K232494 · Siemens Medical Solutions USA, Inc. · LNH · Nov 14, 2023 · Radiology
Device Facts
Record ID
K232494
Device Name
MAGNETOM Avanto fit; MAGNETOM Skyra fit
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Nov 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
AI/ML
Indications for Use
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Device Story
MAGNETOM Avanto and Skyra are MRI systems using magnetic resonance to generate cross-sectional images and spectra of human anatomy. Systems utilize hardware (magnets, gradient systems, RF coils) and syngo MR XA61A software to acquire, reconstruct, and display images. Operated by trained healthcare professionals in clinical settings. New features include GRE_PC sequence, Physiologging for EPI sequences, and Open Recon Framework. Output consists of diagnostic images/spectra interpreted by physicians to assist in clinical diagnosis and interventional guidance. Benefits include enhanced diagnostic visualization and workflow efficiency through updated reconstruction and logging capabilities.
Clinical Evidence
No clinical trials conducted. Evidence consists of bench testing, software verification/validation, and sample clinical images. Performance testing included SNR and image uniformity measurements for new coils and verification of new software features. Clinical publications were referenced to support the use of the GRE_PC feature.
Technological Characteristics
MR diagnostic device; 1.5T/3T magnet systems; RF and gradient systems; new coils (Flex Loop Large, UltraFlex 18, Contour 24); syngo MR XA61A software. Conforms to IEC 60601-1, IEC 60601-2-33, IEC 62304, ISO 14971, and DICOM standards. Connectivity via standard hospital networks.
Indications for Use
Indicated for use as an MR diagnostic device for imaging head, body, or extremities to assist in diagnosis. Applicable to patients requiring MR imaging, including interventional procedures using MR-compatible devices. No specific age or gender contraindications stated beyond standard MR safety precautions.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 14, 2023
Siemens Medical Solutions USA, Inc. Milind Dhamankar Clinical Affairs Professional 40 Liberty Boulevard Malvern, PA 19355
Re: K232494
Trade/Device Name: MAGNETOM Avanto": MAGNETOM Skyrafff Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI, MOS Dated: August 13, 2023 Received: August 17, 2023
Dear Milind Dhamankar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D. R. K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K232494
Device Name
MAGNETOM Avanto fit; MAGNETOM Skyra fit
Indications for Use (Describe)
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92.
#### 1. General Information
| Establishment: | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Malvern, PA 19355, USA<br>Registration Number: 2240869 |
|----------------|-----------------------------------------------------------------------------------------------------------------------|
|----------------|-----------------------------------------------------------------------------------------------------------------------|
- Date Prepared: August 08, 2023
- Manufacturer: Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen Germany Registration Number: 3002808157
#### 2. Contact Information
Milind Dhamankar Clinical Affairs Professional Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355, USA Phone: +1(610) 517-9484 Fax: (610) 448-6547 E-mail: milind.dhamankar@siemens-healthineers.com
#### 3. Device Name and Classification
| Device/ Trade name: | MAGNETOM Avantofit<br>MAGNETOM Skyrafit |
|-----------------------|---------------------------------------------|
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | II |
| Product Code: | Primary: LNH<br>Secondary: LNI, MOS |
{5}------------------------------------------------
## 4. Legally Marketed Predicate and Reference Device
#### Predicate Device for MAGNETOM Skyrafit 4.1.
| Trade name: | MAGNETOM Vida |
|-----------------------|---------------------------------------------|
| 510(k) Number: | K213693 |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | II |
| Product Code: | Primary: LNH<br>Secondary: LNI, MOS |
#### Predicate Device for MAGNETOM Avantofit 4.2.
| Trade name: | MAGNETOM Sola |
|-----------------------|---------------------------------------------|
| 510(k) Number: | K221733 |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | II |
| Product Code: | Primary: LNH |
| | Secondary: LNI, MOS |
#### Reference Device for MAGNETOM Skyrafit 4.3.
| <b>Trade name:</b> MAGNETOM Skyrafit |
|-------------------------------------------------------------------------|
| <b>510(k) Number:</b> K162102 |
| <b>Classification Name:</b> Magnetic Resonance Diagnostic Device (MRDD) |
| <b>Classification Panel:</b> Radiology |
| <b>CFR Code:</b> 21 CFR § 892.1000 |
| <b>Classification:</b> II |
| <b>Product Code:</b> Primary: LNH<br>Secondary: LNI, MOS |
#### 4.4. Reference Device for MAGNETOM Avanto™
| Trade name: | MAGNETOM Avantofit |
|-----------------------|---------------------------------------------|
| 510(k) Number: | K162102 |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | II |
| Product Code: | Primary: LNH<br>Secondary: LNI, MOS |
{6}------------------------------------------------
## 4. Intended Use / Indications for Use
The indications for use for the subject devices are the same as the predicate device:
The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
#### 5. Device Description
The subject devices, MAGNETOM Avanto® and MAGNETOM Skyra™ with software syngo MR XA61A, consist of new and modified software and hardware that is similar to what is currently offered on the predicate device, MAGNETOM Sola with syngo MR XA51A (K221733) and MAGNETOM Vida with syngo MR XA50A (K213693).
A high-level summary of the new and modified hardware and software is provided below:
MAGNETOM Avanto™ with syngo MR XA61A:
#### Hardware
New Hardware:
- Flex Loop Large ।
- -UltraFlex Large 18
- । UltraFlex Small 18
- -Contour 24
Modified Hardware:
- -Host computers ((syngo MR Acquisition Workplace (MRAWP) and syngo MR Workplace (MRWP))
#### Software
New Features and Applications:
- י GRE_PC
- Physiologging .
- । Open Recon Framework
Modified Features and Applications:
- -BEAT_nav (re-naming only)
{7}------------------------------------------------
#### MAGNETOM Skyraft with syngo MR XA61A:
#### Hardware
New Hardware:
- -UltraFlex Large 18
- UltraFlex Small 18 .
- -Contour 24
#### Modified Hardware:
- -Host computers ((syngo MR Acquisition Workplace (MRAWP) and syngo MR Workplace (MRWP))
- MaRS (Measurement and Reconstruction System) computer -
## New Features and Applications:
- । GRE_PC
- . Physiologging
- । Open Recon Framework
Modified Features and Applications:
- -BEAT nav (re-naming only)
- myExam Angio Advanced Assist (Test Bolus)
#### 6. Substantial Equivalence
MAGNETOM Avanto™ and MAGNETOM Skyra™ with software syngo MR XA61A is substantially equivalent to the following predicate device:
| Predicate Device | FDA Clearance Number and<br>Date | Product<br>Code | Manufacturer |
|--------------------------------------------------|----------------------------------|------------------|-------------------------|
| MAGNETOM Sola with <i>syngo</i><br>MR XA51A | K221733 on September 13,<br>2022 | LNH,<br>LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Vida with <i>syngo</i><br>MR XA50A | K213693 on February 25, 2022 | LNH,<br>LNI, MOS | Siemens Healthcare GmbH |
| Reference Device | FDA Clearance Number and<br>Date | Product<br>Code | Manufacturer |
| MAGNETOM Avantofit with<br><i>syngo</i> MR VE11E | K162102 on November 22,<br>2016 | LNH,<br>LNI, MOS | Siemens Healthcare GmbH |
| MAGNETOM Skyrafit with | K162102 on November 22, | LNH, | Siemens Healthcare GmbH |
# 7. Technological Characteristics
The subject devices, MAGNETOM Avanto® and MAGNETOM Skyra™ with software syngo MR XA61A, are substantially equivalent to the predicate devices with regard to the operational environment, programming language, operating system and performance.
The subject devices conform to the standard for medical device software (IEC 62304) and other relevant IEC and NEMA standards.
{8}------------------------------------------------
There are some differences in technological characteristics between the subject device and predicate device, including new and modified hardware/software. Here is summary of differences:
| | Subject Devices | Predicate Device | Reference Devices |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Hardware | MAGNETOM Avantofit<br>MAGNETOM Skyrafit<br>with software syngo MR<br>XA61A | MAGNETOM Vida with<br>syngo MR XA50A<br>(K213693)<br>MAGNETOM Sola with<br>syngo MR XA51A<br>(K221733) | MAGNETOM Avantofit<br>MAGNETOM Skyrafit with<br>syngo MR VE11E1<br>(K162102) |
| Magnet System | Yes | Yes | Yes |
| RF System | Yes | Yes | Yes |
| Transmission<br>technique | Yes | Yes | Yes |
| Gradient System | Yes | Yes | Yes |
| Patient Table | Yes | Yes | Yes |
| Multi-Nuclear<br>Option - Supported<br>Nuclei | Yes<br>Only for MAGNETOM Skyrafit | Yes<br>Only for MAGNETOM Vida | Yes<br>Only for MAGNETOM<br>Skyrafit |
| Computer | Yes<br>Modified compared to<br>predicate device:<br>- New MRAWP and MRWP<br>- New MaRS hardware for<br>MAGNETOM Skyrafit | Yes | Yes |
| Coils | Yes, new coils<br>- Flex Loop Large (only for<br>MAGNETOM Avantofit)<br>- UltraFlex Large 18<br>- UltraFlex Small 18<br>- Contour 24/482 | Yes | Yes |
| Other HW<br>components | Yes | Yes | Yes |
Summary hardware comparison table for the subject and predicate/reference device
Summary software comparison table for the subject and predicate devices
| | Subject Devices | Predicate Device |
|----------|----------------------------------------|--------------------|
| Software | MAGNETOM Avantofit | MAGNETOM Vida with |
| | MAGNETOM Skyrafit | syngo MR XA50A |
| | with software syngo MR XA61A (K213693) | |
<sup>1</sup> Change released by internal documentation based on MAGNETOM Skyraf" with syngo MR E11C (K162102) 2 Contour 24 (K183111, cleared on December 7, 2018 and K173446, was cleared November 17, 2017)
{9}------------------------------------------------
| | | MAGNETOM Sola with |
|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------|
| | | syngo MR XA51A (K221733) |
| Sequences | | |
| SE-based pulse sequence types | Yes | Yes |
| GRE-based/Steady-State pulse<br>sequence types | New or modified pulse sequences:<br>- GRE_PC new pulse sequence<br>- BEAT_NAV pulse sequence re-<br>naming | Yes |
| EPI-based pulse sequence types | New features:<br>- Physiologging for EPI2D_BOLD<br>and EPI2D_PACE | Yes |
| Spectroscopy pulse sequence types | Yes | Yes |
| Feature and Applications | | |
| Other features and<br>applications such as:<br>-Application Suites<br>-myExam Assists<br>-Other Imaging Applications | Modified application feature:<br>- myExam Angio Assist (Test Bolus<br>and Care bolus) for MAGNETOM<br>Skyrafit | Yes |
| User interface and user interaction | Yes | Yes |
| Viewing and post-processing | Yes | Yes |
| Workflow and software utilization | Yes | Yes |
| Patient Management | Yes | Yes |
| Scan Modes and Pulse Sequences | Yes | Yes |
| Scanning | Yes | Yes |
| Reconstruction | New feature:<br>- OpenRecon Framework | Yes |
| Image Display | Yes | Yes |
| File/Data Management | Yes | Yes |
The differences have been tested and the conclusion from the non-clinical data suggests that the features bear an equivalent safety and performance profile to that of the predicate device.
#### 8. Nonclinical Tests
The following performance testing was conducted on the subject devices:
| Performance Test | Tested Hardware or<br>Software | Source/Rationale for test |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Software verification<br>and validation | New or modified software<br>features | Guidance for the Content of Premarket<br>Submissions for Software Contained in<br>Medical Devices |
| Sample clinical images | New or modified software<br>features and coils | Guidance for submission of Premarket<br>Notifications for Magnetic Resonance<br>Diagnostic Devices |
| Image quality<br>assessment by sample<br>clinical images | - new / modified pulse<br>sequence types.<br>- comparison images<br>between the new / modified<br>features and the predicate<br>device features | |
{10}------------------------------------------------
| Physiologging<br>Verification Report | Physiologging | New Feature Introduction |
|--------------------------------------|---------------|--------------------------|
|--------------------------------------|---------------|--------------------------|
The following performance testing for local coils was conducted on the predicate and the reference devices and can be reused for the subject devices:
| Performance Test | Tested Hardware or Software | Source/Rationale for test |
|------------------------|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Performance bench test | - SNR and image uniformity<br>measurements for coils<br>- Heating measurements for coils | Guidance for Submission of<br>Premarket Notifications for<br>Magnetic Resonance Diagnostic<br>Devices |
The results from each set of tests demonstrate that the devices perform as intended and are thus substantially equivalent to the predicate device to which it has been compared.
## 9. Clinical Tests / Publications
No clinical tests were conducted to support substantial equivalence for the subject devices; however, as stated above, sample clinical images were provided.
Furthermore, additional clinical publications were referenced to provide information on the use of the following features and functions:
| Feature | Publications |
|---------|--------------------------------------------------------------------------------------------------------------|
| GRE_PC | [1] Guenthner C. et al. Ristretto MRE: A generalized multi-shot GRE-MRE sequence. NMR Biomed 2019; 32:e4049. |
# 10. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle and continuously throughout the development of the product. Siemens Healthcare GmbH adheres to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
MAGNETOM Avanto® and MAGNETOM Skyra™ with software syngo MR XA61A conforms to the following FDA recognized and international IEC, ISO and NEMA standards:
{11}------------------------------------------------
| Recognitio<br>n Number | Product<br>Area | Title of Standard | Reference Number<br>and date | Standards<br>Development<br>Organization |
|------------------------|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|------------------------------------------|
| 19-4 | General | Medical electrical equipment - part 1:<br>general requirements for basic safety<br>and essential performance | ES60601-<br>1:2005/(R)2012 and<br>A1:2012<br>C1:2009/(R)2012 | AAMI / ANSI |
| 19-8 | General | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety<br>and essential performance - Collateral<br>Standard: Electromagnetic disturbances<br>- Requirements and tests | 60601-1-2 Edition<br>4.0:2014-02 | IEC |
| 12-295 | Radiology | Medical electrical equipment - Part 2-<br>33: Particular requirements for the<br>basic safety and essential performance<br>of magnetic resonance equipment for<br>medical diagnosis | 60601-2-33 Ed. 3.2<br>b:2015 | IEC |
| 5-125 | General | Medical devices - Application of risk<br>management to medical devices | 14971 Third Edition<br>2019-12 | ISO |
| 5-114 | General I<br>(QS/<br>RM) | Medical devices - Part 1: Application of<br>usability engineering to medical devices | 62366-1:2015 | ANSI AAMI IEC |
| 13-79 | Software/<br>Informatics | Medical device software - Software life<br>cycle processes | 62304 Edition 1.1<br>2015-06<br>CONSOLIDATED<br>VERSION | IEC |
| 12-195 | Radiology | NEMA MS 6-2008 (R2014)<br>Determination of Signal-to-Noise Ratio<br>and Image Uniformity for Single-<br>Channel Non-Volume Coils in<br>Diagnostic MR Imaging | MS 6-2008 (R2014) | NEMA |
| 12-342 | Radiology | Digital Imaging and Communications in<br>Medicine (DICOM) Set | PS 3.1 - 3.20<br>(2021e) | NEMA |
| 2-258 | Biocompati<br>bility | Biological evaluation of medical devices<br>- part 1: evaluation and testing within a<br>risk management process.<br>(Biocompatibility) | 10993-1 Fifth<br>edition 2018-08 | AAMI<br>ANSI<br>ISO |
# 11. Conclusion as to Substantial Equivalence
MAGNETOM Avanto® and MAGNETOM Skyra® with software syngo MR XA61A have the same intended use and same basic technological characteristics than the predicate device system, MAGNETOM Vida with syngo MR XA50A and MAGNETOM Sola with syngo MR XA51A, with respect to the magnetic resonance features and functionalities. While there are some differences in technical features compared to the predicate device, the differences have been tested and the
{12}------------------------------------------------
conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device and reference devices.
Siemens believes that MAGNETOM Avanto® and MAGNETOM Skyra™ with software syngo MR XA61A are substantially equivalent to the currently marketed devices MAGNETOM Vida with software syngo MR XA50A (K213693, cleared on February 25, 2022) and MAGNETOM Sola with syngo MR XA51A (K221733 on September 13, 2022).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.