETOM Avanto Fit; MAGNETOM Skyra Fit; MAGNETOM Sola Fit; MAGNETOM Viato.Mobile
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH, LNI, MOS
Dated: May 16, 2025
Received: May 16, 2025
Dear Alina Goodman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250443 - Alina Goodman
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250443 - Alina Goodman
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K250433 | ? |
| Please provide the device trade name(s). | | ? |
| MAGNETOM Avanto Fit;
MAGNETOM Skyra Fit;
MAGNETOM Sola Fit;
MAGNETOM Viato.Mobile | | |
| Please provide your Indications for Use below. | | ? |
| The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that
produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or
spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other
physical parameters derived from the images and/or spectra may also be produced. Depending on the
region of interest, contrast agents may be used. These images and/or spectra and the physical parameters
derived from the images and/or spectra when interpreted by a trained physician yield information that may
assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed
with MR compatible devices such as in-room displays and MR Safe biopsy needles. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K250443
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92.
## 1. General Information
**Establishment:** Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355, USA
Registration Number: 2240869
**Date Prepared:** February 3rd, 2025
**Manufacturer:** Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Registration Number: 3002808157
Siemens Shenzhen Magnetic Resonance LTD.
Siemens MRI Center
Hi-Tech Industrial Park (middle)
Gaoxin C. Ave., 2nd
Shenzhen 518057
P.R. CHINA
Registration Number: 3004754211
## 2. Contact Information
Alina Goodman
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355, USA
Phone: +1(317)371-8593
E-mail: alina.goodman@siemens-healthineers.com
## 3. Device Name and Classification
**Device/ Trade name:** MAGNETOM Avanto Fit
**Classification Name:** Magnetic Resonance Diagnostic Device (MRDD)
**Classification Panel:** Radiology
**CFR Code:** 21 CFR § 892.1000
**Classification:** II
**Product Code:** Primary: LNH
Secondary: LNI, MOS
{5}
Device/ Trade name: MAGNETOM Skyra Fit
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Sola Fit
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name: MAGNETOM Viato.Mobile
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Device/ Trade name:
4. Legally Marketed Predicate and Reference Device
4.1. Predicate Device
Trade name: MAGNETOM Avanto Fit
510(k) Number: K220151
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Skyra Fit
510(k) Number: K220589
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Sola Fit¹
¹ The predicate system MAGNETOM Sola Fit (1.5T) with syngo MR XA51A (K221733) is used in this submission as well as reference system in the Fact Sheets for some systems. However, it will not be listed again as reference system since it is already incorporated into the predicate system overview.
{6}
510(k) Number: K221733
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Viato.Mobile
510(k) Number: K240608
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: syngo.via VB40A
510(k) Number: K191040
Classification Name:
Classification Panel:
CFR Code:
Classification:
Product Code: Primary: LLZ
## 4.2. Reference Device
Trade name: MAGNETOM Cima.X
510(k) Number: K231587
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Sola
510(k) Number: K232535
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Vida
510(k) Number: K213693
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
{7}
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
Trade name: MAGNETOM Aera
510(k) Number: K153343
Classification Name: Magnetic Resonance Diagnostic Device (MRDD)
Classification Panel: Radiology
CFR Code: 21 CFR § 892.1000
Classification: II
Product Code: Primary: LNH
Secondary: LNI, MOS
## 4. Intended Use / Indications for Use
The indications for use for the subject devices are the same as the predicate devices:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
## 5. Device Description
The subject device, MAGNETOM Avanto Fit with software syngo MR XA70A, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Avanto Fit with syngo MR XA50A (K220151).
A high-level summary of the new and modified hardware and software is provided below:
{8}
For MAGNETOM Avanto Fit with syngo MR XA70:
Hardware
New Hardware:
→ myExam 3D Camera
→ BM Head/Neck 20
Modified Hardware:
→ -Sanaflex (cushions for patient positioning)
Software
New Features and Applications:
→ myExam Autopilot Brain
→ myExam Autopilot Knee
→ 3D Whole Heart
→ HASTE_interactive
→ GRE_PC
→ Open Recon
→ Deep Resolve Gain
→ Fleet Reference Scan
→ Physio logging
→ complex averaging
→ AutoMate Cardiac
→ Ghost Reduction
→ BLADE diffusion
→ Beat Sensor
→ Deep Resolve Sharp
→ Deep Resolve Boost and Deep Resolve Boost (TSE)
→ Deep Resolve Boost HASTE
→ Deep Resolve Boost EPI Diffusion
Modified Features and Applications:
→ SPACE improvement (high band)
→ SPACE improvement (incr grad)
→ Brain Assist
→ Eco power mode
→ myExam Angio Advanced Assist (Test Bolus)
The subject device, MAGNETOM Skyra Fit with software syngo MR XA70A, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Skyra Fit with syngo MR XA50A (K220589).
A high-level summary of the new and modified hardware and software is provided below:
{9}
For MAGNETOM Skyra Fit with syngo MR XA70:
Hardware
New Hardware:
→ myExam 3D Camera
Modified Hardware:
→ Sanaflex (cushions for patient positioning)
Software
New Features and Applications:
→ Beat Sensor
→ HASTE_interactive
→ GRE_PC
→ 3D Whole Heart
→ Deep Resolve Gain
→ Open Recon
→ Ghost Reduction
→ Fleet Reference Scan
→ BLADE diffusion
→ HASTE diffusion
→ Physio logging
→ complex averaging
→ Deep Resolve Swift Brain
→ Deep Resolve Sharp
→ Deep Resolve Boost and Deep Resolve Boost (TSE)
→ Deep Resolve Boost HASTE
→ Deep Resolve Boost EPI Diffusion
→ AutoMate Cardiac
→ SVS_EDIT
Modified Features and Applications:
→ SPACE improvement (high band)
→ SPACE improvement (incr grad)
→ Brain Assist
→ Eco power mode
→ myExam Angio Advanced Assist (Test Bolus)
The subject device, MAGNETOM Sola Fit with software syngo MR XA70A, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola Fit with syngo MR XA51A (K221733).
A high-level summary of the new and modified hardware and software is provided below:
For MAGNETOM Sola Fit with syngo MR XA70:
Hardware
New Hardware:
→ myExam 3D Camera
Modified Hardware:
→ Sanaflex (cushions for patient positioning)
Software
New Features and Applications:
{10}
→ GRE_PC
→ 3D Whole Heart
→ Ghost Reduction
→ Fleet Reference Scan
→ BLADE diffusion
→ Physio logging
→ Open Recon
→ Complex averaging
→ Deep Resolve Sharp
→ Deep Resolve Boost and Deep Resolve Boost (TSE)
→ Deep Resolve Boost HASTE
→ Deep Resolve Boost EPI Diffusion
→ AutoMate Cardiac
→ Implant suite
**Modified Features and Applications:**
→ SPACE improvement (high band)
→ SPACE improvement (incr grad)
→ Brain Assist
→ Eco power mode
The subject device, MAGNETOM Viato. Mobile with software syngo MR XA70A, consists of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Viato. Mobile with syngo MR XA51A (K240608).
A high-level summary of the new and modified hardware and software is provided below:
**For MAGNETOM Viato. Mobile with syngo MR XA70:**
**Hardware**
**New Hardware:**
→ n.a.
**Modified Hardware:**
→ Sanaflex (cushions for patient positioning)
**Software**
**New Features and Applications:**
→ GRE_PC
→ 3D Whole Heart
→ Ghost Reduction
→ Fleet Reference Scan
→ BLADE diffusion
→ Physio logging
→ Open Recon
→ Complex averaging
→ Deep Resolve Sharp
→ Deep Resolve Boost and Deep Resolve Boost (TSE)
→ Deep Resolve Boost HASTE
→ Deep Resolve Boost EPI Diffusion
→ AutoMate Cardiac
→ Implant suite
**Modified Features and Applications:**
→ SPACE improvement (high band)
→ SPACE improvement (incr grad)
{11}
Brain Assist
Eco power mode
Furthermore, the following minor updates and changes were conducted for the subject devices:
- Low SAR Protocol minor update (for all subject devices but MAGNETOM Skyra Fit): the goal of the SAR adaptive protocols was to be able to perform knee, spine, heart and brain examinations with $50\%$ of the max allowed SAR values in normal mode for head and whole-body SAR. The SAR reduction was achieved by parameter adaptations like Flip angle, TR, RF Pulse Type, Turbo Factor, concatenations. For cardiac clinically accepted alternative imaging contrasts are used (submitted with K232494).
- Implementation of image sorting prepare for PACS (submitted with K231560).
- Implementation of improved DICOM color support (submitted with K232494).
- Needle intervention AddIn was added all subject device (submitted with K232494).
- Inline Image Filter switchable for users: in the subject device, users have the ability to switch the "Inline image filter" (implicite Filter) on or off. This filter is an image-based filter that can be applied to specific pulse sequence types. The function of the filter remains unchanged from the previous device MAGNETOM Sola with syngo MR XA61A (K232535).
- SVS_EDIT is newly added for MAGNETOM Skyra Fit, but without any changes (submitted with K203443)
- Brain Assist received an improvement and is identical to that of snygo MR XA61A (K232535)
- Open Recon is introduced for all systems. The function of Open Recon remains unchanged from the previous submissions (submitted with K221733).
- Lock TR and FA in Bold received a minor UI update
- Implant Suite is newly introduced for MAGNETOM Sola Fit and MAGNETOM Viato.Mobile, but without any changes (submitted with K232535)
- myExam Autopilot Brain and myExam Autopilot Knee are newly introduced for the subject device MAGNETOM AVANTO Fit and are unchanged from previous submissions (submitted with K221733).
- myExam Angio Advanced Assist (Test Bolus) received a bug fixing and minimal UI improvements
# 6. Substantial Equivalence
MAGNETOM Avanto Fit, MAGNETOM Skyra Fit, MAGNETOM Sola Fit, and MAGNETOM Viato. Mobile, all with software syngo MR XA70A are substantially equivalent to the following devices:
| Predicate Device | FDA Clearance Number and Date | Product Code | Manufacturer |
| --- | --- | --- | --- |
| MAGNETOM Avanto Fit (1.5T) with syngo MR XA50A | K220151, cleared on April 01, 2022 | LNH, LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Skyra Fit (3T) with syngo MR XA50A | K220589, cleared on May 13, 2022 | LNH, LNI, MOS | Siemens Healthcare GmbH |
{12}
# 7. Technological Characteristics
The subject devices, MAGNETOM Avanto Fit, MAGNETOM Skyra Fit, MAGNETOM Sola Fit, and MAGNETOM Viato.Mobile, all with the new software syngo MR XA70A, are substantially equivalent to the predicate devices with regard to the operational environment, programming language, operating system and performance.
The subject devices conform to the standard for medical device software (IEC 62304) and other relevant IEC and NEMA standards.
There are some differences in technological characteristics between the subject devices and predicate devices, including new and modified hardware/software. Here is summary of differences:
Summary hardware comparison table for the subject and predicate device
| Hardware | Subject Devices MAGNETOM Avanto Fit, MAGNETOM Skyra Fit, MAGNETOM Sola Fit, and MAGNETOM Viato.Mobile, all with software syngo MR XA70A | Predicate Devices MAGNETOM Avanto Fit with syngo MR XA50A (K220151), MAGNETOM Skyra Fit with syngo MR XA50A (K220589), MAGNETOM Sola Fit with syngo MR XA51A (K221733), and MAGNETOM Viato.Mobile with syngo MR XA51A (K240608) | Reference Devices MAGNETOM Cima.X (K231587) with syngo MR XA61A; MAGNETOM Sola (K232535) with syngo MR XA61A; MAGNETOM Vida (K213693) with syngo MR XA50A, (K203443) MAGNETOM Aera with syngo MR VE11 |
| --- | --- | --- | --- |
| Magnet System | Yes | Yes | Yes |
| RF System | Yes | Yes | Yes |
| Transmission technique | Yes | Yes | Yes |
| Gradient System | Yes | Yes | Yes |
| Patient Table | Yes | Yes | Yes |
{13}
| Multi-Nuclear Option - Supported Nuclei | No | No | No (Yes for MAGNETOM Cima.X) |
| --- | --- | --- | --- |
| Computer | Yes | Yes | Yes |
| Coils | Yes
New for MAGNETOM Avanto Fit, based on predicate: BM Head/Neck 20 | Yes | Yes |
| Other HW components | Yes
3D camera, cushions modified compared to the respective subject device (see Device Description) | Yes | Yes |
Summary software comparison table for the subject and predicate devices
| Software | Subject Devices
MAGNETOM Avanto Fit, MAGNETOM Skyra Fit, MAGNETOM Sola Fit, and MAGNETOM Viato.Mobile, all with software syngo MR XA70A | Predicate Devices
MAGNETOM Avanto Fit with syngo MR XA50A (K220151), MAGNETOM Skyra Fit with syngo MR XA50A (K220589), MAGNETOM Sola Fit with syngo MR XA51A (K221733), and MAGNETOM Viato.Mobile with syngo MR XA51A (K240608) |
| --- | --- | --- |
| Sequences | | |
| SE-based pulse sequence types | New feature as listed in the Device Description above | Yes |
| GRE-based/Steady-State pulse sequence types | New or modified pulse sequences as listed in the Device Description above | Yes |
| EPI-based pulse sequence types | New features as listed in the Device Description above | Yes |
| Spectroscopy pulse sequence types | Yes | Yes |
| Feature and Applications | | |
| Other features and applications such as:
-Application Suites
-myExam Assists
-Other Imaging Applications | Modified features and applications as listed in the Device Description above | Yes |
| User interface and user interaction | Yes | Yes |
| Viewing and post-processing | New or modified viewing and post-processing features as listed in the Device Description above | Yes |
| Workflow and software utilization | Yes | Yes |
| Patient Management | Yes | Yes |
| Scan Modes and Pulse Sequences | Yes | Yes |
| Scanning | Modified and new features and applications as listed in the Cover letter, Device Description and Substantial Equivalence Comparison Tables | Yes |
| Reconstruction | New feature | Yes |
{14}
The differences have been tested and the conclusion from the non-clinical data suggests that the features bear an equivalent safety and performance profile to that of the predicate device.
# 8. Nonclinical Tests
The following performance testing was conducted on the subject devices:
| Performance Test | Tested Hardware or Software | Source/Rationale for test |
| --- | --- | --- |
| Software verification and validation | New or modified software features | Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices |
| Sample clinical images | New or modified software features | Guidance for submission of Premarket Notifications for Magnetic Resonance |
| Image quality assessment by sample clinical images | - new / modified pulse sequence types. - comparison images between the new / modified features and the predicate device features | Diagnostic Devices |
| myExam 3D Camera Validation and Verification Report | myExam 3D Camera | Hardware |
The following performance testing for local coils was conducted on the predicate and the reference devices and can be reused for the subject device:
| Performance Test | Tested Hardware or Software | Source/Rationale for test |
| --- | --- | --- |
| Performance bench test | - SNR and image uniformity measurements for coils - Heating measurements for coils | Guidance for Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices |
The measurements for the BM Head/Neck 20, which is added with this submission for MAGNETOM Avanto Fit were performed on an older version of the reference device and are also valid for the subject device according to the expert statements.
Overall, the results from each set of tests demonstrate that the devices perform as intended and are thus substantially equivalent to the predicate device to which it has been compared.
# AI Features/Applications training and validation:
The information below shows an executive summary of training and validation dataset of the AI features:
{15}
| | Deep Resolve Boost: | Deep Resolve Sharp: |
| --- | --- | --- |
| Training and Validation data | • TSE: more than 25,000 slices
• HASTE: pre-trained on the TSE dataset and refined with more than 10,000 HASTE slices
• EPI Diffusion: more than 1,000,000 slices
The data covered a broad range of body parts, contrasts, fat suppression techniques, orientations, and field strength. | on more than 10,000 high resolution 2D images.
The data covered a broad range of body parts, contrasts, fat suppression techniques, orientations, and field strength. |
| Test Statistics and Test Results Summary | The impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR) and structural similarity index (SSIM). Most importantly, the performance was evaluated by visual comparisons to evaluate e.g., aliasing artifacts, image sharpness and denoising levels. | The impact of the network has been characterized by several quality metrics such as peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and perceptual loss. In addition, the feature has been verified and validated by inhouse tests. These tests include visual rating and an evaluation of image sharpness by intensity profile comparisons of reconstructions with and without Deep Resolve Sharp. |
| Equipment | 1.5T and 3T MRI systems | |
| Clinical Subgroups | No clinical subgroups have been defined for the collected dataset. | |
| Demographic Distribution | Due to reasons of data privacy, we did not record gender, age and ethnicity during data collection. | |
| Reference Standard | The acquired datasets (as described above) represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further under-sampling of the data by discarding k-space lines, lowering of the SNR level by addition Restricted of noise and mirroring of k-space data. | The acquired datasets represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation. k-space data has been cropped such that only the center part of the data was used as input. With this method corresponding low-resolution data as input and high-resolution data as output / ground truth were created for training and validation. |
# 9. Clinical Tests / Publications
No clinical tests were conducted to support substantial equivalence for the subject devices; however, as stated above, sample clinical images were provided.
{16}
Furthermore, additional clinical publications were referenced to provide information on the use of the following features and functions:
| Feature | Publications |
| --- | --- |
| GRE_PC | [13_1] Christian Guenthner, Sweta Sethi, Marian Troelstra, Ristretto MRE: A generalized multi-shot GRE-MRE sequence NMR Biomed 2019 32:e4049 |
| 3D Whole Heart | [13_2] Henningsson M, Koken P, Stehning C et al.: Whole-heart coronary MR angiography with 2D self-navigated image reconstruction Magn Reson Med, 2012, 67:437-45 [13_3] Prieto C, Doneva M, Usman M et al.: Highly efficient respiratory motion compensated free-breathing coronary MRA using golden-step Cartesian acquisition, J Magn Reson Imaging, 2015 41:738-46 [13_4] Correia T, Ginami G, Cruz G, et al.: Optimized respiratory-resolved motion compensated 3D Cartesian coronary MR angiography, Magn Reson Med, 2018 80:2618-2629 [13_5] Chefd'hotel C, Hermosillo G, Faugeras O Flows of diffeomorphisms for multimodal image registration. Proc. IEEE International Symposium on Biomedical Imaging 2002 753-756 [13_6] Batchelor PG, Atkinson D, Irarrazaval P, et al.: Matrix Description of General Motion Correction Applied to Multishot Images, Magn Reson Med 2005 54:1273-1280 [13_7] Zeilinger MG, Kunze K-P, Munoz C et al.: Non-rigid motion-corrected free-breathing 3D myocardial Dixon LGE imaging in a clinical setting, Eur Radiol 2022 32:4340-4351 |
| SPACE Improvement: high bandwidth IR pulse | [13_8] Wang X, Greer JS, Dimitrov IE, Pezeshk P, Chhabra A, Madhuranthakam AJ: Frequency Offset Corrected Inversion Pulse for B0 and B1 Insensitive Fat Suppression at 3T: Application to MR Neurography of Brachial Plexus, J Magn Reson Imaging, 2018 Oct 48(4):1104-1111. doi: 10.1002/jmri.26021. Epub 2018 Sep 15. PMID: 30218576 |
| SPACE Improvement: increase gradient spoiling | [13_9] Mugler JP 3rd: Optimized three-dimensional fast-spin-echo MRI, J Magn Reson Imaging, 2014 Apr 39(4):745-67. doi: 10.1002/jmri.24542. Epub 2014 Jan 8. PMID: 24399498. |
| Deep Resolve Gain | [13_10] Kellman P. et al: Image Reconstruction in SNR Units: A General Method for SNR Measurement MRM, 2005 54:1439 [13_11] Blu T. et al: The SURE-LET approach to image denoising, IEEE Transactions on Image Processing, 2007 6(11):2778-86 |
| HASTE diffusion | [13_12] Ilica AT, Hidir Y, Bulakbasi N, Satar B, Güvenç I, Arslan HH, Imre N. HASTE diffusionweighted MRI for the reliable detection of cholesteatoma. Diagn Interv |
| | 2012 10:100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 |
{17}
| | Radiol. 2012 Mar-Apr 18(2):153 -8. doi: 10.4261/1 305-3825.DIR. 4246- 11.3. Epub 2011 Sep 29. PMID: 21960134 |
| --- | --- |
| AutoMate Cardiac | [13_13] J. a. Y. S. S. a. S. M. Wetzl, "Al-based Single-Click Cardiac MRI Exam: Initial Clinical Experience and Evaluation in 44 Patients," in ISMRM & ISMRT Annual Meeting & Exhibition, 2023.
[13_14] S. S. a. P. E. a. S. M. a. F. C. a. C. T. a. S. P. a. F. J. L. a. T. C. a. W. J. a. M. A. Yoon, "Automated Cardiac Resting Phase Detection Targeted on the Right Coronary Artery," Machine Learning for Biomedical Imaging, vol. 2, pp. 1-26, 2023.
[13_15] S. S. a. S. M. a. R. M. a. C. T. a. S. P. a. E. T. a. T. C. a. W. R. Yoon, "Validation of a deep learning based automated myocardial inversion time selection for late gadolinium enhancement imaging in a prospective study," in ISMRM & ISMRT Annual Meeting & Exhibition, 2021.
[13_16] R. a. K. T. a. S. Y. a. Y. Y. a. S. Y. S. a. W. J. a. S. M. a. K. T. Ogawa, "Neural network--based fully automated cardiac resting phase detection algorithm compared with manual detection in patients," Acta Radiologica Open, vol. 11, no. 10, p. 20584601221137772, 2022.
[13_17] S. a. W. J. a. S. M. a. B. M. a. Y. S. S. a. G. C. a. B. R. McDermott, "Albased Cardiac Scan Automation: A Prospective Comparison of Highly Automated Scan Workflows in 32 Patients," in Society for Cardiovascular Magnetic Resonance, 2024.
[13_18] S. a. A. M. a. J. A. a. S. R. B. a. Z. T. a. K. M. a. S. J. a. T. E. a. C. E. a. S. C. Bohnen, "Cardiovascular magnetic resonance imaging in the prospective, population-based, Hamburg City Health cohort study: objectives and design," Journal of Cardiovascular Magnetic Resonance, vol. 20, pp. 1-11, 2018.
[13_19] T. a. G. P. a. H. T. a. U. T. a. C. S. a. K. M. a. T. S. a. L. Y. a. M. M. C. a. S. F. Pezel, "Vasodilatation stress cardiovascular magnetic resonance imaging: Feasibility, workflow and safety in a large prospective registry of more than 35,000 patients," Archives of Cardiovascular Diseases, vol. 114, pp. 490-503, 2021.
[13_20] G. a. P. A. U. a. K. K. P. a. N. R. a. H. R. a. W. J. a. Y. S. S. a. S. M. a. N. B. L. a. P. C. a. o. Wood, "Automated detection of cardiac rest period for trigger delay calculation for image-based navigator coronary magnetic resonance angiography," Journal of Cardiovascular |
| Ghost reduction | [13_21] W Scott Hoge, Jonathan R Polimeni: Dual-polarity GRAPPA for simultaneous reconstruction and |
{18}
ghost correction of echo planar imaging data Magnetic Resonance in Medicine 2016 July 76(1):32-44
# 10. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle and continuously throughout the development of the product. Siemens Healthcare GmbH adheres to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
The subject devices conform to the following FDA recognized and international IEC, ISO and NEMA standards:
| Recognition Number | Product Area | Title of Standard | Reference Number and date | Standards Development Organization |
| --- | --- | --- | --- | --- |
| 19-4 | General | Medical electrical equipment - part 1: general requirements for basic safety and essential performance | ES60601-1:2005/(R)2012 and A1:2012 C1:2009/(R)2012 | AAMI / ANSI |
| 19-8 | General | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | 60601-1-2 Edition 4.0:2014-02 | IEC |
| 12-295 | Radiology | Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis | 60601-2-33 Ed. 3.2 b:2015 | IEC |
| 5-125 | General | Medical devices - Application of risk management to medical devices | 14971 Third Edition 2019-12 | ISO |
| 5-114 | General I (QS/RM) | Medical devices - Part 1: Application of usability engineering to medical devices | 62366-1:2015 | ANSI AAMI IEC |
| 13-79 | Software/Informatics | Medical device software - Software life cycle processes | 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION | IEC |
| 12-195 | Radiology | NEMA MS 6-2008 (R2014) Determination of Signal-to-Noise Ratio and Image Uniformity for Single-Channel Non-Volume Coils in Diagnostic MR Imaging | MS 6-2008 (R2014) | NEMA |
| 12-349 | Radiology | Digital Imaging and Communications in Medicine (DICOM) | PS 3.1 - 3.20 (2022d) | NEMA |
{19}
| 2-258 | Biocompatibility | Biological evaluation of medical devices - part 1: evaluation and testing within a risk management process. (Biocompatibility) | 10993-1 Fifth edition 2018-08 | AAMI ANSI ISO |
| --- | --- | --- | --- | --- |
# 11. Conclusion as to Substantial Equivalence
MAGNETOM Avanto Fit, MAGNETOM Skyra Fit, MAGNETOM Sola Fit, and MAGNETOM Viato. Mobile all with software syngo MR XA70A have the same intended use and same basic technological characteristics than the respective predicate device system, MAGNETOM Avanto Fit with syngo MR XA50A, MAGNETOM Skyra Fit with syngo MR XA50A, MAGNETOM Sola Fit with syngo MR XA51, and MAGNETOM Viato. Mobile with syngo MR XA51A, with respect to the magnetic resonance features and functionalities. While there are some differences in technical features compared to the predicate devices, the differences have been tested and the conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device and reference devices.
Siemens believes that the subject devices (MAGNETOM Avanto Fit, MAGNETOM Skyra Fit, MAGNETOM Sola Fit, and MAGNETOM Viato. Mobile all with software syngo MR XA70A) are substantially equivalent to the currently marketed devices MAGNETOM Avanto Fit with syngo MR XA50A (K220151, cleared on April 01, 2022), MAGNETOM Skyra Fit with syngo MR XA50A (K220589, cleared on May 13, 2022), MAGNETOM Sola Fit with syngo MR XA51A (K221733, cleared on September 13, 2022), and MAGNETOM Viato. Mobile with syngo MR XA51A (K240608, cleared on March 29, 2024).